K082888 · Draeger Medical Systems, Inc. · DQA · Nov 26, 2008 · Cardiovascular
Device Facts
Record ID
K082888
Device Name
INFINITY NELLCOR OXIMAX SMARTPOD, MODEL MS23997
Applicant
Draeger Medical Systems, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Nov 26, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Infinity Nellcor OxiMax SmartPod is intended for use under the direct supervision of a licensed healthcare practitioner (i.e. Physicians, Nurses, and Technicians). It is indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor). The Infinity Nellcor OxiMax SmartPod is indicated for use with adult, pediatric, and neonatal patients. The Infinity Nellcor OxiMax SmartPod and accessories are indicated for use with patients who are well or poorly perfused in hospitals and hospital type facilities. The Infinity Nellcor OxiMax SmartPod is not compatible for use in home care environment and in hospitals where CT, Hyperbaric chambers and MRI equipment are in use.
Device Story
Infinity Nellcor OxiMax SmartPod is a pulse oximetry module for Draeger Infinity patient monitoring series; provides continuous non-invasive SpO2 and pulse rate monitoring. Input: SpO2 sensor signals processed via integrated Nellcor Nell-1 circuit board and proprietary Nellcor OxiMax algorithm. Output: SpO2 and pulse rate values displayed on bedside monitor. Used in hospitals/clinical settings by physicians, nurses, and technicians. Device connects externally via RS232/X8 connector; powered by host monitor. Bedside monitor manages alarms, trends, and status messages. Benefits: enables continuous patient oxygenation status tracking to support clinical decision-making.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed that software modifications are safe and effective. Performance testing demonstrated that SpO2 and pulse rate values calculated by the Nellcor OxiMax circuit are identical to values displayed on the Draeger Infinity Delta series monitors. Compliance with IEC 60601-1 and ISO 9919 (2005) standards was verified.
Technological Characteristics
Module integrates Nellcor Nell-1 circuit board for pulse oximetry. Connectivity: RS232 via X8 connector to Infinity Delta series monitors. Power: supplied by host monitor. Standards: IEC 60601-1 (electrical safety), ISO 9919 (pulse oximeter performance). Biocompatibility of sensors per Nellcor Puritan Bennett (K060576).
Indications for Use
Indicated for continuous non-invasive monitoring of functional oxygen saturation (SpO2) and pulse rate in adult, pediatric, and neonatal patients, including those well or poorly perfused. For use by healthcare practitioners in hospitals/hospital-type facilities. Contraindicated for home use and environments with CT, hyperbaric chambers, or MRI equipment.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Dräger medical, along with the text "A Dräger and Siemens Company" underneath. Above the logo, there is handwritten text that reads "K082888 page 1 of 3". The logo is in a bold, sans-serif font, with the word "Dräger" in a larger size than the word "medical".
NOV 2 6 2008
# 3.0 510(k) Summary
as required per 807.92(c)
### 3.1 Name of Submitter, Contact Information and Date Summary Prepared
| Name | Draeger Medical Systems, Inc. |
|----------|---------------------------------|
| Address: | 6 Tech Drive, Andover, MA 01810 |
Official Contact: Pradeep Gupta, Manager Regulatory Affairs (978) 379-8219 Telephone: Fax: (978) 379-8331
Date Prepared: November 5, 2008
### 3.2 Device Trade Name, Common Name& Regulation:
| Trade Name: | Infinity Nellcor OxiMax SmartPod |
|----------------------|----------------------------------|
| Common Name: | Pulse Oximeter |
| Classification Name: | Oximeter |
| Regulation Number: | 21 CFR 870.2700 |
3.3 Device Code & Class:
| Classification Code: | DQA |
|----------------------|----------|
| Class: | Class II |
### 3.4 Legally Marketed Equivalent Device Names
Substantial equivalence is claimed to the Infinity Masimo SET® SpO2 pod 510(k) submission # K061329.
### 3.5 Device Description:
The Infinity Nellcor OxiMax SmartPod is an addition to Draeger Medical System's Infinity patient monitoring series that provides continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adult, pediatric, and neonatal patients. The Infinity Nellcor OxiMax SmartPod is manufactured by Draeger Medical Systems, Inc. and contains the Nellcor OxiMax Pulse Oximeter's Nell-1 circuit board with Nellcor's SpO2 measurement algorithm.
The Infinity Nellcor OxiMax SmartPod works as a component of the Infinity Patient Monitoring series and does not function on its own. The In-
# COMPANY CONFIDENTIAL
DRAEGER MEDICAL SYSTEMS, INC. 6 Tech Drive Andover, MA 01810 Tel: 978-379-8219 Fax:978-379-8331
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the text "K082883 page 2 of 3" at the top. Below that is the logo for Dräger medical. Underneath the logo, it says "A Dräger and Siemens Company". The image appears to be a page from a document.
finity Nellcor OxiMax SmartPod is connected externally via RS232 using X8 connector on the Infinity Delta series monitors. The Infinity Nellcor OxiMax SmartPod is powered by the patient monitor. A Nellcor OxiMax Pulse Oximetry sensor is attached to a patient finger and one end of the patient cable is connected to the sensor and the other end connected to the Infinity Nellcor OxiMax SmartPod. The monitor will begin continuously displaying the patient's pulse rate and SpO2 value. Hi/Low SpO2 and pulse rate alarm limits, alarms, trends, and status messages are all controlled by the bedside monitor.
# 3.6 Intended Use:
The Infinity Nellcor OxiMax SmartPod is intended to provide continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adult, pediatric, and neonatal patients. The Infinity Nellcor OxiMax SmartPod is intended to be used by Healthcare Providers, i.e. Physicians, Nurses, and Technicians in the hospital and hospital type facilities.
# 3.7 Comparison to Predicate Device:
Similar to Infinity Masimo SET® SpO2 pod, the Infinity Nellcor OxiMax SmartPod provides continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adult, pediatric, and neonatal patients.
# 3.8 Assessment of non-clinical performance data for equivalence:
The Infinity Nellcor OxiMax SmartPod was tested in accordance with the applicable standards and internal design control procedures and was determined to be as safe and effective for its intended use as the predicate device. Verification and validation testing performed indicate that the software modifications described in this submission are as safe and effective as previous versions and have not altered the fundamental technology of the device(s). Additionally, testing demonstrates that SpO2 and pulse rate values calculated by the third party (OEM) device (Nellcor OxiMax Pulse Oximeter N-600x) are identical to the values displayed on the Draeger Medical Infinity Delta series monitors.
# COMPANY CONFIDENTIAL
DRAEGER MEDICAL SYSTEMS, INC. 6 Tech Drive Andover, MA 01810 Tel: 978-379-8219 Fax:978-379-8331
{2}------------------------------------------------
### Special 510 (k) Infinity Nellcor OxiMax SmartPod 510(k) Summary - Revised
Image /page/2/Picture/1 description: The image shows the logo for Dräger medical, which is a company that specializes in medical technology. Above the logo is the text "K082888 page 3 of 3". The logo is in black and white, and the text is in a handwritten font.
### 3.9 Biocompatibility:
The biocompatibility of the sensors is provided by the sensor manufacturer Nellcor Puritan Bennett, Inc. (K060576)
### 3.10 Sterilization:
Not Applicable
### 3.11 Standards and Guidance:
Electrical Safety: IEC 60601-1: Medical electrical equipment general requirements for safety and essential performance ISO 9919 (2005): Medical electrical equipment - particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use. Guidance Documents: "Deciding When to Submit a 510(k) for a Change to an Existing Device" released on January 10, 1997 Draft Guidance for Industry and FDA Staff-"Pulse Oximeters - Premarket Notification Submissions [510(k)]" released on July 19, 2007
# COMPANY CONFIDENTIAL
DRAEGER MEDICAL SYSTEMS, INC. 6 Tech Drive Andover, MA 01810 Tel: 978-379-8219 Fax:978-379-8331
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The eagle faces right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Pradeep Gupta Manager Regulatory Affairs Draeger Medical Systems, Incorporated 6 Tech Drive Andover, Massachusetts 01810
NOV 2 6 2008
Re: K082888
Trade/Device Name: Infinity Nellcor OxiMax SmartPod Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: November 6, 2008 Received: November 7, 2008
Dear Mr. Gupta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Gupta
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sytie Y. Michie Davis.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): 长�82888
Device Name: Infinity Nellcor OxiMax SmartPod
The Infinity Nellcor OxiMax SmartPod is intended for use under the direct supervision of a licensed healthcare practitioner (i.e. Physicians, Nurses, and Technicians). It is indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor).
The Infinity Nellcor OxiMax SmartPod is indicated for use with adult, pediatric, and neonatal patients.
The Infinity Nellcor OxiMax SmartPod and accessories are indicated for use with patients who are well or poorly perfused in hospitals and hospital type facilities.
The Infinity Nellcor OxiMax SmartPod is not compatible for use in home care environment and in hospitals where CT, Hyperbaric chambers and MRI equipment are in use.
Prescription Use V AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
mit to
vision Sign-Off) ision of Anesthesiology, General Hospital ction Control, Dental Devices
KOR2888 (k) Number:
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.