CORDIS POWERFLEX PLUS PTA BALLOON CATHETER

K971516 · Cordis Corp. · LIT · Jul 23, 1997 · Cardiovascular

Device Facts

Record IDK971516
Device NameCORDIS POWERFLEX PLUS PTA BALLOON CATHETER
ApplicantCordis Corp.
Product CodeLIT · Cardiovascular
Decision DateJul 23, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Intended Use

The PowerFlex Plus PTA balloon catheter is intended to dilate stenoses in iliac, femoral, iliofemoral, popliteal, infra popliteal and renal arteries, and for treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Device Story

PowerFlex Plus PTA balloon catheter is a dual-lumen catheter with a distal inflatable balloon; used by physicians in clinical settings for percutaneous transluminal angioplasty. Device tracks over a prepositioned guidewire to the target lesion. Balloon inflation/deflation is controlled via the inflation lumen; radiopaque marker bands facilitate accurate placement under fluoroscopy. Device dilates stenoses in peripheral arteries and arteriovenous dialysis fistulae. Benefits include restoration of vessel patency and improved blood flow in treated areas.

Clinical Evidence

No clinical data provided; substantial equivalence is based on bench testing and design comparison to legally marketed predicate devices.

Technological Characteristics

Dual-lumen catheter design; distal inflatable balloon; radiopaque marker bands for visualization; guidewire compatible; inflation lumen for contrast/saline delivery; biocompatible materials.

Indications for Use

Indicated for patients requiring dilation of stenoses in iliac, femoral, iliofemoral, popliteal, infra popliteal, and renal arteries, and for treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # SUMMARY OF SAFETY EFFECTIVENESS ## I. General Provisions PTA Balloon Catheter Common or Usual Name: JUL 2 3 1997 PowerFlex™ Plus PTA Balloon Catheter Proprietary Name: # II. Name of Predicate Devices - Cordis PowerFlex PTA Balloon Catheter - Cordis Opta5 PTA Balloon Catheter - Cordis Small Vessel PTA Balloon Catheter - Cordis Savyy PTA Balloon Catheter - Cordis Mega PTA Balloon Catheter - Meditech UltraThin Balloon Catheter ## III. Classification Class II ### IV. Performance Standards Performance standards have not been established by the FDA under section 514 of the Food, Drug and Cosmetic Act. ### V. Intended Use and Device Description The PowerFlex Plus PTA balloon catheter is intended to dilate stenoses in iliac, femoral, iliofemoral, popliteal, infra popliteal and renal arteries, and for treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The Cordis PowerFlex Plus PTA balloon catheter is a dual lumen design with a distal inflatable balloon. Two radiopaque marker bands indicate the dilating section of the balloon and aid in balloon placement. All balloons distend to sizes above the nominal size at pressures greater than the nominal pressure. The balloon inflation lumen is used to inflate and deflate the balloon. The nominal balloon size is printed on the hub. The guidewire lumen is used to track the catheter over a prepositioned guidewire or to inject contrast medium and/or saline. The maximum injection pressure is 450 psi. The compatible guidewire size, catheter shaft French size and catheter length are printed on the hub. The radiopaque marker bands indicate the stated nominal length of the balloon. {1}------------------------------------------------ い # VI. Biocompatibility All materials used in the PowerFlex Plus PTA balloon catheter are biocompatible. # VII. Summary of Substantial Equivalence The Cordis PowerFlex Plus PTA balloon catheter and the referenced Cordis PowerFlex PTA balloon catheter, Cordis Opta5 PTA balloon catheter, Cordis Small Vessel PTA balloon catheter, Cordis Savvy PTA balloon catheter, Cordis Mega PTA balloon catheter and Meditech UltraThin balloon catheter are similar in their basic design, construction, indication for use and performance characteristics. {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and head. The words "DEPARTMENT OF HEALTH & HUM SERVICES • USA" are arranged in a circular fashion around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JUL 2 3 1997 Ms. Jan Pieter Kappelle Cordis Corporation 14201 N.W. 60th Avenue ... . Miami Lakes, Florida 33014 Re: K971516 PowerFlex™ PTA Balloon Catheter Regulatory Class: II (two) Product Code: 74 LIT Dated: April 25, 1997 Received: April 28, 1997 Dear Ms. Kappelle: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {3}------------------------------------------------ #### Page 2 - Ms. Jan Pieter Kappelle This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbrandinq by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Thomas J. Callahon Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neuroloqical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): Device Name: Cordis PowerFlex™ Plus PTA Balloon Catheter Indications for Use: The PowerFlex Plus PTA Balloon Catheter is intended to dilate stenoses in iliac, femoral, ilio-femoral, popliteal, infra popliteal and renal arteries, and for treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Taut R (Division Sign -Division of C and Neurolo¿ and Neurolog 510(k) Numbe: ..... 179 715112 PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IS NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...