TELEVISIT

K062338 · PHD Medical · DQA · Oct 25, 2006 · Cardiovascular

Device Facts

Record IDK062338
Device NameTELEVISIT
ApplicantPHD Medical
Product CodeDQA · Cardiovascular
Decision DateOct 25, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2

Indications for Use

The TELEVISIT software allows healthcare professionals to conduct remote medical check-ups through interactive sessions with patients who have limited mobility or live in remote locations. The TELEVISIT software is intended for the purposes of collecting physiological data such as: Non-Invasive Blood Pressure; Pulse Oximetry, Temperature and Breath Sounds (Auscultation). The software provides physicians with a tool to schedule appointments with remote patients; to initiate and manage a medical check-up session, to provide data acquisition of medical device data, and to view session reports. The intended use of the TELEVISIT software is in hospital/clinical environments with established specialized care providing access to each patient in their home and/or local community health center. TELEVISIT software is a non-invasive medical device, which does not flag any abnormal results from the medical device inputs. The TELEVISIT software does not monitor, assess or diagnose a disease, a disorder, or an abnormal physical state. The intended use for the software is data acquisition and collection only, with no treatment function or danger to the end user. The data obtained from TELEVISIT software can be used only as an aid in the diagnosis and treatment of the patient.

Device Story

TELEVISIT is a remote physiological data collection system. It consists of a patient/physician terminal, a medical device gateway box, and management software. It interfaces with FDA-cleared accessories to record non-invasive blood pressure, pulse oximetry, temperature, and breath sounds (auscultation). Used in clinical/hospital settings to connect with patients in homes or community health centers. Physicians use the software to schedule appointments, initiate/manage sessions, acquire data, and view reports. The system does not perform automated analysis, flag abnormal results, or provide diagnostic/treatment functions; it serves solely as a data acquisition and transmission tool to assist clinicians in patient management.

Clinical Evidence

Bench testing only. Verification and validation confirmed that physiological data recorded by the patient system is accurately transmitted to and received by the physician's system. Risk analysis performed per ISO 14971:2000.

Technological Characteristics

System comprises a terminal, medical device gateway box, and management software. Interfaces with external FDA-cleared physiological sensors for NIBP, pulse oximetry, temperature, and auscultation. Connectivity involves data transmission between patient and physician systems. Risk management per ISO 14971:2000.

Indications for Use

Indicated for patients with limited mobility or those in remote locations requiring remote medical check-ups. Used by healthcare professionals to collect physiological data (NIBP, pulse oximetry, temperature, breath sounds) to aid in diagnosis and treatment. No contraindications specified.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K062338 ### 510(k) SUMMARY #### Manufacturer's Name, Contact Person, Address, Phone, Fax, Email, Date Prepared | Mfr Name: | PHD Medical | OCT 25 2006 | |----------------------------------------------------------|-------------------------------------------------------------------------|-------------| | Contact Name: | Patrick Haslehurst | | | Postal Address. | 500 Chemin Morgan, Suite 100<br>Baie D'Urfé<br>Québec H9X 3V1<br>Canada | | | Phone: | 514-694-0367 | | | Fax: | 514-694-4280 | | | Email: | patrick@phdmedical.com | | | Date Prepared: | August 2006 | | | Device Name, Common/Generic Name and Classification Name | | | | Proprietary Name: | TELEVISIT | | | Common/Usual Name: | Physiological Data Collection System | | | Classification Name: | Physiological Data Collection System | | | Predicate Device | | | | Device Name: | LifeShirt Real-Time (K043604) | | | Device Name: | March HealthCare Health Monitoring Kit (K060194) | | #### Intended Use of the Device The TELEVISIT software allows healthcare professionals to conduct remote medical check-ups through interactive sessions with patients who have limited mobility or live in remote locations. The TELEVISIT software is intended for the purposes of collecting physiological data such as: Non-Invasive Blood Pressure; Pulse Oximetry, Temperature and Breath Sounds (Auscultation). The software provides physicians with a tool to schedule appointments with remote patients; to initiate and manage a medical check-up session, to provide data acquisition of medical device data, and to view session reports. The intended use of the TELEVISIT software is in hospital/clinical environments with established specialized care providing access to each patient in their home and/or local community health center. TELEVISIT software is a non-invasive medical device, which does not flag any abnormal results from the medical device inputs. The TELEVISIT software does not monitor, assess or diagnose a disease, a disorder, or an abnormal physical state. The intended use for the software is data acquisition and collection only, with no treatment function or danger Page 1 of 2 {1}------------------------------------------------ to the end user. The data obtained from TELEVISIT software can be used only as an aid in the diagnosis and treatment of the patient. #### Technological Characteristics TELEVISIT is consist of the following 3 main components: Televisit Terminal (Patient/Physician), TELEVISIT Medical Device Gateway Box and the TELEVISIT Management Software. FDA Cleared accessories are provided to record physiological data such as: Breath Sounds (Auscultation); Non-Invasive Blood Pressure; Temperature / Thermometry; Pulse Oximetry. #### Performance Data Performance testing included verification and validation that data recorded by the TELEVISIT patient system is accurately transmitted and received by the main computer system and physician's system. Testing confirmed that TELEVISIT meets all of its intended functional requirements. A risk analysis per ISO 14971: 2000 has been performed to document potential hazards and methods of risk mitigation. The results demonstrate that TELEVISIT is safe and effective. #### Substantial Equivalence TELEVISIT is substantially equivalent to the predicate devices in terms of functional design, indications for use, principles of operation, software platform, and hardware requirements. Televisit software is substantially equivalent to the Lifeshirt Real-Time with Vivo Software (K043604) and March HealthCare Health Monitoring Kit (K060194). Safety and effectiveness of TELEVISIT is comparable to the predicate devices. TELEVISIT software collects, transmits, archives and produces reports that contain similar information to the predicate devices such as: Process steps include: Scheduling, Session Start, Data Collection, Data Transmission, Data Transfer, Data Record/Store, Data Analyze, Reporting and Archive. The TELEVISIT system is substantially equivalent in performance and intended use to the predicate devices and does not raise any safety and effectiveness issues. Page 2 of 2 {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 2 5 2006 PHD Medical c/o Ms. Nancy Ruth Associate Director, Regulatory Services CanReg Inc. 4 Innovation Drive Dundas. Ontario L9H 7P3 Canada Re: K062338 Trade Name: PHD Medical TELEVISIT Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II (two) Product Code: DQA Dated: August 9, 2006 Received: August 10, 2006 Dear Ms. Ruth: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Bhimman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use # 510(k) Number (if known): КОС2338 TELEVISIT Device Name: Indications For Use: The TELEVISIT software allows healthcare professionals to conduct remote medical check-ups through interactive sessions with patients who have limited mobility or live in remote locations. The TELEVISIT software is intended for the purposes of collecting physiological data such as: Non-Invasive Blood Pressure; Pulse Oximetry, Temperature and Breath Sounds (Auscultation). The software provides physicians with a tool to schedule appointments with remote patients; to initiate and manage a medical check-up session, to provide data acquisition of medical device data, and to view session reports. The intended use of the TELEVISIT software is in hospital/clinical environments with established specialized care providing access to each patient in their home and/or local community health center. TELEVISIT software is a non-invasive medical device, which i does not flag any abnormal results from the medical device inputs. The TELEVISIT software does not monitor, assess or diagnose a disease, a disorder, or an abnormal physical state. The intended use for the software is data acquisition and collection only, with no treatment function or danger to the end user. The data obtained from TELEVISIT software can be used only as an aid in the diagnosis and treatment of the patient. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Blymmerman Division Sian-Om Division of Cardlovascular Devices 510(k) Ni Page 1 of 17
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...