BRAND NAME 1-STEP PREGNANCY TEST-MAY BE MARKETED UNDER FIRST RESPONSE, ANSWER, OR OTHER BRAND NAMES
Device Facts
| Record ID | K962183 |
|---|---|
| Device Name | BRAND NAME 1-STEP PREGNANCY TEST-MAY BE MARKETED UNDER FIRST RESPONSE, ANSWER, OR OTHER BRAND NAMES |
| Applicant | Armkel, LLC |
| Product Code | LCX · Clinical Chemistry |
| Decision Date | Jul 10, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1155 |
| Device Class | Class 2 |
Intended Use
The Brand Name 1-Step Pregnancy Test is a simple-to-use at-home pregnancy test marketed over-the-counter (OTC) to lay consumers. The test device product is intended for the detection of human chorionic gonadotropin (hCG) in urine as an aid in the detection of early pregnancy.
Device Story
In vitro diagnostic device; immunochemical chromatographic-chromogenic method. Input: urine sample collected via absorbent pad (stream or dip for 5 seconds). Operation: urine moves chromatographically along reagent strip; reconstitutes diffusible reagents; delivers to capture zones. Output: visual result in test window (two pink lines = pregnant; one pink line = not-pregnant). Used at-home by lay consumers. Result provides early pregnancy detection aid.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Plastic stick housing absorbent pad and chromatographic reagent strip. Immunochemical detection principle. Standalone, non-electronic, disposable device.
Indications for Use
Indicated for lay consumers for the detection of human chorionic gonadotropin (hCG) in urine as an aid in the early diagnosis of pregnancy.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
- One-Step CLEARBLUE Easy™ Pregnancy Test
Related Devices
- K961712 — TBD AT-HOME PREGNANCY TEST · Armkel, LLC · May 31, 1996
- K083716 — FIRST RESPONSE EARLY RESULT PREGNANCY TEST (OTC) · Church & Dwight Co., Inc. · Jun 8, 2009
- K180771 — Acro HCG Pregnancy Rapid Test · Acro Biotech, Inc. · May 18, 2018
- K230038 — MissLan® Pregnancy Rapid Test (Strip), MissLan® Pregnancy Rapid Test (Midstream) · Guangzhou Decheng Biotechnology Co., Ltd. · Feb 28, 2023