ATLAS HOME PREGNANCY TEST (MIDSTREAM FORMAT) MODEL 70172001

K102760 · Atlas Medical · LCX · Aug 3, 2011 · Clinical Chemistry

Device Facts

Record IDK102760
Device NameATLAS HOME PREGNANCY TEST (MIDSTREAM FORMAT) MODEL 70172001
ApplicantAtlas Medical
Product CodeLCX · Clinical Chemistry
Decision DateAug 3, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1155
Device ClassClass 2

Intended Use

Atlas Home Pregnancy Test, Midstream (Urine) is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in the urine to aid in the early detection of pregnancy.

Device Story

Device is a midstream lateral flow chromatographic immunoassay for qualitative hCG detection in urine. User immerses absorbent tip in urine stream; sample migrates via capillary action across membrane coated with mouse anti-hCG colloidal gold conjugate and polyclonal anti-hCG antiserum. Presence of hCG results in red band formation at test line; control line confirms proper flow. Used at home by lay users; results visually interpreted at 3 minutes. Provides rapid pregnancy status indication to assist clinical decision-making or personal health monitoring.

Clinical Evidence

Bench testing only. Analytical performance established via precision/reproducibility studies (n=10 replicates per concentration across 3 lots) confirming 25 mIU/mL cutoff. Specificity testing showed no cross-reactivity with LH, FSH, or TSH. Interference testing evaluated pH (4-9), specific gravity (1.000-1.035 g/mL), and various exogenous compounds. Method comparison studies (n=100 and n=51) compared lay-user midstream testing against professional testing on predicate devices, showing 100% agreement. Lay user study (n=40) confirmed ability to correctly interpret results near cutoff.

Technological Characteristics

Midstream format chromatographic immunoassay. Components: membrane with polyclonal anti-hCG antibodies, conjugate pad with colloidal gold-conjugated mouse anti-hCG antibodies, and control line with monoclonal goat anti-mouse antibodies. Detection limit: 25 mIU/ml. Visual readout. No electronic components or software.

Indications for Use

Indicated for qualitative detection of hCG in urine to aid early pregnancy detection in women of childbearing age. For over-the-counter home use.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K10276 D AUG - 3 2011 # 510(k) Summary #### 1- Name and Address of Submitter | The submitter's name: | Atlas Medical. | |-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------| | Address: | Atlas Medical, P.O. Box 204,<br>Sahab Free Zone Area, King Abdullah II Industrial City<br>Amman 11512 Jordan. | | | Tel: +962 6 40 26 468<br>Fax: +962 6 40 22 588<br>Email: quality.assurance4@atlas-site.co.uk<br>Contact person: Noor Al-ahwal (QA Manager) | This 510(k) summary was prepared on April 26th, 2011. #### 2- The name of the device | Trade Name: | Atlas Home Pregnancy Test (Midstream format) | |--------------------------------------------|----------------------------------------------| | Common Name: | Pregnancy Test Kit, OTC. | | Classification Name: | Kit, Test, Pregnancy, hCG, Over The Counter. | | Device classification (Regulation Number): | 21 CFR Part 862.1155. | | Regulatory Class: | II | | Product Code: | LCX. | | Panel: | 75 - Clinical Chemistry. | | Format: | Midstream. | ### 3- Equivalence legally marketed device: Atlas Home Pregnancy Test (Midstream Format) has been determined to be substantially equivalent to ACON™ One Step Test Device (K993317) currently in commercial distribution by Acon Laboratories, Inc. ### 4- Device description: Home Pregnancy Test (Midstream Format) is an OTC single use rapid Atlas chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. King Abdullah The Second Industrial Estate, Street 19, Sahab Free Zone Area, P.O. Box: 201 Amman 11512 Jordan. Tel: +962 (0)6 4026468, Fax: +962 (0)6 4022588 {1}------------------------------------------------ ## 5- Intended use: Atlas Home Pregnancy Test, Midstream (Urine) is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in the urine to aid in the early detection of pregnancy. # 6- Test Format: Atlas Home Pregnancy Test is a Midstream format. # 7- Test Principle: Atlas Home Pregnancy Test (Midstream Format) is a rapid qualitative chromatographic assay, utilizes a combination of antibodies including hCG antibody to detect elevated levels of hCG. The test consists of a membrane coated with a polyclonal anti-hCG Abs as a test line, colored conjugate of mouse anti-hCG Abs colloidal gold conjugate act as a conjugate releasing pad, and the control line which contains a monoclonal goat anti-mouse Abs. When the urine sample is applied it will react with the colored conjugate in the conjugate releasing pad, then the mixture migrates upwards on the membrane chromatographically to react with hCG Abs on the membrane and generates a colored band, which indicates a positive result. As the mixture continues to migrate across the membrane to the goat anti-mouse region colored band will always appear to serve as a procedural control and to show that the test has been performed properly. # 8- Major ingredients of Atlas Home Pregnancy Test (Midstream Format): - Mouse anti- hCG - - -Polyclonal anti-hCG Abs - Monoclonal goat anti-mouse Abs - - -Colloidal gold conjugated with mouse anti-hCG Abs. # 9- Comparison with predicate device Feature Comparison of Atlas Home Pregnancy Test (Midstream Format) with ACONTM Midstream Pregnancy Test. | Parameter | Atlas Home Pregnancy Test<br>(Midstream Format) | ACONTM One Step Pregnancy<br>Test Device. (K993317) | |--------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------| | Intended Use | Home Use (OTC) test for the<br>qualitative detection of human<br>chorionic gonadotropin (hCG) in<br>urine | Home Use (OTC) test for the<br>qualitative detection of human<br>chorionic gonadotropin (hCG) in<br>urine | | Indication for Use | Early detection of pregnancy | Early detection of pregnancy. | | Specimen | Urine | Urine | | Format | Test strip in a midstream device. | Test strip in a plastic device. | | Methodology | Colloidal Gold Immunoassay<br>(Membrane particle assay) | Colloidal Gold Immunoassay<br>(Membrane particle assay) | | Storage | 36-86°F | 59-86°F | | Test Time | 3 minutes. | 3 minutes. | King Abdullah The Second Industrial Estate, Street 19, Sahab Free Zone Area, P.O. Box: 201 Amman 11512 Jordan. Tel: +962 (0)6 4026468, Fax: +962 (0)6 4022588 {2}------------------------------------------------ | Detection limit | 25 mIU/ml | 25 mIU/ml | |-----------------|------------------------------------------------|------------------------------------------------| | Accuracy | ≥99% | ≥99% | | Specificity | No interference when tested with FSH, LH & TSH | No interference when tested with FSH, LH & TSH | | Standardization | WHO Third International Standard. | WHO Third International Standard. | The above comparison is based on the Performance Testing - Bench study and clinical studies results included in this file. The study includes results accuracy conducted for both tests. The results show that Atlas Home Pregnancy Test compares well with the ACON™ One Step Pregnancy Test. ### Conclusion Atlas Home Pregnancy .Test (Midstream Format) is found to be substantially equivalent to the predicate device (ACON1M One Step Pregnancy Test). Risk Analysis as per ISO14971 "Medical Devices - Application of risk management to medical devices" has been conducted on Atlas Home Pregnancy Test (Midstream Format) and had showed that possible risks encountered with the use of the product is very mimimal and the benefits outweigh risks by far. {3}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three figures standing underneath it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circle around the caduceus. The logo is black and white. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 AUG 0 3 2011 Atlas Medical c/o Nour Al-ahwal King Abdullah II Industrial Estate Street 19, Sahab Free Zone Area P.O. Box 201 Amman, Jordan 11512 k102760 Re: > Trade Name: Atlas Home Pregnancy Test (Midstream Format) Regulation Number: 21 CFR §862.1155 Regulation Name: Human Chorionic Gonadotropin (HCG) test system Regulatory Class: Class II Product Code: LCX Dated: July 23, 2011 Received: July 26, 2011 Dear Nour Al-ahwal: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ ## Page 2 - If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to paremarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its than the (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. C.C. Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health : 评论: Enclosure {5}------------------------------------------------ # Indication for Use 510(k) Number (if known): 长 1027 6 0 Device Name: Atlas Home Pregnancy Test. (Midstream Format) Indication for Use: Atlas Home Pregnancy Test (Midstream Format) is a home use (OTC) rapid chromatographic immunoassay for the qualitative detection of human Chorionic Gonadotropin (hCG) in the urine to aid in the early detection of pregnancy. Prescription Use NO (21 CFR Part 801 Subpart D) And/or Over The Counter Use Yes (21 CFR Part 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUO ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Rute Olson Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510 (k) k 102760
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...