SC9000/SC9015 MEDICAL INFORMATION BUS (MIB) PROTOCOL CONVERTER

K973222 · Siemens Medical Solutions USA, Inc. · DQA · Sep 4, 1997 · Cardiovascular

Device Facts

Record IDK973222
Device NameSC9000/SC9015 MEDICAL INFORMATION BUS (MIB) PROTOCOL CONVERTER
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeDQA · Cardiovascular
Decision DateSep 4, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2

Indications for Use

The Medical Information Bus (MIB) Protocol Converter is intended to connect third party medical devices such as Siemens SV300™ ventilator, Baxter Vigilance™ blood gas/continuous cardiac output monitor, Siemens SV900™ ventilator, Puritan Bennett 7200™ ventilator, and the Draeger Evita IITM, Draeger Evita IV™ and Draeger Babylog™ ventilators that do not provide data per the IEEE 1073 Medical Information Protocol Standard to the Siemens SC9000/SC9015 Bedside Monitor for display.

Device Story

Device acts as interface/protocol converter; transforms data from third-party medical devices (ventilators, blood gas/cardiac output monitors) into IEEE 1073 Medical Information Bus (MIB) protocol format. Input: serial RS-232 or analog signals from third-party devices. Output: IEEE 1073 compliant data transmitted to Siemens SC9000/SC9015 bedside monitors. Used in clinical settings (ICU/OR) by healthcare professionals. Enables centralized display of third-party device data on Siemens monitors; facilitates clinical monitoring by integrating disparate device outputs into a single bedside system.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Protocol converter; utilizes RS-232 and analog input ports; IEEE 1073-compliant output port. Complies with IEEE 1073.3.1-1994 (Transport Profile) and IEEE 1073.4.1-1994 (Physical Layer Interface). Not MRI compatible.

Indications for Use

Indicated for use in clinical environments by healthcare professionals (physicians, nurses, technicians) to connect specific third-party ventilators and blood gas/cardiac output monitors to Siemens SC9000/SC9015 bedside monitors for data display.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 3)22 368 Amendment Siemens SC9000/SC9015 Medical Information Bus (MIB) Protocol Converter ## 510(k) AMENDMENT SUMMARY as required per 807.92(c) 2. Submitters Name, Address: - 4 1997 Siemens Medical Systems, Inc. Electromedical Systems Group, PCS Danvers, MA 01923 Tel: (508) 750-7500 Fax: (508) 777-3398 Official Correspondent: David Simard, Director Quality Assurance & Regulatory Affairs Contact person for this submission: Jacqueline E. M. Emery Date amendment to 510(k) K970368 was prepared: August 8, 1997 Trade Name, Common Name and Classification Name: 3. A. Trade Name: (Same as original submission) Siemens SC9000/ SC9015 Series Medical Information Bus (MIB) Protocol Converter [510(k) K970368 cleared May 6, 1997] - B. Common Name, Classification Name, Class and Regulation Number: (Same as original submission) | Common Name | Classification<br>Number | Class | Regulation Number | |---------------------------------------------|--------------------------|-------|-------------------| | Transducer Signal amplifier and conditioner | 73 DQA | II | 21 CFR 870.2060 | - Predicate Device Identification: (Same as original submission) 4. Hewlett-Packard Model M1032A Vuelink Interface Plug In Module cleared under 510(K) number: K923682 - 5. Device Description: (Same as original submission) The Siemens SC9000/SC9015 Series Medical Information Bus (MIB) Protocol Converter transforms data generated by third party devices (e.g. ventilators, SvO2 devices and infusion pumps) into a common format per the IEEE 1073 Medical Information Protocol and transmits this data to the Siemens SC9000/SC9015 Bedside Monitor. Attachment cables specific to third party devices are offered as accessories. #### 6. Intended Use: The Medical Information Bus (MIB) Protocol Converter is intended to connect third party medical devices such as Siemens SV300™ ventilator, Baxter Vigilance™ blood gas/continuous cardiac output monitor, Siemens SV900™ ventilator, Puritan Bennett 7200™ ventilator, and the Draeger Evita IITM, Draeger Evita IV™ and Draeger Babylog™ ventilators that do not provide data per the IEEE 1073 Medical Information Protocol Standard to the Siemens SC9000/SC9015 Bedside Monitor for display. 1 Siemens Medical Systems, Inc. Electromedical Systems Group, PCS 16 Electronics Avenue Danvers, MA 01923 USA Tel: (508) 750-7500 Fax: (508) 777-3398 Telex: 511958 (Siemensm SD) {1}------------------------------------------------ ## 7. Table of Device Similarities and differences to predicate device (Same as original submission) | | Substantial Equivalent Device | Applicant | |----------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------| | | Hewlett Packard Co | Siemens Medical Systems | | | Model MI032A VueLink Module | SC9000/9015 Series Medical<br>Information Bus (MIB) Protocol<br>Converter (Amendment) | | Manufacturer | Hewlett Packard | Siemens Medical Systems -<br>Electromedical Systems Group, PCS | | 510(k) Number | K923682 | K970368 K973222 | | Intended Use | Provides an external device interface<br>capability to third party devices that<br>have a serial RS-232 and/or analog<br>output | Same | | Intended Population | NA<br>Not connected to patients | Same | | Intended Environment | OR or ICU | Same | | Input port | RS-232, Analog, Analog/Digital<br>combination | Same | | Output port | HP specific | IEEE 1073 (MIB connector) | Assessment of non-clinical performance data for equivalence: (Same as original submission) 8. Substantial equivalence is claimed to the Hewlett Packard Model M1032A VueLink Interface Plug-In Module cleared under 510(k) number K923682. - Assessment of clinical performance data for equivalence: Not applicable 9. (Same as original submission) 10. Biocompatability: Not applicable (Same as original submission) 11. Sterilization: Not applicable (Same as original submission) 12. Standards and Guidances: (Same as original submission) Currently, there are no FDA standards for this device. However, the Siemens SC9000/SC9015 Series Medical Information Bus (MIB) Protocol Converter complies with: 1073.3.1 - 1994 IEEE Standard for Medical Device Communications -Transport Profile - Connector Mode - - 1073.4.1 1994 IEEE Standard for Medical Device Communications ﺳ Physical Layer Interface - Cable Conn . Siemens Medical Systems, Inc. Electromedical Systems Group, PCS 16 Electronics Avenue Danvers, MA 01923 USA Tel: (508) 750-7500 Fax: (508) 777-3398 Telex: 511958 (Siemensm SD) <sup>2</sup> {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856 Mr. David Simard ~ · · · · · · Siemens Medical Systems, Inc. 16 Electronics Avenue Danvers, Massachusetts 01923 K973222 Re: Siemens SC9000/SC9015 Series Medical Information Bus (MIB) Protocol Converter for the Puritan Bennett 7200™, the Draeger Evita II™, IV™, Babylog™, and Siemens SV900™ Ventilators Requlatory Class: II (two) Product Code: 73 DQA Dated: August 18, 1997 Received: August 21, 1997 Dear Mr. Cohen: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ # Page 2 - Mr. David Simard This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html". Sincerely yours, Thomas J. Callahon Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ Siemens SC9000/SC9015 Series Medical Information Bus (MIB) Device Name:__________________________________________________________________________________________________________________________________________________________________ Protocol Converter Indications for Use:-- ...... ................................................................................................................................................ The Siemens SC9000/SC9015 Series Medical Information Bus (MIB) Protocol Converter is indicated for use in an environment where patient care is provided by healthcare professionals (Physician, Nurse, Technician) when the professional determines that the Siemens SV300™ ventilator, Baxter Vigilance™ blood gas/continuous cardiac output monitor, Siemens SV900™ ventilator, Draeger Evita II™, Evita IVTM, and Draeger Babylog™ ventilators, and Puritan Bennett 7200™ ventilator should be connected to the Siemens SC9000/SC9015 Bedside Monitor for display. MRI Compatibility Statement: The Siemens SC9000/SC9015 Series Medical Information Bus (MIB) Protocol Converter is not compatible for use in an MRI magnetic field. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | | Concurrence of CDRH, Office of Device Evaluation (ODE) | |----------------------|--------------------------------------------------------| | Prescription Use | | | (Per 21 CFR 801.109) | | | | OR | | | Over-The-Counter Use | | | (Division Sign-Off) | | | Division of Cardiovascular, Respiratory, | | | and Neurological Devices | 510(k) Number_________________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...