K023195 · Schiller Medical S.A. · MWI · Jun 3, 2003 · Cardiovascular
Device Facts
Record ID
K023195
Device Name
MAGLIFE C/MAGLIFE C PLUS
Applicant
Schiller Medical S.A.
Product Code
MWI · Cardiovascular
Decision Date
Jun 3, 2003
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
MAGLIFE C is intended for monitoring electrocardiogram (ECG), pulse oximetry (SpO2), pulse rate, partial carbon dioxide pressure at the end of expiration (EtCO2), nitrous oxide concentration (%N20), partial pressure of inspired carbon dioxide (min inspired CO2), respiratory rate (RR), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), concentration of gas inspired and expired (isoflurane, enflurane, desflurane, sevoflurane), and fraction of inspired oxygen (FiO2) and temperature tracking in the immediate vicinity of Magnetic Resonance Imagers for the surveillance of patients undergoing MRI examinations.
Device Story
Multi-parameter patient monitor for MRI environment; console placed outside magnet bore; uses non-metallic sensors and fiber optic signal transmission to prevent burns and minimize MRI interference; Faraday cage enclosure protects electronics. Inputs: ECG, SpO2, pulse rate, EtCO2, N2O, inspired CO2, respiratory rate, NIBP, IBP, anesthetic gases, FiO2, temperature. Outputs: real-time physiological data displayed on console for clinician surveillance during MRI. Benefits: enables continuous patient monitoring within MRI suite without compromising image quality or patient safety.
Clinical Evidence
Bench testing only. Validation included unit and system-level testing for compliance with IEC 601-1 and IEC 601-1-2. Testing in 1.5 Tesla MRI unit verified performance stability, lack of ferrous material movement, and no degradation of MRI image homogeneity at distances as close as 60 cm from the magnet bore.
Technological Characteristics
Multi-parameter monitor; line-powered console with Faraday cage shielding; non-metallic sensors; fiber optic signal transmission. Dimensions/form factor: console-based. Connectivity: standalone. Sterilization: N/A. Software: embedded firmware for physiological signal processing.
Indications for Use
Indicated for monitoring ECG, SpO2, pulse rate, EtCO2, N2O, inspired CO2, respiratory rate, NIBP, IBP, anesthetic gas concentrations, FiO2, and temperature in patients undergoing MRI examinations.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{1}------------------------------------------------
K023195
page 1 of 3
# 510(k) Summary for MAGLIFE C/MAGLIFE C Plus
## 1. Date this summary was prepared: February 5, 2003
#### 2. Submitter's Name and Address
SCHILLER MEDICAL 4, rue Louis Pasteur ZAE Sud BP 90050 67162 WISSEMBOURG CEDEX FRANCE
#### 3. Contact person
M. Court GOEHRY Regulatory Affairs & Vigilance Manager SCHILLER MEDICAL 4. rue Louis Pasteur Z.A.E Sud BP 90050 67162 WISSEMBOURG CEDEX France
Tel: + 33 3 88 63 36 00 Fax: + 33 3 88 94 12 82 E-mail: court.goehry@schiller.fr
## 4. Device Name
Maglife C/Maglife C Plus Trade/Proprietary Name:
MR Safe Patient Monitor Common Name:
Clasification Names: Patient physiological monitor
## 5. . Predicate Devices
The legally marketed devices to which equivalence is being claimed are:
- MAGLIFE, marketed by O.D.A.M. (K950264) .
- MRI-Compatible Model 9500 manufactured by Magnetic Resonance Corp. (K954120)
- Omni Trak 3100 MRI manufactured by In Vivo Research (K864889) ●
- . Model 3100 LUXTRON CK
{2}------------------------------------------------
K023195
page 2 of 3
## 6. Device Description
The Maglife C is a multi-parameter patient monitor which is indicated for monitoring electrocardiogram (ECG), pulse oximetry (SpO2), pulse rate, partial carbon dioxide pressure at the end of expiration (EtCO-), nitrous oxide concentration (%N2O), partial pressure of inspired carbon dioxide (min inspired CO2), respiratory rate (RR), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), concentration of gas inspired and expired (isoflurane, halothane, enflurane, desflurane, sevoflurane), and fraction of inspired oxygen (FiO2) and temperature tracking.
The monitor consists of line-powered console (with optional back-up battery) which is placed outside the magnet boren, and appropriate electrodes, transducers, cables and tubes to allow the patient to be monitored during the examination. The console enclosure is a Faraday cage for the protection of sensitive electronic circuits.
## 7. Intended Use
MAGLIFE C is intended for monitoring electrocardiogram (ECG), pulse oximetry (SpO2), pulse rate, partial carbon dioxide pressure at the end of expiration (EtCO2), nitrous oxide concentration (%N20), partial pressure of inspired carbon dioxide (min inspired CO2), respiratory rate (RR), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), concentration of gas inspired and expired (isoflurane, enflurane, desflurane, sevoflurane), and fraction of inspired oxygen (FiO2) and temperature tracking in the immediate vicinity of Magnetic Resonance Imagers for the surveillance of patients undergoing MRI examinations.
## 8. Comparison of Technological Characteristics
All four MRI compatible patient monitors employ established measurement methods combined with non-metallic sensors and fiber optic signal transmission to minimize the risk of burns and special shielding and filtering to minimize interference to and from the MRI unit.
## 9. Nonclinical Tests Used in Determination of Substancial Equivalence
The design of the Madife C has been thoroughly validated at the unit and system level. Non-clinical test were conducted to demonstrate compliance with the following standards:
- IEC 601-1 ●
- IEC 601-1-2 ●
Test were conducted in an MRI unit to determine the maximum field strength that Maglife C can withstand without degradation of performance and to verify that the amount of ferrous material in the Maglife C is sufficiently small so that movement of the unit due to magnetic attraction is not possible.
{3}------------------------------------------------
Tests were conducted to assess the effect of the Maglife C on the Homogeneity of the magnetic field inside the magnet bore. These tests, using a 1.5 Tesla MR unit, showed that the Maglife could be placed as close as 60 cm from the front of the opening from the magnet without causing any visible difference in the images.
## 10. Conclusions From Nonclinical Testing
The testing of the Maglife C demonstrates that the performance is substantially equivalent to predicate devices cited above.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, formed by a series of curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 3 2003
Schiller Medical SA c/o Mr. Court Goehry Regulatory Affairs Specialist 4, rue Louis Pasteur Z.A.E Sud BP 90050 67162 Wissembourg Cedex France
Re: K023195
Trade Name: Maglife C/Maglife C Plus Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm). Regulatory Class: Class II (two) Product Code: MWI Dated: May 2, 2003 Received: May 8, 2003
Dear Mr. Goehry:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
•
{5}------------------------------------------------
Page 2 - Mr. Court Goehry
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D. Zuckerman, M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): __________
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications For Use:
The Maglife C is a multi-parameter patient monitor which is indicated for monitoring electrocardiogram (ECG), pulse oximetry (SpO2), pulse rate, partial carbon dioxide pressure at the end of expiration (EtCO2), nitrous oxide concentration (%N₂O), partial pressure of inspired carbon dioxide (min inspired CO2), respiratory rate (RR), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), concentration of gas inspired (isoflurane, halothane, enflurane, desflurane, sevoflurane), and fraction of inspired oxygen (FiO2) and temperature tracking in the immediate vicinity of a Magnetic Resonance Imagers for the surveillance of patients undergoing MRI examinations.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|---------|
| (Division Sign-Off) | |
| Division of Cardiovascular Devices | |
| 510(k) Number | K023195 |
| Prescription Use | X | OR | Over-The-Counter Use |
|----------------------|---|----|----------------------|
| (Per 21 CFR 801.109) | | | |
I / I
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.