SC9000/SC9015 MEDICAL INFORMATION BUS (MIS) PROTOCOL CONVERTER

K970368 · Siemens Medical Solutions USA, Inc. · DQA · May 6, 1997 · Cardiovascular

Device Facts

Record IDK970368
Device NameSC9000/SC9015 MEDICAL INFORMATION BUS (MIS) PROTOCOL CONVERTER
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeDQA · Cardiovascular
Decision DateMay 6, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2

Indications for Use

The Medical Information Bus (MIB) Protocol Converter is intended to connect third party medical devices that do not provide data per the IEEE 1073 Medical Information Protocol Standard to the Siemens SC9000/SC9015 Bedside Monitor for display.

Device Story

Device acts as interface/protocol converter; transforms data from third-party medical devices (ventilators, infusion pumps, SvO2 monitors) into IEEE 1073 Medical Information Bus (MIB) format. Inputs include serial RS-232 or analog signals from third-party equipment. Output is IEEE 1073-compliant data transmitted to Siemens SC9000/SC9015 bedside monitors for display. Used in OR or ICU environments by clinicians. Facilitates centralized monitoring of disparate medical equipment data on a single display, aiding clinical decision-making by integrating patient data from multiple sources.

Clinical Evidence

No clinical data; not applicable.

Technological Characteristics

Protocol converter; inputs: RS-232, analog; output: IEEE 1073 (MIB) connector. Complies with IEEE 1073.3.1-1994 (Transport Profile) and IEEE 1073.4.1-1994 (Physical Layer Interface).

Indications for Use

Indicated for connecting third-party medical devices (e.g., ventilators, SvO2 devices, infusion pumps) to Siemens SC9000/SC9015 bedside monitors. No specific patient population or contraindications defined; device is not connected to patients.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Submission Summary (Full Text)

{0} 510 (K) Notification Siemens SC9000/SC9015 Medical Information Bus (MIB) Protocol Converter K970368 # 510(k) SUMMARY as required per 807.92(c) MAY - 6 1997 ## 2: Submitters Name, Address: Siemens Medical Systems, Inc. 16 Electronics Avenue Danvers, MA 01923 Tel: (508) 750-7500 Fax: (508) 777-3398 Official Correspondent: David Simard, Director Quality Assurance and Regulatory Affairs Contact person for this submission: Jacqueline E. M. Emery Date submission was prepared: January 7, 1997 ## 3: Trade Name, Common Name and Classification Name: A. Trade Name: Siemens SC9000/ SC9015 Series Medical Information Bus (MIB) Protocol Converter B. Common Name, Classification Number, Class and Regulation Number: | Common Name | Classification Number | Class | Regulation Number | | --- | --- | --- | --- | | Transducer Signal amplifier and conditioner | 73 DQA | II | 21CFR870.2060 | ## 4: Predicate Device Identification: Hewlett-Packard Model M1032A Vuelink Interface Plug In Module cleared under 510(K) number: K923682 ## 5. Device Description: The Siemens SC9000/SC9015 Series Medical Information Bus (MIB) Protocol Converter transforms data generated by third party devices (e.g. ventilators, SvO2 devices and infusion pumps) into a common format per the IEEE 1073 Medical Information Protocol and transmits this data to the Siemens SC9000/SC9015 Bedside Monitor. Attachment cables specific to third party device are offered as accessories. ## 6. Intended Use: The Medical Information Bus (MIB) Protocol Converter is intended to connect third party medical devices that do not provide data per the IEEE 1073 Medical Information Protocol Standard to the Siemens SC9000/SC9015 Bedside Monitor for display. Siemens Medical Systems, Inc. Electromedical Group 16 Electronics Avenue Danvers, MA 01923 USA Tel: (508) 750-7500 Fax: (508) 777-3398 Telex: 511958 (Siemensm SD) {1} 510 (K) Notification Siemens SC9000/SC9015 Medical Information Bus (MIB) Protocol Converter 7. Table of device similarities and differences to predicate device | | Substantial Equivalent Device Hewlett Packard Co Model M1032A VueLink Module | Applicant Siemens Medical Systems SC 9000/9015 Series Medical Information Bus (MIB) Protocol Converter | | --- | --- | --- | | Manufacturer | Hewlett Packard | Siemens Medical Systems - Electromedical Group | | 510K Number | K923682 | To be assigned | | Intended Use | Provides an external device interface capability to third party devices that have a serial RS-232 and/or analog output | Same | | Intended Population | NA Not connected to patients | Same | | Intended Environment | OR or ICU | Same | | Input port | RS-232, Analog, Analog/Digital combination | Same | | Output Port | HP specific | IEEE 1073 (MIB Connector) | 8. Assessment of non-clinical performance data for equivalence: Substantial equivalence is claimed to the Hewlett Packard Model M1032A Vuelink Interface Plug-In Module cleared under 510(K) number K923682. 9. Assessment of clinical performance data for equivalence: Not applicable 10. Biocompatibility: Not applicable 11. Sterilization: Not applicable 12. Standards and Guidelines: Currently there are no FDA standards for this device. However, the Siemens SC9000/SC9015 Series Medical Information Bus (MIB) Protocol Converter complies with: - 1073.3.1-1994 IEEE Standard for Medical Device Communications - Transport Profile- Connector Mode - 1073.4.1-1994 IEEE Standard for Medical Device Communications - Physical Layer Interface - Cable Connected Siemens Medical Systems, Inc. Electromedical Group 16 Electronics Avenue Danvers, MA 01923 USA Tel: (508) 750-7500 Fax: (508) 777-3398 Telex: 511958 (Siemensm SD)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...