SIEMENS MEDICAL INFORMATION BUS (MIB) PROTOCOL CONVERTER

K991661 · Siemens Medical Solutions USA, Inc. · DQA · May 28, 1999 · Cardiovascular

Device Facts

Record IDK991661
Device NameSIEMENS MEDICAL INFORMATION BUS (MIB) PROTOCOL CONVERTER
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeDQA · Cardiovascular
Decision DateMay 28, 1999
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.2700
Device ClassClass 2

Indications for Use

Siemens Medical Information Bus (MIB) Protocol Converter is intended to connect third party medical devices that do not provide data per the IEEE1073 Medical Information Protocol Standard to the Siemens INFINITY Modular Bedside Monitors for display, such as: Siemens SV 300 ventilator Baxter Vigilance blood gas/continuous cardiac output monitor Siemens SV900 ventilator Draeger Evita II ventilator Draeger Evita IV ventilator Draeger Babylog ventilator Puritan Bennett 7200 ventilator Draeger Narkomed II Anesthesia System Draeger Narkomed IV Anesthesia System Draeger Julian Anesthesia Machine Ohmeda 7900 / Modulus CD Anesthesia Machine Abbott Oximetrix 3 Oximeter A VL Medical Instruments: Opti Critical Care Analyzer, Portable Blood Gas Analyzer Optical Sensors Inc .: OSI - Optical CAM VIA Medical: VIA V-ABG1 Blood Gas Chemistry Monitor

Device Story

The MIB Protocol Converter acts as a communication bridge; it translates proprietary data signals from third-party medical devices (ventilators, anesthesia machines, blood gas analyzers) into a format compatible with Siemens INFINITY Modular Bedside Monitors. The device uses specific accessory cables to interface with external equipment. It is operated by healthcare professionals in clinical settings. The converter enables the central display of parameters from non-IEEE 1073 compliant devices on the bedside monitor. This integration allows clinicians to view data from multiple sources on a single interface, potentially improving monitoring efficiency and clinical decision-making. The monitor may generate alarms for MIB-transmitted anesthesia parameters independently of the source machine.

Clinical Evidence

No clinical data. Bench testing only.

Technological Characteristics

The device functions as a signal amplifier and conditioner (21 CFR 870.2060). It utilizes IEEE 1073.3.1-1994 (Transport Profile) and IEEE 1073.4.1-1994 (Physical Layer Interface) standards for communication. It is a hardware-based protocol converter with associated accessory cables for physical connectivity to third-party medical devices. It is not MRI compatible.

Indications for Use

Indicated for use in clinical environments by healthcare professionals (physicians, nurses, technicians) to connect third-party medical devices (ventilators, anesthesia systems, oximeters, blood gas analyzers) that lack IEEE 1073 protocol support to Siemens INFINITY Modular Bedside Monitors for data display. Not MRI compatible.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ ## Special 510(k): Device Modification SIEMENS Medical Information Bus (MIB) Protocol Converter K991661 # 510(k) SUMMARY as required per 807.92(c) Submitters Name, Address: Siemens Medical Systems, Inc. Electromedical Systems Group, PCS Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879 Official Correspondent: David Simard, Director, QA/RA Contact person for this submission: Penelope H. Greco Date submission was prepared: May 5, 1999 Trade Name, Common Name and Classification Name: - A. Trade Name: Siemens Medical Information Bus (MIB) Protocol Converter B. Common Name, Classification Name, Class and Regulation Number: | Common Name | Classification<br>Number | Class | Regulation Number | |---------------------------------------------|--------------------------|-------|-------------------| | Transducer Signal amplifier and conditioner | 73 DQA | II | 21 CFR 870.2060 | ## Legally Marketed Device Identification: Siemens Medical Information Bus (MIB) Protocol Converter: 510(k) K970368 and K973222. ### Description of Modification: The Medical Information Bus (MIB) Protocol Converter has received two 510(k) clearances. - 510(k) K970368 was cleared for interface with Siemens SV300 ventilator and 1. the Baxter Vigilance blood gas/continuous cardiac output monitor. - 510(k) K973222 was cleared for interface with Puritan Bennett 7200 ventilator, 2. the Draeger Evita II, Draeger Evita IV, and Draeger Babylog ventilators, and Siemens SV900 ventilator. Minor software modifications have been made to Siemens Medical Information Bus (MIB) Protocol Converter and device specific accessory cables are now available that allow interface connections for the following devices to the INFINITY modular bedside monitors (SC9000/SC9000XL/SC8000). The monitor may alarm for some MIB anesthesia parameters independent of the anesthesia machine's alarms. #### Anesthesia Systems - Dräger Narkomed II - Dräger Narkomed IV - · Dräger Julian - Ohmeda 7900 / Modulus CD Anesthesia Machine ### Oximeter - · Abbott Oximetrix 3 Point of Care Blood Gas Analyzers AVL Medical Instruments - · Opti Critical Care Analyzer, Portable Blood Gas Analyzer Optical Sensors Inc. • OSI – Optical CAM # VIA Medical - VIA V-ABG1 Blood Gas Chemistry Monitor l ## COMPANY CONFIDENTIAL ## Siemens Medical Systems, Inc. Electromedical Systems Group, PCS 16 Electronics Avenue Danvers, MA 01923 USA Tel: (978) 907-7500 Fax: (978) 750-6879 Telex: 511958 (Siemensm SD) {2}------------------------------------------------ #### Intended Use: Siemens Medical Information Bus (MIB) Protocol Converter is intended to connect third party medical devices that do not provide data per the IEEE1073 Medical Information Protocol Standard to the Siemens INFINITY Modular Bedside Monitors for display, such as: > Siemens SV 300 ventilator Baxter Vigilance blood gas/continuous cardiac output monitor Siemens SV900 ventilator Draeger Evita II ventilator Draeger Evita IV ventilator Draeger Babylog ventilator Puritan Bennett 7200 ventilator Draeger Narkomed II Anesthesia System Draeger Narkomed IV Anesthesia System Draeger Julian Anesthesia Machine Ohmeda 7900 / Modulus CD Anesthesia Machine Abbott Oximetrix 3 Oximeter A VL Medical Instruments: Opti Critical Care Analyzer, Portable Blood Gas Analyzer Optical Sensors Inc .: OSI - Optical CAM VIA Medical: VIA V-ABG1 Blood Gas Chemistry Monitor Assessment of non-clinical performance data for equivalence: Section K Assessment of clinical performance data for equivalence: Not applicable Biocompatability: Not applicable Sterilization: Not applicable Standards and Guidances: 1073.3.1-1994 IEEE Standard for Medical Device Communications Transport Profile - Connector Mode 1073.4.1 - 1994 IEEE Standard for Medical Device Communications Physical Layer Interface - Cable Connected 2 # COMPANY CONFIDENTIAL # Siemens Medical Systems, Inc. Electromedical Systems Group. PCS 16 Electronics Avenue Danvers, MA 01923 USA Tel: (978) 907-7500 Fax: (978) 750-6879 Telex: 511958 (Siemensm SD) {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is written around the top half of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 2 8 1999 Mr. David Simard Siemens Medical Systems, Inc. 16 Electronics Avenue Danvers, MA 01923 K991661 Re: Siemens Medical Information Bus (MIB) Protocol Converter Regulatory Class: II (two) Product Code: 73 DQA May 5, 1999 Dated: Received: May 14, 1999 Dear Mr. Simard: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 - Mr. David Simard This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahan Ph.D. Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neuroloqical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): _ Device Name: Siemens Medical Information Bus (MIB) Protocol Converter Indications for Use: The Medical Information Bus (MIB) Protocol Converter is indicated for use in an environment where patient care is provided by healthcare professionals (Physician, Nurse, Technician) when the professional determines that third party medical devices that do not provide data per the IEEE 1073 Medical Information Protocol Standard should be connected to the Siemens INFINITY Modular Bedside Monitors for display, such as: > Siemens SV 300 ventilator Baxter Vigilance blood gas/continuous cardiac output monitor Siemens SV900 ventilator Draeger Evita II ventilator Draeger Evita IV ventilator Draeger Babylog ventilator Puritan Bennett 7200 ventilator Draeger Narkomed II Anesthesia System Draeger Narkomed IV Anesthesia System Draeger Julian Anesthesia Machine Ohmeda 7900 / Modulus CD Anesthesia Machine Abbott Oximetrix 3 Oximeter AVL Medical Instruments: Opti Critical Care Analyzer, Portable Blood Gas Analyzer Optical Sensors Inc .: OSI - Optical CAM VIA Medical: VIA V-ABG1 Blood Gas Chemistry Monitor At. A. Carlousk. 510k) Num MRI Compatibility Statement: The Medical Information Bus (MIB) Protocol Converter is not compatible for use in a MRI magnetic field. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | <div> <img alt="Checkmark" src="checkmark.png"/> </div> | |----------------------|---------------------------------------------------------| | (Per 21 CFR 801.109) | | OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...