Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031404020 ); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031406020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031408020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031410020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031412020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031415020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H749390314

K241683 · Boston Scientific Corporation · LIT · Jul 10, 2024 · Cardiovascular

Device Facts

Record IDK241683
Device NameSterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031404020 ); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031406020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031408020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031410020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031412020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031415020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H749390314
ApplicantBoston Scientific Corporation
Product CodeLIT · Cardiovascular
Decision DateJul 10, 2024
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Sterling Monorail PTA Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature, including iliac, femoral, popliteal, renal, and carotid arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilatation of balloon expandable and selfexpanding stents in the peripheral vasculature.

Device Story

Monorail rapid-exchange PTA balloon dilatation catheter; coaxial shaft design; semicompliant balloon at distal tip; tapered tip for stenosis navigation; radiopaque marker bands for fluoroscopic positioning. Used by physicians in clinical settings for peripheral vascular interventions. Input: guidewire (0.014 in / 0.018 in) for catheter advancement. Operation: inflation of balloon via outer lumen to dilate stenotic lesions or expand stents. Output: mechanical dilation of vasculature. Benefit: restoration of blood flow in peripheral arteries and dialysis fistulae.

Clinical Evidence

Bench testing only. Performance data included dimensional verification, simulated use, balloon inflation/deflation time, catheter bond strength, flexibility/kink testing, torque strength, shaft/bond burst, catheter extension/deflation, packaging testing, and design validation.

Technological Characteristics

Monorail rapid-exchange catheter; coaxial shaft; semicompliant balloon; 200 cm effective length; radiopaque marker bands; compatible with 0.014 in / 0.018 in guidewires. Sterilization method same as predicate.

Indications for Use

Indicated for patients requiring percutaneous transluminal angioplasty in peripheral vasculature (iliac, femoral, popliteal, renal, carotid arteries) or treatment of obstructive lesions in native/synthetic arteriovenous dialysis fistulae; also indicated for post-dilatation of balloon-expandable and self-expanding stents.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size below. July 10, 2024 Boston Scientific Corporation Michael Mcdonagh Fellow Regulatory Specialist Two Scimed Place Maple Grove, Minnesota 55311 #### Re: K241683 Trade/Device Name: Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031404020, H74939031406020, H74939031408020, H74939031410020, H74939031412020, H74939031415020, H74939031420020, H74939031504020, H74939031506020, H74939031508020, H74939031510020, H74939031512020, H74939031515020, H74939031520020. H74939031604020. H74939031606020. H74939031608020. H74939031610020, H74939031612020, H74939031615020, H74939031620020, H74939031704020. H74939031706020. H74939031708020. H74939031710020. H74939031712020, H74939031715020, H74939031720020) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: June 11, 2024 Received: June 11, 2024 Dear Michael Mcdonagh: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of {1}------------------------------------------------ Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Ariel G. Ash- Digitally signed by Ariel G. Ash-shakoor -S shakoor -S Date: 2024.07.10 16:56:10 -04'00' {2}------------------------------------------------ For Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use Submission Number (if known) #### K241683 Device Name | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031404020); | |------------------------------------------------------------------------| | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031406020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031408020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031410020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031412020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031415020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031420020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031504020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031506020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031508020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031510020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031512020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031515020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031520020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031604020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031606020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031608020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031610020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031612020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031615020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031620020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031704020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031706020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031708020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031710020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031712020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031715020); | | Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031720020) | | Indications for Use (Describe) | Indications for Use (Describe) The Sterling Monorail PTA The Sterling Monorail PTA Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature, including iliac, femoral, popliteal, renal, and carotid arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilatation of balloon expandable and selfexpanding stents in the peripheral vasculature. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. {4}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ # 510(k) Summarv #### 1. Submitter Boston Scientific Corporation Two Scimed Place Maple Grove MN 55311 United States Contact Person: Michael McDonagh Michael.McDonagh@bsci.com +353 91 517044 Tel Ballybrit Business Park, Galway, Ireland Date Prepared: June 11, 2024 ## 2. Device Name of Device: Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031404020 ); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031406020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031408020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031410020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031412020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031415020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031420020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031504020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031506020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031508020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031510020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031512020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031515020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031520020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031604020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031606020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031608020); {6}------------------------------------------------ Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031610020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031612020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031615020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031620020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031704020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031706020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031708020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031710020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031712020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031715020); Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (H74939031720020) Common or Usual Name: Percutaneous Catheter Classification Name: Catheter, Angioplasty, Peripheral, Transluminal Regulatory Class: II Product Code: LIT # 3. Predicate Device Predicate Device: Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter (K162350) Reference Devices: Sterling™ OVER-THE-WIRE PTA Balloon Dilatation Catheter (K141112, K132430, K053116); Sterling™ SL OVER-THE-WIRE PTA Balloon Dilatation Catheter (K093720) ## 4. Device Description The Sterling Balloon Dilatation Catheter is a Monorail brand rapid exchange catheter with a semicompliant balloon fixed at the distal tip. The balloon catheter has a coaxial shaft design. The outer lumen is used for inflation of the balloon, and the wire lumen permits the use of guidewires 0.014 in / 0.018 in to facilitate advancement of the catheter to and through the stenosis to be dilated. The product's catheter includes a tapered tip to facilitate advancement of the catheter to and through the stenosis. Two radiopaque marker bands (one proximal and one distal), in conjunction with fluoroscopy, enable accurate positioning of the balloon. The effective length of the balloon catheter is 200 cm; catheter markers indicate the exit of the dilatation catheter tip out of the guiding catheter (one at 90 cm and two at 100 cm). ## 5. Indications for Use The Sterling Monorail PTA Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature, including iliac, femoral, popliteal, renal, and {7}------------------------------------------------ carotid arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilatation of balloon expandable and self-expanding stents in the peripheral vasculature. ## 6. Comparison of Technological Characteristics with the Predicate Device The Sterling Monorail PTA Balloon Dilatation Catheter has the same intended use/ indication for use, principle of operation, sterilization method, as the predicate Sterling MONORAIL PTA Balloon Dilatation Catheter (K162350). The following technological differences exist between the subject and predicate devices: - . Longer catheter length of 200cm - . Increase in catheter outer diameter and sheath compatibility Differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness. The results of bench testing provide reasonable assurance of substantial equivalence to the predicate device. # 7. Performance Data The following performance data were provided in support of the substantial equivalence determination. #### Sterilization Sterilization Testing #### Bench Testing - Dimensional Verification - Simulated Use - Balloon Inflation & Deflation Time - Catheter Bond Strength - Flexibility & Kink test - Torque Strength - Shaft & Bond Burst - Catheter Extension & Deflation - . Packaging Testing - . Design Validation Testing ## 8. Conclusions The subject and predicate devices share the same indications for use and fundamental scientific technology, including principle of operation. Differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness. Non {8}------------------------------------------------ clinical performance evaluations support substantial equivalence of the Subject Device to the predicate Sterling MONORAIL PTA Balloon Dilatation Catheter (K162350).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...