K964108 · Direct Access Diagnostics · LCX · Nov 18, 1996 · Clinical Chemistry
Device Facts
Record ID
K964108
Device Name
FACT PLUS PREGNANCY TEST
Applicant
Direct Access Diagnostics
Product Code
LCX · Clinical Chemistry
Decision Date
Nov 18, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Intended Use
The modified Fact PLUS Pregnancy Test (Cup and Dropper) is an over-the counter in vitro diagnostic immunoassay intended for the qualitative detection of human chorionic gonadotropin (hCG), in urine. A labeling modification to the currently marketed Fact PLUS® Pregnancy Test enables the test results to be read in three minutes or less without altering the accuracy, sensitivity or specificity of the test.
Device Story
Fact PLUS Pregnancy Test is a disk-shaped, lateral flow immunoassay for qualitative hCG detection in urine. Device consists of a molded plastic disk housing a test strip with conjugate, antibody, and control sites. User applies urine to 'Urine Well'; urine migrates along strip. If hCG is present, hCG/conjugate complex binds to anti-beta hCG antibody (vertical bar) and polyclonal antibodies (horizontal bar), forming a pink 'plus' sign. If hCG is absent, conjugate binds only to horizontal bar, forming a 'minus' sign. Control Window (CW) displays red color to confirm sufficient sample volume. Results are read by the consumer in 3 minutes or less. Device is intended for home use. Output provides visual confirmation of pregnancy status, aiding in early pregnancy detection.
Clinical Evidence
Laboratory validation testing compared the modified device to currently marketed versions using hCG-positive and hCG-negative clinical urine specimens across four lots. Results showed 100% agreement between 3-minute and 5-minute read times. A single-blind consumer study (n=150) with female volunteers aged 18-45 demonstrated 100% accuracy in test performance and interpretation using the modified instructions.
Technological Characteristics
Disk-shaped molded plastic housing containing a lateral flow test strip. Sensing principle: immunochromatographic assay using anti-alpha hCG antibody conjugate and anti-beta hCG/polyclonal antibody capture sites. Visual readout via colorimetric reaction (pink plus/minus sign). Built-in control window (red color) confirms sample volume. Standalone, non-electronic, disposable device.
Indications for Use
Indicated for consumer use to detect pregnancy as early as the first day of a missed period, any time of the day.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
K962521 — FACTPLUS ONE STEP PREGNANCY TEST · Direct Access Diagnostics · Jul 29, 1996
K970082 — ABBOTT ADVISOR ONE-STEP PREGNANCY TEST · Abbott Laboratories · Jan 24, 1997
K042280 — FACT PLUS ONE-STEP PREGNANCY TEST KIT · Unipath , Ltd. · Sep 10, 2004
K193318 — Distinct® Early Detection Pregnancy Test · ACON Laboratories, Inc. · Aug 20, 2020
K150063 — One Step hCG Pregnancy Test · Coretests, Inc. · Oct 2, 2015
Submission Summary (Full Text)
{0}
K964108
Direct Access Diagnostics Division
Premarket Notification [510(k)]
FactPlus® Pregnancy Test (modification)
NOV 18 1995
APPENDIX
510(k) SUMMARY
{1}
DIRECT ACCESS DIAGNOSTICS
# 510(k) SUMMARY per 21 CFR 807.92(c)
**Trade name**
Fact PLUS Pregnancy Test
**Common name**
Consumer Use Home Pregnancy Test
**Classification name**
Human chorionic Gonadotropin (hCG) Test System
## Statement
Information supporting claims of substantial equivalence, as defined under the Federal Food Drug and Cosmetic Act, respecting safety and effectiveness is summarized below. For the convenience of the reviewer, this summary is formatted in accordance with the Agency's final rule..."510(k) Summaries and 510(k) Statements." (21CFR 807) and can be used to provide a substantial equivalence summary to anyone requesting it from the Agency.
- Modified device name: Fact PLUS Pregnancy Test
- Predicate device name(s): currently marketed Fact PLUS Pregnancy Test and Fact PLUS One Step Pregnancy Test
## Intended Use
The modified Fact PLUS Pregnancy Test (Cup and Dropper) is an over-the counter *in vitro* diagnostic immunoassay intended for the qualitative detection of human chorionic gonadotropin (hCG), in urine. A labeling modification to the currently marketed Fact PLUS® Pregnancy Test enables the test results to be read in three minutes or less without altering the accuracy, sensitivity or specificity of the test.
## Indications
The Fact PLUS Pregnancy Test is indicated for consumer use to detect pregnancy as early as first day of a missed period, any time of the day. Negative test results are confirmed after three minutes, some positives results can be read as soon as one minute.
## Safety
Home pregnancy tests for detection of hCG in urine have been safely used and in commercial distribution for over twenty years. *In vitro* diagnostic hCG urine tests which do not come in contact with the body are not generally associated with safety issues. Home-use pregnancy tests are categorized as "waived" status under the 1988 Clinical Laboratory and Improvement Amendment (CLIA).
440 Rt. 22 East, Bridgewater, N.J. 08807 ▼ TEL: 908 218-7300 FAX: 908 253-6412
{2}
510(k) SUMMARY (continued)
## Device Description
The modified Fact PLUS Pregnancy Test is a disk-shaped ("Test Disk") device constructed of two pieces of molded plastic. The test chemistries are coated on a test strip housed within the test disk. The test strip contains three reactive sites: a conjugate site, an antibody site and a control site. Three openings on the upper surface of the test disk correspond to these sites: "Urine Well," "Result Window," "Control Window" (CW) respectively.
When urine is added via the Urine Well it wets the conjugate (anti-α hCG antibody) mobilizing it. If hCG molecules are present in the urine specimen, an hCG/conjugate complex is formed. This complex and any remaining unbound conjugate migrate along the strip attaching to the anti-β hCG antibody contained on the vertical bar (positive) and also to the polyclonal antibodies on the intersecting horizontal bar (negative) causing a pink reaction. The appearance of pink color on the horizontal and vertical bars forms the "plus" sign or positive "pregnant" result.
When no hCG molecules are present in the urine specimen, the conjugate remains unbound and attaches only to the horizontal or "minus" bar causing it to turn pink. The appearance of only one pink horizontal bar constitutes the negative or ("not pregnant") result.
The urine flow continues on to the Control Window where it activates a special dye causing a red color to appear. The red color in the CW tells the user that the enough urine was added. It is can also play an important role in helping the Consumer Service representatives on our toll free number to evaluate consumer reported problems. It takes approximately five minutes for the urine to reach the Control Window. The appearance of a "plus" or a "minus" in the Result Window however, appears sooner usually within three minutes or less. Some positive results can appear as soon as one minute. The appearance of a plus or minus in the Result Window and red color in the Control Window are the built-in controls which indicate the test chemistries worked properly and that a sufficient volume of urine was added.
## Substantial Equivalence
The modified Fact PLUS Pregnancy Test is substantially equivalent to the marketed version of Fact PLUS® Pregnancy Test with a labeling modification for reading the results sooner. Additionally modified Fact PLUS (Cup & Dropper) is substantially equivalent to Fact PLUS® One Step (Direct) Pregnancy test which also provides results within 3 minutes.
The proposed and the marketed devices have the same intended use, performance characteristics, and utilize the same chemistries, materials, and test analyte. All three pregnancy tests can be used the first day of missed period, at any time of day.
{3}
3
# 510(k) SUMMARY (continued)
## Performance Data
Substantial equivalence was demonstrated in laboratory validation testing. Fact PLUS Pregnancy Test when performed according to the modified instructions was compared to the currently marketed Fact PLUS Pregnancy Tests for accuracy specificity, and sensitivity.
Percent laboratory accuracy was determined by testing hCG-positive and hCG-negative female clinical urine specimens across three different lots of Fact PLUS Pregnancy Test and one lot of Fact PLUS One Step. Data compared test results read according to the modified labeling instructions (three minutes) to test results read according to the current labeling instructions (End of Test Window /approximately 5 minutes). There was 100% agreement for all four lots at the three minute and five minute time frames.
## Consumer Testing
A single-blind, consumer, clinical study was conducted to verify that consumers could obtain accurate test results when performing the Fact PLUS Pregnancy test and reading the results according to the modified directions. The consumer population consisted of 150 evaluable female volunteers recruited from a Central New Jersey location. On hundred and fifty-two subjects had been enrolled, however two subjects were dropped from the study for protocol violations. The participants were between 18 to 45 years of age, from various income levels, and educational and employment backgrounds.
Subjects were provided with a randomized hCG positive or hCG negative urine sample, the Fact Pregnancy Test and the modified instructions. All subjects who tested positive specimens recorded the results as positive and all subjects who tested negative specimens recorded the results as negative. The overall consumer accuracy rating was 100% for this study.
## Conclusion
The Fact PLUS Pregnancy Test with modified directions has been shown to be substantially equivalent to the currently marketed Fact PLUS® and the Fact PLUS® One Step Pregnancy Test in laboratory validation testing demonstrating 100% agreement for accuracy, sensitivity, specificity. Additionally, consumer testing demonstrated that the FactPLUS Pregnancy Test can be performed according to the modified directions and correctly interpreted by consumers with an overall accuracy rating of 100%.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.