DOLPHIN MEDICAL 2150 HANDHELD PULSE OXIMETER

K030629 · Dolphin Medical, Inc. · DQA · Sep 4, 2003 · Cardiovascular

Device Facts

Record IDK030629
Device NameDOLPHIN MEDICAL 2150 HANDHELD PULSE OXIMETER
ApplicantDolphin Medical, Inc.
Product CodeDQA · Cardiovascular
Decision DateSep 4, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2
AttributesPediatric

Indications for Use

The Dolphin Medical 2150 Handheld Pulse Oximeter is indicated for use for spot checking and/or continuous noninvasive monitoring of fractional oxygen saturation of arterial hemoglobin, pulse rate, and patient temperature. The 2150 is for use with adult pediatric and neonatal patients in hospitals and hospital type facilities.

Device Story

Handheld pulse oximeter; utilizes Dolphin ONE OEM 701 module; inputs include light absorption signals from two-LED sensors and temperature data from YSI 400 series thermistor probes. Digital signal processing (128 MHz RISC processor) calculates SpO2, pulse rate, and temperature. Output displayed on LCD; includes infrared printer interface. Used in acute-care settings (OR, Recovery, ICU) by physicians, nurses, or respiratory therapists. Provides real-time patient data for clinical monitoring; supports decision-making regarding oxygenation and thermal status. Benefits include noninvasive, continuous assessment of patient vitals across diverse perfusion and motion conditions.

Clinical Evidence

Clinical validation performed via breathe-down protocol at VA Hospital of Wisconsin. Sample included conscious volunteers with SaO2 ranging 70-100%. Primary endpoint: accuracy (ARMS) compared to functional SaO2. Results: <2.0% ARMS for adults, pediatrics, and infants; <3.5% ARMS for neonates. Testing confirmed performance during motion and no-motion conditions, and across varying perfusion levels.

Technological Characteristics

Handheld form factor; 128 MHz RISC processor; OEM 701 oximetry engine. Interfaces with YSI 400 series thermistor probes. Power: internal rechargeable battery or medical-grade AC supply. Connectivity: infrared printer interface. Complies with EN 60601-1, EN 60601-1-2, ASTM 1415:1992, and E1112-00. Biocompatibility testing performed for prolonged skin contact.

Indications for Use

Indicated for spot checking and continuous noninvasive monitoring of SpO2, pulse rate, and patient temperature in adult, pediatric, and neonatal patients in hospital settings.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the text "K030629" in a handwritten style. The text is oriented diagonally, starting from the bottom left and rising to the upper right. The characters are bold and slightly distorted, giving them a casual, handwritten appearance. # 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS # Dolphin Medical 2150 Handheld Pulse Oximeter ## Submitting by Dolphin Medical Inc. 12525 Chadron Avenue Hawthorne, CA 90250 ### Submission Correspondent Bill Curnan Regulatory Specialist 9433 S. Morning Glory Lane Highlands Ranch, CO 80130 Phone: 720-939-6482 Fax: 786-551-8221 ## Common, Classification & Proprietary Names | Common Name: | oximeter | |----------------------|----------------------------------------------| | Classification Name: | oximeter | | Proprietary Name: | Dolphin Medical 2150 Handheld Pulse Oximeter | # Predicate Devices The Dolphin Medical 2150 Handheld Pulse Oximeter is substantially equivalent to the following devices: | K024235 | Dolphin Stand Alone Pulse Oximeter and Accessories | |---------|----------------------------------------------------| | K020075 | Dolphin Stand-Alone Pulse Oximeter and Accessories | | K953817 | Aristo Medical Model 2101 Patient Monitor | # Device Description The Dolphin Medical 2150 Pulse Oximeter consists of the Dolphin ONE™ OEM 701 Module technology, and works with existing Dolphin ONE extension cables, and oximetry sensors to non-invasively calculate the functional saturation of arterial hemoglobin (SpO2) and pulse rate. It also includes a thermistor port which accepts off the shelf approved YSI 400 series patient temperature sensors for the reading of patient temperature. It features an easy-to-read display that presents patient data and status information: an LCD (liquid crystal display) display that shows the SpO2, pulse rate values, patient temperature, and other messages as appropriate. {1}------------------------------------------------ A Dolphin ONE Extension Cable connects the sensor to the 2150. The cable is available in two configurations, three foot or eight in length. The oximeter can be operated off either an internal rechargeable battery or with use of a medical grade AC power supply which is fumished with the unit. The Dolphin Medical Model 2150 Pulse Oximeter will be used for continuous noninvasive monitoring and spot checking of SpO2, pulse rate, and patient temperature for adult, pediatric, infant and neonatal patients in hospital-type facilities. Typical acute-care uses are the Operating Room, Recovery Room and the Adult and Neonatal Intensive Care Units. The Dolphin Medical Model 2150 Pulse Oximeter and accessories will be used for adult, pediatric and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused. The user responsible for the operation of the Dolphin Medical Model 2150 will normally be a licensed clinical professional such as a physician, nurse, or respiratory therapist. ## Intended Use The Dolphin Medical 2150 Handheld Pulse Oximeter is indicated for use for spot checking and/or continuous noninvasive monitoring of fractional oxygen saturation of arterial hemoglobin, pulse rate, and patient temperature. The 2150 is for use with adult pediatric and neonatal patients in hospitals and hospital type facilities. ### Technological Characteristics Comparison The Dolphin 2150Handheld Pulse is a redesign of the Dolphin 2100 Stand Alone Pulse Oximeter (K002036) into a handheld configuration with the addition of a temperature sensor port. The Dolphin Model 2150 Pulse Oximeter incorporates the Dolphin OEM 701 Oximeter Engine in a handheld Oximeter with patient temperature monitoring capability and an infra red printer interface. Dolphin Medicals OEM 701 Module is a complete pulse oximeter engine, based on a patented Digital Signal Processing technology with a high speed 128 MHz RISC processor. This circuit board interfaces directly with the Dolphin ONE™ family of sensors and the host platform to calculate the SpO2 and pulse rate. It provides the required sensor drive currents, accepts the sensor data, runs the algorithms to calculate SPO2. Pulse Rate, and patient temperature and provides a digital output to the host unit for display. Additionally, the board contains all necessary support for direct connection to off the shelf YSI 400 series thermistor patient temperature probes. The OEM 701 is a redesign of the OEM 601 which is the oximetry engine for the Dolphin Model 2100(k002036). The Dolphin ONE Sensors measure light absorption of blood from two light emitting diodes. Oxygen saturated blood absorbs light differently as compared to unsaturated blood. The amount of light absorbed by the blood is used to calculate the ratio of oxygenated hemoglobin to total hemoglobin in arterial blood. {2}------------------------------------------------ # Performance Testing #### Biocompatibility t Biocompatibility tests, appropriate for skin-contacting devices for prolonged exposure, have been performed on all Dolphin ONE sensors #### Electrical Safety 트 The Dolphin Medical 2150 Handheld Pulse Oximeter was designed to comply with the following standards: - 1. CSA C22.2 No. 601.1 - 2. EN 60601-1, and Amendments 1 and 2 - 3. EN 60601-1-2, 2002 - 4. EN 475 1995 - 5. EN 865: 1997 - 6. FDA Guidance Document for Pulse Oximeters: 977/1992 - 7. ASTM 1415:1992. - 8. UL2601-1: Second Edition, 1997 - 9. IEC 6068-2-6 - 10. IEC 6068-2-27 - 11. IEC 6068-2-64 - 12. ISTA Procedure 2A - 13. E1112-00 # Clinical Testing The 2150 was validated using Dolphin ONE Sensors in breathe-down protocols at the VA Hospital of Wisconsin - Milwaukee, (Dr. Phillip Clifford, MD.). Scientific accuracy was demonstrated by statistically comparing Dolphin 2150 SpO2 values to functional SaO2 values. Volunteers participated in the breathe-down protocol while fully conscious at SaO2 values ranging from 70-100%. Data was analyzed to determine the ARMS for each probe. Clinical Validation for the Dolphin ONE Reusable, Adult disposable, Neonatal disposable resulted in an accuracy determination of less then 2.0% Arws in the range of 70-100% Sa02 for adults, pediatrics, and infants and less than 3.5% Arms in the range of 70-100 for Neonates. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a stylized eagle-like emblem. The emblem consists of three abstract human profiles facing right, with flowing lines beneath them, possibly representing water or movement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP - 4 2003 Mr. Bill Curnan Regulatory Specialist Dolphin Medical Incorporated 9433 S. Morning Glory Lane Highlands Ranch, Colorado 80130 Re: K030629 Trade/Device Name: Dolphin ONE 2150 Pulse Oximeter Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: June 17, 2003 Received: June 18, 2003 Dear Mr. Curnan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ #### Page 2 - Mr. Curnan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### 4. STATEMENT OF INDICATIONS FOR USE (FDA FORM) 510(k): KO30629 Device: Dolphin ONE 2150 Pulse Oximeter # Indications for Use: The Dolphin Medical 2150 Handheld Pulse Oximeter is indicated for use for spot checking and/or continuous noninvasive monitoring of fractional oxygen saturation of arterial hemoglobin, pulse rate, and patient temperature. The 2150 is for use with adult pediatric and neonatal patients in hospitals and hospital type facilities. (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number:_ Prescription device \/
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...