Fingertip Pulse Oximeter MD300C is a portable non-invasive, spot-check, oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care and etc). Not for continuously monitoring.
Device Story
Fingertip Pulse Oximeter MD300C is a portable, non-invasive device for spot-checking arterial hemoglobin oxygen saturation (SpO2) and pulse rate. It utilizes a dual-wavelength LED (660 nm red, 940 nm infrared) and a photodetector to measure light absorption in the arteriolar vascular bed of a finger. The device processes these signals via an internal CPU to calculate SpO2 based on the ratio of light absorption during systole and diastole. It features a digital display for results, a low-battery alarm, and an automatic power-off function. Powered by two AAA batteries, it is intended for use by clinicians in hospitals (e.g., surgery, anesthesia, intensive care) or by patients at home. The device provides immediate readings to assist healthcare providers in assessing patient oxygenation status. It is not intended for continuous monitoring, life support, or implantation.
Clinical Evidence
Clinical testing conducted at Beijing Friendship Hospital and Wulanchabu City Center Hospital in accordance with ISO 9919:2005. The study confirmed the accuracy of the MD300C pulse oximeter and its safety for use in clinical environments. No specific numerical sensitivity/specificity metrics were provided in the summary, but the device was found to be compliant with safety and effectiveness requirements.
Technological Characteristics
Fingertip form factor; dual-wavelength LED (660nm/940nm) and photodetector sensor; CPU-based signal processing; powered by 2 AAA batteries. Biocompatible skin-contact materials (ISO 10993 compliant). Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Non-sterile, reusable.
Indications for Use
Indicated for spot-check measurement of SpO2 and pulse rate in adult and pediatric patients in home and hospital settings. Not for continuous monitoring.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Model 9550 Onyx® II Finger Pulse Oximeter (K051107)
K101577 — MODIFICATION TO:FINGERTIP PULSE OXIMETER, MODEL MD300C SERIES · Beijing Choice Electronic Technololgy Co., Ltd. · Dec 2, 2010
K072825 — FINGERTIP PULSE OXIMETER, MODEL MD300I · Beijing Choice Electronic Technololgy Co., Ltd. · Jan 14, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
# AUG 1 3 2007
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 9807.92.
The Assigned 510(k) Number is:
K07037
# 1. Applicant Device Information
510(k) Preparation Date: Jan 23, 2007 Trade/Proprietary Name: Fingertip Pulse Oximeter MD300C Common Name: Oximeter Classification Name: Oximeter Device Class: II Product Code: DQA Regulation Number: 870.2700 Intended Use: Fingertip Pulse Oximeter MD300C is a portable non-invasive, spot-check, oxygen saturation of
arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care and etc). Not for continuously monitoring.
# 2. Submitter Information
#### Manufacturer Name:
Beijing Choice Electronic Technology Co., Ltd. Bailangyuan Building B 1127-1128, Fuxing Road A36 Beijing, China 100039
## Contact Person of the Submission:
Ms. Diana Hong Mr. Eric Chen Suite 8D, Zhongxin Zhongshan Mansion, No.19, Lane 999, Zhong Shan Nan Er Road Shanghai, China 20020 Phone: +86-21-64264467 x 152 Fax: +86-21-64264468 x 809 Email: Diana.hong@mid-link.net
A 1980 M E 1998 States Comments of Children
{1}------------------------------------------------
## 3. Predicate Device
The Legally Marketed Contact Lens Case as predicate devices is identified as followings:
Model 9550 Onyx® II Finger Pulse Oximeter
K-number: K051107
Product Code: DQA
Intended Use: The Nonin® Onyx® II Model 9550 Finger Pulse Oximeter is a small, lightweight portable device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO 2) and pulse rate. It is intended for spotchecking of adult and pediatric patients on fingers (other than the thumb) between 0.3 - 1.0 inch (0.8 - 2.5 cm) thick. The index finger is the recommended site.
#### Manufactured by:
Nonin Medical, Inc. 13700 1 st Ave. N. Plymouth, MN 55441-5443
## 4. Device Description
The applicant device of Fingertip Pulse Oximeter MD300C is a fingertip device, which can display % SpO2, pulse rate value.
The applicant device consists of detector and emitter LED, signal amplify unit, CPU, data display unit and power unit.
The Pulse oximetry works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660 nm, which is red light; the other is 940 nm, which is ultra red light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photodetector in fingersensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastele is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.
The applicant device has low battery voltage alarm function and automatically power of function. The power source of the applicant device is 2 AAA alkaline or rechargeable batteries.
The applicant device is not for life-supporting or life-sustaining, not for implant. The device or transducers are not sterile and the transducer does not need sterilization and the transducer is reusable but does not need re-sterilization since it is not sterile. The device is not for prescription. The device does contain drug or biological products.
{2}------------------------------------------------
The device is electrically operated and the Electrical Safety Test report of R-006-2006 and Electromagnetic Compatibility Test report of BJ690108-1 following IEC 60601-1-2 with was conducted as the environmental test for the home use. Please see the Attachment 2 Electrical Safety and EMC Test.
The device is software-driven and the software validation is provided in Chapter VIII Software Validation.
The Performance Test reports regarding with safety and effectiveness test of the safety mechanism preventing the excess current from leading to burning injury to user (Report No. MD300C-01-001) and Low-Voltage Alarm System (Report No. MD MD300C-01-002) are presented in Attachment 3 Performance Bench Test.
The Clinical Test reports following ISO 9919:2005, Medical electrical equipment - Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use are conducted in Beijing Friendship Hospital and Wulanchabu City Center Hospital in Attachment 4 Clinical Test Reports.
All applicable standards are listed in Chapter II Standards.
The device is not kit.
{3}------------------------------------------------
## 5. Effectiveness and Safety Considerations
#### Effectiveness:
The Clinical Test reports following ISO 9919:2005, Medical electrical equipment - Particular requirements for the safety and essential performance of pulse oximeter
The accuracy of MD300C pulse oximeter equipment is compliance to the requirement, and the product is safe during the used in the clinical environment. It is substantially equivalent to other pulse oximeter product with the same effectiveness and safety.
#### Safety Considerations:
The Performance Test reports regarding with safety and effectiveness test of the safety mechanism preventing the excess current from leading to burning injury to user (Report No. MD300C-01-001) and Low-Voltage Alarm System (Report No. MD MD300C-01-002) are presented in Appendix 3 Performance Bench Test.
Safety mechanism preventing the excess current from leading to burning injury to user is effective. And low voltage alarm system meet the function requirement of the design input.
The test results of biocompatibility of all the skin-contacting material are presented as Table IV-2 for the consideration of Biological Specifications. Please see Appendix 1 Biocompatibility Reports.
The Biological Evaluation Tests are in compliance with the standards of ISO 10993, "Biological Evaluation of Medical Devices". The compatibility of the skin-contact component material in the finished product meets the requirement of Biocompatibility.
The applicant device is compliance with IEC60601-1, Medical electrical equipment - Part 1: General requirements for safety and IEC60601-1-2, Medical electrical equipment - Part 1-2: General requirements for safety -Collateral standard: Electromagnetic compatibility -. Requirements and tests
The Software Validation is in compliance with FDA Guidance to Compliance with FDA Guidance for the Content of PreMarket Submissions for Software Contained in Medical Devices.
{4}------------------------------------------------
## 6. Substantially Equivalence Determination
### Comparison Analysis
The applicant device has same classification information, same indications and intended use, same design principle, similar product design and specifications, same performance effectiveness, performance safety as the predicate device. The only difference is working time Operating temperature, PR display range, Relative humidity, Atmosphere pressure. These differences are slight and do not effluence the effectiveness and safety of the device and don't raise new question of effectiveness and safety.
#### Conclusion:
The applicant device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the applicant device is determined as safe and effectiveness.
S
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle.
AUG 1 3 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Beijing Choice Electronic Technology Limited C/O Ms. Diana Hong General Manager Shanghai MIDLINK Business Consulting Company, Limited Suite 8D, Zhongxin Zhongshan Mansion No.19, Lane 999, Zhong Shan Nan Er Road Shanghai, 20020 CHINA
Re: K070371
Trade/Device Name: Fingertip Pulse Oximeter MD300C Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: July 23, 2007 Received: July 26, 2007
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 -Ms. Hong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number: _______________________________________________________________________________________________________________________________________________________________ Pending
Device Name: Fingertip Pulse Oximeter MD300C
## Indications for Use:
Fingertip Pulse Oximeter MD300C is a portable non-invasive, spot-check, oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care and etc). Not for continuously monitoring.
Prescription Use __ √ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
\$\sim \underset{}{\bigcirc} \sim\$
(Division Sign-Off) / actry B.C. Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page _1_ of __ 1
510(k) Number: K070371
l
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.