The Pirouette 014 PTA Catheter is intended for balloon dilation of the illac, femoral, popliteal, and renal arteries.
Device Story
Pirouette 014 is an over-the-wire (OTW), semi-compliant, coaxial PTA catheter; features balloon mounted on distal tip; hydrophilic coating; radiopaque markers for fluoroscopic visualization. Used by physicians in clinical settings for peripheral artery dilation. Device tracks over .014 inch guidewires; inflation/deflation via manifold connector. Provides mechanical dilation of stenotic vessels; improves blood flow. Benefits patient by restoring vessel patency in iliac, femoral, popliteal, and renal arteries.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by bench testing, including dimensional verification, balloon fatigue, burst pressure, bond strength, flexibility, torque, radiopacity, coating integrity, and particulate evaluation. Biocompatibility testing (ISO 10993-1) was leveraged from the predicate device.
Technological Characteristics
OTW, semi-compliant, coaxial PTA catheter. Materials: same as predicate. Dimensions: 1.25-4.0 mm balloon diameter, 15-300 mm length, 90-150 cm effective length. Compatible with .014 inch guidewires. Features hydrophilic coating, dual-lumen manifold, and radiopaque marker bands. Sterilization: Ethylene Oxide (EO).
Indications for Use
Indicated for balloon dilation of iliac, femoral, popliteal, infra-popliteal, and renal arteries in patients requiring percutaneous transluminal angioplasty.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a family of three figures.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 22, 2016
ArraVasc Ltd. Ms. Aoife Donoghue Regulatory and Design Assurance Manager 2 Ballybrit Business Park Galway, Ireland
Re: K162316
Trade/Device Name: Pirouette 014 Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: October 25, 2016 Received: October 31, 2016
Dear Ms. Donoghue:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Brian D. Pullin -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162316
Device Name Pirouette 014
Indications for Use (Describe)
The Pirouette 014 PTA Catheter is intended for balloon dilation of the illac, femoral, popliteal, and renal arteries.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
#### 5.1 Submitter
| Submitter Address: | ArraVasc Limited<br>2 Ballybrit Business Park,<br>Galway<br>Ireland |
|--------------------|---------------------------------------------------------------------|
| Phone Number: | +353-91-758939 |
| Fax Number: | +353-91-758930 |
| Contact Person: | Aoife Donoghue Regulatory & Design Assurance Manager |
| Date Prepared: | 04 Aug 2016 |
#### 5.2 Device
| Device Trade Name: | Pirouette 014 |
|------------------------|----------------------------------------------|
| Common Name: | OTW PTA catheter |
| Classification Name: | Peripheral Transluminal Angioplasty Catheter |
| Classification number: | 21 CFR 870.1250 |
| Product code: | LIT |
| Class: | II |
| Classification Panel: | Cardiovascular |
#### 5.3 Predicate Device
| Predicate: | Pirouette 018 OTW PTA Catheter<br>ArraVasc Ltd.<br>510(K) number: K151153 |
|-------------------|---------------------------------------------------------------------------------|
| Reference Device: | Bantam α PTA Catheter<br>ClearStream Technologies Ltd<br>510(K) number: K093139 |
#### 5.4 Device Description
Device Description:
The Pirouette 014 Percutaneous Transluminal Angioplasty (PTA) Catheter Family are standard over the wire (OTW), semi-compliant, coaxial design catheters with a balloon mounted on the distal tip. The distal portion of the catheter has a hydrophilic coating.
The manifold connector and shaft design consists of a guidewire lumen allowing the catheter to track over a guidewire and an inflation lumen, used to inflate and deflate the balloon. Radiopaque markers are positioned on the shaft within the balloon to enable visualization of the catheter/balloon under fluoroscopy. The catheter is compatible with .014 inch (0.356 mm) wire guides.
The Pirouette 014 PTA Catheter Family includes multiple balloon sizes ranging from 1.25 mm to 4.0 mm in diameter and 15 mm to 300 mm in length. The nominal balloon diameter (mm) and the balloon length (mm) are inscribed on the guidewire hub of the manifold. The effective lengths of the catheter are 90 cm, 130 cm and 150 cm.
{5}------------------------------------------------
Percutaneous Transluminal Angioplasty Catheter (PTA Catheter). Physical Description:
#### 5.5 Indication for use
The Pirouette 014 PTA Catheter is intended for balloon dilation of the iliac, Indications for Use femoral, popliteal, infra-popliteal, and renal arteries. Statement:
#### 5.6 Comparison with the predicate device
The Pirouette 014 is substantially equivalent to the predicate device in design, materials, packaging, fundamental scientific technology, manufacturing, sterilization and intended use. The following table provides a summary of general and technical characteristics as compared to the primary predicate device and predicate device.
| Parameter | Characteristics |
|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Classification | Class II, 21 CFR 870.1250<br>Same Classification as predicate devices. |
| Intended Use | Same intended use: balloon dilatation of the iliac, femoral, popliteal<br>and infra-popliteal arteries and renal arteries. |
| Balloon material | Same material |
| Balloon diameter | Comparable range of balloon diameters: 1.25 mm - 4.0 mm versus<br>2.0 mm - 9.0 mm for the predicate device |
| Balloon length | Comparable range of balloon lengths: 15 mm - 300 mm versus 20<br>mm - 300 mm for the predicate device |
| Nominal pressure (atm) | Same nominal pressure: 8 atm. |
| Rated burst pressure<br>(atm) | Same rated burst pressure: 16 atm |
| Radioopaque<br>markerbands | Two Markerbands, one at distal and one at proximal side of the<br>balloon, with the same function |
| Outer shaft | Same material |
| Catheter shaft<br>outer diameter | Comparable catheter diameter: 2-7-3.6 Fr versus: 3.6 – 4.2 Fr for the<br>Predicate |
| Catheter Usable<br>lengths (cm) | Comparable range of usable lengths: 90 cm - 150 cm versus 45 cm -<br>150cm for the Predicate device |
| Recomended<br>Introducer Sheath Size<br>compatibility | Comparable size: 4 Fr versus 4 - 5 Fr for the Predicate device |
| Recommended<br>Guidewire diameter | Maximum 0.014" versus; maximum 0.018" |
| Catheter strain relief &<br>manifold material | Same material, with the same function |
| Catheter manifold<br>design | Same design, dual lumen Y design, with the same function |
| Catheter coating | Same coating |
| Sterilization Method | Same method; Ethylene Oxide |
| Single Use / Reusable | Single Use |
# Table 1 Comparison of Pirouette 014 to Pirouette 018 (Predicate)
{6}------------------------------------------------
#### 5.7 Performance Data
As per ArraVasc Risk Analysis procedures, the Pirouette 014 was thoroughly tested on the bench to evaluate and verify that it meets the required performance specifications and to support a determination of substantial equivalence. The bench testing plan was developed with the recommendations outlined in the applicable FDA guidance documents, tests recommended in ISO 10555-1, Intravascular catheters-Sterile and single-use catheters - Part 1: General Requirements, and 10555-4, Intravascular catheters-Sterile and single-use catheters - Part 4: Balloon dilatation catheters. Testing performed on Pirouette included the following:
### Non-Clinical Tests:
- . Dimensional verification
- Balloon preparation, pushability, trackability, deployment, withdrawal and balloon reinsertion .
- Balloon rated burst pressure (RBP) .
- Balloon fatigue (repeated balloon inflations)
- . Balloon Compliance at nominal and rated burst inflation pressure
- Balloon Inflation/Deflation Time ●
- Catheter Bond Strength (Tip to balloon, balloon to proximal shaft, shaft to manifold) ●
- . Flexibility and Kink test
- Torque Strength ●
- Radiopacity ●
- . Coating Integrity
- . Particulate evaluation
- Guide wire compatibility ●
- . Introducer sheath compatibility
The results showed that the Pirouette 014 met the pre-determined acceptance criteria. No new safety or effectiveness issues were raised during this testing.
### Biocompatibility Tests:
Per ISO10993-1:2009, Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process, the Pirouette 014 is classified as an externally communicating device, which contacts circulating blood during a limited contact duration (≤24hours). For reason that the Pirouette 014 has the same indication, general design, materials, and packaging, the Biocompatibility tests with the Pirouette 018 were considered valid for the Pirouette 014. Biocompatibility testing with the Pirouette 018 was conducted per ISO 10993-1:2009 were completed and passed:
- MTT Cytotoxicity Study ●
- ISO Maximization Sensitization Study .
- ISO Intracutaneous Study .
- ISO Systemic Toxicity Study ●
- ASTM Hemolysis .
- C3a Complement Activation Assay .
- SC5b-9 Complement Activation Assay ●
- ASTM Partial Thromboplastin Time ●
- In Vivo Thromboresistance Study - Jugular Vein
- . Pyrogenicity
{7}------------------------------------------------
Sterilization, Shelf life tests and Packaging validation:
- EO sterilzation validation by an adoption validation using a sub-lethal cycle approach; .
- EO/ECH residues of Pirouette 035 (K161427) were tested (largest surface area of Pirouette . family, therefore worst case);
- . Shelf life testing was leveraged from the data on the Pirouette 018;
- . Packaging validation was leveraged on the data of the Pirouette 018.
The results show that the Pirouette 014 is sterility is maintained by the packaging during the entire shelf life of the device. From the results it was also concluded that the device meets the criteria for Residual Testing, i.e. EO/ECH residues.
### Clinical Performance Data:
- . No clinical studies were performed for the purpose of obtaining safety and effectiveness data.
- . The Pirouette 014 is approved for marketing in the European Union (CE certified).
#### 5.8 Conclusions
Based upon the intended use, fundamental scientific technology, performance characteristics, non-clinical performance testing, and comparison to legally marketed devices, it is concluded that the Pirouette 014 is appropriate for its intended use, and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.