SpO2 Sensor

K183277 · Shenzhen Upnmed Equipment Co., Ltd. · DQA · May 15, 2019 · Cardiovascular

Device Facts

Record IDK183277
Device NameSpO2 Sensor
ApplicantShenzhen Upnmed Equipment Co., Ltd.
Product CodeDQA · Cardiovascular
Decision DateMay 15, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2
AttributesPediatric

Indications for Use

This SpO2 Sensor is intended for the continuous noninvasive monitoring and spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR, measured by an SpO2 sensor) for use with the finger of adult and pediatric patients. It acts as re-usable accessory of a Patient Monitor.

Device Story

SpO2 sensor (models U401-A/B/C/D) functions as reusable accessory for patient monitors; utilizes red (660 nm) and infrared (905 nm) LEDs to transmit light through finger tissue; photodetector measures light absorption variations caused by blood pulse and hemoglobin oxygenation; microprocessor calculates SpO2 and pulse rate values. Used in clinical settings by healthcare providers; output displayed on host patient monitor to assist in patient status assessment and clinical decision-making regarding oxygenation levels.

Clinical Evidence

Clinical validation conducted per ISO 80601-2-61:2011 Annex EE. Study involved healthy volunteers (ages 18-55) with varied skin pigmentations, including three darkly pigmented subjects. Accuracy validated against arterial blood samples measured via CO-oximeter. Results demonstrate SpO2 accuracy of ±2% and pulse rate accuracy of ±2 bpm.

Technological Characteristics

Reusable finger-clip SpO2 sensor; 2-wavelength optical absorption (660 nm/905 nm); photodetector sensing. Complies with ISO 10993-1 (biocompatibility), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and ISO 80601-2-61 (performance). No embedded software.

Indications for Use

Indicated for continuous or spot-check monitoring of functional arterial hemoglobin oxygen saturation (SpO2) and pulse rate in adult and pediatric patients via finger sensor. Contraindications: none stated.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. Shenzhen Upnmed Equipment Co., Ltd. Fu Jian Management Representative 4th Floor, Building #1 East, Huihuang Industrial Area Xitian Community, Gongming Town. Guangming District Shenzhen, 518107 Cn Re: K183277 Trade/Device Name: SpO2 Sensor U401-A, U401-B, U401-C, U401-D Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: April 8, 2019 Received: April 16, 2019 Dear Fu Jian: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Michael Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K183277 Device Name SpO2 Sensor, models U401-A, U401-B, U401-C, U401-D #### Indications for Use (Describe) This SpO2 Sensor is intended for the continuous nonitoring and spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR, measured by an SpO2 sensor) for use with the finger of adult and pediatric patients. It acts as re-usable accessory of a Patient Monitor. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------| | <span style="font-size: 10pt;"> <span style="font-family: Wingdings;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size: 10pt;"><span style="font-family: Wingdings;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary # K183277 # Prepared in accordance with the requirements of 21 CFR Part 807.92 | 1. Submitter: | Shenzhen Upnmed Equipment Co., Ltd.<br>4th Floor, Building #1 East, Huihuang Industrial Area, Xitian Community,<br>Gongming Town, Guangming District, 518107, Shenzhen, P.R. China<br>Tel.: +86 -755-29728789<br>Fax: +86 -755-29423789 | |---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Fu Jian | | Prepare date: | 2019-04-08 | | 2. Device name<br>and classification: | Device Name: SpO2 Sensor<br>Models: U401-A, U401-B, U401-C, U401-D<br>Classification Name: 21 CFR 870.2700 Oximeter<br>Product code: DQA<br>Regulatory Class: Class II | | 3. Reason for<br>Submission | New Application. | | 4. Predicate<br>Device(s): | Shenzhen Med-link Electronics Tech Co., Ltd., Shenzhen Med-Link Pulse Oximeter<br>Probe, model S0136J-L/ K113727 | | 5. Device<br>Description: | The SpO2 Sensor consists of compatible connectors, cable, patient side sensor. And<br>one side of sensor is designed to locate light emitting diodes (LEDs) and a light<br>detector (called a photo-detector). Red and Infrared lights are shone through the<br>tissues from one side of the probe to the other. Then parts of the light emitted<br>absorbed by blood and tissues. The light absorbed by the blood varies with the<br>oxygen saturation of haemoglobin. After that, the photo-detector detects the light<br>volume transmitted through the tissues which depends on blood pulse, Hereafter, the<br>microprocessor calculates a value for the oxygen saturation (SpO2).<br><br>The U401 series SpO2 Sensor acts as re-usable accessory of Mindray PM-8000<br>Express Patient Monitor, which has been cleared by K070791. And the function<br>alarms and interface accessories are controlled by the monitor, and such information<br>is not available in this submission. | {4}------------------------------------------------ This SpO2 Sensor is intended for the continuous noninvasive monitoring and 6. Indications for spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and Use: pulse rate (PR, measured by an SpO2 sensor) for use with the finger of adult and pediatric patients. It acts as re-usable accessory of a Patient Monitor. # 7. Predicate Device Comparison Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness as the predicate device. Please refer to following table to find differences between the subject device and predicate device. All the differences do not affect the basic design principle, usage, effectiveness of the subject device. And the differences do not raise different questions of safety of effectiveness. | ITEM | Proposed Device<br>U401 Series SpO2 Sensor | Predicate Device<br>S0136J-L/K113727 | Comparison<br>Result | |-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------| | Manufacture | Shenzhen Upnmed Equipment<br>Co., Ltd. | Shenzhen Med-link Electronics<br>Tech Co., Ltd. | --- | | Indications for Use | This SpO2 Sensor is intended<br>for the continuous noninvasive<br>monitoring and spot-checking<br>of functional oxygen saturation<br>of arterial hemoglobin (SpO2)<br>and pulse rate (PR, measured<br>by an SpO2 sensor) for use<br>with adult and pediatric<br>patients. | Shenzhen Med-link disposable<br>probe, model S0136J-L, is<br>indicated for single patient use<br>when continuous, non-invasive<br>arterial oxygen saturation and<br>pulse rate monitoring are<br>required for adult patients or<br>pediatric patients weighing more<br>than 30 kg. The S0136J-L is<br>contraindicated for use with<br>patients during motion<br>condition. | Different 1 | | Intended patient<br>population | Pediatric, Adult | Pediatric, Adult | Same | | Intended<br>application site | Finger | Finger | Same | | use under motion<br>and low perfusion<br>conditions | No | No | Same | | Measurement<br>Principles | 2-wavelength Relative<br>Optical Absorption | 2-wavelength Relative<br>Optical Absorption | Same | | Light Emitting<br>Diodes (LEDs)<br>wavelengths | RED: 660 nm, nominal<br>IRED: 905nm, nominal | RED: 660 nm, nominal<br>IRED: 905 nm, nominal | Same | | Signal Detection<br>Method | Photodetector | Photodetector | Same | | SpO2 Range | 70% -100% | 70%-100% | Same | | SpO2 Accuracy | ±2% | ±3% | Different 2 | | Pulse Rate Range | 20 bpm - 250 bpm | 20 bpm – 250 bpm | Same | | Pulse Rate<br>Accuracy | ±2 bpm | ±3 bpm | Different | | Measuring Mode | spot-checking or continuous<br>monitoring | spot-checking or continuous<br>monitoring | Same | | Shipped Sterile | No | No | Same | | Storage and<br>Transport | Temperature: -10℃to 50°C<br>Atmospheric Pressure: 50 kPa<br>to 106 kPa<br>Relative Humidity: ≤93% (no<br>condensation) | Temperature: -10℃to 40°C<br>Atmospheric Pressure: 86 kPa to<br>106 kPa<br>Relative Humidity: 0%-80% (no<br>condensation) | Different 3 | | Operating<br>Temperature | Temperature: 5°C to 40°C<br>Atmospheric Pressure: 50 kPa<br>to 106 kPa<br>Relative Humidity: ≤85% (no<br>condensation) | Temperature: -10℃to 40°C<br>Atmospheric Pressure: 86 kPa to<br>106 kPa<br>Relative Humidity: 0%-80% (no<br>condensation) | | | Physical Specifications | | | | | Cable Length | 1.0 m | 0.9 m | Different 4 | Table 1 Comparison between main predicate S0136J-L and the subject device {5}------------------------------------------------ ### Justification for the differences: - 1) Different Indications for Use As indicated in the comparison table, the difference of indications for Use is caused by different language, so the difference do not raise different questions of safety and effectiveness. - 2) Different SpO2 Accuracy The accuracy of the subject device is ±2%, which of the predicate is ±3%, which means the performance of the subject is better, and such specification is verified per the international standard ISO 80601-2-61, so the different accuracy will be acceptable for the subject probe. - 3) Different Storage and Transport Minor difference to Storage & Transport and operation environments (including Temperature, Atmospheric Pressure and Relative Humidity) for the subject device, but the system has been proved to be safe and effective since the testing was conducted under the suggested environment; Moreover, environment testing data shows the device can work as declared under the suggested conditions. So those changes do not raise different questions of safety and effectiveness. - 4) Different Physical Specifications The subject and the predicate are of different size but proximity. Moreover, such engineering design has been verified during the international standards, so such minor differences do not raise different questions of safety and effectiveness. As seen in the comparison tables, the subject and predicate devices have same design principle, similar design features and performance specifications. The main differences between the subject and predicate devices are minor, including only the operating/storage environment, cable length which do not raise different questions of safety or effectiveness. Moreover, as demonstrated in the non-clinical, the different technological characteristics do not raise different questions of safety and effectiveness. # 8. Performance Testing: Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below. # Non-Clinical Data: {6}------------------------------------------------ The following performance data were provided in support of the substantial equivalence determination. ### Biocompatibility testing The biocompatibility evaluation for the SpO2 Sensor was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the battery of testing included the following tests: - Cytotoxicity - Skin Sensitization - Skin Irritation ### Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted, and the results show that the subject device complies with the IEC 60601-1: 2005+CORR. 1 (2006)+CORR. 2 (2007)+AM1 (2012) Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety and the IEC 60601-1-2: 2007 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests standard for EMC. # Bench Testing Bench testing was conducted and the results show that the subject device complies with the ISO 80601-2-61: 2011 Medical electrical equipment — Part 2-61: Particular requirements for basic safety and essential performance of Pulse Oximeter Equipment standard for performance effectiveness. ### Software Verification and Validation Testing Not applicable, no software is embedded. # Clinical data: Clinical testing is conducted per Annex EE Guideline for evaluating and documenting SpO2 ACCURACY in human subjects of ISO 80601-2-6:2011 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment. The SpQ2 accuracy has been validated in human studies against arterial blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically distributed, only about two-thirds of the measurements can be expected to fall within the specified accuracy compared to CO-oximeter measurements. The volunteer population in the studies composed of local healthy men and women from age 18 to 55, with variations of skin pigmentations and per FDA's guidance for Pulse Oximeters, darkly pigmented subjects are included in the desaturation study is three. # Summary Based on the non-clinical performance and clinical data as documented in the device development, the subject devices were found to be as safe and as effective as the predicate device. # 9. Conclusion: Verification and validation testing was conducted on the subject device and all testing passed pre-specified criteria. This premarket notification submission demonstrates that the subject device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...