K210341 · Spd Swiss Precision Diagnostics GmbH · LCX · Mar 10, 2022 · Clinical Chemistry
Device Facts
Record ID
K210341
Device Name
One Step Pregnancy Test
Applicant
Spd Swiss Precision Diagnostics GmbH
Product Code
LCX · Clinical Chemistry
Decision Date
Mar 10, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Intended Use
One Step Pregnancy Test is an over-the-counter lateral flow immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine. The visual test is intended for home use as an aid in early detection of pregnancy. The test is indicated for use from 4 days before the expected period (5 days before the missed period).
Device Story
One Step Pregnancy Test is a lateral flow immunochromatographic sandwich assay; detects human chorionic gonadotropin (hCG) in urine. User applies urine to test cassette via pipette; capillary action draws sample across nitrocellulose membrane. Conjugated monoclonal anti-alpha-hCG antibodies bind hCG to form gold-labeled complexes; immobilized anti-beta-hCG antibodies at test line capture complexes to produce visible colored line. Control line confirms test validity. Used at home by lay users; results read visually within 3-10 minutes. Positive result (two lines) indicates pregnancy; negative result (one line) indicates absence of detectable hCG. Early detection claim allows testing 4 days before expected period. Benefits include rapid, convenient, and accurate pregnancy status confirmation.
Clinical Evidence
Bench testing included precision, specificity, interference, and hook effect studies. Clinical evidence includes a study of 272 women (ages 18-55) evaluating early pregnancy detection from days -9 to 0 relative to expected period. A lay user study (n=120) demonstrated 100% agreement between lay users and professional results. A specificity study in 433 women across pre-, peri-, and post-menopausal groups showed 0% false-positive rate. Method comparison with predicate device (n=300) showed 100% agreement.
Technological Characteristics
Lateral flow sandwich immunoassay; monoclonal antibodies against alpha and beta hCG subunits; nitrocellulose membrane; gold particle conjugate; visual result display. Sensitivity: 25mIU/mL. Sample: urine. Connectivity: None. Energy source: None (passive capillary action).
Indications for Use
Indicated for qualitative detection of hCG in urine to aid in early pregnancy detection. Intended for home use by women of childbearing age. No specific contraindications listed.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
K212418 — hCG One Step Pregnancy Test · Healstone Biotech, Inc. · Jun 30, 2022
K081150 — RAPIDAN OPTIMA EARLY PREGNANCY TEST AND TOYO PREGNANCY TEST · Turklab Medical Devices, Inc. · Jun 16, 2009
K241394 — iHealth® Early Pregnancy Test; iHealth® Early Pregnancy Test Strip · Andon Health Co, Ltd. · Dec 20, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box. Underneath that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 10, 2022
SPD Swiss Precision Diagnostics GmbH % Kamila Przedmojska Senior Regulatory Specialist - Product Lifecycle SPD Development Company Limited Priory Business Park, Stannard Way Bedford, Bedfordshire MK44 3UP United Kingdom
Re: K210341
Trade/Device Name: One Step Pregnancy Test Regulation Number: 21 CFR 862.1155 Regulation Name: Human Chorionic Gonadotropin (hCG) Test System Regulatory Class: Class II Product Code: LCX Dated: December 1, 2021 Received: December 10, 2021
Dear Kamila Przedmojska:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Marianela Perez-Torres, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) k210341
Device Name One Step Pregnancy Test
#### Indications for Use (Describe)
One Step Pregnancy Test is an over-the-counter lateral flow immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine. The visual test is intended for home use as an aid in early detection of pregnancy. The test is indicated for use from 4 days before the day of the expected period (5 days before the day of the missed period).
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for Swiss Precision Diagnostics GmbH. The logo features the letters "SPD" in a stylized font, with the letters arranged in a circular shape. The circle transitions from blue at the top to green at the bottom. To the right of the circle, the text "Swiss Precision Diagnostics GmbH" is written in a simple, sans-serif font.
# 510(k) Summary
| A. Submitted By: | SPD Swiss Precision Diagnostics GmbH<br>47 Route de Saint-Georges<br>Petit-Lancy<br>CH-1213<br>Geneva<br>Switzerland<br>Telephone: +41 580048741 |
|---------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| B. Contact Person: | Kamila Przedmojska<br>Principal Regulatory Specialist<br>SPD Development Company Limited<br>Priory Business Park<br>Bedford<br>MK44 3UP<br>United Kingdom |
| C. Date Prepared: | 09 March 2022 |
| D. Device Name: | One Step Pregnancy Test |
| Product Code:<br>Common name:<br>Classification:<br>Regulation Description:<br>Regulation number:<br>510(k) number: | LCX<br>Kit, Test, Pregnancy, hCG, over the counter<br>Class II<br>Human chorionic gonadotropin (hCG) test system<br>21CFR 862.1155<br>k210341 |
- K012215, QUIK-CHECK™ Home Pregnancy Test E. Predicate Device:
test
{4}------------------------------------------------
#### F. Indication for Use
One Step Pregnancy Test is an over-the-counter lateral flow immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine. The visual test is intended for home use as an aid in early detection of pregnancy. The test is indicated for use from 4 days before the day of the expected period (5 days before the day of the missed period).
#### G. Device Description
The One Step Pregnancy Test is an over-the-counter, visual pregnancy test with sensitivity of 25mIU/mL hCG (human chorionic gonadotropin) and is indicated for use up to 4 days before the expected period (5 days before the missed period). The device eeeploys an immunochromatographic sandwich assay to detect hCG on a lateral flow test strip housed in a cassette. Following sample application using the supplied pipette, the user is required to visually interpret the results displayed as lines in a result window.
#### H. Substantial Equivalence Information
Predicate device name:
QUIK-CHECK™ Home Pregnancy Test
Predicate (k) number:
K012215
{5}------------------------------------------------
## Comparison with predicate:
# Similarities and differences between the One Step Pregnancy Test and the predicate QUIK-CHECK™ Home Pregnancy Test
| Component | One Step Pregnancy Test<br>(Proposed Device) | QUIK-CHECK™ Home<br>Pregnancy Test<br>(Predicate Device) |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Similarities | | |
| Intended Use | Over-the-counter lateral flow<br>immunoassay for the<br>qualitative detection of human<br>chorionic gonadotropin (hCG)<br>in urine. The visual test is<br>intended for home use as an<br>aid in early detection of<br>pregnancy. | Intended for non-professional<br>/ over-the-counter use for<br>the qualitative identification<br>of hCG (human chorionic<br>gonadotropin) in urine to aid<br>in the determination of<br>pregnancy. |
| Target User | Over-the-counter use | Same |
| Device format | Cassette | Same |
| Analyte | hCG | Same |
| Sensitivity | 25mIU/mL | Same |
| Test Principle | Lateral flow qualitative<br>chromatographic immunoassay<br>with visual result display | Same |
| Sample<br>application | Sample applied as drops to<br>sample well via supplied<br>pipette | Same |
| Time to results | 3 minutes | Same |
| Result Display | Visual interpretation of<br>coloured lines in result window | Same |
| Differences | | |
| Traceability | WHO 4th International<br>Standard for hCG | WHO 3rd International<br>Standard for hCG |
| Early Claim | Pregnancy can be detected as<br>early as 4 days before the date<br>of the expected period (5 days<br>before missed period) | Not tested for this device |
{6}------------------------------------------------
# I. Test Principle
The One Step Pregnancy Test is a lateral flow sandwich immunoassay employing monoclonal antibodies that are directed against the alpha and beta sub-units of hCG.
To use the test, the user collects their urine in a clean dry container and uses the supplied pipette to draw up and deposit 4 drops of urine into the sample well.
If hCG is present in the urine sample, it is bound by a conjuqated monoclonal anti α-hCG antibody forming an anti hCG-antibody-gold complex. The sample is drawn through the nitrocellulose membrane by capillary action. At the test (T) line, any hCG is bound by immobilised antiß hCG antibody forming a sandwich, immobilising gold particles as the test line. This sandwich gives rise to a coloured line. If no hCG is present in the sample, there is no binding at the test line and subsequently, no coloured line.
At the control (C) line, rabbit IgG sensitised gold particles encounter the immobilised control antibody line (goat anti-rabbit antibody). The rabbit IgG is bound at the control (C) line within the result window and a coloured line forms to show that the test has worked correctly.
At the read time the appearance of two coloured lines in the result window, the test (T) line and control (C) line, indicates a Pregnant result. The appearance of one coloured line in the result window, the control (C) line, indicates a Not Pregnant result. If there is no control (C) line, the test has not run correctly and is classed as an Invalid result.
# J. Performance characteristics
#### 1. Analytical Performance
#### a) Precision/Reproducibility/Sensitivity
A pooled negative urine was spiked with hCG to provide eight urine standards with the hCG concentrations of <0.5, 12.5, 15, 18, 25, 50, 100 and 200mIU/mL. The eight standards were each tested with 180 devices from four different batches. Each batch was tested by 3 operators over 3 non-consecutive days
{7}------------------------------------------------
The results of the precision and sensitivity study at each concentration of hCG are summarised in the tables below. The results demonstrated that the analytical sensitivity of the One Step Pregnancy Test is 25mIU/mL.
| hCG Standard<br>(mIU/ml) | Total (n) | Pregnant<br>(n) | Not Pregnant<br>(n) | Pregnant<br>Results (%) |
|--------------------------|-----------|-----------------|---------------------|-------------------------|
| <0.5 | 180 | 0 | 180 | 0.0 |
| 12.5 | 180 | 30 | 150 | 16.7 |
| 15 | 180 | 76 | 104 | 42.2 |
| 18 | 180 | 145 | 35 | 80.6 |
| 25 | 180 | 180 | 0 | 100.0 |
| 50 | 180 | 180 | 0 | 100.0 |
| 100 | 180 | 180 | 0 | 100.0 |
| 200 | 180 | 180 | 0 | 100.0 |
| Total | 1440 | 971 | 469 | N/A |
Overall Precision Results for One Step Pregnancy Test
Percentage Pregnant Results for each hCG standard by Batch
| hCG<br>Standard<br>(mIU/<br>ml) | CYG00011 | | CYG00012 | | CYG00013 | | CYG00016 | |
|---------------------------------|----------------------------------|-------------------------|----------------------------------|-------------------------|----------------------------------|-------------------------|----------------------------------|-------------------------|
| | Pregnant/ Not<br>Pregnant<br>(n) | Pregnant<br>Results (%) | Pregnant/ Not<br>Pregnant<br>(n) | Pregnant<br>Results (%) | Pregnant/ Not<br>Pregnant<br>(n) | Pregnant<br>Results (%) | Pregnant/ Not<br>Pregnant<br>(n) | Pregnant<br>Results (%) |
| <0.5 | 0/45 | 0.0 | 0/45 | 0.0 | 0/45 | 0.0 | 0/45 | 0.0 |
| 12.5 | 7/38 | 15.6 | 8/37 | 17.8 | 6/39 | 13.3 | 9/36 | 20.0 |
| 15 | 16/29 | 35.6 | 22/23 | 48.9 | 19/26 | 42.2 | 19/26 | 42.2 |
| 18 | 35/10 | 77.8 | 38/7 | 84.4 | 37/8 | 82.2 | 35/10 | 77.8 |
| 25 | 45/0 | 100 | 45/0 | 100 | 45/0 | 100 | 45/0 | 100 |
| 50 | 45/0 | 100 | 45/0 | 100 | 45/0 | 100 | 45/0 | 100 |
| 100 | 45/0 | 100 | 45/0 | 100 | 45/0 | 100 | 45/0 | 100 |
| 200 | 45/0 | 100 | 45/0 | 100 | 45/0 | 100 | 45/0 | 100 |
The study indicates that the results returned by the One Step Pregnancy Test are not significantly affected by operator, day of testing or batch across the standards tested in the study. The results demonstrated that the analytical sensitivity of the One Step Pregnancy Test is 25mIU/mL.
b) Linearity/assay reportable range:
{8}------------------------------------------------
Not applicable. This is a qualitative device.
#### c) High dose hook effect study:
Negative pooled urine was spiked with hCG to concentrations of <0.5, 25, 250000, 500000, 100000mIU/mL and tested with 5 replicates per each of 3 batches. No hook effect was observed at tested concentrations.
### d) Traceability
The devices are calibrated aqainst the WHO 4th International Standard for Chorionic Gonadotropin (hCG) (NIBSC code 75/589).
#### e) Stability
The claimed shelf life of the device stored in the sealed pouches at room temperature is 24 months. Real time stability studies are ongoing.
#### f) Detection limit (sensitivity)
See Precision / Reproducibility / Sensitivity section
g) Analytical specificity
Structure not-related compounds
#### Interfering substances
The One Step Pregnancy Test devices were tested with potentially interfering substances. Each interfering substance was spiked into nonpregnant pooled urine (<0.5mIU/mL) and 25mIU/mL hCG urine standards.
Each condition was tested with 5 devices from each of three batches for each of the two urine standards. No interference effect was observed at the tested concentration shown in table below:
| Interfering Substance | Concentration |
|--------------------------------|--------------------------|
| Acetylsalicylic acid | 1.0mg/mL |
| Acetone | 1.0mg/mL<br>(0.125% v/v) |
| Albumin | 5mg/mL |
| Interfering Substance | Concentration |
| Ampicillin | 200µg/mL |
| Ascorbic acid | 150µg/mL |
| Atropine | 200µg/mL |
| Bilirubin | 200µg/mL |
| Blood | 0.3% v/v |
| Caffeine | 1.2mg/mL |
| Cannabinol | 100µg/mL |
| Clomiphene citrate | 24µg/mL |
| Cotinine | 40µg/mL |
| Ethanol | 5% v/v |
| E3G | 620ng/mL |
| Glucose | 20mg/mL |
| Gentisic acid | 200µg/mL |
| Haemoglobin | 100µg/mL |
| Hydrochloric acid | 1.25mM |
| Ibuprofen | 100µg/mL |
| Leukocytes | 1x106 cells/mL |
| Oxytetracycline | 300µg/mL |
| Paracetamol<br>(Acetaminophen) | 600µg/mL |
| P3G | 40µg/mL |
| Phenylpropanolamine | 200µg/mL |
| Semen | 5% v/v |
| Sodium hydroxide | 1.25mM |
| Tetracycline | 300µg/mL |
| Urea | 30mg/mL |
| Uric acid | 750µg/mL |
| Urobilinogen | 100µg/mL |
{9}------------------------------------------------
{10}------------------------------------------------
#### Structure related compounds
The One Step Pregnancy Test devices were tested with 3 potential cross reactants. Each potential cross reactant was spiked into non-pregnant pooled urine <0.5mIU/mL and 25mIU/mL hCG urine standard at the following concentration:
| Cross Reactant | Concentration |
|--------------------------------------|---------------|
| Follicle – Stimulating Hormone (FSH) | 1000 mIU/mL |
| Luteinizing Hormone (LH) | 500 mIU/mL |
| Thyroid – Stimulating Hormone (TSH) | 1mIU/mL |
Specificity of ≥99% was demonstrated when tested with potential cross reactants.
#### Effects of urine pH
Effect of urine pH was performed by adjusting non-pregnant pooled urine (<0.5mIU/mL) and 25mIU/mL hCG urine standard to a pH range of 4, 5, 6 (unadjusted control), 7, 8, and 9. Each urine standard was tested with 15 devices, 5 from each of 3 batches per pH condition. The results demonstrated that One Step Pregnancy Test will continue to return a correct result when tested with a urine sample in the pH range of 4 - 9.
#### Effect of urine specific gravity
To test the effect of specific gravity, device was challenged with negative (<0.5mIU/mL) and 25mIU/mL urine standards with the specific gravity of 1.000, 1.005, 1.016, 1.029 and 1.035. The results showed One Step Pregnancy Test will continue to return a correct result in response to changes in specific gravity within the range from 1.000 to ≤1.035.
#### Effect of hCG beta-core fragment (hCGβcf)
To evaluate the effect of hCGβcf, 11 conditions were tested with 5 devices from each of 3 batches, totalling 165 devices. Pooled pregnant urine collected from 6-7 weeks and 9-12 weeks pregnancy were tested with and without hCGβcf spiked up to 1μM. Pooled negative urine spiked with hCG to a concentration representative of 6-7 weeks and 9-12 weeks pregnant
{11}------------------------------------------------
urine samples were tested with and without hCGBcf spiked up to 1µM. Negative pooled urine spiked to 25mIU/mL hCG then spiked with 375pM hCGßcf (a concentration 5 times the molar concentration of intact hCG) was tested. Positive (neqative pooled urine spiked to 25mIU/mL) and nonpregnant (<0.5mIU/mL) controls were also tested.
The results show that the performance of the One Step Pregnancy Test is not affected by high concentrations of hCG ß-core fragment.
h) Assay cut-off
See Limit of Detection section.
# 2. Comparison Study
- a. Method comparison with predicate device:
150 urine samples from pregnant women and 150 urine samples from not pregnant women were tested by trained technicians using devices across three batches of the One Step Pregnancy Test. The same samples were also tested using the predicate device QUIK-CHECK™ Home Pregnancy Test.
The One Step Pregnancy Test had 100% agreement with the QUIK-CHECK™ Home Pregnancy Test and 100% agreement with the clinical status of the volunteers' urine samples.
b. Matrix comparison:
Not Applicable. The device is intended for urine sample only.
# 3. Clinical Studies
- a. Clinical Sensitivity:
Not Applicable
b. Clinical Specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable)
{12}------------------------------------------------
# Detection of hCG in Early Pregnancy Clinical Samples
1197 One Step Pregnancy Test devices were tested across 3 batches with early pregnancy urine samples collected from days 0 to -9 relative to the day of the missed period.
The early pregnancy detection results are summarised in table below:
| Days Relative<br>to Missed<br>Period | Total<br>(n) | Pregnant<br>(n) | Pregnant<br>(%) |
|--------------------------------------|--------------|-----------------|-----------------|
| -9 | 51 | 0 | 0.0 |
| -8 | 66 | 0 | 0.0 |
| -7 | 90 | 12 | 13.3 |
| -6 | 132 | 57 | 43.2 |
| -5 | 132 | 85 | 64.4 |
| -4 | 132 | 121 | 91.7 |
| -3 | 132 | 124 | 93.9 |
| -2 | 132 | 130 | 98.5 |
| -1 | 132 | 131 | 99.2 |
| 0 | 132 | 132 | 100 |
| Not Pregnant<br>Control | 66 | 0 | 0.0 |
#### Lay User Study
Pregnant and not pregnant women volunteers with diverse educational and professional backgrounds and aged over 18 years old participated in the Lay User Study. They tested their own urine sample with the One Step Pregnancy Test according to the instructions for use (IFU) provided.
The same urine sample was also tested by a technician. Volunteer results were compared to their clinical pregnancy status and to the results obtained from trained technicians.
The agreement between lay-user volunteer results and their clinical status with the One Step Pregnancy Test was 100%. There was also 100% agreement between lay-user volunteer results and technician results.
{13}------------------------------------------------
The results are summarised in tables below:
Aqreement Between Volunteer Device Results and Clinical Pregnancy Status
| | Volunteer Result | | | |
|--------------------|------------------|----------|--------------|-------|
| | | Pregnant | Not Pregnant | Total |
| Clinical<br>Status | Pregnant | 21 | 0 | 21 |
| | Not Pregnant | 0 | 99 | 99 |
| | Total | 21 | 99 | 120 |
Agreement Between Volunteer and Technician Results
| | | Volunteer Result | | |
|-----------------|--------------|------------------|--------------|-------|
| | | Pregnant | Not Pregnant | Total |
| Clinical Status | Pregnant | 21 | 0 | 21 |
| | Not Pregnant | 0 | 99 | 99 |
| | Total | 21 | 99 | 120 |
Overall, this study demonstrated that volunteers were able to use the product following the IFU to obtain a valid result.
# Lay User Study on Biobank Samples
This study supports the lay user study by presenting examples of lay users reading genuine pregnant results from samples with a confirmed clinical pregnancy status.
Women volunteers aged over 18 years old participated in the Lay User Study on Biobank Samples. Volunteers tested and read the result of a single randomly selected sample (either clinically confirmed pregnant or not pregnant).
The agreement between lay-user volunteer results and the clinical status of the biobank sample with the One Step Pregnancy Test was 100%.
{14}------------------------------------------------
The results are summarised in table below:
Agreement Between Volunteer Device Results and Clinical Pregnancy Status
| | | Pregnant | Not Pregnant | Total |
|-----------------|--------------|----------|--------------|-------|
| Clinical Status | Pregnant | 70 | 0 | 70 |
| | Not Pregnant | 0 | 9 | 9 |
| | Total | 70 | 9 | 79 |
The overall accuracy for the One Step Pregnancy Test in the hands of volunteers was therefore 100% relative to the clinical pregnancy status of the samples tested.
# Specificity study to determine false-positive result rate
A study was performed to determine the incidence of false positive results among non-pregnant women of peri-menopausal age (41-55 years) and post-menopausal age (>55 years).
150 urine samples were collected from individual women of each cohort and tested by technicians with three batches of the One Step Pregnancy Test.
The results are summarised in table below:
| Cohort | Specificity<br>(Proportion) | Specificity<br>(%) |
|-----------------|-----------------------------|--------------------|
| Peri-menopausal | 150 | 100.0 |
| Post-menopausal | 150 | 100.0 |
The incidence of false positive results among non-pregnant women of premenopausal age were evaluated as part of the Comparison to the Predicate Study.
133 urine samples from not pregnant women aged 18-40 were tested by trained technicians usinq devices across three batches of the One Step Pregnancy Test. The results are summarised in table below:
{15}------------------------------------------------
| Cohort | Specificity<br>(Proportion) | Specificity<br>(%) |
|----------------|-----------------------------|--------------------|
| Pre-menopausal | 133 | 100.0 |
### Clinical Cut-off
Not applicable.
#### Expected value / Reference range
Not applicable.
#### Conclusions
The conclusions drawn from the nonclinical tests demonstrate that the device is safe, effective and performs in accordance with its intended use. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.