MASIMO RAINBOW SET RADCHECK PULSE CO-OXIMETER, RADCHECK, RAD 57 SPOT CHECK
Applicant
Masimo Corporation
Product Code
DQA · Cardiovascular
Decision Date
Oct 10, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Masimo Rainbow SET® RadCheck Pulse CO-Oximeter and Accessories are indicated for noninvasive spot checking of functional saturation of arterial hemoglobin (SpO2), pulse rate, and total hemoglobin (SpHb). The Masimo Rainbow SET® RadCheck Pulse CO-Oximeter and Accessories are indicated for use, by trained personnel, with adult and pediatic patients during both no motion and motion conditions, and for patients who are well or poorly perfused in clinical and non-clinical settings.
Device Story
RadCheck is a portable, microprocessor-controlled pulse CO-oximeter for spot-checking vital signs. It utilizes 12-wavelength optical sensor technology to measure arterial oxygen saturation (SpO2), pulse rate, and total hemoglobin (SpHb). Input signals from the sensor are processed to output SpO2, pulse rate, SpHb, Perfusion Index (PI), and Total Arterial Oxygen Content (SpOC). The device is operated by trained personnel in clinical or non-clinical settings. Output is displayed on the device screen, allowing clinicians to assess patient status immediately. The device provides a noninvasive alternative to laboratory blood analysis, facilitating rapid clinical decision-making regarding oxygenation and hemoglobin levels.
Clinical Evidence
Clinical validation performed on healthy adult volunteers and surgical patients with varying skin pigmentation. Accuracy for SpO2 and pulse rate validated against laboratory CO-oximeter and ECG monitors under no-motion, motion, and low-perfusion conditions. SpHb accuracy validated in the 8-17 g/dl range. Bench testing conducted using Biotek Index 2 and Masimo simulators. Results demonstrate performance within specified tolerances (±1 SD).
Technological Characteristics
Microprocessor-controlled; 12-wavelength optical sensing; Polycarbonate/ABS blend housing; internally powered by non-rechargeable batteries; Type BF-applied part; IPX1 ingress protection. Connectivity via Masimo Rainbow sensors. Complies with EN 60601-1-2 (EMC) and IEC/UL 60601-1 (Electrical Safety).
Indications for Use
Indicated for noninvasive spot checking of SpO2, pulse rate, and total hemoglobin (SpHb) in adult and pediatric patients. Suitable for use in clinical and non-clinical settings by trained personnel under motion/no-motion conditions and varying perfusion states.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Masimo Rainbow SET Rad 57t Pulse CO-Oximeters and Accessories (K080238)
Submission Summary (Full Text)
{0}------------------------------------------------
K08.2052
| Submitted by: | Masimo Corporation<br>40 Parker<br>Irvine, CA 92618<br>949-297-7000<br>FAX 949-297-7001 |
|---------------------------------------|-----------------------------------------------------------------------------------------|
| Company Contact: | Marguerite Thomlinson, Manager of Regulatory Affairs |
| Date Summary Prepared: | July 16, 2008 |
| Trade Name | Masimo Rainbow SET RadCheck Pulse CO-Oximeter and Accessories |
| Common Name | Oximeter Sensor |
| Classification Name and Product Code: | Oximeter (74DQA) (870.2700) Cable, Transducer and Electrode (74DSA) (870.2900) |
| Substantially Equivalent Devices: | Masimo Rainbow SET Rad 57t Pulse CO-Oximeters and Accessories, 510(k) Number K080238 |
### Device Description
The Masimo Rainbow SET® RadCheck Pulse CO-Oximeter and Accessories (RadCheck) have modified intended use/indications for use in comparison to the Rad 57t Pulse CO-Oximeters and Accessories (Rad 57t) in the K080238 filing. The main difference is that the RadCheck in this filing is for spot checking, whereas the Rad 57t in K0808238 is for continuous monitoring. The keypad membrane and the software of the RadCheck also differ from the Rad 57t.
The RadCheck in this filing is similar in construction to the Rad 57t in the K0808238 filing, including the 12-wavelength technology for the measurement of total hemoglobin. The performance of the Radcheck is similar to the Rad 57t. Similar to the Rad 57t, the RadCheck provides noninvasive monitoring of arterial oxygen saturation (%SpO2), pulse rate, and total hemoglobin concentration (g/d) SpHb). Other information displayed by the RadCheck include: Low Signal IQ (Low SIQ), Perfusion Index (PI), Total Arterial Oxygen Content (SpOC), battery life, and sensor status.
{1}------------------------------------------------
## Predicate Device
The predicate device used in this filing is the Masimo Rainbow SET® Rad 57t Pulse CO-Oximeter (Rad 57t) and Accessories, 510(k) Number K080238.
### Intended Use
The Masimo Rainbow SET® RadCheck Pulse CO-Oximeter and Accessories are indicated for noninvasive spot checking of functional saturation of arterial hemoglobin (SpO2), pulse rate, and total hemoglobin (SpHb). The Masimo Rainbow SET® RadCheck Pulse CO-Oximeter and Accessories are indicated for use, by trained personnel, with adult and pediatic patients during both no motion and motion conditions, and for patients who are well or poorly perfused in clinical and non-clinical settings.
## Technology Comparison
The Masimo Rainbow SET® RadCheck Pulse CO-Oximeter (Rad 57 Spotcheck) is substantially equivalent in design, principles of operation, materials, and performance to predicate device (the Rad 57t).
Similar to the Rad 57t, the RadCheck is designed, and manufactured for full compatibility with Masimo Rainbow pulse CO-Oximeter sensors with SpHb capabilities (K080238).
The RadCheck performance is equivalent to those of the Rad 57t, as following:
| FEATURES | SPECIFICATIONS |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Display Ranges | Saturation (SpO2): 0% - 100%<br>Pulse Rate (bpm): 25 - 240 bpm<br>Total Hemoglobin (SpHb): 0-25 g/dl<br>Total Oxygen Concentration (SpOC): 0-35 ml/dl<br>Perfusion Index: 0.02% - 20% |
| Accuracy: SpO2 and Pulse Rate | See Footnotes 1, 2, 3, 4, and 5 |
| Accuracy – SpO2<br>During No Motion Conditions | Adults, Pediatrics: 60% - 80% ± 3%<br>Adults, Pediatrics: 70% - 100% ±2%<br>Adults, Pediatrics: 0% - 69% unspecified |
| Accuracy – SpO2<br>During Motion Conditions | Adults, Pediatrics: 70% - 100% ± 3%<br>Adults, Pediatrics: 0% - 69% unspecified |
| Accuracy – SpO2<br>Low Perfusion | Adults, Pediatrics: 70% - 100% ±2%<br>Adults, Pediatrics: 0% - 69% unspecified |
| Accuracy – Pulse Rate<br>During No Motion Conditions | Adults, Pediatrics: 25 - 240 ± 3 bpm |
| Accuracy – Pulse Rate<br>During Motion Conditions | Adults, Pediatrics: 25 - 240 ± 5 bpm |
| Accuracy – Pulse Rate<br>Low Perfusion | Adults, Pediatrics: 25 - 240 ± 3 bpm |
| FEATURES | SPECIFICATIONS |
| Accuracy: SpHb | |
| Accuracy - SpHb<br>During No Motion Conditions | Adults, Pediatrics: 8 - 17 g/dl +1 g/dl<br>SpHb accuracy has been validated on healthy adult male and<br>female volunteers and on surgical patients with light to dark skin<br>pigmentation in the range of 8 - 17 g/dl SpHb against a<br>laboratory CO-oximeter. This variation equals plus or minus<br>one standard deviation which encompasses 68% of the<br>population. The SpHb accuracy has not been validated with<br>motion or low perfusion. |
| General | |
| Resolution | SpO2: 1%<br>Pulse Rate: 1 bpm<br>SpHb: 0.1 g/dl |
| Measurements | Low Signal IQ<br>Perfusion Index (PI)<br>Total Oxygen Concentration (SpOC) |
| Interfering Substances | Elevated levels of Methemoglobin (MetHb) may lead to<br>inaccurate SpO2 measurements Elevated levels of Carboxyhemoglobin (COHb) may lead to<br>inaccurate SpO2 measurements. Severe anemia may cause erroneous SpO2 readings. Dyes, or any substance containing dyes, that change usual<br>blood pigmentation may cause erroneous readings. Elevated levels of total bilirubin may lead to inaccurate SpO2,<br>and SpHb readings |
| Electrical | See Footnote 6 |
| Batteries | Non-Rechargeable |
| Circuitry | Microprocessor controlled |
| Firmware | Rainbow SET technology, MX-1 Board/Circuitry |
| Mechanical | |
| Material | Polycarbonate/ABS Blend |
| Environmental | |
| Operating Temperature | 41°F to + 104°F (5°C to +40°C) |
| Storage Temperature | -40°F to + 158°F (-40°C to +70°C) |
| Relative Humidity | 5% to 95% noncondensing |
| Operating Altitude | Operating Altitude: 500 mbar to 1,060 mbar pressure; -1,000 ft to<br>18,000 ft (-304 m to 5,486m) |
| Mode & Sensitivity | |
| Averaging Mode - SpO2 | Maximum sensitivity mode fixes perfusion limit to 0.02% |
| Alarms | |
| System | System failure |
| Battery Alarm | Low battery |
| FEATURES | SPECIFICATIONS |
| Display and Indicators | |
| Data Display | SpO2 (%) |
| | Pulse rate (bpm) |
| | SpHb (g/dl) |
| | Perfusion index (%) |
| | SpOC (ml/dl) |
| | Signal IQ |
| | Pulse indicator |
| | Sensor life indicator |
| | Sensor status |
| | Status messages |
| | Battery status |
| Compliance | |
| EMC Compliance | EN 60601-1-2, Class B |
| Electrical Safety | IEC 60601-1, |
| | UL 60601-1 |
| Type of Protection (battery power) | Internally Powered |
| Degree of Protection-Patient Cable | Type BF-Applied Part |
| Enclosed Degree of Ingress | IPX1 |
| Protection from Solids/ Liquids | |
| Mode of Operation | Spot check |
{2}------------------------------------------------
{3}------------------------------------------------
## Footnotes
- SpO2 accuracy was determined by testing on healthy adult volunteers in the range of 1 60-100% SpO2 against a laboratory CO-Oximeter.
- 2 The Masimo sensors have been validated for no motion accuracy in human blood studies on healthy adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory CO-Oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population weight.
- ന The Masimo sensors have been validated for motion accuracy in human blood studies on healthy adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies while performing rubbing and tapping motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory CO-oximeter and ECG monitor. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- 4 The Masimo SET Technology has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator and Masimo's simulator with signal strengths of greater than 0.02% and transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
{4}------------------------------------------------
- The Masimo sensors have been validated for pulse rate accuracy for the range of 5 25-240 bpm in bench top testing against a Biotek Index 2 simulator. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- 6 If the batteries are to be stored for extended periods of time, it is recommended that they be stored between -20 to +30°C, and at a relative humidity less than 85%. If stored for a prolonged period at environmental conditions beyond these limits, overall battery capacity may be diminished, and lifetime of the batteries may be shortened.
## Test Summary
The RadCheck complies with the voluntary standards as detailed in this submission. The following quality assurance measures were applied to the development of the RadCheck:
- . Risk Analysis
- . Design Reviews
- Biocompatibility Testing .
- Performance Testing .
- Safety Testing .
- . Environmental Testing
- . Clinical Testing
### Conclusions
The information in this 510(k) submission demonstrates that the Masimo Rainbow SET® RadCheck Pulse CO-Oximeter and Accessories are substantially equivalent to the predicate device, with respect to safety, effectiveness, and performance.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three overlapping wings, representing the department's mission to protect the health of all Americans. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# OCT 1 0 2008
Ms. Marguerite Thomlinson Manager of Regulatory Affairs Masimo Corporation 40 Parker Irvine, California 92618
Re: K082052
Trade/Device Name: Masimo Rainbow SET RadCheck Pulse CO-Oximeter and Accessories Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: July 16, 2008 Received: July 21, 2008
Dear Ms. Thomlinson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Ms. Thomlison
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Thlammels-Lund, mt for 11
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Masimo Rainbow SET RadCheck Pulse CO-Oximeter and Accessories Device Name:
### Indications For Use:
The Masimo Rainbow SET RadCheck Pulse CO-Oximeter and Accessories are indicated for noninvasive spot checking of functional saturation of arterial hemoglobin (SpO2), pulse rate, and total hemoglobin (SpHb). The Masimo Rainbow SET RadCheck Pulse CO-Oximeter and Accessories are indicated for use, by trained personnel, with adult and pediatric patients during both no motion and motion conditions, and for patients who are well or poorly perfused in clinical and non-clinical settings.
A.R. (MD)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
| Prescription Use<br>X | AND/OR | Over-The-Counter Use |
|--------------------------------|--------|--------------------------------|
| (Per 21 CFR 801.109 Subpart D) | | (Per 21 CFR 801.109 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 10 of 259 Masimo RadCheck 510(k)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.