K031740 · Welch Allyn Protocol, Inc. · DQA · Aug 26, 2003 · Cardiovascular
Device Facts
Record ID
K031740
Device Name
VSM - VITAL SIGNS MONITOR, MODEL 53000 SERIES
Applicant
Welch Allyn Protocol, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Aug 26, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The model 53000 Series of monitors are intended to be used by clinicians and medically qualified personnel for monitoring of noninvasive blood pressure, pulse rate, body temperature, noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and body temperature in normal and axillary modes of neonatal, pediatric and adult patients. The most likely locations for patients to be monitored are general med/surg. floors, general hospital and alternate care environments. This device is available for sale only upon the order of a physician or licensed health care professional.
Device Story
Vital Signs Monitor 53000 Series provides multi-parameter patient monitoring; inputs include noninvasive blood pressure (NIBP), pulse rate, SpO2, and body temperature. Device integrates established measurement modules; temperature module is identical to SureTemp Plus. Used by clinicians/medically qualified personnel in general hospital/med/surg floors. Output displayed for real-time clinical assessment; assists providers in patient status evaluation. Benefits include consolidated vital sign tracking for neonatal, pediatric, and adult populations.
Clinical Evidence
Bench testing only. Performance verified per Test Plan/Report #831-0719-00 and Clinical Validation Report #831-0752-00 using production-equivalent units. Risk analysis and mitigation documented per product development procedures. Quality system conforms to 21CFR820, ISO 9001, EN 46001, and ISO 13485.
Technological Characteristics
Multi-parameter monitor for NIBP, SpO2, pulse, and temperature. Integrates SureTemp Plus temperature module. Connectivity and specific software architecture not detailed. Quality system certified to ISO 9001, EN 46001, ISO 13485.
Indications for Use
Indicated for neonatal, pediatric, and adult patients requiring monitoring of noninvasive blood pressure, pulse rate, body temperature, and functional oxygen saturation (SpO2). Used by clinicians in general hospital, med/surg floors, and alternate care settings. Prescription use only.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Welch Allyn CVSM vital signs monitors model 52STP (K951193)
Protocol Systems Propaq Encore 200 Series (K951246)
{1}------------------------------------------------
AUG 26 2003
K031740
i i
## 510(k) Summary of Safety and Effectiveness
May 30, 2003
#### Submitter
Welch Allyn Protocol. Inc. 8500 S.W. Creekside Place Beaverton, OR 97008-7107 USA
Telephone: (503) 530-7500 Fax: (503) 526-4901
Contact: Mr. James W. Sandberg, Director of Regulatory Affairs & Compliance Engineering
#### Device Name
Vital Signs Monitor, Model 53000 Series Trade Name: Common Name: Vital Signs Measurement Device
Classification Name: Noninvasive Blood Pressure Measurement System, 21CFR870.1130, April 1.2002. The VSM model 53000 Series also offers a Pulse Oximetry channel (SpO2) - reference 21CFR870-2700, April 1, 2002, and a patient temperature channel, Clinical electronic thermometer, reference 21CFR880.2910, April 1, 2002.
Classification: Class II
#### Predicate Devices
The predicate devices for the model 53000 Series monitors are the
- Welch Allyn CVSM vital signs monitors model 52STP, K951193, and .
- Protocol Systems Propaq Encore 200 Series, K951246, and .
- Nellcor Puritan Bennett Pulse Oximeter module MP506 cleared with Pulse Oximeter . Model N-550 K021090, and
- Welch Allyn SureTemp™ Plus, K030580.
#### Device Description
The model 53000 Series of monitors are intended to be used by clinicians and medically qualified personnel for monitoring of noninvasive blood pressure, pulse rate, body temperature, noninvasive functional oxygen saturation of arteriolar hemoglobin (SpOz), and body temperature in normal and axillary modes of neonatal, pediatric and adult patients.
The most likely locations for patients to be monitored are general med/surg. floors, general hospital and alternate care environments. This device is available for sale only upon the order of a physician or licensed health care professional.
{2}------------------------------------------------
#### Indications for Use
The model 53000 Series of monitors are intended to be used by clinicians and medically qualified personnel for monitoring of noninvasive blood pressure, pulse rate, body temperature, noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO₂), and body temperature in normal and axillary modes of neonatal, pediatric and adult patients.
The most likely locations for patients to be monitored are general med/surg. floors, general hospital and alternate care environments. This device is available for sale only upon the order of a physician or licensed health care professional.
#### Technological Comparison to the Predicate Devices
It is Welch Allyn Protocol's conclusion that the vital signs monitors model 53000 Series are substantially equivalent to the Welch Allyn Tycos vital signs monitor CVSM Model 52SP-E1, the Protocol Systems Propaq Encore 200 Series monitors, the Nellcor Puritan Bennett Pulse Oximeter, Model OxiMAX N-550, and the Welch Allyn Clinical electronic thermometer, Model SureTemp™ Plus.
- . The noninvasive blood pressure measurement specifications and performance are equivalent to the Welch Allyn Tycos CVSM monitor and the Protocol Systems Propaq Model 200 Series monitors, and
- The pulse rate specifications and performance derived from either noninvasive blood . pressure or SpO2 are equivalent to the Welch Allyn Tycos CVSM monitor, the Protocol Systems Propaq, Model 200 Series monitors, and the Nellcor Puritan Bennett Pulse Oximeter Model N-550, and
- . The Sp0, specifications and performance are equivalent to the Nellcor Puritan Bennett Model N-550 Pulse Oximeter, and
- . The Welch Allyn Clinical temperature meter SureTemp™ Plus is the same temperature module used in the Welch Allyn model 53000 Series monitors.
#### Summary of Performance Testing
The Welch Allyn vital signs monitors, model 53000 Series will be tested in accordance with the Test Plan / Report, #831-0719-00, and Clinical Validation Report #831-0752-00 included with the submission using production equivalent units prior to release to market.
A risk analysis identifying potential hazards and documenting mitigation of the hazards has been developed and applied as part of Welch Allyn Protocol's product development procedure. Welch Allyn Protocol's Quality System conforms to 21CFR820 and is certified by TÜV Product Service to ISO 9001, EN 46001 and ISO 13485.
#### Conclusions
As stated above, Welch Allyn Protocol's conclusion is that the vital signs monitor Model 53000 Series are safe and effective and comply with the appropriate medical device standards and are substantially equivalent to the earlier identified predicate devices.
#### This 510(k) Summary of Safety and Effectiveness may be copied and submitted to interested parties as required by 21CFR807.92.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with text around the perimeter. Inside the circle is a stylized image of an eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 6 2003
Welch Allyn Protocol, Inc. c/o Mr. James W. Sandberg Director of Regulatory Affairs & Compliance Engineering 8500 S.W. Creekside Place Beaverton, OR 97008-7107
Re: K031740
Trade Name: VSM Vital Signs Monitor, Model 53000 Series Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II (two) Product Code: DOA Dated: May 30, 2003 Received: June 4, 2003
Dear Mr. Sandberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. James W. Sandberg
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D. Luckerman, M.D. Director Division of Cardiovascular Device Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 3/740
# INDICATIONS FOR USE
### Applicant:
Welch Allyn Protocol, Inc. 8500 S.W. Creekside Place Beaverton, OR 97008-7107 USA
Telephone: (503) 530-7500 Fax: (503) 526-4901
510(k) Number:
Device Name: Vital signs monitors, model 53000 Series
#### Indications for Use:
The VSM series of monitors are intended to be used by clinicians and medically qualified personnel for monitoring of noninvasive blood pressure, pulse rate, body temperature. noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and body temperature in normal and axillary modes of neonatal, pediatric and adult patients.
The most likely locations for patients to be monitored are general med/surg, floors, general hospital and alternate care environments. This device is available for sale only upon the order of a physician or licensed health care professional.
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | ✓ |
|----------------------|---|
| OR Over-The-Counter | |
| (Per 21 CFR 801.109) | |
(Division Sign-Off)
Division of Cardiovascul.
| 510(k) Number | K031740 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.