FINGERTIP PULSE OXIMETER, MODEL M70, AND HANDHELD PULSE OXIMETER, MODEL M700
Applicant
Guangdong Biolight Meditech Co., Ltd.
Product Code
DQA · Cardiovascular
Decision Date
Sep 12, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The M70 fingertip pulse oximeter is intended to measure functional arterial oxygen saturation (SpO2) and pulse rate of adult patients in hospital type facilities as well as in the home care environment. The oximeter is not suitable to monitor patient continuously for long term. The M700 handheld pulse oximeter is indicated for spot checking of functional arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric and neonatal patients in hospital, hospital type facilities as well as in the home care environment.
Device Story
M70 (fingertip) and M700 (handheld) pulse oximeters measure functional arterial oxygen saturation (SpO2) and pulse rate. Devices use spectrophotometry based on double-wavelength Lambert-Beer's Law; detect specific absorbance of oxygenated hemoglobin and deoxyhemoglobin. M70 is a self-contained fingertip unit with LED emitter/detector and OLED display. M700 is a handheld unit with a main frame (control/SpO2 modules) and detachable sensors (adult, pediatric, neonatal). Operated by clinicians or patients in hospitals or home settings. Output includes %SpO2, pulse rate, and waveform pulse amplitude bar. Data assists in spot-check clinical assessments; not for continuous monitoring or life support.
Clinical Evidence
Laboratory and clinical testing conducted. Bench testing verified electrical safety, EMC, and biocompatibility compliance. Clinical testing performed for M700 to validate design specifications across adult, pediatric, and neonatal populations.
Technological Characteristics
Spectrophotometric pulse oximetry using double-wavelength LED emitters. M70: Fingertip form factor, OLED display, battery-powered. M700: Handheld, detachable sensors, control/SpO2 modules. Non-sterile, non-implantable. Prescription use.
Indications for Use
Indicated for spot-checking SpO2 and pulse rate. M70: Adult patients in hospital or home care. M700: Adult, pediatric, and neonatal patients in hospital or home care. Not for continuous long-term monitoring.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a sequence of handwritten digits and letters. The sequence starts with the letters 'Ko', followed by the numbers '81712'. The characters are written in a dark ink, and the image has a white background.
Premarket Notification 510(k) Chapter III 510(k) Summary Report No : A2008-002-050
# Chapter III 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
The Assigned 510(k) Number is:
#### 1. Sponsor
Guangdong Biolight Meditech Co., Ltd Innovation First Road, Technology Innovation Coast Zhuhai, Guangdong, 519085, China Contact Person: Mr. Tianbao Li, Chief Engineer +86-756-3399963 Tel: Fax: +86-756-3399989 E-mail: li tb@blt.com.cn
#### 2. Submission Correspondent
Ms. Diana Hong Mr. Lee Fu Suite 8D, Zhongxin Zhongshan Mansion, No.19, Lane 999, Zhong Shan Nan Er Road Shanghai, China 20020 Phone: +86-21-64264467 Fax: 760-466-5084 Email: Diana.hong@mid-link.net
#### Subject Device Information 3.
#### M70 Fingertip Pulse Oximeter a.
(1)Classification Name: Oximeter
(2)Regulation Number:870.2700
(3)Product Code:DQA
(4)Class: II
(5)Review Panel: Anesthesiology
#### M700 Handheld Pulse Oximeter b.
(1)Classification Name: Oximeter
{1}------------------------------------------------
(2)Regulation Number:870.2700 (3)Product Code: DQA (4)Class: II (5)Review Panel: Anesthesiology
#### Predicate Device 4.
- a. Fingertip Oximeter MD300C
K-number: K070371 Product Code: DQA
### Intended Use:
Fingertip Pulse Oximeter MD300C is a portable non-invasive, spot-check, oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care and etc). Not for continuously monitoring.
### Manufactured by:
Beijing Choice Electronic Technology Co., Ltd. Room 1127-1128 Building B, Bailangyuan Fuxing Road , No. A36 Beijing, CHINA 100039
### b. PM-60 Pulse Oximeter
K-number: K072581
Product Code: DQA
### Intended Use:
The PM-60 handheld Pulse Oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult, pediatric and neonatal patients in hospitals, out-of-hospital transport and home care.
### Manufactured by:
Shenzhen Mindray Bio-medical Electronics Co., LTD. Mindray Building, Keji 12th Road South, Hi-tech Industrial Park Nanshan, Shenzhen, 518057, P.R.China
{2}------------------------------------------------
#### 5. M70 Fingertin Pulse Oximeter
#### 5.1 Device Description
The subject device of M70 Fingertip Pulse Oximeter is a fingertip device, which can display %SpO2, pulse rate value, waveform pulse amplitude bar indication.
The subject device consists of detector and emitter LED, OLED display module, CPU, driving circuit, power supply circuit and battery.
The operational principle of subject device uses the method as mensurated optical density by spectrophotometer. It is base on the double wavelength Lambert Beer's Law and uses the specific absorbance of oxygenation hemoglobin and deoxyhemoglobin to calculate.
The subject device is not for life-supporting or life-sustaining, not for implant. The device is not sterile and does not need sterilization or re-sterilization. The device is for prescription. The device does contain drug or biological product.
#### 5.2 Statement of Intended Use:
The M70 fingertip pulse oximeter is intended to measure functional arterial oxygen saturation (SpO2) and pulse rate of adult patients in hospital type facilities as well as in the home care environment.
The oximeter is not suitable to monitor patient continuously for long term.
#### 5.3 Testing
Laboratory testing was conducted to validate and verify that M70 Fingertip Pulse Oximeter met all design specifications, including electrical safety, EMC, biocompatibility, specification. Results of these tests demonstrate compliance to the requirements of all consensus standards
#### 5.4 Substantially Equivalence Determination
#### Comparison Analysis
The subject device has same classification information, same indications and intended use, same design principle, similar product design and specifications, same performance effectiveness, performance safety as the predicate device. The only difference is Measurement Wavelengths, Working time, Operating temperature, PR Measuring range, Atmosphere pressure.
{3}------------------------------------------------
#### Conclusion:
The subject device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the subject device is determined as safe and effectiveness.
#### M700 Handheld Pulse Oximeter 6.
#### 6.1 Device Description
The subject device of M700 Handheld Pulse Oximeter is a handheld device, whose mainly function are display %SpO2, pulse rate value, pulse amplitude bar indication.
The operational principle of subject device uses the method as mensurated optical density by spectrophotometer. It is base on the double wavelength Lambert Beer's law and uses the specific absorbance of oxygenation hemoglobin and deoxyhemoglobin to calculate.
The subject device consists of detachable SpO2 sensor, Main Frame and Battery. The Main Frame consist two parts, which is Control Module and SpO2 Module.
The subject device has 6 model sensors as the accessory, they are listed below:
A0212-SA103PV (for Adult) A0212-SP103PV (for Pediatric) A0212-SW103PU (for Neonate) RSJ091DA (for Adult) RSJ091DI (for Pedicatio) RSY091DN (for Neonate)
### 6.2 Statement of Intended Use
The M700 handheld puls oximeter is indicated for spot checking of functional arterial oxygen saturation (SpO2) will pulse rate of adult, pediatric and neonatal patients in hospital, hospital type facilities as in the home care environment.
#### 6.3 Testing
Laboratory and Clinical insting was conducted to validate and verify that M700 Handheld Pulse Oximeter met all design specifications, including electrical safety, EMC, biocompatibility, specific n. Results of these tests demonstrate compliance to the requirements of all consensus standards
{4}------------------------------------------------
# 6.4 Substantially Equivalence Determination
### Comparison Analysis
The subject device has same classification information, same indications and intended use, same design principle, similar product design and specifications, same performance effectiveness, performane safety as the predicate device. The only difference is Measurement Wavelengths, Display Unit, Power, Working time, Operating temperature, PR Measuring range, Relative humidity, Atmosphere pressure.
# Conclusion:
The subject device is Sul clantially Equivalent (SE) to the predicate device which is US legally market device. The subject device is determined as safe and effectiveness.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features an abstract eagle design with three stylized wing segments. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
# SEP 1 2 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Guangdong Biolight Meditech Company, Limited C/O Ms. Diana Hong General Manager Shanghai Mid-Link Business Consulting Company, Limited Suite 8D, Zhongshan Zhongxin Masion No.19, Lane 999, Zhongshan No.2 Road(S) Shanghai, 200030 CHINA
Re: K081712
Trade/Device Name: M700 Handheld Pulse Oximeter M70 Fingertip Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: June 13, 2008 Received: June 17, 2008
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Ms. Hong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of vour device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Thlmelt Kirdx>
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{7}------------------------------------------------
# Indication for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: M70 Fingertip Pulse Oximeter_____________________________________________________________________________________________________________________________________
Indications for Use:
The M70 fingertip pulse oximeter is intended to measure functional arterial oxygen saturation (SpO2) and pulse rate of adult patients in hospital type facilities as well as in the home care environment.
The oximcter is not suitable to monitor patient continuously for long term.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
hul
Page _1_ of __2
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K081712
{8}------------------------------------------------
Premarket Notification 510(k) Chapter II Indication for Use Statement Report No.: A2008-002-050 Comments of the country
# Indication For Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: __ M700 Handheld Pulse Oximeter
Indications for Use:
The M700 handheld pulse oximeter is indicated for spot checking of functional arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric and neonatal patients in hospital, hospital type facilities as well as in the home care environment.
Prescription Use __ J ________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
:
AND/OR
Over-The-Counter Use __________ (21 CFR 801 Subpart C)
and the contract and the country of the
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
and the country of the country of the country of the county of the county of
Concurrence of CDRH, Office of Device Evaluation (ODE)
e
Page _2_ of _ 2_
(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital Division Control, Dental Devices
510(k) Number: K081712