K110433 · SPD Development Company Limited · LCX · May 12, 2011 · Clinical Chemistry
Device Facts
Record ID
K110433
Device Name
CLEARBLUE EASY PREGNANCY TEST
Applicant
SPD Development Company Limited
Product Code
LCX · Clinical Chemistry
Decision Date
May 12, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1155
Device Class
Class 2
Intended Use
Clearblue® Plus Pregnancy Test is an over-the-counter urine hCG test which is intended for the detection of pregnancy. The test is indicated for use from four days before the expected period (5 days before the missed period).
Device Story
Rapid, visually read, in vitro diagnostic immunoassay; detects hCG in urine; indicates pregnancy. Modifications: wick dimensions increased for easier midstream sampling; read time extended from 2 to 3 minutes; anti-flood pad added to mitigate excessive urine application; scavenger antibody added to reduce non-specific binding to latex particles; housing and window shapes modified; result key enlarged. Device used by consumers for home pregnancy testing. Output: visual result in window. Benefit: early pregnancy detection.
Clinical Evidence
No clinical data provided; device relies on bench testing to demonstrate analytical performance and substantial equivalence to the predicate.
Technological Characteristics
Lateral flow immunoassay; qualitative detection of hCG in urine. Device utilizes antibody-antigen binding principle. Form factor is a handheld test stick. No electronic components, software, or connectivity. Sterilization not applicable for this diagnostic test type.
Indications for Use
Indicated for qualitative detection of hCG in urine for early pregnancy detection in women.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
K041404 — CLEARBLUE EASY EASY READ PREGNANCY TEST · Unipath , Ltd. · Jun 8, 2004
K023699 — CLEARBLUE EASY EARLY RESULT PREGNANCY TEST · Unipath , Ltd. · Nov 27, 2002
K240025 — Clearblue® Early Detection Pregnancy Test · Spd Swiss Precision Diagnostics GmbH · Jan 31, 2024
K040341 — CLEARBLUE EASY EARLIEST RESULTS PREGNANCY TEST · Unipath , Ltd. · Mar 8, 2004
K013372 — CLEARBLUE EASY EARLY RESULT PREGNANCY TEST · Unipath , Ltd. · Nov 9, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is an abstract symbol of a human figure, represented by three stylized profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
MAY 1 2 2011
SPD Development Company Limited c/o Ms. Tanya Fair Regulatory Affairs Manager Stannard Way, Priory Business Park Bedford, Bedfordshire United Kingdom MK44 3UP
Re: k110433
Trade/Device Name: Clearblue Plus Pregnancy Tesst Regulation Number: 21 CFR 862.1155 Regulation Name: Human Chorionic Gonadotropin (HCG) Test System Regulatory Class: Class II Product Code: LCX Dated: April 28, 2011 Received: May 03, 2011
Dear Ms. Fair:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{1}------------------------------------------------
Page 2 --
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolli-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours.
C.C.
Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use Form
510(k) Number (if known): K110433
Device Name: Clearblue® Plus Pregnancy Test
Indications for Use:
Clearblue® Plus Pregnancy Test is an over-the-counter urine hCG test which is intended for the detection of pregnancy. The test is indicated for use from four days before the expected period (5 days before the missed period).
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) 、
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k): K110433
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.