MissLan™ Digital Pregnancy Rapid Test

K222305 · Guangzhou Decheng Biotechnology Co., Ltd. · LCX · Nov 30, 2022 · Clinical Chemistry

Device Facts

Record IDK222305
Device NameMissLan™ Digital Pregnancy Rapid Test
ApplicantGuangzhou Decheng Biotechnology Co., Ltd.
Product CodeLCX · Clinical Chemistry
Decision DateNov 30, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1155
Device ClassClass 2

Intended Use

MissLan™ Digital Pregnancy Rapid Test is intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy. It is intended for use by people who would like to find out whether they are pregnant in a home environment. Only for use outside the body. For over the counter use.

Device Story

MissLan™ Digital Pregnancy Rapid Test is a lateral flow immunoassay for qualitative hCG detection in urine; used in dip or midstream mode. Device consists of a test strip in a plastic housing with an absorbent tip. Principle of operation: lateral flow immunoassay combined with light reflection; LED light source detects light intensity on the test strip; results displayed on an integrated LCD screen. Used in home environments by lay users. Output is a qualitative digital result (pregnant/not pregnant). Benefits include early pregnancy detection via simple, self-administered testing.

Clinical Evidence

Bench testing only. Analytical performance: sensitivity 25 mIU/mL; no hook effect up to 500,000 mIU/mL; no cross-reactivity with hLH, hFSH, hTSH, or hCG β-core fragment; no interference from common substances, pH (4-9), or density (1.000-1.035). Method comparison study (n=200) showed 100% conformity with predicate. Lay person study (n=400 total across two studies) demonstrated 100% agreement between lay user results and professional results, confirming ease of use and accurate interpretation.

Technological Characteristics

Lateral flow immunoassay; plastic housing with absorbent tip. LED light source and optical reflection sensor for line detection. Qualitative detection of hCG; 25 mIU/mL cutoff. Traceable to 5th WHO International Standard (NIBSC 07/364). Standalone digital display. No external connectivity.

Indications for Use

Indicated for qualitative detection of hCG in urine to aid early pregnancy detection. Intended for over-the-counter home use by individuals seeking pregnancy status.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. November 30, 2022 Guangzhou Decheng Biotechnology Co., Ltd. % Joe Shia, Director LSI International 504 E Diamond Ave. Suite I Gaithersburg, MD 20877 ## Re: K222305 Trade/Device Name: MissLan™ Digital Pregnancy Rapid Test Regulation Number: 21 CFR 862.1155 Regulation Name: Human Chorionic Gonadotropin (hCG) Test System Regulatory Class: Class II Product Code: LCX Dated: July 31, 2022 Received: August 2, 2022 ## Dear Joe Shia: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Digitally signed by Paula____ Paula Caposino -S Caposino -S Date: 2022.11.30 17:02:33 -05'00' Paula Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) k222305 Device Name MissLanTM Digital Pregnancy Rapid Test Indications for Use (Describe) MissLan™ Digital Pregnancy Rapid Test is intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy. It is intended for use by people who would like to find out whether they are pregnant in a home environment. Only for use outside the body. For over the counter use. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|---------------------------------------------| | Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY K22305 | 1. | Date: | November 29, 2022 | |----|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 2. | Submitter: | Guangzhou Decheng Biotechnology Co., Ltd.<br>Building 2, No. 68, 1st Nanxiang Road, Science<br>City, Huangpu District, 510000 Guangzhou,<br>Guangdong, China | | 3. | Contact person: | Joe Shia<br>LSI International Inc.<br>504 East Diamond Ave., Suite I<br>Gaithersburg, MD 20877<br>Telephone: 240-505-7880<br>Fax: 301-916-6213<br>Email: shiajl@yahoo.com | | 4. | Device Name:<br>Classification:<br>Product Code:<br>CFR: | MissLan™ Digital Pregnancy Rapid Test<br>Class II<br>LCX<br>862.1155 | | 5. | Predicate Devices: | Preview® Digital Pregnancy Test (K173229) | #### 6. Intended Use MissLan™ Digital Pregnancy Rapid Test is intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy. It is intended for use by people who would like to find out whether they are pregnant in a home environment. Only for use outside the body. For over the counter use. #### 7. Device Description MissLan™ Digital Pregnancy Rapid Test is used for in vitro qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine, and is designed to be tested in dip or midstream mode. The test device consists of a single test strip assembled in a plastic housing, with an absorbent tip. The device is in a ready-touse format. #### 8. Substantial Equivalence Information | Similarities | | | |--------------|------------------|------------------| | Item | Candidate device | Predicate device | {4}------------------------------------------------ | Intended use | Early detection of<br>pregnancy | Early detection of<br>pregnancy | |--------------------|---------------------------------|---------------------------------| | Specimen | Urine | Urine | | Assay technical | Immunochromatographic<br>assay | Immunochromatographic<br>assay | | Sensitivity | 25 mIU/mL | 25 mIU/mL | | Results | Qualitative | Qualitative | | Target user | Over the counter use | Over the counter use | | Sample application | Midstream and dip<br>methods | Midstream and dip methods | | Readout | Digital/LCD screen | Digital/LCD screen | | Differences | | | | Item | Device | Predicate | | Appearance | 155.5 x 21.5 x 14.5 mm | 150 x 25 x 15 mm | #### 9. Test Principle MissLan™ Digital Pregnancy Rapid Test uses lateral flow immunoassay and light reflection for the detection of the HCG in urine specimens. The test would detect the light intensity by using the LED as the light source. After that, the result can be displayed on the display screen. ## 10. Performance Characteristics ## A. Analytical performance ## a. Precision/Reproducibility/Sensitivity Negative female urine was spiked with hCG standard (Traceable to the 5th WHO) to hCG concentrations of 0, 12.5, 15, 18.75, 22.5, 25, 50, 100 and 200 mIU/mL. Each sample was tested by both dip and midstream methods in 10 replicates per day for 5 days for each device lot. Total of three device lots were tested. Tests were performed by three different operators for each sample concentration. The results are summarized in the table below: | hCG<br>Concentration<br>(mIU/mL) | Operator<br>1<br>Lot 1 | | Operator<br>2<br>Lot 2 | | Operator<br>3<br>Lot 3 | | Total<br>result | | %<br>Negative | %<br>Positive | |----------------------------------|------------------------|----|------------------------|----|------------------------|----|-----------------|-----|---------------|---------------| | | - | + | - | + | - | + | - | + | | | | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100% | 0% | | 12.5 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100% | 0% | | 15 | 23 | 27 | 24 | 26 | 24 | 26 | 71 | 79 | 47% | 53% | | 18.75 | 12 | 38 | 11 | 39 | 11 | 39 | 34 | 116 | 23% | 77% | ## Midstream Testing {5}------------------------------------------------ | 22.5 | 5 | 45 | 5 | 45 | 6 | 44 | 16 | 134 | 11% | 89% | |------|---|----|---|----|---|----|----|-----|-----|------| | 25 | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0% | 100% | | 50 | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0% | 100% | | 100 | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0% | 100% | | 200 | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0% | 100% | ## Dip Testing | hCG<br>Concentration | Operator<br>1 | | Operator<br>2 | | Operator<br>3 | | Total<br>result | | %<br>Negative | %<br>Positive | |----------------------|---------------|----|---------------|----|---------------|----|-----------------|-----|---------------|---------------| | (mIU/mL) | - | + | - | + | - | + | - | + | | | | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100% | 0% | | 12.5 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100% | 0% | | 15 | 24 | 26 | 24 | 26 | 25 | 25 | 73 | 77 | 49% | 51% | | 18.75 | 12 | 38 | 11 | 39 | 12 | 38 | 35 | 115 | 23% | 77% | | 22.5 | 4 | 46 | 5 | 45 | 5 | 45 | 14 | 136 | 9% | 91% | | 25 | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0% | 100% | | 50 | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0% | 100% | | 100 | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0% | 100% | | 200 | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0% | 100% | ## Overall Testing | hCG<br>Concentration<br>(mIU/mL) | Lot 1 | | Lot 2 | | Lot 3 | | Total<br>result | | %<br>Negative | %<br>Positive | |----------------------------------|-------|-----|-------|-----|-------|-----|-----------------|-----|---------------|---------------| | | - | + | - | + | - | + | - | + | | | | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 300 | 0 | 100% | 0% | | 12.5 | 100 | 0 | 100 | 0 | 100 | 0 | 300 | 0 | 100% | 0% | | 15 | 47 | 53 | 48 | 52 | 49 | 51 | 144 | 156 | 48% | 52% | | 18.75 | 24 | 76 | 22 | 78 | 23 | 77 | 69 | 231 | 23% | 77% | | 22.5 | 9 | 91 | 10 | 90 | 11 | 89 | 30 | 270 | 10% | 90% | | 25 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 300 | 0% | 100% | | 50 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 300 | 0% | 100% | | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 300 | 0% | 100% | | 200 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 300 | 0% | 100% | MissLan™ Digital Pregnancy Rapid Test exhibited reproducible results. Based on the above results, the sensitivity of MissLan™ Digital Pregnancy Rapid Test is demonstrated to be 25 mIU/mL. ## b. Linearity/assay reportable range: Linearity is not applicable since this is a qualitative test. The test device was evaluated for high dose or hook effect. {6}------------------------------------------------ ## Hook effect test: Negative urine samples were spiked with varying hCG concentrations (6,250 mIU/mL, 12,500 mIU/mL, 25,000 mIU/mL, 50,000 mIU/mL, 100,000 mIU/mL, 200,000 mIU/mL and 500,000 mIU/mL). All tested concentrations gave a positive result. The results demonstrated that no hook effect was observed at hCG concentration up to 500,000 mIU/mL. ## c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: MissLan™ Digital Pregnancy Rapid Test is calibrated against reference material traceable to WHO International Standard 5th edition, NIBSC code 07/364. ## Stability: A 32-month real time stability test is planned to verify the shelf-life stability of the device. Three batches of products in sealed foil pouch are currently stable for 24 months at 2°C and 30°C, and the real time stability study is still on going. ## d. Specificity and cross reactivity To evaluate specificity, 300 urine samples were collected from healthy, nonpregnant female in pre-menopausal (ages 18~40 years old), peri-menopausal (41~55 years old) and post-menopausal (>55 years old) groups. 100 people for each age group. Both dip and midstream testing are evaluated. No false positive results were observed for any of the age groups. To evaluate cross-reactivity, negative and positive urine samples (0, 5 and 25 mIU/mL hCG) were spiked with potential cross reactants (500 mIU/mL hLH, 1000 mIU/mL hFSH, 1000 uIU/mL hTSH). No cross-reactivity was observed at tested concentration. To evaluate the effect of the hCG ß-core fragment, Negative urine samples (0 and 5 mIU/mL hCG) and positive urine samples (25 and 20,000 mIU/mL hCG) were spiked with hCG ß-core fragment (hCGBcf) at concentrations of 50,000 pmol/L, 125,000 pmol/L, 250,000pmol/L and 500,000pmol/L. The performance of MissLan™ Digital Pregnancy Rapid Test is not affected by hCG ß-core fragment concentrations up to 500,000 pmol/L. ### e. Interfering substance To evaluate potential interferers with MissLan™ Digital Pregnancy Rapid Test, urine samples containing 0, 5 and 25 mIU/mL hCG were spiked with the interfering substance to obtain the certain desired test concentration. No interference effect was observed at the tested concentration shown in table below: | Substance | Concentration | |-----------|---------------| | Glucose | 2000 mg/dL | | Albumin | 2000 mg/dL | | Bilirubin | 40 mg/dL | {7}------------------------------------------------ | Hemoglobin | 1000 mg/dL | |---------------------|------------| | Uric acid | 23.5 mg/dL | | Acetaminophen | 20 mg/dL | | Amoxicillin | 20 mg/dL | | Aspirin | 80 mg/dL | | Gentisic acid | 20 mg/dL | | Salicylic Acid | 20 mg/dL | | Ascorbic acid | 20 mg/dL | | Folic acid | 0.03 mg/dL | | Vitamin B1 | 80 mg/dL | | Atropine | 20 mg/dL | | Caffeine | 20 mg/dL | | Tetracycline | 20 mg/dL | | Ampicillin | 20 mg/dL | | Ibuprofen | 40 mg/dL | | Pregnanediol | 1.5 mg/dL | | ẞ-hydroxybutyrate | 2000 mg/dL | | EDTA | 80 mg/dL | | Ethanol | 1% | | Ketone | 20 mg/dL | | Thiophene | 20 mg/dL | | Benzoylecgonine | 10 mg/dL | | Cannabinol | 10 mg/dL | | Ephedrine | 20 mg/dL | | Phenylpropanolamine | 20 mg/dL | | Phenothiazine | 20 mg/dL | To evaluate the effect of urine pH on the results of MissLan™ Digital Pregnancy Rapid Test, urine samples containing 0, 5 and 25 mIU/mL hCG were tested at pH values of 4, 5, 6, 7, 8 and 9. The results indicated that urine pH ranges between 4 and 9 does not affect the performance of MissLan™ Digital Pregnancy Rapid Test. To evaluate the effect of urine density on the results of MissLan™ Digital Pregnancy Rapid Test, urine samples containing 0, 5 and 25 mIU/mL hCG were tested at density values of 1.000, 1.005, 1.010, 1.015, 1.020, 1.025, 1.030 and 1.035. The results indicated that urine with a relative density of 1.000 to 1.035 does not affect the performance of MissLan™ Digital Pregnancy Rapid Test. ## B. Method comparison study Method comparison with predicate device The performance of the new device was compared to the predicate test. Urine samples were collected from 200 women presenting to test for pregnancy. Approximately half of the 200 women were suspected to be pregnant and most of {8}------------------------------------------------ them are in the early stage of less than 5 weeks. All samples were tested with candidate and predicate devices at three POC sites (3 different professionals using the candidate device and 1 professional using the predicate device at each site). | | Midstream method | Predicate device | | | |------------------|------------------|------------------|----------|-------| | Candidate device | | Positive | Negative | Total | | | Positive | 55 | 0 | 55 | | | Negative | 0 | 45 | 45 | | | Total | 55 | 45 | 100 | Summary midstream testing results | | Predicate device | | | | |------------------|------------------|----------|-------|-----| | Dip method | Positive | Negative | Total | | | Candidate device | Positive | 47 | 0 | 47 | | | Negative | 0 | 53 | 53 | | | Total | 47 | 53 | 100 | ### Summary dip testing results The conformity between MissLan™ Digital Pregnancy Rapid Test (midstream method / dip method) and the predicate device is 100%. ## C. Lay person study First study: 200 women's individual pregnancy status was self-tested. Individuals with varying educational and occupational backgrounds from three sites were chosen for the study. Each subject tested her own urine sample using the device according to the package insert and provided a sample for professional testing. Summary | Midstream method | Professional | | | | |------------------|--------------|----------|-------|-----| | | Positive | Negative | Total | | | Layperson | Positive | 55 | 0 | 55 | | | Negative | 0 | 45 | 45 | | | Total | 55 | 45 | 100 | | Dip method | | Professional | | | |------------|----------|--------------|-------|-----| | | Positive | Negative | Total | | | Layperson | Positive | 47 | 0 | 47 | | | Negative | 0 | 53 | 53 | | | Total | 47 | 53 | 100 | From the above tables, the lay person results showed 100% positive and 100% negative conformity with the professional results. Second study: {9}------------------------------------------------ 200 women's individual pregnancy status was self-tested. Negative urine sample pools were spiked with 5 mIU/mL hCG and 25 mIU/mL hCG. All aliquots were blind labeled by the person who prepared the samples and didn't take part in the sample testing. Both laypersons and professionals use dip method to test the above samples. 100 laypersons tested the 5 mIU/mL hCG aliquots and 100 laypersons tested the 25 mIU/mL hCG aliquots. Each testing site had a study administrator to observe or monitor the studies by laypersons without providing assistance to the participants. #### Summary | hCG<br>Concentration<br>(mIU/mL) | Lay person result | | Professional result | | The percentage of<br>correct results (%) | |----------------------------------|--------------------|--------------------|---------------------|--------------------|------------------------------------------| | | No. of<br>Positive | No. of<br>Negative | No. of<br>Positive | No. of<br>Negative | | | 5 | 0 | 100 | 0 | 100 | 100% | | 25 | 100 | 0 | 100 | 0 | 100% | Each lay person was given a questionnaire to assess the readability of the labeling. The results of the questionnaire reflected that the consumers found the test easy to use and that they did not have trouble understanding the labeling and interpreting the results. ## 11. Conclusion Based on the test principle and performance characteristics of the device including precision, cut-off, interference, specificity, method comparison and lay-user studies of the device, it's concluded that MissLan M Digital Pregnancy Rapid Test is substantially equivalent to the predicate.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...