MRI COMPATIBLE MONITOR AND ACCESSORIES

K980362 · Datex-Engstrom, Inc. · DPS · Jun 30, 1998 · Cardiovascular

Device Facts

Record IDK980362
Device NameMRI COMPATIBLE MONITOR AND ACCESSORIES
ApplicantDatex-Engstrom, Inc.
Product CodeDPS · Cardiovascular
Decision DateJun 30, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2340
Device ClassClass 2

Intended Use

The Datex-Engstrom MRI Patient Monitor and accessories is indicated for monitoring of ECG waveform for Heart Rate and Asystole detection, SpO22, Non-invasive blood pressure (NIBP), airway gases, anesthetic agents, airway flow, airway pressure and airway volume on patients during MR imaging. It is indicated for use inside the MRI room mounted outside the 20 Gauss line and with magnets of field strength 1.5 Tesla or less. SpO2monitoring is indicated only with the use of Datex-Engstrom sensors specifically designed for use in the MRI environment. ECG monitoring is indicated only with use of the MRI ECG lead set.

Device Story

Device monitors physiological parameters during MRI procedures; inputs include ECG waveforms, SpO2, NIBP, airway gases, anesthetic agents, and airway mechanics (flow, pressure, volume). Operates inside MRI room; must be positioned outside 20 Gauss line; compatible with 1.5T magnets. Designed for use by clinicians; provides real-time data to assist in patient management during imaging. Benefits include continuous patient safety monitoring in high-magnetic field environments.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

MRI-compatible patient monitor; operates in 1.5T magnetic fields; requires placement outside 20 Gauss line. Utilizes proprietary MRI-specific sensors and ECG lead sets. Class II device.

Indications for Use

Indicated for monitoring patients undergoing MR imaging (1.5 Tesla or less). Parameters: ECG (HR, asystole), SpO2, NIBP, airway gases, anesthetic agents, airway flow, pressure, and volume. Requires specific MRI-compatible sensors and lead sets.

Regulatory Classification

Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HI" on the left side and "IN SERVICES - USA" on the right side. In the center of the logo is a stylized image of a bird. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 3 0 1998 Mr. Joel C. Kent Datex Engstrom, Inc. 3 Highwood Drive Tewksbury, MA 01876 K980362 Re : MRI Patient Monitor and Accessories Requlatory Class: II (two) Product Code: 74 DPS Dated: June 11, 1998 Received: June 12, 1998 Dear Mr. Kent: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through-periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {1}------------------------------------------------ Page 2 - Mr. Joel C. Kent This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.qov/cdrh/dsma/dsmamain.html". Sincerely yours Thomas J. Callahan Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Page 1 of 1 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: MRI Patient Monitor and Accessories Indications For Use: The Datex-Engstrom MRI Patient Monitor and accessories is indicated for monitoring of ECG waveform for Heart Rate and Asystole detection, SpO22, Non-invasive blood pressure (NIBP), airway gases, anesthetic agents, airway flow, airway pressure and airway volume on patients during MR imaging. It is indicated for use inside the MRI room mounted outside the 20 Gauss line and with magnets of field strength 1.5 Tesla or less. SpO2monitoring is indicated only with the use of Datex-Engstrom sensors specifically designed for use in the MRI environment. ECG monitoring is indicated only with use of the MRI ECG lead set. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use (Per 21 CFR 801.109) | <div style="display:inline-block;width:30px;text-align:center;">✓</div> | |---------------------------------------|-------------------------------------------------------------------------| | | OR | | Over-The-Counter Use | | (Optional Format 1-2-96) | (Division Sign-Off) | | |-------------------------------------------------------------------|---------| | Division of Cardiovascular, Respiratory, and Neurological Devices | | | 510(k) Number | R980362 | Mark Kunn
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%