FLASH PTA BALLON DILATATION CATHETER-5MM X 19MM X 135CM

K121175 · Ostial Corporation · LIT · Jun 29, 2012 · Cardiovascular

Device Facts

Record IDK121175
Device NameFLASH PTA BALLON DILATATION CATHETER-5MM X 19MM X 135CM
ApplicantOstial Corporation
Product CodeLIT · Cardiovascular
Decision DateJun 29, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Flash PTA Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature at aorto-ostial locations, including iliac, renal and carotid arteries. This device is also indicated for post-dilatation of balloon expandable stents in the peripheral vasculature.

Device Story

Flash PTA Balloon Dilatation Catheter is a .014" guidewire-compatible, rapid exchange (RX) angioplasty catheter (135cm length) for stenotic ostial lesions. Features dual-balloon design: compliant anchoring balloon prevents distal migration; semicompliant higher-pressure balloon performs luminal dilatation and post-stent expansion. Used in clinical settings by physicians for peripheral vascular interventions. Output is mechanical dilatation of vessels/stents; provider monitors via fluoroscopy. Benefits include stable positioning during angioplasty and effective lesion/stent management.

Clinical Evidence

Bench testing only. No clinical data. Performance verified via in-vitro testing including balloon crossing profile, burst pressure, compliance, inflation/deflation times, fatigue (with/without stent), bond/tip pull/torque strength, and radiopacity. Biocompatibility testing per ISO-10993 (MEM elution, hemolysis, complement activation, thromboresistance, pyrogenicity, sensitization, systemic injection, intracutaneous reactivity).

Technological Characteristics

Rapid exchange (RX) catheter; .014" guidewire compatible; 135cm length. Dual-balloon system: compliant anchoring balloon and semicompliant high-pressure dilatation balloon. Materials evaluated per ISO-10993. Mechanical actuation via fluid inflation. Radiopaque markers for visualization.

Indications for Use

Indicated for patients requiring percutaneous transluminal angioplasty in peripheral vasculature at aorto-ostial locations (iliac, renal, carotid arteries) and post-dilatation of balloon-expandable stents.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### SECTION 5.0: 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS A. Submitter Information Submitter's Name: Address: > Telephone: Fax: Contact Person: Date of Preparation: B. Subject Device Trade Name: Common/Usual Name: Classification Name: C. Predicate Device Name(s) # Primary Predicate: Trade Name(s): Classification Name: Secondary Predicate: Trade Name(s): Classification Name: Ostial Corporation 510 Clyde Avenue Mountain View, CA 94043 650-903-9100 x 232 650-903-9119 Kaitlin von Hoffmann Clinical and Regulatory Associate April 17, 2012 Flash PTA Balloon Dilatation Catheter Balloon Catheter Catheter, Angioplasty, Peripheral, Transluminal (21 CFR 870.1250, Product Code LIT) Sterling PTA Balloon Dilation Catheter, K053118 Catheter, Percutaneous (21 CFR 870.1250, Product Code DQY) Flash PTA Balloon Dilatation Catheter, K120738 Catheter, Angioplasty, Peripheral, Transluminal (21 CFR 870.1250, Product Code LIT) ### D. Device Description: The Flash PTA Balloon Dilatation Catheter is designed for dilation of stenotic ostial lesions in the peripheral vasculature. The Flash PTA Balloon Dilatation Catheter is a .014" guidewire-compatible, rapid exchange (RX) angioplasty balloon catheter with proximal anchoring and a working length of 135cm. The Flash PTA Balloon Dilatation Catheter uses a dual balloon design that features a compliant anchoring balloon, which prevents distal migration of the balloon during angioplasty. The second semicompliant higher-pressure balloon allows for luminal dilatation of de novo lesions and post deployment stent expansion. ### E. Intended Use: The Flash PTA Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature at aorto-ostial locations, including iliac, renal and carotid arteries. This device is also indicated for post-dilatation of balloon expandable stents in the peripheral vasculature. ### F. Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use: The purpose of this Traditional 510(k) is to seek an expanded indication for two of the Flash PTA Balloon Dilatation Catheter models cleared by 510(k) #K120738 on April 3, 2012. In accordance with the current thinking of the FDA as reflected by The 510k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] Draft Guidance dated December 27, 2011, Ostial Corporation is claiming two predicates for the subject device. The expanded indication for the JUN 2 9 2012 K1211 15 {1}------------------------------------------------ 510(k) Notification: Flash PTA Balloon Dilatation Catheter subject Flash PTA Balloon Dilatation Catheter is consistent with the overall intended use of the predicate devices, namely balloon dilatation within the peripheral vasculature. The predicate Flash PTA Balloon Dilatation Catheter was cleared with an indication that is a subset of the proposed expanded indication. The expanded indication adds that the device is also indicated for post-dilatation of balloon expandable stents in the peripheral vasculature. The predicate and subject Flash PTA Balloon Dilatation Catheters are the exact same device. No design modifications or changes to packaging, manufacturing or sterilization have been made since the clearance of the predicate Flash device. Additionally, substantial equivalence was established between the first iteration of the Flash PTA Balloon Dilatation Catheter and the primary predicate Sterling PTA Balloon Dilatation Catheter via the original Traditional 510(k) for the device product line, #K102482 dated February 25, 2011. The Sterling PTA Balloon Dilatation Catheter is indicated for use in Percutaneous Transluminal Angioplasty in the peripheral vasculature, including iliac, femoral, ilio-femoral, popliteal, renal and carotid arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilatation of balloon expandable and self-expanding stents in the peripheral vasculature. With respect to the subject device, the Sterling catheter features an equivalent design, packaging, fundamental technology, manufacturing and sterilization. The scope of the subject device's indication for use is narrower than that of the Sterling catheter, and both are indicated for post-dilatation of balloon expandable stents. The subject device and predicate devices are substantially equivalent in terms of intended use, fundamental scientific technology, target population, and operating principles. ### G. Performance Data: Biocompatibility testing has previously been completed on the Flash PTA Balloon Dilatation Catheter, Requirements for biological evaluation of the proposed device were based on the Blue Book Memorandum issued on May 1, 1995, Use of International ISO-10993, "Biological Evaluation of Medical Devices, Part 1:Evaluation and Testing." The biocompatibility test results show that the materials used in the design and manufacture of the components of the proposed device are non-toxic and non-sensitizing to biological tissues consistent with its intended use. The following biocompatibility tests were completed: . - ISO MEM Elution Assay . . - ASTM Hemolysis Assay . - . Complement Activation C3a and SC5b-9 Assay - . Thromboresistance Evaluation - Materials Mediated Rabbit Pyrogen ● - ISO Guinea Pig Maximization Sensitization ● - ISO Acute Systemic Injection Test . - . ISO Intracutaneous Reactivity Test - Pyrogen (LAL) Chromogenic . The Flash PTA Balloon Dilatation Catheter was evaluated using the following in-vitro performance bench testing to confirm the performance characteristics as compared to the predicate device: - Balloon Crossing Profile . - · Catheter Shaft Diameter - . Angioplasty Balloon Rated Burst Pressure - Anchoring Balloon Burst Volume . - . Angioplasty Balloon Compliance - . Balloon Inflation Time - Balloon Deflation Time � - . Angioplasty Balloon Rated Burst Pressure (in Stent) - Anchoring Balloon Burst Volume (in Stent) � - Angioplasty Balloon Fatigue . - Anchoring Balloon Fatigue ● - Catheter Bond Strength ● - Catheter Tip Pull Strength � - Catheter Torque Strength � - Simulated Use/Flexibility/Kink . - Radiopacity . - Angioplasty Balloon Fatigue (in Stent) ● - Anchoring Balloon Fatigue (in Stent) {2}------------------------------------------------ All test results demonstrate that the device materials, the manufacturing process, and the design for the Flash PTA Balloon Dilatation Catheter met the established specifications necessary for consistent performance according to its intended use. ### H. Conclusions: The Flash PTA Balloon Dilatation Catheter met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, test protocols, and/or customer inputs. The Flash PTA Balloon Dilatation Catheter is substantially equivalent to the legally marketed predicate devices and does not raise any new safety or effectiveness questions. 0 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JUN 2 9 2012 Ostial Corporation c/o Kaitlin von Hoffmann Clinical and Regulatory Associate 510 Clyde Avenue Mountain View, CA 94043 Re: K121175 Trade/Device Name: Flash PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT, DQY Dated: April 17, 2012 Received: April 18, 2012 Dear Ms. von Hoffmann: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. However, we remind you that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {4}------------------------------------------------ Page 2 – Ms. Kaitlin von Hoffmann or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. M.G. Wilhelm Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # SECTION 4.0: INDICATIONS FOR USE STATEMENT 510(k) Number: Device Name: ، _، K121175 Flash PTA Balloon Dilatation Catheter Indication For Use: The Flash PTA Balloon Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature at aorto-ostial locations, including iliac, renal and carotid arteries. This device is also indicated for post-dilatation of balloon expandable stents in the peripheral vasculature. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) M.R. Hilleman (Division Sign-Off) Division of Cardiovascular Devices 11-12-11-75 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...