MD300C1 Fingertip Pulse Oximeter is a portable non-invasive, spot-check, oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care and etc). Not for continuously monitoring.
Device Story
Fingertip pulse oximeter; measures SpO2 and pulse rate via non-invasive optical sensor. Device uses dual-wavelength LEDs (660 nm red, 940 nm ultra-red) and photodetector to measure light absorption in pulsating arteriolar vascular bed. CPU calculates SpO2 based on ratio of light absorption during systole and diastole. Operated by clinicians or patients for spot-checks in home or hospital environments. Features low battery alarm and auto-power-off. Output displayed on integrated unit; assists healthcare providers in assessing patient oxygenation status. Benefits include portable, non-invasive monitoring without need for continuous observation.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 9919 (pulse oximeter performance), and ISO 10993 (biocompatibility). Software validation performed per FDA guidance.
Technological Characteristics
Fingertip form factor; dual-wavelength LED (660nm/940nm) sensor; CPU-based signal processing; powered by 2 AAA batteries. Reusable transducer; non-sterile. Complies with IEC 60601-1, IEC 60601-1-2, ISO 9919, and ISO 10993.
Indications for Use
Indicated for adult and pediatric patients for non-invasive, spot-check measurement of arterial hemoglobin oxygen saturation (SpO2) and pulse rate. Used in home and hospital settings (e.g., surgery, anesthesia, intensive care). Not for continuous monitoring.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
K093757
# 510(k) Summary
# MAR - 3 2010
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
The Assigned 510(k) Number is: Pending
## 1. Submitter Information
- . Manufacturer Name and Address: Beijing Choice Electronic Technology Co., Ltd. Bailangyuan Building B, Rm. 1127-1128, Fuxing Road, A36 Beijing, China 100039
Beijing Choice Electronic Technology Co., Ltd. No.9 Shuangyuan Road., Badachu Hi-tech Zone, Shijingshan District Beijing, China 100041
#### ● Contact Person:
Ms. Yajing Li North Building 3F, No. 9 Shuangyuan Road, Badachu Hi-tech Zone, Shijingshan District Beijing China 100041 Phone: +86-10-88790480 x 6046 Fax:+86-10-88798860 Email: liyajing@choicemmed.com & cc@choicemmed.com
- . Date prepared: October 22, 2009
#### 2. Applicant Device Information
- Trade/Proprietary Name: MD300C1 Fingertip Pulse Oximeter ●
- . Common Name: Pulse Oximeter
- Classification: 21CFR 870.2700 Oximeter Class: II
# 3. Legally Marketed Predicate Device
Fingertip Pulse Oximeter model MD300C K-number: K070371 Beijing Choice Electronic Technology Co., Ltd
{1}------------------------------------------------
# 4. Device Description
The applicant device of MD300C1 Fingertip Pulse Oximeter is a fingertip device, which can display % SpO2, pulse rate value.
The applicant device consists of detector and emitter LED, signal amplify unit, CPU, data display unit and power unit.
The Pulse oximetry works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660 nm, which is red light; the other is 940 nm, which is ultra red light. Skin, bone, tissue, and p venous vessels normally absorb a constant amount of light over time. The photodetector in fingersensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.
The applicant device has low battery voltage alarm function and automatically power of function. The power source of the applicant device is 2 AAA alkaline or rechargeable batteries.
The applicant device is not for life-supporting or life-sustaining, not for implant. The device or transducers are not sterile and the transducer is reusable and does not need sterilization or re-sterilization. The device is for prescription. The device does not contain drug or biological products.
#### 5. Intended use
MD300C1 Fingertip Pulse Oximeter is a portable non-invasive, spot-check, oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care and etc). Not for continuously monitoring.
### 6. Effectiveness and Safety Considerations
The applicant device is compliance with IEC60601-1, Medical electrical equipment - Part 1: General requirements for safety and IEC60601-1-2, Medical electrical equipment - Part 1-2: General requirements for safety -Collateral standard: Electromagnetic compatibility -.
{2}------------------------------------------------
Requirements and tests.
The applicant device is compliance with ISO 9919, Medical electrical equipment - Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use.
The Software Validation is in compliance with FDA Guidance to Compliance with FDA Guidance for the Content of PreMarket Submissions for Software Contained in Medical Devices.
The compatibility of the skin-contact component material in the finished product meets the requirement of Biocompatibility. The Biological Evaluation Tests are in compliance with the standards of ISO 10993, "Biological Evaluation of Medical Devices".
# 7. Substantially Equivalence Determination
The applicant device MD300C1 Fingertip Pulse Oximeter has same classification information, same intended use, same design principle, same product materials and performance effectiveness as the predicate device. These are no obvious differences to influence the effectiveness and safety of the device.
# 8. Conclusion
The applicant device MD300C1 Fingertip Pulse Oximeter is Substantially Equivalent (SE) to the predicate device which is US legally market device. Performance test results do not raise new questions of safety and effectiveness when compared to the legally marketed devices. Therefore, the applicant device is determined as safe and effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized symbol consisting of three overlapping shapes that resemble a human figure.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 . Silver Spring, MD 20993-0002
Ms. Yajing Li Beijing Choice Electronic Technology Company, Limited North Building 3F, No. 9 Shuangyuan Road Badachu Hi-tech Zone, Shijingshan District Beijing CHINA 100041
MAR - 3 2010
Re: K093757
Trade/Device Name: MD300C1 Fingertip Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: February 4, 2010 Received: February 4, 2010
Dear Ms. Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Li
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indication for Use
510(k) Number (if known): __ Pending
Device Name: __MD300C1 Fingertip Pulse Oximeter
Indications for Use:
MD300C1 Fingertip Pulse Oximeter is a portable non-invasive, spot-check, oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care and etc). Not for continuously monitoring.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schulten
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
_1_ of _______________________________________________________________________________________________________________________________________________________________________ Page
510(k) Number: K093757
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.