K172512 · Acro Biotech, Inc. · LCX · Feb 2, 2018 · Clinical Chemistry
Device Facts
Record ID
K172512
Device Name
ACRO HCG Pregnancy Rapid Test
Applicant
Acro Biotech, Inc.
Product Code
LCX · Clinical Chemistry
Decision Date
Feb 2, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1155
Device Class
Class 2
Intended Use
The ACRO hCG Pregnancy Rapid Test is used by the consumer for the visual qualitative detection of Human Chorionic Gonadotropin (hCG) in urine. It is used to detect the pregnancy of HCG in urine as an indication of pregnancy.
Device Story
Lateral flow immunoassay for qualitative hCG detection in urine; modified version of MiniStick Pregnancy Test (k960733). Modifications include: addition of absorbent sample application tip; updated storage conditions (35-86°F); adjusted read time (3-10 minutes); renamed to ACRO HCG Pregnancy Rapid Test. Device used for point-of-care or home pregnancy testing. User applies urine sample to absorbent tip; hCG presence indicated by visual color change on test strip. Results assist in pregnancy confirmation.
Clinical Evidence
Bench testing only; design control activities and risk analysis performed to validate modifications.
Technological Characteristics
Lateral flow sandwich immunochromatographic assay. Components include nitrocellulose membrane, colloidal gold conjugate pad, and sample pad. Uses monoclonal hCG antibodies. Form factor is a midstream test strip in plastic housing with optional absorbent tip. Qualitative visual readout. Traceable to WHO 4th International Standard.
Indications for Use
Indicated for the qualitative detection of human chorionic gonadotropin (hCG) in urine for the early detection of pregnancy.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
MiniStick Pregnancy Test (k960733)
Related Devices
K233624 — HCG One Step Pregnancy Test Strip OTC, HCG One Step Pregnancy Test Strip Rx, HCG One Step Pregnancy Test Midstream OTC, HCG One Step Pregnancy Test Midstream Rx · Hangzhou Aichek Medical Technology Co., Ltd. · Apr 12, 2024
K180771 — Acro HCG Pregnancy Rapid Test · Acro Biotech, Inc. · May 18, 2018
K192843 — INNOVITA HCG Pregnancy Rapid Test Strip, INNOVITA HCG Pregnancy Rapid Test Cassette, INNOVITA HCG Pregnancy Rapid Test Midstream · Innovita (Tangshan) Biological Technology Co., Ltd. · Nov 22, 2019
K150063 — One Step hCG Pregnancy Test · Coretests, Inc. · Oct 2, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 2, 2018
ACRO Biotech, Inc. % Feng-Yu Lee Principal Regulatory Consultant IVDD Regulatory Consultant 29222 Rancho Viejo Road, Suite 218 San Juan Capistrano, CA 92675
Re: K172512
Trade/Device Name: ACRO HCG Pregnancy Rapid Test Regulation Number: 21 CFR 862.1155 Regulation Name: Human chorionic gonadotropin (HCG) test system Regulatory Class: Class II Product Code: LCX Dated: January 12, 2018 Received: January 26, 2018
Dear Feng-Yu Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR
{1}------------------------------------------------
Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Kellie B. Kelm -S
for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) k172512
Device Name ACRO HCG Pregnancy Rapid Test
Indications for Use (Describe)
The ACRO hCG Pregnancy Rapid Test is used by the consumer for the visual qualitative detection of Human Chorionic Gonadotropin (hCG) in urine. It is used to detect the pregnancy of HCG in urine as an indication of pregnancy.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR \$807.92.
The assigned 510(k) number is: k172512
- 1. Submitter's Identification: Acro Biotech, Inc 9500 Seventh Street, Suite M Rancho Cucamonga, CA 91730 Phone Number: 1-909-4666892 FAX Number: 1-909-4666892
c/o IVDD Regulatory Consultant 29222 Rancho Viejo Road, Suite 218 San Juan Capistrano, CA 92675 Contact Person: Feng-Yu Lee Phone Number: 1-949-218-0929 Fax Number: 1-949-218-0928
Date Summary Prepared: February 2nd, 2018
- 2. Name of the Device: ACRO HCG Pregnancy Rapid Test,
- 3. Common or Usual Name: Blood Glucose Monitoring System
| Product Code | Classification | Regulation Section | Panel |
|---------------------------------------------------------|----------------|----------------------------------------------------------------------|--------------------------|
| LCX; Kit, Test,<br>Pregnancy, HCG, Over-<br>the-Counter | Class II | 21 CFR 862.1155<br>Human Chorionic Gonadotropin<br>(HCG) test system | Clinical<br>Chemistry 75 |
#### 4. Device Description:
The ACRO HCG Pregnancy Rapid Test is designed to be tested in midstream mode. The ACRO HCG Pregnancy Rapid Test consists of a single test strip encased in plastic device housing, with or without an additional absorbent tip. The result is generated by immersing the tip in the urine stream for a sufficient amount of time to absorb an adequate sample volume. It is a rapid, one-step chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the detection of pregnancy. The test utilizes a combination of antibodies including a monoclonal hCG antibody to selectively detect elevated levels of hCG.
- 5. Intended Use:
{4}------------------------------------------------
## Acro Biotech, Inc
The ACRO hCG Pregnancy Rapid Test is used by the consumer for the visual qualitative detection of Human Chorionic Gonadotropin (hCG) in urine. It is used to detect the pregnancy of HCG in urine as an indication of pregnancy.
- 6. Predicate Device Information:
The subject devices are substantially equivalent to the brand of Rightest Blood Glucose Monitoring System noted below.
MiniStick Pregnancy Test Name: Device Company: ACRO Biotech, Inc. 510(K) Number: K960733
- 7. Comparison to Predicate Devices:
The ACRO HCG Pregnancy Rapid Test is a modified model of a cleared device, the MiniStick Pregnancy Test, k960733. The changes to the cleared device include addition of an absorbent tip, result reading time and device storage condition.
There is no change to the indications for use nor to the fundamental scientific technology of the device.
| Similarities | | |
|-------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|----------------------------|
| Item | Candidate | Predicate Device (k960733) |
| Intended Use | Rapid qualitative detection of<br>hCG to aid in the early detection<br>of pregnancy. | Same |
| Cutoff | 25 mIU/ml | Same |
| Test Principle | Lateral flow Sandwich<br>Immunochromatographic Assay | Same |
| Positive result | 2 colored lines | Same |
| Negative result | 1 colored line | Same |
| Detection reagent | Colloidal gold | Same |
| Specificity | Negative at:<br>hLH: 500mIU/ml<br>hFSH: 1000mIU/mL<br>hTSH: 1000 $ \mu $ IU/mL | Same |
| Traceability | WHO 4th International Standard | Same |
| Critical Raw<br>Materials (source)<br>and Preparation<br>Formulas | Antibody and Antigen<br>Nitrocellulose Membrane<br>Colloidal Gold Conjugate Pad<br>Sample Pad | Same |
| Component Strip | Width and Length | Same |
| Critical Production<br>Process | Membrane spray/coating<br>Conjugate Pad<br>Sample pad treatment | Same |
Substantial Equivalence Comparison
{5}------------------------------------------------
| | Differences | |
|---------------------|------------------------------------------------------------------------|---------------------------------|
| Storage Temperature | 35 - 86°F (2 – 30 °C) | Below 86°F (30 °C) |
| Read time | 3 to 10 minutes | After 5 minutes |
| Device Format | Midstream without Absorbent Tip<br>and<br>Midstream with Absorbent Tip | Midstream without Absorbent Tip |
- 8. Summary of Verification and Validation activities Lay user performance, Precision, Time Flex, Storage and Stability Studies were conducted and passed acceptance criteria to support the changes.
#### 9. Conclusions from Verification and Validation Activities:
The results of all verification and validation activities demonstrate that the candidate device is substantially equivalent to the cleared predicate MiniStick Pregnancy Test (k960733). The differences between the modified device and the cleared predicate device do not raise any new issue of safety and effectiveness.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.