Pirouette 018

K151153 · Arravasc Limited · LIT · Aug 20, 2015 · Cardiovascular

Device Facts

Record IDK151153
Device NamePirouette 018
ApplicantArravasc Limited
Product CodeLIT · Cardiovascular
Decision DateAug 20, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Intended Use

Pirouette 018 is intended for balloon dilation of the iliac, femoral, popliteal, infra-popliteal and renal arteries.

Device Story

Pirouette 018 is an over-the-wire (OTW) percutaneous transluminal angioplasty (PTA) catheter; used by physicians in clinical settings for balloon dilation of peripheral arteries. Device consists of a catheter shaft with a balloon at the distal end and a manifold at the proximal end. Physician inserts catheter over a guidewire (max 0.018 inch) into the target vessel; balloon is inflated to dilate the artery; radiopaque marker bands facilitate positioning under fluoroscopy. Device is single-use and sterilized via ethylene oxide. Output is mechanical dilation of stenotic vessels; helps restore blood flow. Benefits include minimally invasive treatment of arterial disease.

Clinical Evidence

No clinical studies were performed. Safety and effectiveness were established through extensive non-clinical bench testing, including dimensional verification, balloon fatigue, burst pressure, bond strength, and biocompatibility (cytotoxicity, sensitization, systemic toxicity, hemolysis, and thromboresistance).

Technological Characteristics

OTW PTA catheter; dual-lumen Y-design manifold; balloon diameters 2-9mm; lengths 20-300mm; nominal pressure 8 atm; rated burst pressure 12-16 atm. Features two radiopaque marker bands. Materials include standard catheter polymers; hydrophilic coating. Sterilization via Ethylene Oxide (EtO).

Indications for Use

Indicated for balloon dilation of iliac, femoral, popliteal, infra-popliteal, and renal arteries in patients requiring percutaneous transluminal angioplasty.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other to create a sense of depth and connection. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 20, 2015 ArraVasc Limited c/o Jan-Paul van Loon Qserve Group US Inc. P.O Box 940 Charlestown, NH 03603 Re: K151153 Trade/Device Name: Pirouette 018 Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: July 9, 2015 Received: July 13, 2015 Dear Jan-Paul van Loon: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {1}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Kenneth J. Cavanaugh -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K151153 Device Name Pirouette 018 Indications for Use (Describe) Pirouette 018 is intended for balloon dilation of the iliac, femoral, popliteal and renal arteries. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary #### 1. Submitter | Submitter Address: | ArraVasc Limited<br>2 Ballybrit Business Park,<br>Galway<br>Ireland | |--------------------|---------------------------------------------------------------------| | Phone Number: | +353-91-758939 | | Fax Number: | +353-91-758930 | | Contact Person: | Máiréad McCaffrey | | Date Prepared: | 15 April 2015 | #### 2. Device | Device Trade Name: | Pirouette 018 | |------------------------|----------------------------------------------| | Common Name: | OTW PTA catheter | | Classification Name: | Peripheral Transluminal Angioplasty Catheter | | Classification number: | 21 CFR 870.1250 | | Product code: | LIT | | Class: | II | | Classification Panel: | Cardiovascular | #### 3. Predicate Devices | Primary predicate: | Bantam and Bantam OTW PTA Catheter<br>Clearstream Technologies Ltd.<br>510(K) number: K112335/K093139 | |----------------------|-------------------------------------------------------------------------------------------------------| | Secondary predicate: | SABER IM PTA Dilation Catheter<br>Cordis Corporation<br>510(K) number: K133843 | #### 4. Device Description Physical Description: Percutaneous Transluminal Angioplasty Catheter (PTA Catheter). {4}------------------------------------------------ #### 5. Indication for use | Indications for Use | Pirouette 018 is intended for balloon dilation of the iliac, femoral,<br>popliteal, infra-popliteal and renal arteries. | |---------------------|-------------------------------------------------------------------------------------------------------------------------| | Statement: | | #### 6. Comparison of technological characteristics with the predicate device | Summary of<br>Technological<br>Characteristics | The Pirouette 018 incorporates<br>substantially equivalent design, packaging, fundamental technology,<br>materials, manufacturing, sterilization and intended use as those<br>featured in the primary predicate Bantam and Bantam OTW PTA<br>catheters and contains a hydrophilic coating substantially equivalent to<br>the secondary predicate SABERTM PTA catheter. | |------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| ### Table 1 Summary of general and technical characteristics against the predicate devices. | Parameter | Characteristics | |--------------------------|------------------------------------------------------------------------------| | Classification | Class II, 21 CFR 870.1250 | | | Same Classification as predicate devices. | | Intended Use | Same intended use of balloon dilatation of the femoral, popliteal and | | | infra-popliteal arteries with additional vessels; iliac and renal indicated. | | Balloon material | Same type of material | | Balloon diameter | Same range of balloon diameters: 2 - 9 mm | | Balloon length | Comparable range of balloon lengths: 20 - 300mm | | Nominal pressure (atm) | Same or higher nominal pressure: 8 atm. | | Rated burst pressure | Comparable rated burst pressure: 16atm (2.0-5.0mm balloon | | (atm) | diameter) / 14atm (6.0mm balloon diameter) / 12atm( 7.0-9.0mm | | | balloon diameter) | | Radiopaque | Two Markerbands, one at distal and one at proximal side of the | | Marker bands | balloon, with the same function. | | Outer shaft | Same type material | | | | | Catheter shaft | Same diameter or lower: 3.6 . 4.2 Fr | | outer diameter | | | Catheter Usable | Comparable range of usable lengths: 45 - 150cm | | Lengths (cm) | | | Recommended | Comparable size or lower: 4 . 5 Fr | | Introducer Sheath | | | compatibility | | | Recommended | Same compatibility; maximum 0.018+ | | Guidewire diameter | | | Catheter strain relief & | Same type of material, with the same function. | | manifold material | | | Catheter manifold | Same design, dual lumen Y design, with the same function. | | design | | | Catheter coating | Similar type of coating. | | Sterilization Method | Same method; Ethylene Oxide. | | Single Use / Reusable | Single Use | {5}------------------------------------------------ #### 7. Performance Data The Pirouette 018 was thoroughly tested on the bench to evaluate and verify that it meets the required performance specifications and to support a determination of substantial equivalence. The bench testing plan was developed with the consideration of the recommendations outlined in the applicable FDA quidance documents, tests recommended in ISO 10555-1. Intravascular catheters- Sterile and single-use catheters - Part 1: General Requirements, and 10555-4, Intravascular catheters- Sterile and single-use catheters - Part 4: Balloon dilatation catheters. Testing performed on the Pirouette included the following: ### Non-Clinical Tests: - Dimensional verification - Balloon preparation, pushability, trackability, deployment,withdrawal and balloon reinsertion ● - Balloon rated burst pressure (RBP) ● - Balloon fatigue (repeated balloon inflations) . - Balloon Compliance at nominal and rated burst inflation pressure - . Balloon Inflation/Deflation Time - Catheter Bond Strength (Tip to balloon, balloon to proximal shaft, shaft to manifold) - Flexibility and Kink test ● - Torque Strength ● - Radiopacity . - Coating Integrity ● - . Particulate evaluation - . Guide wire compatibility - Introducer sheath compatibility The results showed that the Pirouette 018 met the pre-determined acceptance criteria. ### Biocompatibility Tests: Per ISO10993-1:2009, Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process, the Pirouette 018 is classified as an externally communicating device, which contacts circulating blood during a limited contact duration (m24hours). Biocompatibility testing conducted on Pirouette 018 per ISO 10993-1:2009 included the following: - . MTT Cytotoxicity Study - ISO Maximization Sensitization Study ● - ISO Intracutaneous Study ● - ISO Systemic Toxicity Study ● - ASTM Hemolysis - C3a Complement Activation Assay . - SC5b-9 Complement Activation Assay ● - ASTM Partial Thromboplastin Time ● - In Vivo Thromboresistance Study Jugular Vein ● - . Pyrogenicity The test results show that the Pirouette 018 is biocompatible. {6}------------------------------------------------ Sterilization, Shelf life tests and Packaging validation: - . EtO sterilzation validation - . EtO/ECH residue determination - . Shelf life testing - . Packaging validation Test results show that the Pirouette 018 is sterily is maintained by the packaging during the entire shelf life of the device. From the results it was also concluded that the device meets the criteria for Residual Testing, i.e. EtO/ECH residues. ### Clinical Performance Data: - No clinical studies were performed for the purpose of obtaining safety and effectiveness data. . - The Pirouette 018 has been approved for marketing in the European Union (CE certified) in 2014. . and has been marketed in the EU for since then. #### 8. Conclusions The performance testing presented above establishes that the Pirouette 018 is as safe and effective for its intended use as the predicate device. The comparison tabulated above demonstrates that the Pirouette 018 device is substantially equivalent to the predicate devices. Information contained within this submission demonstrates that the Pirouette 018: - . Has legally marketed predicate devices; - Has identical indications for use as the identified primary predicate devices; ● - Incorporates the same fundamental technology, and uses accepted scientific methods and international standards to evaluate device safety and effectiveness; - Demonstrates that the design and performance of the Pirouette 018 has equivalent safety and performance characteristics of predicate devices, and does not raise different questions of safety and effectiveness. Based upon the; intended use, design, performance characteristics, non-clinical performance testing performed, and comparison to legally marketed devices, it is concluded that the Pirouette 018 is appropriate for its intended use, and is substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...