The Datex-Ohmeda S/5™ Nellcor Compatible Saturation Module, E-NSAT is intended for use with the Dates-Ohmeda modular multiparameter patient monitoring arterial oxygen saturation of hospitalized patients. The Datex-Ohmeda S/5™ Nellcor Compatible Saturation Module, E-NSAT, and accessories are indicated for monitoring arterial oxygen saturation of hospitalized patients. The device is indicated for use by qualified medical personnel only.
Device Story
E-NSAT is a single-width plug-in parameter module for Datex-Ohmeda S/5 modular monitoring systems; used for continuous non-invasive measurement of pulse rate and oxygen saturation in hospitalized patients. Device utilizes Nellcor pulse oximetry technology (MP100 electronic measurement board) to process real-time pulse oximetry signals. Outputs include calculated SpO2 values and pulse rate, displayed numerically or alongside pulse oximetry waveforms on the host monitor. System provides visual and auditory alarms for high/low pulse rate and saturation limits. Operated by qualified medical personnel in clinical settings. Healthcare providers use output for patient monitoring and clinical decision-making regarding oxygenation status. Benefits include continuous, non-invasive assessment of patient oxygen saturation.
Clinical Evidence
Bench testing only. Device verified against specifications and safety standards including IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, ISO 9919, and EN 865. No clinical data presented.
Technological Characteristics
Single-width plug-in module for S/5 modular monitors. Utilizes Nellcor MP100 pulse oximetry electronic measurement board. Connectivity via Datex-Ohmeda interface board to host monitor. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, ISO 9919, and EN 865. Software-based signal processing for SpO2 and pulse rate calculation.
Indications for Use
Indicated for monitoring arterial oxygen saturation of hospitalized patients. For use by qualified medical personnel only.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
K020479 — DATEX-OHMEDA S/5 NELLCOR COMPATIBLE SATURATION MODULE, M-NSAT AND ACCESSORIES · Datex-Ohmeda · Mar 26, 2002
K081870 — MASIMO COMPATIBLE SATURATION MODULE, E-MASIMO · GE Healthcare Finland Oy · Jul 31, 2008
K011670 — DATEX-OHMEDA OXYGEN SATURATION MODULE, M-OSAT AND ACCESSORIES (WITH TRUTRAK+MOTION CORRECTION PERFORMANCE · Datex-Ohmeda · Jun 26, 2001
K082888 — INFINITY NELLCOR OXIMAX SMARTPOD, MODEL MS23997 · Draeger Medical Systems, Inc. · Nov 26, 2008
K142865 — Nellcor Bedside SpO2 Patient Monitoring System · Covidien, LLC · Mar 26, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
## OCT 1 3 2005
KOS2753
### Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda S/57M Nellcor Compatible Saturation Module, E-NSAT and accessories
### GENERAL COMPANY INFORMATION as required by 807.92(a){1}
### COMPANY NAME/ADDRESS/PHONE/FAX:
GE Healthcare 86 Pilgrim Road Needham. MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344
### NAME OF CONTACT:
#### Mr. Joel Kent
#### DATE:
September 28, 2005
### DIATICE NAME as required by 807.92(a)(21
### TRADE NAME
Datex-Ohmeda S/5™ Nellcor Compauble Saturation Module, E-NSAT and accessories
#### COMMON NAME:
Pulse Oximeter
### CLASSIFICATION NAME:
The following Class II classification appears applicable:
| Product Code | Classification Name | CFR Section |
|--------------|---------------------|-------------|
| DQA | Oximeter | 870.2700 |
| DPZ | Ear Oximeter | 870.2710 |
### NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)
The Datex-Ohmeda S 5781 Nellcor Compatible Saturation Module, F-NSAT is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda M NSAT Module (K020479).
{1}------------------------------------------------
### DEVICE DESCRIPTION as required by 807.92(a)(4)
The Datex-Ohmeda S'57M E-NSAT module is a single-width plug-in parameter module for a Datex-Ohmeda S 5 modular monitoring system. E-NSAT is used for monitoring arterial oxygen saturation of hospitalized patients.
The Datex-Ohmeda Nelleor Compatible Saturation Module, E-NSAT can be used with the following Datex-Ohmeda modular monitors:
- S/5TM Anesthesia Monitor (AM) with main software S-STD94(A) or S ARK94(A) or . newer
- S/5TM Compact Anesthesia Monitor (CAM) with main S-STD94(A) or S ARK 94(A) or . newer
- S/5TM Critical Care Monitor (CCM) with main software S-ICU97(A) or newer. .
- S/5TM Compact Critical Care Monitor (CCCM) with main software S-IC U97(A) or . newer.
The E-NSAT module provides continuous non-invasive measurement of the pulse rate and oxygen saturation. Visual and auditory alarms are given for high/low pulse rate and high/low saturation values. The user can adjust the alarm limits.
The E-NSAT module is designed using the Nelleor pulse oximetry technology. The real-time pulse oximetry signal is displayed without any scale. The calculated Sp()2 value and pulse rate can be shown in a separate numberfield, or beside the pulse oximetry signal. The parameters can be trended and the trends can be printed with the Recorder module in the momitor or with a separate laser printer.
The accessories are all new for E-NSAT. The accessories are manufactured and tested by Nellcor Puritan Bennet Inc. and distributed by GE Healthcare.
### INTENDED USE as required by 807.92(a)(5)
### Intended Use:
The Datex-Ohmeda S/5™ Nelleor Compatible Saturation Module. E-NSAT is intended for use with the Dates-Ohmeda modular multiparameter patient monitoring arterial oxygen saturation of hospitalized patients.
### Indication for use:
The Datex-Ohmeda S 50% Nellcor Compatible Saturation Module, E-NSAT, and accessories are indicated for monitoring arterial oxygen saturation of hospitalized patients. The device is indicated for use by qualified medical personnel only.
### SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a){6}
The Datex-Ohmeda S/5™ Nelleor Compatible Saturation Module, F-NSAT is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda M NSAT Module (K020479).
The E-NSAT module has the following similarities compared to the predicate M-NSAT (K()20479):
- Identical intended use and indications for use .
- . Use the same operating principle
- The same fundamental scientific technology .
- Can be used with the same pulse oximetry -specific monitor software. (Have the same t user interface at the monitor)
{2}------------------------------------------------
- The Customer and parameter specifications are the same .
- Have the same safety and effectiveness .
- Are manufactured using the same processes .
The main differences between the new E-NSAT and the predicate M-NSAT (K020479) is primarily due to fact that the new E-NSAT module has the following changes
- New color, shape, and size and thus differing mechanies .
- The front panel and labeling have changed .
- The Nellcor OEM Pulse Oximetry electronic measurement board with software, provided . by Nelleor Puritan Bennet Inc has changed from MP404 to MP100. The MP100 is an enhanced version of the MP404, and has been previously cleared with the Sandman Express (K040113).
- The Datex-Ohmeda Interface board connecting the measurement board to the monitor . have been revised to accommodate the changes to the electronic measurement board
- The E-NSAT connector coding differs to allow use of the accessories specified by . Nellcor for the MP100 board
- . The accessories (provided by Nellcor) have changed
Based on the analysis and other documentation included in this 510(k) notification and attachments it is evident that the main features and indications for use of the Datex-Ohmeda S/5™ Nellcor Compatible Saturation Module, E-NSAT are substantially equivalent to the predicate Datex-Ohmeda M-NSAT Module (K020479).
### SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)
The Datex-Ohmeda S:500 Nelleor Compatible Saturation Module, F:-NSAT and accessories has been assessed against the standards below. The device has been thoroughly tested through validation and verification of specifications.
- COUNCH. DIRECTIVE 93/42/EEC of 14 June 1993 concerning medical devices .
- FDA:DCRND Reviewer Guidance for Premarket Notification Submissions, November 1993 .
- 1EC 60601-1:1988 + Amdt. 2:1991 + Andt. 2:1995 (Part 1: General requirements for safety) .
- EN 60601-1:1990+ A1:1993 + A13:1996 + A2:1995 (identical to IE:C60601-1:1988 + Amdt. . 1:1991 + Anidt. 2:1995)
- CAN/CSA C22.2 No. 601.1-M90 + S1:1994 (Canadian deviations to IFC 60601-1:1988 + . Andt. 1:1991) + S2:1998 (=IFC Andt 2:1995)
- UI, 2601-1, October 24, 1997 (U.S. deviations to IEC 60601-1:1988 + Amdt. . 2:1995)
- IEC 60601-1-2:2001 (Electromagnetic compatibility Requirements and tests) .
- 1EC 60601-1-4:2000 (Programmable medical systems) .
- ISO 9919 (1994) Pulse oximeters for medical use safety requirements .
- EN 865 (1997) Pulse oximeters Particular requirements .
- AAMI ES1-1993 (Safe current limits for electromedical apparatus) .
- FDA-ODE Guidance for Content of Premarket Submission for Software Contained in Medical . Devices. (May 11, 2005)
### CONCLUSION.
The summary above shows that the Datex-Ohmeda S:57M Nelleor Compatible Saturation Module, I :- NSAT is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda M NSAT Module (K020479).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three curved lines representing its body and wings. The graphic is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA". The text is arranged around the circumference of the circle, framing the graphic element.
001 1 3 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Joel Kent Manager, Quality and Regulatory Affairs GE Healthcare 86 Pilgrim Road Needham, Massachusetts 02492
Re: K052755
Trade/Device Name: Datex-Ohmeda S/5™ Nellcor Compatible Saturation Module, E-NSAT and accessories Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA, DPZ Dated: September 29, 2005 Received: October 3, 2005
Dear Mr. Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Mr. Kent
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours,
Chris Liu, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Datex-Ohmeda S/5™ Nellcor Compatible Saturation Module, E-NSAT and accessories.
Indications for use:
The Datex-Ohmeda S/5™ Nellcor Compatible Saturation Module, E-NSAT, and accessories are indicated for monitoring arterial oxygen saturation of hospitalized patients. The device is indicated for use by qualified medical personnel only.
Over-The-Counter Use _ Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
Bivision Sign-Off) sion Sign-Om)
ion of Anesthesiology, General Hospital, Division of Anesthoolarda Devices
510(k) Number: K052755
Page __ of __
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.