PTA Balloon Dilatation Catheter

K221245 · Kossel Medtech (Suzhou) Co., Ltd. · LIT · Dec 14, 2022 · Cardiovascular

Device Facts

Record IDK221245
Device NamePTA Balloon Dilatation Catheter
ApplicantKossel Medtech (Suzhou) Co., Ltd.
Product CodeLIT · Cardiovascular
Decision DateDec 14, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Intended Use

The PTA balloon dilatation catheter is indicated for the percutaneous transluminal angioplasty (PTA) of the peripheral vascular system, including the iliac, femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Device Story

PTA balloon dilatation catheter; OTW (over-the-wire) design. Components: tip tube, inner tube, balloon, marker bands, outer tube, stress release tube, hub. Hydrophilic coating on distal portion. Used by physicians in clinical settings for peripheral vascular interventions. Input: guide wire passage through inner tube; inflation medium via hub. Operation: catheter advanced over guide wire to stenosis; balloon inflated to specific pressure to dilate vessel. Marker bands allow fluoroscopic positioning. Output: mechanical dilation of stenotic lesions. Benefits: restores vessel patency; improves blood flow in peripheral arteries and dialysis fistulae.

Clinical Evidence

Bench testing only. Testing included visual inspection, dimensional verification, hydration, leakage, inflation/deflation time, fatigue, radiopacity, simulated use, bond/torque/pull strength, burst pressure, compliance, flexibility/kinking, coating integrity, particulate evaluation, flow rate, corrosion, and packaging/sterility. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, reactivity, systemic toxicity, hemocompatibility, pyrogenicity). All results met acceptance criteria.

Technological Characteristics

Nylon12 balloon; hydrophilic coating; OTW catheter type; 0.018" guidewire compatibility; 6F sheath compatibility; semi-compliant balloon; EO sterilization. Dimensions: 90/150cm length, 2.0-7.0mm diameter. Pressure ratings vary by size (10-16 atm RBP).

Indications for Use

Indicated for percutaneous transluminal angioplasty (PTA) of peripheral vascular system (iliac, femoral, popliteal, infra-popliteal, renal arteries) and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 14, 2022 Kossel Medtech (Suzhou) Co., Ltd. Ron Lv Regulatory Affair Engineer BLDG 6, No. 8, Jinfeng Road Suzhou, Jiangsu Province 215163 China ## Re: K221245 Trade/Device Name: PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: November 1, 2022 Received: November 7, 2022 ### Dear Ron Lv: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Digitally signed by Gregory W. Gregory W. O'connell - S O'connell -S Date: 2022.12.14 13:02:23 -05'00' Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K221245 Device Name PTA Balloon Dilatation Catheter #### Indications for Use (Describe) The PTA balloon dilatation catheter is indicated for the percutaneous transluminal angioplasty (PTA) of the peripheral vascular system, including the iliac, femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Type of Use (Select one or both, as applicable) | <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | |-------------------------------------------------------------| | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY | Submitter: | Kossel Medtech (Suzhou) Co., Ltd.<br>BLDG 6, No. 8 Jinfeng Road<br>Suzhou New District, Jiangsu Province, China | |---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Ron Lv<br>F3, BLDG 6, No. 8 Jinfeng Road<br>Suzhou New District,<br>Jiangsu Province, China<br>Phone: +86 150 5142 9102<br>Mail:lvxin@kosselmed.com | | Date Prepared: | November 22, 2022 | | Trade Name: | PTA Balloon Dilatation Catheter | | Common Name: | Percutaneous Transluminal Angioplasty (PTA) Catheter | | | Classification Name: Catheter, Angioplasty, Peripheral, Transluminal (21 CFR 870.1250), Class II<br>(special controls) | | Product Code: | LIT | | Predicate Device: | Predicate Device:<br>SABER .018 Percutaneous Transluminal Angioplasty (PTA) Dilatation<br>Catheter (K133843, cleared May 19, 2014) | | Device Description: | The 0.0180" PTA balloon dilatation catheter is mainly composed of tip tube,<br>inner tube, balloon, marker bands, outer tube, stress release tube and hub.<br>Among them the balloon is the most important part of the catheter. In order to<br>dilate different stenosis, the balloon should be dilated to difference dimension<br>by inflating different pressure. The soft tip at the end of the balloon is to make<br>the balloon catheter more easily to push to the stenosis position. The inner<br>tubing which connects to the tip tubing is for guide wire passage and the<br>pushing road. The two marker bands which wrapping on the inner tubing are<br>for positioning the balloon location by cooperating in vitro monitoring<br>equipment. The distal of the outer tube is connected with the balloon by a<br>variable-diameter stretching structure, which can make the balloon get smaller<br>winding diameter. The proximal of the outer tube is connected with the hub as<br>the inflation passage and the pushing shaft. The hub is for connecting the outer<br>pressure filling equipment. The stress release tubing is for protecting the outer<br>tube from bending.<br>The outer surface of the catheter is coated with a hydrophilic coating that ranges<br>400mm in length from the tip tube to the proximal end of the balloon, including<br>part of the outer tube. | {4}------------------------------------------------ Image /page/4/Figure/1 description: The image shows a diagram of a medical device, specifically a balloon catheter. The different parts of the catheter are labeled, including the tip tube, balloon tube, outer tube, hub, stylet, marker band, stress release, and inner tube. The catheter has a long, thin shape with a balloon at the end, and the hub is located at the opposite end. Intended Use/ The PTA balloon dilatation catheter is indicated for the percutaneous Indications transluminal angioplasty (PTA) of the peripheral vascular system, including the for Use: iliac, femoral, ilio-femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. - Comparisons of the new and predicate devices show that the technological Technological Characteristics: characteristics such as product performance, design, and intended use are substantially equivalent to the currently marketed predicate devices. The following table lists the material and size comparisons between the subject device and the predicate device. | Technological<br>Characteristics | PTA Balloon Dilatation Catheter | Predicate Device<br>SABER® PTA Dilatation Catheter<br>K133843 | Comparison | |----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Balloon material | Nylon12 | Nylon12 | Identical | | Coating | Hydrophilic | Hydrophilic | Identical | | Balloon Diameters<br>(mm) | 2.0~7.0 | 2.0~10.0 | Equivalent | | Balloon lengths<br>(mm) | 20~300 | 20~300 | Identical | | Balloon compliance | Semi-Compliant | Semi-Compliant | Identical | | Catheter Type | OTW | OTW | Identical | | Guide wire size (inch) | 0.018" | 0.018" | Identical | | Sheath Compatibility | 6F | 6F | Identical | | Proximal shaft diameter | 4.0F(1.32mm)/4.7F(1.55mm) | 3.9F/4.7F | Equivalent | | Distal shaft diameter | 4.0F(1.32mm)/4.7F(1.55mm) | 3.9F/4.7F | Equivalent | | Catheter length<br>(cm) | 90/150 | 90/150 | Identical | | Rated burst Pressure<br>(atm) | 16atm (ø2.0-2.5×20-120mm);<br>14atm (ø2.0-2.5×150-300mm,<br>ø5.0-6.0×20-120mm);<br>12atm (ø5.0-6.0×150-300mm, ø7.0×20-120mm);<br>10atm (ø7.0×150-300mm) | 18atm (ø2.0-2.5mm, ø3.0-4.0×20-100mm);<br>16atm(ø5.0-6.0×20-100mm);<br>14atm (ø7.0mm, ø3.0-6.0×150-300mm);<br>12atm (ø8.0mm);<br>10atm (ø9.0-10.0mm) | Equivalent | The following table lists the material and device and the predicate device. {5}------------------------------------------------ | Technological<br>Characteristics | PTA Balloon Dilatation Catheter | Predicate Device | | Comparison | |----------------------------------|---------------------------------|-------------------------------------------|--|------------| | | | SABER® PTA Dilatation Catheter<br>K133843 | | | | Nominal pressure<br>(atm) | 6 | 6/8 | | Equivalent | | Sterilization Method | EO | EO | | Identical | The PTA Balloon Dilatation Catheter and predicate device have following same technological elements: - Operating principle balloon dilatation of stenotic portion by pressurization of inflation ● medium - Fundamental catheter design balloon, shaft, radiopaque marker, hub, coating - Balloon Length - Balloon compliance - Semi-Compliant - Catheter type - OTW - Guide wire size - Sheath Compatibility - Catheter effective length - Sterilization Ethylene oxide There are following minor technological differences between PTA Balloon Dilatation Catheter and predicate device: - Combination of balloon diameter - Proximal/Distal shaft diameters - Pressure - Nominal Pressure and Rated Burst Pressure Any differences that exist between the PTA Balloon Dilatation Catheter and the predicate device did not raise any new questions of safety or effectiveness. {6}------------------------------------------------ Performance Data: Bench testing was performed on the subject device to determine substantial equivalence. The testing performed is as follows: - Visual inspection - Dimensional verification - Hydration test - Leakage test - Balloon inflation and deflation time - . Balloon fatigue - Radiopacity - Simulated use test/guidewire and introducer sheath compatibility test - . . . . . Tip pull strength - Catheter bond strength - Torque strength - Balloon rated burst pressure - Balloon compliance - Catheter body burst pressure - . . . . Flexibility and kinking - Coating integrity - Particulate evaluation - Infusion flow rate - Corrosion resistance test - Packaging performance Test (Visual inspection, Sealing strength, - Sealing integrity, Sterility) - Shelf life testing ● The results met all acceptance criteria and ensure that the Balloon Dilatation Catheter design and construction are suitable for its intended use as recommended by the draft Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions (FDA; January 13, 2020). - Biocompatibility To demonstrate the biological safety of the body-contacting materials, the following biocompatibility testing, which was performed in accordance with 'Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk mangement process'; draft 'Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions' were followed. Cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemocompatibility (hemolysis, partial thromboplastin time, complement activation, and in vivo thromboresistance), materialmediated pyrogenicity were conducted. The results of the testing show that the subject device included in this submission met all acceptance criteria and the subject device is biocompatible. - Conclusion: This information supports a determination of substantial equivalence between the PTA Balloon Dilatation Catheter and the predicate device described above.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...