Guardian Angel GA1000 Digital Vital Sign Monitoring System
K162580 · Taiwan Aulisa Medical Devices Technologies, Inc. · DQA · Mar 3, 2017 · Cardiovascular
Device Facts
Record ID
K162580
Device Name
Guardian Angel GA1000 Digital Vital Sign Monitoring System
Applicant
Taiwan Aulisa Medical Devices Technologies, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Mar 3, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Guardian Angel GA1000 Digital Vital Sign Monitoring System is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patients. It is indicated for spot-checking and / or continuous monitoring of patients during non-motion and under well-perfused conditions. The intended environment of use is hospital. This system is a reusable device.
Device Story
The Guardian Angel GA1000 is a digital vital sign monitor consisting of a wrist-worn sensor module and a portable, table-top wireless display unit. It uses red and infrared LED transmittance technology to measure functional arterial hemoglobin oxygen saturation (SpO2) and pulse rate. Data is transmitted via Bluetooth v4.0 from the sensor to the display unit. The display unit features an LCD panel and speakers to provide visual and audio alarms for SpO2, pulse rate, low battery, and sensor disconnection/faults. The device is intended for use by clinicians in hospital settings. By providing real-time monitoring and alerts, the system assists healthcare providers in identifying abnormal vital signs, potentially enabling timely clinical intervention and improving patient safety.
Clinical Evidence
Clinical performance was verified in a study of healthy adult subjects under controlled induced hypoxia (70%-100% SpO2 range) per ISO 80601-2-61:2011. Over 200 data points were collected per sensor. The Arms (accuracy root-mean-square difference) for each sensor was below 3%, meeting FDA guidance requirements for pulse oximeters.
Technological Characteristics
Transmittance-based pulse oximetry using red (660 nm) and infrared (905 nm) LEDs. Components: wrist-worn sensor module, wireless display unit. Connectivity: Bluetooth v4.0. Power: 3.7V Lithium battery (sensor), Lithium battery/AC (display). Standards: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 10993-5/10 (biocompatibility), ISO 80601-2-61 (pulse oximetry), IEC 60601-1-8 (alarms), IEC 62304 (software).
Indications for Use
Indicated for adult and pediatric patients for spot-checking or continuous monitoring of functional SpO2 and pulse rate in hospital environments during non-motion and well-perfused conditions.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Avant Digital Pulse Oximetry System, Model 4000, 4100, Nonin Medical, Inc. (K052669)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is a circular logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling an emblem. The emblem consists of three stylized human profiles facing right, layered on top of each other to create a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 3, 2017
Taiwan Aulisa Medical Devices Technologies, Inc. % Don Mizota Consultant 725 Morninghome Rd Danville, California 94526
Re: K162580
Trade/Device Name: Guardian Angel GA1000 Digital Vital Sign Monitoring System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: January 26, 2017 Received: January 30, 2017
Dear Don Mizota:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Lori A. Wiggins -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K162580
Device Name
Guardian Angel GA1000 Digital Vital Sign Monitoring System
Indications for Use (Describe)
The Guardian Angel GA1000 Digital Vital Sign Monitoring System is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patients. It is indicated for spot-checking and / or continuous monitoring of patients during non-motion and under well-perfused conditions. The intended environment of use is hospital. This system is a reusable device.
| Type of Use (Select one or both, as applicable) | <span> <svg height="12" width="12"> <rect fill="white" height="12" stroke="black" stroke-width="1" width="12"></rect> <line stroke="black" stroke-width="1" x1="0" x2="12" y1="0" y2="12"></line> <line stroke="black" stroke-width="1" x1="0" x2="12" y1="12" y2="0"></line> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <svg height="12" width="12"> <rect fill="white" height="12" stroke="black" stroke-width="1" width="12"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### Section 5: 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of 21 CFR 807.92
#### 5.1. General Information
| | Date Summary Prepared: September 13th, 2016 |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by: | Taiwan Aulisa Medical Devices Technologies, Inc.<br>Rm. 1052, Bldg. H, 10F., No.3-2, YuanQu St.,<br>Nangang Dist., Taipei City 115 TW<br>Phone: 886-2-2655-7297<br>FAX: 886-2-2655-7260 |
| Contact Person: | Jenny Lai<br>RA<br>Taiwan Aulisa Medical Devices Technologies, Inc.<br>Rm. 1052, Bldg. H, 10F., No.3-2, YuanQu St., |
Nangang Dist.,Taipei City 115 TW Phone: 886-2-2655-7297 FAX: 886-2-2655-7260 Email: jenny.lai@aulisa.com
#### 5.2. Name of Device
| Trade Name: | Guardian Angel GA1000 Digital Vital Sign Monitoring<br>System |
|-----------------|---------------------------------------------------------------|
| Common Name: | Pulse Oximeter |
| Classification: | Oximeter CFR §870.2700 |
| Class: | Class II |
| Product Code: | DQA |
#### 5.3. Legally Marketed Predicate Devices for Claimed Equivalence
K052669, Avant Digital Pulse Oximetry System, Model 4000, 4100, Nonin Medical, Inc.
{5}------------------------------------------------
#### 5.4. Device Description
Aulisa's GA1000 is a digital vital sign monitoring system that measures and displays a patient's pulse rate and oxygen saturation level. The system includes a self-contained wrist-worn Sensor Module and a portable, table-top wireless Display Unit. The Guardian Angel GA1000 Digital Vital Sign Monitoring System is also equipped with an alarm system that alerts the caregiver when a patient's pulse rate is too low or too high and when the patient's SpO2 is too low or too high.
Guardian Angel GA1000 Digital Vital Sign Monitoring System contains the following components.
- Sensor Module ●
- Display Unit ●
- Finger Sensor ●
- Sensor Module wristband ●
- Sensor Module Charging Adaptor ●
- Display Unit Charging Adaptor ●
- . Display Unit Stand
The GA1000 measures SpO2 and pulse rate based on transmittance technology, measuring the absorbance of red and infrared light passed through the tissue.
The GA1000 uses Bluetooth v4.0 to transmit data between the wrist-worn Sensor Module and the wireless Display Unit. Both the Sensor Module and the Display Unit is Bluetooth® 4.0 compatible.
The GA1000 uses non-invasive red and infrared LED sensors to measure the functional blood oxygen saturation and pulse rate. The measurements are wirelessly transmitted to the Display Unit, which displays the measurements using a Liquid-Crystal-Display (LCD) panel. The system provides adjustable visual and audio oxygen saturation, and pulse rate alarms through the LCD panel and speakers. Additional alarms are featured, including low battery on the Sensor Module, Sensor Cable disconnection, Sensor Module disconnection, Sensor Cable Probe fault, Sensor Probe detached from patient, Display Unit battery low.
#### 5.5. Intended Use
The Guardian Angel GA1000 Digital Vital Sign Monitoring System is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patients. It is indicated for spot-checking and / or continuous monitoring of
{6}------------------------------------------------
patients during non-motion and under well-perfused conditions. The intended environment of use is hospital. This system is a reusable device.
#### Technological Characteristics 5.6.
The comparison table of technological characteristics for Guardian Angel GA1000 Digital Vital Sign Monitoring System and Nonin Avant Digital Pulse Oximetry System is shown as follows:
## Table 5.1 - Comparison Table to Predicate Device
| | Subject Device | Predicate Device | | |
|-------------------------------------------------|---------------------------------------------------------------------------------------|-----------------------------------------------|-----------------------------------------------------------------------|------------------------------------------------------------|
| Product code | DQA | DQA | | |
| Classification | Class II | Class II | | |
| Principles of<br>Operation | Red and Infra-red<br>transmittance technology | Red and Infra-red<br>transmittance technology | | |
| System<br>Components | Sensor Module and<br>Display Unit | Patient Module and<br>Display Unit | | |
| Wireless<br>technology/<br>Data<br>Transmission | Bluetooth v4.0 | Bluetooth 1.1 | | |
| Software level<br>of concern | Medium | Medium | | |
| SpO2<br>Specifications<br>(No Motion) | Displayed<br>Range: | 1-100% | Displayed<br>Range: | 0-100% |
| | Declared<br>Range: | 70-100% | Declared<br>Range: | 70-100% |
| | Accuracy:<br>(Pediatrics :<br>SC100M,<br>ENSC100M;<br>Adults:<br>SC100L,<br>ENSC100L) | Adults/<br>Pediatrics:<br>3 digits | Accuracy:<br>(Finger Clip<br>Flexi,<br>Flexifoam,<br>8000R,<br>8000Q) | Adults/<br>Pediatrics:<br>2 digits<br>3 digits<br>4 digits |
| | Range: | 30-290 bpm | Range: | 18-300 bpm |
| Pulse Rate<br>Specifications<br>(No motion) | Accuracy:<br>( SC100M,<br>SC100L,<br>ENSC100M,<br>ENSC100L) | Adults/<br>Pediatrics:<br>±3 % | Accuracy:<br>(Finger Clip<br>Flexi,<br>Flexifoam,<br>8000R,<br>8000Q) | Adults/<br>Pediatrics:<br>±3% |
{7}------------------------------------------------
| Measurement<br>Wavelengths<br>and Output<br>Power | Red 660 nm @ 1.8 mw<br>nominal<br>Infrared 905 nm @ 2 mw<br>nominal | Red 660 nm @ 3 mw nominal<br>Infrared 910 nm @ 3 mw<br>nominal |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| System<br>Components | Sensor Module and<br>Display Unit | Patient Module and<br>Display Unit |
| Power Supply | Sensor Module: 3.7V<br>Lithium battery<br>Display Unit: Lithium<br>battery, AC adaptor | Patient Module: AA battery<br>Display Unit: NiMH battery,<br>AC adaptor |
| Sensor Form | Soft Finger Sensor | Clip and Wrap Sensor |
| Alarms | Audible and visual pulse<br>rate and oxygen saturation<br>alarms.<br>Low battery and critical<br>battery, disconnection<br>alarms. | Audible and visual pulse rate,<br>oxygen saturation, and<br>perfusion alarms.<br>Low battery and<br>disconnection alarms. |
| Indications For<br>Use | The Guardian Angel<br>GA1000 Digital Vital Sign<br>Monitoring System is<br>indicated for use in<br>measuring and displaying<br>functional oxygen<br>saturation of arterial<br>hemoglobin (SpO2) and<br>pulse rate of adult and<br>pediatric patients. It is<br>indicated for spot-checking<br>and / or continuous<br>monitoring of patients<br>during non-motion and<br>under well-perfused<br>conditions. The intended<br>environment of use is<br>hospital. This system is a<br>reusable device. | The NONIN® Avant®<br>4000 Digital Pulse Oximetry<br>System is indicated for<br>measuring and displaying<br>functional oxygen saturation<br>of arterial hemoglobin (SpO2)<br>and pulse rate of adult,<br>pediatric, and infant patients.<br>It is indicated for spot<br>checking and/or continuous<br>monitoring of patients during<br>both motion and no motion<br>conditions, and for patients<br>who are well or poorly<br>perfused. |
#### Substantial Equivalence 5.7.
The Guardian Angel GA1000 Digital Vital Sign Monitoring System uses technology which is equivalent to the legally marketed predicate device cleared under K052669. The subject device is indicated for monitoring adults and pediatric patients whereas the predicate is indicated for use for monitoring adults, pediatric, and infant patients. In addition, the subject
{8}------------------------------------------------
device is indicated to monitor during non-motion conditions and in poorly perfused patients. These differences are not critical to the intended diagnostic use of the device. The indications of the subject device encompass a smaller patient population and less conditions of use; thus, the differences do not affect the safety and effectiveness of the device when used as labeled.
#### 5.8. Summary of Performance Testing
Accuracy of pulse oximeter, electrical, mechanical, electromagnetic compatibility, software and biocompatibility testing have been performed on the GA1000.
The results of the testing demonstrate that the Taiwan Aulisa Medical Technologies Guardian Angel GA1000 Digital Vital Sign Monitoring System is as safe and effective, as the legally marketed predicate device.
#### 5.8.1. Non-clinical Testing
The following non-clinical testing has been conducted for the GA1000 Digital Vital Sign Monitoring System, performed according to GA1000 product requirement specification and quality system to illustrate substantial equivalence.
| Performance Testing | ISO 80601-2-61:2011 |
|-------------------------------------------------|----------------------------------------------------------------------------------------|
| Electrical Safety<br>testing | IEC 60601-1:2005 |
| Biocompatibility<br>Testing | ISO 10993-10: 2010, ISO 10993-5:2009 |
| Electrical Safety<br>Testing and EMC<br>Testing | IEC 60601-1-2:2007 |
| Alarm System | IEC 60601-1-8:2006 |
| Wireless Coexistence | FDA Guidance Radio Frequency Wireless<br>Technology in Medical Devices (2013) |
| Software Verification | IEC 62304:2006<br>FDA Guidance Off-The-Shelf Software Use<br>in Medical Devices (1999) |
| Pulse Oximetry<br>Testing | ISO 80601-2-61:2011<br>FDA Pulse Oximeters Guidance (2013) |
## Table 5.2 - Utilization of Standards table
The tests performed on the GA1000 include:
- · Functional Test
{9}------------------------------------------------
- Push, Impact and Drop Test ●
- Particulate and Water Ingress Test .
- . Temperature and Humidity Test
- Battery Life Test ●
- Radio Frequency Compatibility ●
- Emissions and Immunity Tests ●
- Biocompatibility Test .
- Alarm System Test ●
The subject device underwent bench tests following standards listed above. Software Validation and Verification was also performed following IEC 62304:2006. Biocompatibilities of patient contact materials have been tested, showing the materials to be biocompatible. The tests show compliance to IEC 60601-1, IEC 60601-1-2, IEC 60601-8, IEC 62304, ISO 10993-5, ISO 10993-10, and ISO 80601-2-61.
#### 5.8.2. Clinical Testing
Clinical data were collected to verify the performance accuracy of GA1000 on healthy adult subjects over the range of 70%-100% SpO through controlled induced hypoxia. The data was performed according to Clause 201.12.1.101.2 and Annex EE.2 of ISO 80601-2-61:2011 Standard.
Over 200 data points were collected for each sensor. The Arms of each sensor is below 3%, compliant with FDA quidance on Pulse Oximeter -Premarket notification submissions [510(k)]: Guidance for Industry and Food and Drug Administration staff, Issued March 4t0, 2013.
#### Conclusion 5.9.
Based on the clinical data and non-clinical testing summarized in 510(k) submission, the result demonstrates Guardian Angel GA1000 Digital Vital Sign Monitoring System is in compliance with design specifications. applicable standards, and the intended use. Performance test data demonstrate substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.