The Clearblue® Early Pregnancy Test is an over-the-counter chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine. This test is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the day of the missed period, i.e. as early as five (5) days before the day of the expected period. The test is intended for home use.
Device Story
Clearblue® Early Pregnancy Test is an OTC lateral flow chromatographic immunoassay for qualitative hCG detection in urine. Device uses monoclonal antibodies against hCG alpha and beta subunits. Features proprietary FSH-modulated hCG scavenger system upstream of test line to enhance specificity by removing hCG in samples with high FSH levels. User urinates on absorbent wick or dips into collected sample; capillary action draws urine through conjugate pad, mobilizing antibody-coated gold-sol particles. hCG/FSH binds to gold-sol label; sample moves across scavenger zone (immobilized anti-beta FSH), then test line (anti-beta hCG), and control line (polyclonal anti-rabbit). Results displayed visually: two lines indicate 'Pregnant', one line indicates 'Not Pregnant'. Used at home by lay users. Provides rapid pregnancy status confirmation, aiding early detection.
Clinical Evidence
Clinical performance evaluated via lay-user study (n=295) and early pregnancy clinical sample testing. Lay-user study showed 100% agreement with clinical pregnancy status and technician results (PPV, NPV, sensitivity, specificity, accuracy all 100%). Early pregnancy detection study (n=204 samples per day) demonstrated detection capability from days -10 to 0 relative to missed period. Analytical studies confirmed 10mIU/ml sensitivity, no high-dose hook effect up to 1,000,000 mIU/ml, and no interference from common substances, cross-reactants (LH, TSH, FSH), or hCG beta-core fragments.
Technological Characteristics
Lateral flow sandwich immunoassay; nitrocellulose membrane; gold-sol conjugate; monoclonal antibodies against hCG alpha/beta subunits. Includes FSH-modulated hCG scavenger system. Visual readout. Single-use device. Calibrated to WHO 4th International Standard for hCG. Operates via capillary action. No electronic components or software.
Indications for Use
Indicated for qualitative detection of hCG in urine to aid in early pregnancy detection. Intended for home use by individuals as early as 6 days before the missed period (5 days before expected period).
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
FIRST RESPONSE™ Early Result Pregnancy Test (k123436)
Related Devices
K200913 — Clearblue® Early Digital Pregnancy Test · Spd Swiss Precision Diagnostics GmbH · Aug 13, 2021
K023699 — CLEARBLUE EASY EARLY RESULT PREGNANCY TEST · Unipath , Ltd. · Nov 27, 2002
K040341 — CLEARBLUE EASY EARLIEST RESULTS PREGNANCY TEST · Unipath , Ltd. · Mar 8, 2004
K240025 — Clearblue® Early Detection Pregnancy Test · Spd Swiss Precision Diagnostics GmbH · Jan 31, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.
September 21, 2022
SPD Swiss Precision Diagnostics GmbH % Kamila Przedmojska Principal Regulatory Affairs Specialist SPD Development Company Limited Priory Business Park, Stannard Way Bedford, Bedfordshire MK44 3UP United Kingdom
Re: K213379
Trade/Device Name: Clearblue® Early Pregnancy Test Regulation Number: 21 CFR 862.1155 Regulation Name: Human Chorionic Gonadotropin (hCG) Test System Regulatory Class: Class II Product Code: LCX Dated: June 24, 2022 Received: July 5, 2022
Dear Kamila Przedmojska:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Paula Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) k213379
Device Name Clearblue® Early Pregnancy Test
The Clearblue® Early Pregnancy Test is an over-the-counter chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine. This test is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the day of the missed period, i.e. as early as five (5) days before the day of the expected period.
The test is intended for home use.
| Type of Use ( <i>Select one or both, as applicable</i> ) | <table><tr><td><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</td></tr><tr><td><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | | | |
| <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for Swiss Precision Diagnostics GmbH. The logo features the letters "SPD" in a stylized font, with the letters arranged in a circular shape. The circle is colored with a gradient that transitions from blue at the top to green at the bottom. To the right of the circle, the text "Swiss Precision Diagnostics GmbH" is written in a simple, sans-serif font.
# 510(k) Summary
| A. Submitted By: | SPD Swiss Precision Diagnostics GmbH<br>47 Route de Saint-Georges<br>Petit-Lancy<br>CH-1213<br>Geneva<br>Switzerland<br>Telephone: +41 580048741 |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| B. Contact Person: | Kamila Przedmojska<br>Principal Regulatory Affairs Specialist<br>SPD Development Company Limited<br>Priory Business Park<br>Bedford<br>MK44 3UP<br>United Kingdom<br>Telephone: +44 1234835504 |
| C. Date Prepared: | 24 June, 2022 |
| D. Device Name: | Clearblue® Early Pregnancy Test |
| Product Code: | LCX |
| Common name: | Kit, Test, Pregnancy, hCG, over the counter |
| Classification: | Class II |
| Product code: | LCX |
| Regulation Description: | Human chorionic gonadotropin (hCG) test system |
| Regulation number: | 21CFR 862.1155 |
| E. Predicate Device: | FIRST RESPONSE™ Early Result Pregnancy Test<br>(k123436) |
{4}------------------------------------------------
## F. Indication for Use
Clearblue® Early Pregnancy Test is an over-the-counter The chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine. This test is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the day of the missed period, i.e. as early as five (5) days before the day of the expected period.
The test is intended for home use.
## G. Device Description
The Clearblue® Early Pregnancy Test is an over-the-counter (OTC), visual pregnancy test and is indicated for use up to 6 days before the missed period (5 days before expected period). The device employs an immunochromatographic sandwich assay to detect hCG on a lateral flow test strip.
The test incorporates a proprietary, FSH modulated, hCG scavenger system positioned upstream of the hCG test line to maintain high specificity to pregnancy. The scavenger system captures hCG when there are high levels of FSH in the sample. This ensures that hCG is removed from samples with elevated levels of FSH, reducing the chance of false positive results which while rare, may occur in some women.
The result is displayed to the user in the test window as two lines for a 'Pregnant' result and one line for a `Not Pregnant' result.
## H. Substantial Equivalence Information
Predicate device name: FIRST RESPONSE™ Early Result Pregnancy Test
Predicate (k) number: k123436
Comparison with predicate:
Table 1 Similarities and differences between Clearblue® Early Pregnancy Test and the predicate FIRST RESPONSE™ Early Results Pregnancy Test
{5}------------------------------------------------
| Component | Clearblue® Early Pregnancy<br>Test (Proposed Device) | FIRST RESPONSE™ Early<br>Result Pregnancy Test<br>(Predicate Device) |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| | | Similarities |
| Intended Use | Qualitative detection of human<br>hCG for an aid in early detection<br>of pregnancy | Same |
| 'Early Test'<br>Claim | in some cases as early as six (6)<br>days before the day of the missed<br>period, i.e. as early as five (5)<br>days before the day of the<br>expected period. | Same |
| Target User | Over-The-Counter use | Same |
| Device format | Single Use | Same |
| Sample Matrix | Urine | Same |
| Analyte | hCG | Same |
| Sample<br>application | In-stream and dip | Same |
| hCG<br>Sensitivity | 10mIU/ml | Same |
| Traceability | WHO 4th International Standard<br>for hCG | Same |
| Test Principle | Lateral flow sandwich immuno-<br>chromatographic assay | Same |
| Assay Mobile<br>Phase | Gold conjugate | Same |
| Test Type | Qualitative | Same |
| Control<br>Mechanism | Visual | Same |
| Results<br>Display | Visual Parallel Line<br>2 Lines = Pregnant<br>1 Line = Not Pregnant | Same |
| | | Differences |
| Analyte<br>Detection | Detects intact hCG.<br>Scavenger system to remove<br>intact hCG in the presence of<br>FSH. | Recognises: intact hCG<br>hyperglycosylated hCG<br>hCG ẞ-subunit<br>hCG ẞ-core fragment |
| Time to<br>results | 5 minutes | 3 minutes |
{6}------------------------------------------------
## I. Test Principle
The Clearblue® Early Pregnancy test is a lateral flow sandwich immunoassay employing monoclonal antibodies that are specifically directed against the alpha and beta sub-units of hCG.
To use the Clearblue® Early Pregnancy test, the user either urinates directly onto the absorbent wick or collects a sample in a container and dips the absorbent wick into the collected sample.
Buffer salts in the wick are dissolved by the sample, normalising the pH and ionic strength to provide suitable conditions for the down-stream immunoassay. Upon wetting of the wick, urine is drawn by capillary action into the conjugate pad. As the sample moves from the wick through the conjugate release pad, the antibody coated gold-sol particles are mobilized and transported along the test strip. Any hCG and/or FSH in the sample will bind to the test gold-sol label via their common alpha sub-unit.
On reaching the nitrocellulose membrane, the sample is drawn across the plotted line of immobilised anti-beta FSH antibody in the scavenger zone which is not visible to the user as it is located within the plastic case moulding. The sample then progresses across the monoclonal anti-beta hCG antibody test line and polyclonal anti-rabbit antibody control line and on to the distal end of the test strip into the sink pad.
## J. Performance characteristics
## 1. Analytical Performance
## a) Precision/Reproducibility
1. A not-pregnant pooled urine was spiked with hCG traceable to the 4th WHO international standard with hCG concentrations of <1, 2, 3, 5, 7, 10, 15 and 20 mIU/ml. Each standard was tested with devices from three different batches using both dip and simulated in-stream sampling methods. The study was performed by three technicians over five nonconsecutive days.
The results are summarised in the tables below:
{7}------------------------------------------------
| | Clearblue® Early Pregnancy Test Overall Results | | | | | | | |
|-----------------|-------------------------------------------------|-----------------|----------------------------|----------------------------|-----------------|----------------------------|-----------------|--|
| hCG<br>Standard | Dip method | | | Simulated in stream method | | | Total | |
| (mIU /<br>ml) | Not<br>Pregnant<br>(n) | Pregnant<br>(n) | Pregnant<br>Results<br>(%) | Not<br>Pregnant<br>(n) | Pregnant<br>(n) | Pregnant<br>Results<br>(%) | Pregnant<br>(%) | |
| <1 | 135 | 0 | 0.0 | 135 | 0 | 0.0 | 0.0 | |
| 2 | 135 | 0 | 0.0 | 135 | 0 | 0.0 | 0.0 | |
| 3 | 102 | 33 | 24.4 | 100 | 35 | 25.9 | 25.2 | |
| 5 | 43 | 92 | 68.1 | 42 | 93 | 68.9 | 68.5 | |
| 7 | 16 | 119 | 88.1 | 18 | 117 | 86.7 | 87.4 | |
| 10 | 0 | 135 | 100.0 | 0 | 135 | 100.0 | 100 | |
| 15 | 0 | 135 | 100.0 | 0 | 135 | 100.0 | 100 | |
| 20 | 0 | 135 | 100.0 | 0 | 135 | 100.0 | 100 | |
Overall Precision Results of Clearblue® Early Pregnancy Test
## Percentage Pregnant Results for Each hCG Standard by Technician
| hCG<br>Standard<br>(mIU/<br>ml) | Technician 1 | | Technician 2 | | Technician 3 | |
|---------------------------------|-------------------------------------|----------------------------|-------------------------------------|----------------------------|-------------------------------------|----------------------------|
| | Pregnant/<br>Not<br>Pregnant<br>(n) | Pregnant<br>Results<br>(%) | Pregnant/<br>Not<br>Pregnant<br>(n) | Pregnant<br>Results<br>(%) | Pregnant/<br>Not<br>Pregnant<br>(n) | Pregnant<br>Results<br>(%) |
| <1 | 0/90 | 0.0 | 0/90 | 0.0 | 0/90 | 0.0 |
| 2 | 0/90 | 0.0 | 0/90 | 0.0 | 0/90 | 0.0 |
| 3 | 21/69 | 23.3 | 23/67 | 25.6 | 24/66 | 26.7 |
| 5 | 59/31 | 65.6 | 62/28 | 68.9 | 64/26 | 71.1 |
| 7 | 77/13 | 85.6 | 78/12 | 86.7 | 81/9 | 90.0 |
| 10 | 90/0 | 100 | 90/0 | 100 | 90/0 | 100 |
| 15 | 90/0 | 100 | 90/0 | 100 | 90/0 | 100 |
| 20 | 90/0 | 100 | 90/0 | 100 | 90/0 | 100 |
{8}------------------------------------------------
| hCG<br>Standard<br>(mIU/<br>ml) | Batch 1 | | Batch 2 | | Batch 3 | |
|---------------------------------|-------------------------------------|----------------------------|-------------------------------------|----------------------------|-------------------------------------|----------------------------|
| | Pregnant/<br>Not<br>Pregnant<br>(n) | Pregnant<br>Results<br>(%) | Pregnant/<br>Not<br>Pregnant<br>(n) | Pregnant<br>Results<br>(%) | Pregnant/<br>Not<br>Pregnant<br>(n) | Pregnant<br>Results<br>(%) |
| <1 | 0/90 | 0.0 | 0/90 | 0.0 | 0/90 | 0.0 |
| 2 | 0/90 | 0.0 | 0/90 | 0.0 | 0/90 | 0.0 |
| 3 | 25/65 | 27.8 | 22/68 | 24.4 | 21/69 | 23.3 |
| 5 | 63/27 | 70.0 | 63/27 | 70.0 | 59/31 | 65.6 |
| 7 | 79/11 | 87.8 | 80/10 | 88.9 | 77/13 | 85.6 |
| 10 | 90/0 | 100 | 90/0 | 100 | 90/0 | 100 |
| 15 | 90/0 | 100 | 90/0 | 100 | 90/0 | 100 |
| 20 | 90/0 | 100 | 90/0 | 100 | 90/0 | 100 |
## Percentage Pregnant Results for Each hCG standard by Batch
## Percentage Pregnant Results for Each hCG standard by Day
| hCG<br>Standard<br>(mIU /<br>ml) | Day 1 | | Day 2 | | Day 3 | | Day 4 | | Day 5 | |
|----------------------------------|-------------|----------------------------|-------------|----------------------------|-------------|----------------------------|-------------|----------------------------|-------------|----------------------------|
| | P/NP<br>(n) | Pregnant<br>Results<br>(%) | P/NP<br>(n) | Pregnant<br>Results<br>(%) | P/NP<br>(n) | Pregnant<br>Results<br>(%) | P/NP<br>(n) | Pregnant<br>Results<br>(%) | P/NP<br>(n) | Pregnant<br>Results<br>(%) |
| < 1 | 0/54 | 0.0 | 0/54 | 0.0 | 0/54 | 0.0 | 0/54 | 0.0 | 0/54 | 0.0 |
| 2 | 0/54 | 0.0 | 0/54 | 0.0 | 0/54 | 0.0 | 0/54 | 0.0 | 0/54 | 0.0 |
| 3 | 14/40 | 25.9 | 14/40 | 25.9 | 11/43 | 20.4 | 15/39 | 27.8 | 14/40 | 25.9 |
| 5 | 39/15 | 72.2 | 36/18 | 66.7 | 37/17 | 68.5 | 35/19 | 64.8 | 38/16 | 70.4 |
| 7 | 47/7 | 87.0 | 49/5 | 90.7 | 46/8 | 85.2 | 44/10 | 81.5 | 50/4 | 92.6 |
| 10 | 54/0 | 100 | 54/0 | 100 | 54/0 | 100 | 54/0 | 100 | 54/0 | 100 |
| 15 | 54/0 | 100 | 54/0 | 100 | 54/0 | 100 | 54/0 | 100 | 54/0 | 100 |
| 20 | 54/0 | 100 | 54/0 | 100 | 54/0 | 100 | 54/0 | 100 | 54/0 | 100 |
{9}------------------------------------------------
- b) Linearity/assay reportable range:
Not applicable. This is a qualitative device.
- c) High dose hook effect study:
A not-pregnant pooled urine was spiked with hCG to concentrations of <1, 10, 10,000 and 1,000,000mIU/ml and tested with 5 replicates from each of three batches. No hook effect was observed at the tested concentrations.
## d) Traceability
The tests are calibrated against the WHO 4th International Standards for human Chorionic Gonadotropin (hCG).
- e) Stability
The claimed shelf life of the device stored in the sealed foil pouches at room temperature is 39 months.
- f) Detection Limit (Sensitivity)
See Precision/Reproducibility section.
- g) Analytical Specificity
#### Structurally not-related compounds
#### Interfering substances
The Clearblue® Early Pregnancy Test devices were tested with potential interfering substances. Each interfering substance was spiked into nonpregnant pooled urine and 10mIU/ml hCG urine standards.
Each condition was tested with 5 devices from each of three batches of the Clearblue® Early Pregnancy Test for each of the two urine standards according to the dip sampling method. No interference effect was observed at the tested concentrations shown in the table below:
{10}------------------------------------------------
| Interfering Substance | Concentration |
|-----------------------------|---------------|
| Acetylsalicylic acid | 1.0mg/ml |
| Acetone | 1.0mg/ml |
| Albumin | 5mg/ml |
| Ampicillin | 20 mg/dL |
| Ascorbic acid | 150µg/ml |
| Atropine | 200 µg/mL |
| Bilirubin | 20 mg/dL |
| Blood | 0.3% v/v |
| Caffeine | 1.2mg/ml |
| Cannabinol | 10 mg/dL |
| Clomiphene citrate | 24µg/ml |
| Cotinine | 40 µg/ml |
| Ethanol | 1% v/v |
| E3G | 620ng/ml |
| Gentisic acid | 20 mg/dL |
| Glucose | 20mg/ml |
| Haemoglobin | 100µg/ml |
| Hydrochloric acid | 1.25mM |
| Ibuprofen | 100µg/ml |
| Leukocytes | x106 cells/ml |
| Oxytetracycline | 300µg/ml |
| Paracetamol (Acetaminophen) | 600µg/ml |
| Phenylpropanolamine | 20 mg/dL |
| P3G | 40µg/ml |
| Semen | 5% v/v |
| Sodium hydroxide | 1.25mM |
| Tetracycline | 300µg/ml |
| Urea | 30mg/ml |
| Uric acid | 750µg/ml |
{11}------------------------------------------------
### Structurally related compounds
#### Effects of cross reactants
The results in the table below show that all the devices tested returned the correct result when tested with potential cross reactants at the concentrations shown below.
| | Result | | |
|------------------------------------------------|--------------|---------------------|-----------|
| Standards | Pregnant (n) | Not Pregnant<br>(n) | Pass/Fail |
| <1 hCG mIU/ml | 0 | 15 | Pass |
| 10 hCG mIU/ml | 15 | 0 | Pass |
| <1 hCG mIU/ml, 500 LH mIU/ml | 0 | 15 | Pass |
| 10 hCG mIU/ml, 500 LH mIU/ml | 15 | 0 | Pass |
| <1 hCG mIU/ml, 1 TSH mIU/ml | 0 | 15 | Pass |
| 10 hCG mIU/ml, 1 TSH mIU/ml | 15 | 0 | Pass |
| <1 hCG mIU/ml, 1000 FSH mIU/ml | 0 | 15 | Pass |
| 10 hCG mIU/ml, 15 FSH mIU/ml | 15 | 0 | Pass |
| 4 hCG mIU/ml, 100 FSH mIU/ml | 0 | 30 | Pass |
| 4 hCG mIU/ml, 1000 FSH mIU/ml | 0 | 30 | Pass |
| 4 hCG mIU/ml, 100 FSH mIU/ml, 1 TSH<br>mIU/ml | 0 | 30 | Pass |
| 4 hCG mIU/ml, 100 FSH mIU/ml, 500<br>LH mIU/ml | 0 | 30 | Pass |
#### Effects of urine pH
Effect of urine pH was performed by adjusting negative (<1 mIU/ml) and 10mIU/ml hCG urine standards to a pH range of 4, 6 and 9. Each urine standard was tested with 5 devices from each of 3 batches by dip sampling method. The results demonstrated that Clearblue® Early Pregnancy Test will continue to return a correct result when tested with a urine sample in the pH range of 4 – 9.
#### Effect of urine specific gravity
To test the effect of specific gravity, the device was challenged with nonpregnant pooled urine and positive (10mIU/ml hCG) urine standards with specific gravity of 1.000, 1.01, 1.015, 1.03 and 1.035. Each urine standard
{12}------------------------------------------------
was tested with 5 devices from each of 3 batches by dip sampling method. The results showed the Clearblue® Early Pregnancy Test will continue to return a correct result in response to changes in specific gravity within the range from 1.000 to 1.035.
### Effect of hCG beta core fragment (hCGβcf)
To evaluate the effect of hCGβcf, 11 conditions were tested with 5 devices from each of 3 batches, totalling 165 devices.
Pooled pregnant urine collected from 6-7 weeks and 9-12 weeks pregnancy were tested with and without hCGßcf spiked up to 1µM.
Pooled neqative urine spiked with hCG to a concentration representative of 6-7 weeks and 9-12 weeks pregnant urine samples were tested with and without hCGβcf spiked up to 1μM.
A non-pregnant pooled urine was spiked to 10mIU/ml hCG then spiked with 150pM hCGβcf (a concentration 5 times the molar concentration of intact hCG) was tested. Positive (negative pooled urine spiked to 10mIU/ml) and non-pregnant (<1.0 mIU/ml) controls were also tested.
The results show that the performance of the Clearblue® Early Pregnancy Test is not affected by high concentrations of hCG β-core fragment.
h) Assay cut-off
See Precision/Reproducibility Section.
## 2. Comparison Study
- a. Method comparison with predicate device:
The Lay User Usage study collected the lay users and technician results and compared them against physician determined clinical pregnancy status, hence the comparison to the predicate device was not performed.
b. Matrix comparison:
Not Applicable. The device is intended for urine sample only.
{13}------------------------------------------------
- 3. Clinical Studies
- a. Clinical Sensitivity:
Not Applicable
- b. Clinical Specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable)
Detection of hCG in Early Pregnancy Clinical Samples
Early pregnancy urine samples from days -10 to 0 relative to the day of the missed period were collected. Each sample was tested using both methods of sampling across three batches of devices.
The early pregnancy detection results are summarised in table below:
| Day Relative to<br>Missed Period | Total Samples<br>(n) | Number<br>Pregnant Result<br>(n) | Percent<br>Pregnant Result<br>(%) |
|----------------------------------|----------------------|----------------------------------|-----------------------------------|
| -10 | 42 | 0 | 0.0 |
| -9 | 72 | 0 | 0.0 |
| -8 | 120 | 6 | 5.0 |
| -7 | 204 | 58 | 28.4 |
| -6 | 204 | 158 | 77.5 |
| -5 | 204 | 191 | 93.6 |
| -4 | 204 | 200 | 98.0 |
| -3 | 204 | 204 | 100 |
| -2 | 204 | 204 | 100 |
| -1 | 204 | 204 | 100 |
| 0 | 204 | 204 | 100 |
{14}------------------------------------------------
## Lay User Study
Pregnant and not pregnant women volunteers with diverse educational and professional backgrounds and ages between 18 and 55 years old participated in the Lay User Usage Study. Their results were compared to their clinical pregnancy status and to the results obtained from trained technicians testing the same urine samples in the same sampling method (either simulated in-stream or dipping).
The study confirmed that PPV, NPV, sensitivity, specificity and accuracy for the Clearblue® Early Pregnancy Test in the hands of lay-user volunteers was 100%, for both dip and in-stream testing methods.
The agreement between lay-user volunteer results and their clinical status with the Clearblue® Early Pregnancy Test was 100%. There was also 100% agreement between all lay-user volunteer results and technician results.
The results are summarised in tables below:
Volunteer (both in-stream and dip results combined) vs clinical pregnancy status.
| Clinical Pregnancy<br>Status | Volunteer Result | | |
|------------------------------|------------------|-----------------|-------|
| | Pregnant | Not<br>Pregnant | Total |
| Pregnant | 152 | 0 | 152 |
| Not Pregnant | 0 | 143 | 143 |
| Total | 152 | 143 | 295 |
Volunteer (In-stream) results vs clinical pregnancy status
| Clinical Pregnancy<br>Status | Volunteer Result | | |
|------------------------------|------------------|-----------------|-------|
| | Pregnant | Not<br>Pregnant | Total |
| Pregnant | 59 | 0 | 59 |
| Not Pregnant | 0 | 57 | 57 |
| Total | 59 | 57 | 116 |
{15}------------------------------------------------
Volunteer (Dip) results vs clinical pregnancy status
| Clinical Pregnancy | Volunteer Result | | |
|--------------------|------------------|-----------------|-------|
| Status | Pregnant | Not<br>Pregnant | Total |
| Pregnant | 93 | 0 | 93 |
| Not Pregnant | 0 | 86 | 86 |
| Total | 93 | 86 | 179 |
Volunteer results vs technician result
| Technician Test<br>Results | Volunteer Result | | |
|----------------------------|------------------|-----------------|-------|
| | Pregnant | Not<br>Pregnant | Total |
| Pregnant | 152 | 0 | 152 |
| Not Pregnant | 0 | 143 | 143 |
| Total | 152 | 143 | 295 |
Lay User Spiked Standard Study
A study was performed to analyse the performance of the Clearblue® Early Pregnancy Test when read by lay user according to the Instructions for Use. A range of the hCG urine standards at 0, 2, 3, 5, 10 and 15mIU/ml were sampled by dip method of sampling and read by lay users. The results are shown in table below:
| hCG Standard<br>(mIU/ml) | Total (n) | Number<br>Pregnant<br>(n) | Percent<br>Pregnant<br>(%) |
|--------------------------|-----------|---------------------------|----------------------------|
| 0 | 107 | 0 | 0.0 |
| 2 | 107 | 2 | 1.9 |
| 3 | 103 | 28 | 27.2 |
| 5 | 106 | 74 | 69.8 |
| 10 | 105 | 105 | 100 |
| 15 | 108 | 108 | 100 |
{16}------------------------------------------------
## Specificity study to determine false-positive result rate
A study was performed to determine the incidence of false positive results among not-pregnant women of pre-menopausal age (18-40 years), perimenopausal age (41-55 years) and post-menopausal age (>55 years). Urine samples were collected from individual women of each cohort and were tested by technicians with three batches of the Clearblue® Early Pregnancy Test devices by both dip and simulated in-stream method of sampling.
The results (combined method of sampling) are summarised in table below:
| Cohort | Not Pregnant<br>(n) | Samples<br>(n) | Specificity<br>(%) |
|------------------|---------------------|----------------|--------------------|
| Pre-menopausal | 300 | 300 | 100 |
| Peri-menopausal | 299 | 299 | 100 |
| Post-menopausal | 300 | 300 | 100 |
| All Not Pregnant | 899 | 899 | 100 |
Clinical Cut-off
Not applicable.
#### Expected value
Not applicable.
## K. Conclusion
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.