VECTOR PTA BALLOON DILATATION CATHETER

K131329 · R4 Vascular, Inc. · LIT · Jun 6, 2013 · Cardiovascular

Device Facts

Record IDK131329
Device NameVECTOR PTA BALLOON DILATATION CATHETER
ApplicantR4 Vascular, Inc.
Product CodeLIT · Cardiovascular
Decision DateJun 6, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Vector4 TM PTA Balloon Dilatation Catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, popliteal, and renal arteries and for the treatment of obstructive lesions of native or synthetic arterio-venous dialysis fistulae. The Vector Catheter is not indicated for use in the coronary arteries or for placement of or post dilatation of stents.

Device Story

Vector4 PTA Balloon Dilatation Catheter is a high-pressure, non-compliant, coaxial dual-lumen over-the-wire catheter. Features radio-opaque coating on balloon working length for fluoroscopic visualization without contrast; Parylene C coating for lubricity. Used by physicians in clinical settings for peripheral artery angioplasty and dialysis fistula lesion treatment. Input: guide wire (up to 0.035 inch). Operation: catheter advanced to stenosis; balloon inflated via outer lumen (saline or contrast) to dilate lesion. Output: mechanical dilation of vessel/fistula. Benefits: improved visualization during placement/inflation; reduced contrast requirement; atraumatic tip for easier advancement.

Clinical Evidence

Bench testing only. Performance testing included rated burst pressure, compliance, diameter sizing, inflation/deflation times, repeat inflation, removal/refold/reuse, crossing profile, coating integrity/particulates, balloon protector, guide wire/introducer compatibility, and balloon burst. Testing performed before and after accelerated aging. All devices met specifications.

Technological Characteristics

Coaxial dual-lumen over-the-wire catheter; high-pressure composite balloon; Parylene C coating; radio-opaque balloon coating. Compatible with 0.035" guide wires and 6-7F introducers. Rated burst pressure 30 atm. Biocompatible, latex-free materials. Sterilized via ethylene oxide (SAL 10^-6).

Indications for Use

Indicated for patients requiring Percutaneous Transluminal Angioplasty (PTA) of femoral, iliac, popliteal, and renal arteries, or treatment of obstructive lesions in native or synthetic arterio-venous dialysis fistulae. Contraindicated for coronary artery use and stent placement or post-dilatation.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### 510(k) Summary JUN 0 6 2013 #### Primary Contact Person Laurie Lewandowski, Consultant Honkanen Consulting, Inc. 738 Saddle Wood Drive Eagan, MN 55123 Phone: 612-770-4038 Fax: N/A (763-494-8484) Email: lalew207@gmail.com #### Sponsor's Contact Steve Allex r4 Vascular, Inc. 7550 Meridian Circle North, Suite 150 Maple Grove, MN 55369 Phone: 763-494-8400 x 20 Fax: 763-494-8484 Email: sallex@r4vascular.com #### Device Name Trade Name Vector4TM Common Name Percutaneous Catheter Classification Name Percutaneous Catheter Classification II Product Code: LIT #### Predicate Device The device is substantially equivalent to the r4 Vascular Vector. "PTA Balloon Dilatation Catheter (K 121385). #### Indications for Use/Intended Use The Vector 4 " PTA Balloon Dilatation Catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, popliteal, and renal arteries and for the treatment of obstructive lesions of native or synthetic arterio-venous dialysis fistulae. The Vector Catheter is not indicated for use in the coronary arteries or for placement of or post dilatation of stents. #### Device Description The Vector" PTA Balloon Catheter is a high pressure, non-compliant balloon catheter with a radio-opaque coating on the balloon. The radio-opaque coating on the working length of Vector 4 PTA Balloon Dilatation Catheter ર-દ {1}------------------------------------------------ the balloon minimizes or eliminates the need for using contrast solution to inflate the balloon. The radio-opaque coating delineates the working length of the balloon and provides continuous fluoroscopic visibility during balloon placement. Saline may be used to inflate the balloon without affecting visualization. Contrast mixture may be used at the discretion of · the physician. #### Technological Characteristics The Vector !TM P.TA Balloon Dilatation Catheter was cleared under K121385. The fundamental scientific technology and materials for the Vector" PTA Balloon Catheter remain the same. It is a coaxial dual lumen over the wire balloon dilatation catheter. The central lumen, accommodates guide wires up to 0.035 inches in diameter while the outer lumen is the inflation'lument or the balloon. The catheter includes an atraumatic tip to ease advancement of the catheter, to and through the stenosis. The distal tip is visible under fluoroscopy. The balloon is a high pressure composite balloon with a radio-opaque coating on the working length to help visualize placement within the lesion during both placement and inflation; without the use of contrast. The balloon's rated burst pressure is 30 atm. The catheter is supplied sterile and non-pyrogenic. Different balloon sizes (5 mm to 10 mm) and catheter lengths (50 to 135 cm) are available. The proximal end of the catheter contains a hub with two female luer locks. The device is compatible with 6-7F introducers. The introducer compatibility for each device size is listed on the device label itself. The materials used in the construction are biocompatible and latex free. The modifications to the Vector TM PTA Balloon Dilatation Catherer are as follows. - 1) The balloon will be coated with Parylene C coating, referred to throughout as Parylene, providing a natural lubricity for insertion and withdrawal through the introducer. - 2) Each balloon catheter hub will be laser marked to include the balloon and introducer size, rated burst pressure and the product name, Vector" PTA. Balloon Catheter for customer convenience. - 3) The balloon protector ID for the 5x40mm balloon was increased slightly. There are no changes in the balloon catheter dimensions or significant changes in weight with these modifications .. #### Performance Data Based on the risk analysis of the modified device, and testing was conducted on the modified Vector" PTA Balloon Dilatation Catheter. The testing performed on the modified devices was: Vector 4" PTA Balloon Dilatation Catheter {2}------------------------------------------------ - Balloon rated burst pressure - Balloon compliance . - Balloon Diameter Sizing - Inflation/deflation times � - Repeat inflation . - Removal, Refold and Reuse . - . Crossing Profile - Coating integrity/Particulates - Balloon Protector . - Guide Wire and Introducer Compatibility . - Balloon Burst . - Aged Balloon Testing: Rated Burst Pressure, Balloon Compliance, Balloon Diameter Sizing, Inflation/deflation times, Repeat Inflation, Removal/Refold & Reuse, Flexibility / Kink, Crossing Profile, Particulate/Coating Integrity, Balloon Protector, Guide Wire/Introducer Compatibility and Balloon Burst All devices met the performance specifications. The device is sterilized by ethylene oxide to an SAL 10- level and the device is biocompatible meeting the requirements of ISO 10993-1. After accelerated aging of the device, a sub-set of the above listed tests that may be affected by aging were repeated !. The bench testing demonstrated that the device meets specifications before and after aging indicating that the device is as safe and effective as the predicate device. #### Substantial Equivalence The modified Vector " PTA Balloon Catheter is substantially equivalent to the Vector "" P.TA catheter (K121385). They have the identical intended use, and treat the same target. population. Both devices are intended to treat peripheral arteries. The manner in accessing and treating lesions in all these arteries is identical. The devices are identical in design with the exception of the addition of Parylene coating, the laser marked hub and a slight increase in the ID of the balloon protector. #### Conclusion The modified Vector 4 PTA Balloon Catheters and the predicate Vector 4 PTA Balloon Catheters have identical intended use and similar technological/characteristics and are therefore substantially equivalent. The bench testing of the Vector "" PTA Balloon Dilatation Catheters met all the product specifications. These tests demonstrate that the Viector PTA Balloon Dilatation Catheters are as safe and effective as the predicate device. {3}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the eagle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 6, 2013 r4 Vascular, Inc. c/o Ms. Laurie Lewandowski, Consultant Honkanen Consulting, Inc. 738 Saddle Wood Drive Eagan, MN 55123 Re: K131329 Trade/Device Name: Vector 4TM PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: May 7, 2013 Received: May 8, 2013 Dear Ms. Lewandowski: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ #### Page 2 - Ms. Laurie Lewandowski forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation' (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # Bram D. Zuckerman -S Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT K 131329 510(k) Number (if known): Device Name: Vector4 TM PTA Balloon Dilatation Catheter #### Indications for Use: The Vector4 TM PTA Balloon Dilatation Catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, popliteal, and renal arteries and for the treatment of obstructive lesions of native or synthetic arterio-venous dialysis fistulae. The Vector Catheter is not indicated for use in the coronary arteries or for placement of or post dilation of stents. Prescription Use: X AND/OR Over-the-Counter Use: Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) 2-1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...