Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter
Applicant
Masimo Corporation
Product Code
DQA · Cardiovascular
Decision Date
Dec 28, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
The Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter is intended to be used with compatible Philips Intellivue Patient Monitors. The indications for use as specified for the Intellivue Patient Monitors applies. The Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused. The Masimo Rainbow SET® Intellivue Module Pulse CO-Oximeter is not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.
Device Story
Module connects to Philips Intellivue patient monitors to provide continuous, noninvasive physiological measurements. Inputs: multi-wavelength optical signals from Masimo sensors (SpO2, SpCO, SpMet, SpHb, PI, PVI, SpOC) and acoustic airflow signals (RRa). Transformation: spectrophotometry for blood constituents; photoplethysmography for arterial blood volume changes; acoustic signal processing for respiration rate. Output: processed data sent to host monitor for display, alarm, and trend analysis. Used in clinical environments where Philips monitors are deployed; operated by healthcare professionals. Device lacks internal power, display, or alarms, relying entirely on host monitor. Output assists clinicians in patient assessment and intervention decisions. Benefits: continuous monitoring of hemoglobin fractions and respiratory status without invasive sampling.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by non-clinical bench testing, including electrical safety, EMC, usability, software verification, and mechanical/environmental testing.
Technological Characteristics
Plastic enclosure; 102.5 x 99.6 x 36 mm; 195g. Optical sensing (multi-wavelength spectrophotometry) and acoustic sensing (airflow). Powered by host monitor via 12-pin circular connector. Defibrillation-proof type CF-applied part. Compliant with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for noninvasive monitoring of SpO2, pulse rate, SpCO, SpMet, SpHb, and/or RRa in patients who are well or poorly perfused, during motion or no-motion conditions. Not for use as sole diagnostic basis for carbon monoxide poisoning.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
K020479 — DATEX-OHMEDA S/5 NELLCOR COMPATIBLE SATURATION MODULE, M-NSAT AND ACCESSORIES · Datex-Ohmeda · Mar 26, 2002
K081937 — PICOSAT II AND M3002A MULTIMEASUREMENT AND M1020B PLUG-IN MODULES SPO2 PULSE OXIMETRY MODULE · Philips Medizin Systeme Boeblingen GmbH · Aug 29, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an emblem consisting of three stylized human profiles facing to the right, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 28, 2016
Masimo Corporation Marguerite Thomlinson Senior Director, Regulatory Affairs 52 Discovery Irvine, California 92618
Re: K162675
Trade/Device Name: Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: November 23, 2016 Received: November 28, 2016
Dear Marguerite Thomlinson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image contains a signature and a name. The signature is a complex series of loops and curves. To the right of the signature is the name "Tina Kiang-S", written in a simple, sans-serif font. The name is stacked vertically, with "Tina" on top and "Kiang-S" below.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K162675
#### Device Name
Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter
#### Indications for Use (Describe)
The Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter is intended to be used with compatible Philips Intellivue Patient Monitors. The indications for use as specified for the Intellivue Patient Monitors applies.
The Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter is indicated for the nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter is indicated for use during both no motion conditions, and for patients who are well or poorly perfused.
The Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter is not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 20px;">□</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the logo is the text "MASIMO CORPORATION" in a smaller, gray font, followed by the address "52 Discovery, Irvine, CA 92618".
#### Section 5. 510(k) Summary
| Submitter and Address of<br>Manufacturing Facility: | Masimo Corporation<br>52 Discovery<br>Irvine, CA 92618<br>Phone: (949) 297-7683<br>FAX: (949) 297-7592 |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Date: | December 23, 2016 |
| Contact: | Marguerite Thomlinson<br>Senior Director, Regulatory Affairs |
| Trade Name: | Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter |
| Common Name: | Pulse CO-oximeter |
| Classification Regulation/<br>Product Code: | 21 CFR 870.2700, Class II/DQA |
| Establishment<br>Registration Number: | 2031172 |
| Reason for Premarket<br>Notification: | New Device |
| Predicate Device: | K150975, Philips Patient Monitors<br>K040259, Masimo SET Intellivue Pulse Oximeter Module |
| Performance Standards | No performance standards for the above device have been<br>promulgated pursuant to Section 514. |
### Device Description
The subject device, Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter (Rainbow IVM), resulted from simply taking the same Masimo Rainbow SET technology included in the primary predicate, Philips Multi-Measurement Module (MMS), cleared under K150975 and putting the Masimo Rainbow SET technology into the same form factor as the reference predicate, Masimo Intellivue Pulse Oximeter Module (IVM), cleared under K040259. The resulting Rainbow IVM provides continuous, noninvasive measurements of functional oxygen arterial hemoglobin (SpO2), pulse rate, perfusion index (PI), pleth variability index (PVI), carboxyhemoglobin (SpCO), methemoglobin (SpMet), oxygen content (SpOC) and respiration rate (RRa). The Rainbow IVM is intended to be used with compatible host monitors, Philips Monitors (K150975). Hence, the Rainbow IVM can be used in the same environment that the Philips Monitors are used. Furthermore, the subject device does not
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red symbol on the left, followed by the word "Masimo" in a bold, sans-serif font. To the right of the word "Masimo" are the words "MASIMO CORPORATION" in a smaller, sans-serif font, followed by the address "52 Discovery, Irvine, CA 92618".
# Section 5. 510(k) Summary
have its own power. monitoring alarms or display and must rely on the host monitor for those capabilities.
The Rainbow IVM is intended to be used with the compatible host monitors, Philips Monitors, which include the following models:
- IntelliVue MP40 in internal slots
- IntelliVue MP50 in internal slots ●
- IntelliVue MX500 in internal slots ●
- IntelliVue MX550 in internal slots
- IntelliVue MP60 via FMS-4 or FMS-8
- IntelliVue MP70 via FMS-4 or FMS-8 ●
- IntelliVue MP80 via FMS-4 or FMS-8 .
- IntelliVue MP90 via FMS-4 or FMS-8 ●
- IntelliVue MX600 via FMS-4 or FMS-8
- . IntelliVue MX700 via FMS-4 or FMS-8
- IntelliVue MX800 via FMS-4 or FMS-8 ●
See the table below for the Rainbow IVM specifications.
| RAINBOW IVM SPECIFICATIONS | |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FEATURE | SPECIFICATION |
| Measurement Range | Arterial Oxygen Saturation (SpO2): 0-100% |
| | Pulse Rate (PR): 25-240 beats per minute (bpm) |
| | Perfusion Index (PI): 0.02-20% |
| | Pleth Variability Index (PVI): 0-100% |
| | Carboxyhemoglobin Saturation (SpCO): 0-99% |
| | Methemoglobin Saturation (SpMet): 0-99.9% |
| | Total Hemoglobin (SpHb): 0-25.0 g/dL |
| | Total Oxygen Concentration (SpOC): 0-35ml/dL of O2 dL of blood |
| | Respiratory Rate (RRa): 0-70 breaths per minute |
| Display resolution | SpO2: 1% |
| | PR: 1 bpm |
| | SpCO: 1% |
| | SpMet: 0.1% |
| | SpHb: 0.1 g/dL |
| | RRa: 1breath per minute |
| RAINBOW IVM SPECIFICATIONS | |
| FEATURE | SPECIFICATION |
| Measurement Accuracy | |
| SpO2, No Motion | 60-80%, 3%, adults/pediatrics/infants<br>70-100%, 2%, adults/pediatrics/infants |
| SpO2, Motion | 70-100%, 3%, neonates |
| SpO2, Low Perfusion | 70-100%, 3%, adults/pediatrics/infants/neonates |
| Pulse Rate, No Motion | 70-100%, 2%, adults/pediatrics/infants/neonates |
| Pulse Rate, Motion | 25-240 bpm, 3 bpm, adults/pediatrics/infants/neonates |
| Pulse Rate, Low Perfusion | 25-240 bpm, 5 bpm, adults/pediatrics/infants/neonates |
| SpCO | 25-240 bpm, 3 bpm, adults/pediatrics/infants/neonates |
| SpMet | 1-40%, 3%, adults/pediatrics/infants |
| SpHb | 1-15%, 1%, adults/pediatrics/infants/neonates |
| RRa | 8-17 g/dL, 1 g/dL, adults/pediatrics |
| Display/Indicator for Host Monitor | 4-70 breaths per minute, 1 breath per minute, adults/pediatrics |
| | SpO2%, pulse rate, alarm status, alarm silenced status, perfusion index,<br>carboxyhemoglobin (SpCO%), methemoglobin (SpMet%), total hemoglobin<br>(SpHb g/dL or mmol/l), SpOC, Pleth Variability Index (PVI), Acoustic<br>respiration rate, (RRa), total oxygen content (SpOC), |
| Modes | |
| SpO2 Averaging mode | 2, 4, 8, 10, 12, 14 and 16 seconds |
| Sensitivity mode | Normal, maximum and APOD |
| SpHb averaging mode | Long, medium, short |
| SpHb monitoring mode | Arterial or venous |
| SpHb precision mode | 1,0.5, 0.1 (g/dL or mmol/L) |
| PI averaging (smooth PI) | Automatically set based on SpO2 averaging mode<br>SpO2 Averaging mode: 2-8 seconds = Short<br>SpO2 Averaging mode: 10-16 seconds = Long |
| PVI | Automatically set based on SpO2 averaging mode<br>SpO2 Averaging mode: 2-6 seconds = Short<br>SpO2 Averaging mode: 10-16 seconds = Long |
| RRa averaging | None, fast, medium, slow, trending |
| RRa pause time | 15, 20, 25, 30, 35, and 40 seconds |
| RRa freshness timeout | 0, 1, 5, 10, and 15 minutes |
| Line frequency | 50/60Hz |
| SmartTone | On/Off |
| Desaturation index alarm (Desat<br>Index) | % Threshold: 2-10%, in 1% increments<br>Occurrence count: 1-25 in 1 step increments<br>Period: 1-4 hours in 1 hour increments |
| Perfusion index delta alarm (PI Delta) | % Delta: 10-100%, in 2% increments<br>Duration: 1, 5, or 30 minutes, 1 hour, 4-48 hours (4 hour increments), infinite. |
| Electrical | |
| Power requirement | Power supplied by Philips Monitor |
| Interface | |
| Power connector | Interface Philips Monitor standard 12-pin circular connector |
| Patient cable connector | Interface Masimo patient cable, which connects to Masimo sensor |
| | RAINBOW IVM SPECIFICATIONS |
| FEATURE | SPECIFICATION |
| Indicator/ Control | |
| Indicator | 1 LED indicator – green color |
| Control | 1 push button switch |
| Mechanical | |
| Enclosure material | Plastic |
| Dimensions/weight | 102.5 x 99.6 x 36 mm (4.0 x 3.9 x 1.4 in.)/ 195g (7.0 oz.) |
| Environmental | |
| Operating temperature | 0 to 55°C, ambient humidity (32 to 131°F) |
| Transport/storage temperature | -40 to 70°C, ambient humidity (-40 to +158°F) |
| Operating humidity | 95% RH max at 40°C |
| Transport/storage humidity | 95% RH max at 65°C |
| Operating altitude | Up to 4,600 m. (15,000 ft.) |
| Transport/storage altitude | -380 to 5,560 m (-1,250 to 18250 ft) |
| Compliance | |
| Electrical safety and EMC | IEC 60601 compliant |
| Degree of protection -electrical shock | Defibrillation proof type CF-applied part |
| Mode of Operation | Continuous operation |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the logo is the text "MASIMO CORPORATION" and the address "52 Discovery, Irvine, CA 92618".
# Section 5. 510(k) Summary
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white line running through it, followed by the word "Masimo" in a bold, sans-serif font. To the right of the logo, the text "MASIMO CORPORATION" is displayed in a smaller, sans-serif font, followed by the address "52 Discovery, Irvine, CA 92618".
#### Section 5. 510(k) Summary
### Intended Use
The Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter is intended as a measurement module for use with compatible Philips Intellivue Patient Monitors (Philips Monitor).
### Indications For Use
The Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter is intended to be used with compatible Philips Intellivue Patient Monitors. The indications for use as specified for the Intellivue Patient Monitors applies.
The Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Rainbow SET Intellivue Module Pulse CO-Oximeter is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.
The Masimo Rainbow SET® Intellivue Module Pulse CO-Oximeter is not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white line running through it, followed by the word "Masimo" in a bold, black font. To the right of the word "Masimo" is the text "MASIMO CORPORATION" in a smaller, gray font, followed by the address "52 Discovery, Irvine, CA 92618".
#### Section 5. 510(k) Summary
### Technological Characteristics
### Principle of Operation
Optical Measurements (SpO2, Pulse Rate, SpCO, SpMet, SpHb, PI, PVI, SpOC) – The Masimo Rainbow SET technology utilizes the general principles of pulse oximetry for noninvasive optical measurements of SpO2, pulse rate, SpCO, SpMet and SpHb measurements. The perfusion index (PI), pleth variability index (PVI) and oxygen content (SpOC) values are derived from these measurements.
The Masimo Rainbow SET pulse CO-oximetry technology is governed by the following principles:
- Oxyhemoglobin (oxygentated blood) and deoxyhemoglobin (non-oxygenated . blood), carboxyhemoglobin (blood with carbon monoxide content), methemoglobin (blood with oxidized hemoglobin) and blood plasma constituents differ in their absorption of visible and infrared light (using spectrophotometry).
- The amount of arterial blood in tissue changes with pulse (photoplethysmography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.
The Rainbow Pulse CO-oximetry technology uses a multi-wavelength sensor to distinguish between oxygenated blood, deoxygenated blood, blood with carbon monoxide, oxidized blood and blood plasma.
Acoustic Measurement (RRa) – The Masimo Rainbow Acoustic Monitoring (RAM) technology uses acoustic signals for respiration rate (RRa) measurements. RRa measures a patient's respiration rate based on airflow sounds generated in the upper airway. RRa translates airflow sounds generated in the upper airway to an electrical signal that can be processed to produce a respiration rate, measured as breaths per minute.
### Mechanism of Action for Achieving the Intended Effect
The subject device is connected to a compatible host/backboard device, Philips Monitor. A sensor is connected to Rainbow IVM on one end and applied to the patient at the other end. Monitoring begins when the power on the host/backboard device is turned on. The sensor collects physiological signals from the patient and sends the signals to the subject device for processing. The processed signals or measurements are then sent to the host/backboard for display. The clinician can use
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark-like shape inside, followed by the word "Masimo" in bold, black letters. To the right of the logo, the text "MASIMO CORPORATION" is displayed above the address "52 Discovery, Irvine, CA 92618".
# Section 5. 510(k) Summary
the information to assess the patient condition and help determine if intervention is required. When use is complete, the sensor is removed from the patient and the power is turned off.
### Summary of Technological Characteristics of Subject Device Compared to Predicates
The subject device and primary predicate (K150975) have the following key similarities:
- Both are intended for use as measurement modules; ●
- Both rely on connected host monitor for power, display and alarm functions; ●
- Both are intended to be used with compatible models of Philips Monitors (K150975):
- Both include the same Masimo Rainbow SET technology: ●
- Both include the same principles of operation for the Masimo Rainbow SET technology; and
- . Both are compatible with cleared Masimo sensors/cables for the measurement of pulse CO-oximeter parameters including SpO2, pulse rate, SpCO, SpMet and SpHb and respiration rate parameters.
The subject device and primary predicate (K150975) have the following key differences:
- . The subject device only includes the Masimo Rainbow SET technology, whereas the predicate includes other technologies such as NIBP and ECG in addition to the Masimo Rainbow SET technology; and
- . The subject device and the primary predicate have different form factor;
The subject device and reference predicate (K040259) have the following key similarities:
- Both are intended for use as measurement modules;
- Both rely on connected host monitor for power, display and alarm functions; ●
- Both are intended to be used with compatible models of Philips Monitors (K150975); ●
- Both include the same Masimo technology for pulse oximetry; ●
- Both include the same principles of operation for the Masimo pulse oximetry technology; and
- Both are compatible with cleared Masimo sensors/cables for the measurement of ● pulse oximetry parameters.
The subject device and reference predicate (K040259) have the following key differences:
- . The subject device includes Masimo Rainbow SET technology, whereas the predicate
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark-like shape inside, followed by the word "Masimo" in a bold, sans-serif font. To the right of the logo is the text "MASIMO CORPORATION" in a smaller font, followed by the address "52 Discovery, Irvine, CA 92618".
## Section 5. 510(k) Summary
only includes Masimo pulse oximetry technology, a subset of the Masimo Rainbow SET technology; and
- The subject device is compatible with cleared Masimo sensors/cables for . measurements of pulse CO-oximeter parameters including SpO2. pulse rate. SpCO. SpMet and SpHb and respiration rate parameters, whereas the predicate is compatible with Masimo sensors/cables only for the measurement of pulse oximetry parameters.
### Non-clinical Testing
The following tests, as applicable, were performed for the qualification of the subject devices, Rainbow IVM, in accordance with the requirements of the design control regulations and established quality assurance processes to demonstrate substantial equivalence with its predicates:
- Electrical safety testing per IEC-60601-1
- EMC testing per IEC-60601-1-2 ●
- Usability testing per FDA Human Factors and Usability Draft Guidance ●
- Software verification, including integration testing with host monitors, per FDA ● Software Guidance
- Mechanical testing per ISTA-2A and MIL-STD 810E
- Environmental testing per IEC-60601-1 .
### Clinical Testing
No clinical testing was done.
### Conclusion
The results of the testing demonstrate that all requirements and performance specifications were satisfied, and that the subject device is substantially equivalent to its predicates. Thus, it is concluded that the subject device, Rainbow IVM, is substantially equivalent to the predicate based on the non-clinical tests discussed above.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.