K110889 · Ind Diagnostics, Inc. · LCX · Jun 20, 2011 · Clinical Chemistry
Device Facts
Record ID
K110889
Device Name
IND ONE STEP HCG URINE PREGNANCY TEST (STRIP)
Applicant
Ind Diagnostics, Inc.
Product Code
LCX · Clinical Chemistry
Decision Date
Jun 20, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Intended Use
IND One Step hCG Urine Pregnancy Tests Device is intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine to help in the early determination of pregnancy. The device is designed for over-the-counter use.
Device Story
Lateral flow immunochromatographic strip test; detects human chorionic gonadotropin (hCG) in urine. Device contains goat anti-mouse IgG, mouse monoclonal anti-hCG antibody A, and colloidal gold-conjugated mouse monoclonal anti-hCG antibody B. User immerses test strip in urine sample; hCG binds to gold-labeled antibody conjugate; complex migrates via lateral flow. Positive result: colored test and control lines; negative result: colored control line only. Designed for over-the-counter use by lay-users; no instrumentation required. Provides rapid visual qualitative result to assist in pregnancy determination.
Clinical Evidence
No clinical data provided; substantial equivalence established via bench testing and performance characteristics typical of lateral flow hCG immunoassays.
Technological Characteristics
Lateral flow immunochromatographic sandwich assay. Materials: goat anti-mouse IgG, mouse monoclonal anti-hCG antibodies, colloidal gold conjugate. Format: test strip. Standardized to WHO 4th International Standard for Chorionic Gonadotropin. Cut-off: >20 mIU/mL hCG. No instrumentation required.
Indications for Use
Indicated for qualitative detection of hCG in urine to aid in early pregnancy determination. Intended for over-the-counter use by individuals seeking pregnancy status.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Related Devices
K974059 — ONE STEP MIDSTREAM HCG PREGNANCY TEST · International Newtech Development, Inc. · Apr 9, 1998
K083716 — FIRST RESPONSE EARLY RESULT PREGNANCY TEST (OTC) · Church & Dwight Co., Inc. · Jun 8, 2009
K984079 — INSTANT-VIEW PREGNANCY URINE DIP-STRIP TEST · Alfa Scientific Designs, Inc. · Feb 16, 1999
K131236 — UCP PREGNANCY CASSETTE TEST, DIP/CARD STRIP, HOME PREGNANCY CASSETTE TEST, MIDSTREAM TEST AND DIP/CARD STRIP · Ucp Biosciences, Inc. · Nov 7, 2013
K230038 — MissLan® Pregnancy Rapid Test (Strip), MissLan® Pregnancy Rapid Test (Midstream) · Guangzhou Decheng Biotechnology Co., Ltd. · Feb 28, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
IND Diagnostic Incorporated c/o Kai Lou Regulatory Assistant 1629 Fosters Way Delta, British Columbia Canada, V3M 6S7
Food & Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993
JUN 2 0 2011
Re: k110889
> Trade Name: IND One Step hCG Urine Pregnancy Test (Strip) Regulation Number: 21 CFR §862.1155 Regulation Name: Human Chorionic Gonadotropin (hCG) Test System Regulatory Class: Class II Product Codes: LCX Dated: May 12, 2011 Received: May 16, 2011
Dear Kai Lou:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{1}------------------------------------------------
#### Page 2 -
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
C.C.
Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
: ذكت :
Enclosure
{2}------------------------------------------------
# Indications for Use
## 510(k) Number (if known): K110889
#### Device Name
IND One Step hCG Urine Pregnancy Test (Strip Format)
### Indications for Use
IND One Step hCG Urine Pregnancy Tests Device is intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine to help in the early determination of pregnancy. The device is designed for over-the-counter use.
Prescription Use _ (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use _ X (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K110889
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.