K050018 · Mammendorfer Institute For Physics & Medicine GmbH · DQA · Jul 1, 2005 · Cardiovascular
Device Facts
Record ID
K050018
Device Name
TESLAOXYSAT, MODEL EFOXYSATE
Applicant
Mammendorfer Institute For Physics & Medicine GmbH
Product Code
DQA · Cardiovascular
Decision Date
Jul 1, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The TeslaOxySat Pulse Oximeter is indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The pulse oximeter is indicated for use with adult, pediatric, and neonatal patients. The TeslaOxySat PulseOximeter is designed for use in an MRI-environment at a maximum magnetic field strength of 20mT.
Device Story
Tesla OxySat is a pulse oximeter designed for use in MRI environments. It utilizes a fiber optic finger probe and specialized shielding to function within magnetic and RF fields. The device monitors SpO2 and pulse rate in adult, pediatric, and neonatal patients. It must be positioned at least 1.5m from the magnet core with field strength not exceeding 20 mT. Operated by clinicians qualified in MRI and pulse oximetry procedures, the device provides continuous monitoring data to assist in the care of intensive care or sedated patients during MRI scans. The system mitigates risks associated with MRI-induced interference through its specific construction and shielding, ensuring patient safety and monitoring continuity during imaging.
Clinical Evidence
Laboratory testing using human subjects was conducted to validate functional and accuracy specifications of the pulse oximeter and sensors. Testing confirmed performance in both normal and MRI environments, demonstrating equivalency to the predicate device.
Technological Characteristics
Fiber optic finger probe; specialized shielding for MRI compatibility. Complies with EN 60601-1 (IEC 601-1) safety standards, EN 60601-1-2 EMC requirements, and ISO 10993-1 biocompatibility standards. Designed for operation in magnetic fields up to 20 mT.
Indications for Use
Indicated for continuous noninvasive monitoring of SpO2 and pulse rate in adult, pediatric, and neonatal patients. Designed for use in MRI environments up to 20mT.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Masimo SET® /Quartz 2500 Pulse Oximeter and Accessories (K993555)
Related Devices
K052119 — TESLAGUARD, MODEL EFT G04 · Mammendorfer Institute For Physics & Medicine GmbH · Jan 3, 2006
K071415 — DIGITAL PULSE OXIMETER, MODEL 7500FO · Nonin Medical, Inc. · Aug 8, 2007
K980362 — MRI COMPATIBLE MONITOR AND ACCESSORIES · Datex-Engstrom, Inc. · Jun 30, 1998
K153739 — MRI Patient Monitor TeslaDUO · Mipm Mammendorfer Institut Fur Physik Und Medizin · Jun 28, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
| 510(k) Number: | K050018 |
|----------------|---------|
| Date: | |
Page 1 of 2
# 510(k) Summary
JUL 0 1 2005
### Introduction
This summary is intended to comply with requirements of the SMDA and 21CFR8807.92. FDA may make this summary available to the public within 30 days following a finding of substantial equivalence.
# 510(k) Applicant
Mammendorfer Institut für Physik und Medizin GmbH Oskar-von-Miller-Strasse 6 82291 Mammendorf, Germany
# 510(k) Correspondent
Robert N. Clark, President and Senior Consultant Medical Device Regulatory Advisors 13605 West 7th Ave., Golden, CO USA Tel: 303-234-9412 / Fax: 303-234-9413
# Date Prepared
December 30, 2004
# Trade Name of Device
Tesla OxySat®
# Common Name of Device
Pulse Oximeter and Sensor
# Classification Name
Oximeter
# 510(k) Classification
21CFR§870.2700 / Class II Product Code: DQA
# Device Description and Intended Use
The Tesla OxySat® design allows examination of intensive care or sedated patients while in an MRI-scanner. The Tesla OxySat® construction, fiber optic finger probe, and additional shielding make it possible to use the device within the magnetic and RF fields of the MRI examination room. During use, the unit must be positioned in a way that the maximum field strength is not higher than 20 mT, and the distance to the magnet core is at least 1.5m.
Image /page/0/Picture/21 description: The image shows the characters 'E 000002' in a bold, sans-serif font. The 'E' is larger than the numbers that follow it. The numbers are all zeros except for the last digit, which is a '2'.
{1}------------------------------------------------
| 510(k) Number: | |
|----------------|--|
| Date: | |
Page 2 of 2
The TeslaOxySat Pulse Oximeter is indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The pulse oximeter is indicated for use with adult, pediatric, and neonatal patients.
### Predicate Devices
Masimo SET® /Quartz 2500 Pulse Oximeter and Accessories (K993555)
# Safety & Effectiveness
#### Safety Testing
The Tesla OxySat complies with the Safety Testing requirements of EN 60601-1 (IEC 601-1).
#### EMC Compliance
The Tesla OxySat complies with the EMC requirements of standard EN 60601-1-2.
#### Biocompatibility
The patient contact portions of the Tesla OxySat fiber optic finger probe comply with the biocompatibility requirements of standard ISO 10993-1.
#### MRI Compatibility Tests
Testing was completed to determine the influence of the Tesla OxySat on the MRI system, and the influence of the MRI system on the Tesla OxySat®.
#### Functional Testing
Function and accuracy of the Tesla OxySat was tested in both normal environment (Non MRI) and in MRI environment.
#### Equivalency Testing
Laboratory testing using human subjects was conducted to validate the functional and accuracy specifications of the pulse oximeter and sensors, and to demonstrate equivalency to the predicate device.
### Risk Management
This device has been designed to either completely eliminate or mitigate known health hazards associated with the use of the device. Health hazard risk reduction has been accomplished by rigorous application of a risk management program.
The user must be qualified in MRI and pulse oximeter procedures, and must be familiar with all labeling and instructions for use associated with the device.
Mammendorfer Institut für Physik und Medizin GmbH believes that the Tesla OxySat® is safe and effective when used as instructed by knowledgeable and trained personnel, and is substantially equivalent to the legally marketed predicate devices.
# 000003
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
---
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mammendorfer Institute for Physics and Medicine GMBH C/o Mr. Robert N. Clark Medical Device Regulatory Advisors 13605 West 7th Avenue Golden, Colorado 80401
JUL 0 1 2005
Re: K050018
Trade/Device Name: Tesla OxySat® Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: May 6, 2005 Received: May 10, 2005
Dear Mr. Clark:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Image /page/2/Picture/12 description: The image is a black and white emblem for the U.S. Department of Health & Human Services. The emblem features a stylized caduceus, a symbol often associated with healthcare, consisting of three parallel lines that curve and flow together. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus, framing the symbol.
{3}------------------------------------------------
Page 2- Mr. Robert N. Clark
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
.
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number:
Device Name:
TeslaOxySat®
Indications for Use:
The TeslaOxySat Pulse Oximeter is indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The pulse oximeter is indicated for use with adult, pediatric, and neonatal patients. The TeslaOxySat PulseOximeter is designed for use in an MRI-environment at a maximum magnetic field strength of 20mT.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aun Sylionn
Division Sign-Off) Division of Anesthestotogy, General Hospital, Infection Control, Dental Devices
510(k) Number:
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter-Use
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.