FLASH PTA BALLOON DILATATION CATHETER 5MM X 19MM X 135CM; FLASH PTA BALLOON DILATATION CATHETER 6MM X 19MM X 135CM
Applicant
Ostial Corporation
Product Code
LIT · Cardiovascular
Decision Date
Feb 25, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Flash PTA Balloon Dilatation Catheter is indicated for use in Percutaneous Transluminal Angioplasty in the peripheral vasculature at aorto-ostial locations, including iliac, renal and carotid arteries. This catheter is not intended for use in coronary arteries.
Device Story
Flash PTA Balloon Dilatation Catheter is a .014" guidewire-compatible, rapid exchange (RX) angioplasty catheter; features dual balloon design: compliant anchoring balloon for precise positioning at aorto-ostial anatomies to prevent distal migration; semi-compliant high-pressure balloon for luminal dilatation. Used by physicians in clinical settings for peripheral vascular procedures. Operator positions catheter using anchoring balloon, then inflates dilatation balloon to treat obstructive lesions. Output is mechanical dilation of vessel. Benefits include improved stability and precise placement at ostial locations compared to standard single-balloon catheters.
Clinical Evidence
Bench testing included balloon crossing profile, burst pressure, compliance, inflation/deflation times, fatigue, bond/tip pull/torque strength, and radiopacity. Biocompatibility testing followed ISO 10993 (MEM elution, hemolysis, complement activation, thromboresistance, rabbit pyrogen, sensitization, systemic injection, intracutaneous reactivity). In-vivo testing performed in a swine model to assess simulated angioplasty procedure and downstream/cognitive effects. No clinical human data provided.
Technological Characteristics
Rapid exchange (RX) angioplasty catheter; .014" guidewire compatible. Dual balloon system: compliant anchoring balloon and semi-compliant high-pressure dilatation balloon. Usable length 135 cm; diameters 5mm and 6mm; lengths ~20mm. Materials evaluated per ISO 10993. Mechanical operation; no software or energy source.
Indications for Use
Indicated for patients requiring percutaneous transluminal angioplasty in peripheral vasculature at aorto-ostial locations, including iliac, renal, and carotid arteries. Not for use in coronary arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{1}------------------------------------------------
#### SECTION 5.0: 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
K102482
CONFIDENTIAL
| A. Submitter Information | |
|--------------------------|------------------------------------------------------------------------|
| Submitter's Name: | Ostial Corporation |
| Address: | 510 Clyde Avenue<br>Mountain View, CA 94043 |
| Telephone: | 650-903-9100 |
| Fax: | 650-903-9119 |
| Contact Person: | Eric Ankerud<br>Executive Vice President Clinical, Regulatory, Quality |
| Date of Preparation: | August 25, 2010 |
| B. Subject Device | |
| Trade Name: | Flash PTA Balloon Dilatation Catheter |
| Common/Usual Name: | Balloon Catheter |
Classification Name:
B
Catheter, Angioplasty, Peripheral, Transluminal (21 CFR 870.1250, Product Code LIT)
# C. Predicate Device Name(s)
Trade Name(s): Classification Name:
Sterling PTA Balloon Dilation Catheter, K053118 Catheter, Percutaneous (21 CFR 870.1250, Product Code DQY)
### D. Device Description:
The proposed Flash PTA Balloon Dilatation Catheter is designed for dilation of aorto-ostial lesions of peripheral vessels in the arterial system. The Flash PTA Balloon Dilatation Catheter is a .014" guidewirecompatible, rapid exchange (RX) angioplasty balloon catheter with proximal anchoring. The Flash PTA Balloon Dilatation Catheter uses a dual balloon design that features a compliant anchoring balloon that enables the operator to precisely position the catheter at aorto-ostial anatomies and prevent distal migration of the balloon during angioplasty. The second semi-compliant high pressure balloon allows for luminal dilatation.
### E. Intended Use:
The Flash PTA Balloon Dilatation Catheter is indicated for use in Percutaneous Transluminal Angioplasty in the peripheral vasculature at aorto-ostial locations, including iliac, renal and carotid arteries. This catheter is not intended for use in coronary arteries.
### F. Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use:
The proposed Flash PTA Balloon Dilatation Catheter and the predicate Sterling PTA Balloon Dilation Catheter have the same intended use. Both are indicated for treatment of obstructive lesions by high pressure dilation in the arterial system. The indication statement for the Flash PTA Balloon Dilatation Catheter is a subset of the broader indication statement of the Sterling PTA Balloon Dilation Catheter.
The proposed Flash PTA Balloon Dilatation Catheter and the Sterling PTA Balloon Dilation Catheter contain an inflatable semi-compliant balloon for dilation of obstructive lesions. The proposed Flash PTA Balloon Dilatation Catheter includes a second compliant balloon for locating and anchoring the device at ostial vessel locations.
The usable length of the proposed Flash PTA Balloon Dilatation Catheter is 135 cm which is the same usable length as the predicate Sterling Balloon Dilation Catheter. Both the proposed and predicate devices are offered in 5mm and 6mm balloon diameter sizes and approximately 20mm balloon lengths.
{2}------------------------------------------------
The proposed device and predicate device are substantially equivalent in terms of intended use, fundamental scientific technology, target population, and operating principles.
### G. Performance Data:
Biocompatibility testing on the proposed Flash PTA Balloon Dilatation Catheter has been completed. Requirements for biological evaluation of the proposed device were based on the Blue Book Memorandum issued on May 1, 1995, Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices, Part 1:Evaluation and Testing." The biocompatibility test results show that the materials used in the design and manufacture of the components of the proposed device are non-toxic and non-sensitizing to biological tissues consistent with its intended use. The following biocompatibility tests were completed:
- ISO MEM Elution Assay .
- . ASTM Hemolysis Assay
- Complement Activation C3a and SC5b-9 Assay .
- . Thromboresistance Evaluation
- Materials Mediated Rabbit Pyrogen .
- . ISO Guinea Pig Maximization Sensitization
- ISO Acute Systemic Injection Test .
- . ISO Intracutaneous Reactivity Test
- Pyrogen (LAL) Chromogenic �
The proposed Flash PTA Balloon Dilatation Catheter was evaluated using the following in-vitro performance bench testing to confirm the performance characteristics as compared to the predicate device:
- Balloon Crossing Profile .
- Catheter Shaft Diameter �
- Balloon Rated Burst Pressure (Angioplasty) .
- Balloon Burst Volume (Anchoring) .
- . Angioplasty Balloon Compliance
- . Balloon Inflation Time
- Balloon Deflation Time .
- Angioplasty Balloon Fatigue .
- Anchoring Balloon Fatigue .
- Catheter Bond Strength .
- Catheter Tip Pull Strength .
- . Catheter Torque Strength
- Simulated Use/Flexibility/Kink .
- Radiopacity .
In-vivo testing was completed using a swine model. A simulated angioplasty procedure was performed on test and control comparator groups. Post procedure animals were survived and observed for a predetermined period to assess for downstream and cognitive effects.
All test results demonstrate that the materials chosen, the manufacturing process, and the design utilized for the Flash PTA Balloon Dilatation Catheter met the established specifications necessary for consistent performance according to its intended use.
### H. Conclusions:
The Flash PTA Balloon Dilatation Catheter met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, test protocols, and/or customer inputs. The Flash PTA Balloon Dilatation Catheter is substantially equivalent to the legally marketed predicate device and does not raise any new safety or effectiveness questions.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
FEB 2 5 2511
Ostial Corporation c/o Mr. Mark Smutka Clinical, Regulatory, and Quality Consultant 510 Clyde Avenue Mountain View, CA 94043
Re: K102482
Trade/Device Name: Flash PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT, DQY Dated: February 15, 2011 Received: February 16, 2011
Dear Mr. Smutka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the idence) on for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendmonts, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract lightlity warranties. However, we remind you that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Mark Smutka
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# SECTION 4.0: INDICATIONS FOR USE STATEMENT
510(k) Number:
K102482.
Device Name:
Indication For Use:
Flash PTA Balloon Dilatation Catheter
The Flash PTA Balloon Dilatation Catheter is indicated for use in Percutaneous Transluminal Angioplasty in the peripheral vasculature at aorto-ostial locations, including iliac, renal and carotid arteries. This catheter is not intended for use in coronary arteries.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off
Division of Cardiovascular Devices
510(k) Number K102482
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.