HCG One Step Pregnancy Test Strip OTC, HCG One Step Pregnancy Test Strip Rx, HCG One Step Pregnancy Test Midstream OTC, HCG One Step Pregnancy Test Midstream Rx
HCG One Step Pregnancy Test Strip OTC, HCG One Step Pregnancy Test Strip Rx, HCG One Step Pregnancy Test Midstream OTC, HCG One Step Pregnancy Test Midstream Rx
Applicant
Hangzhou Aichek Medical Technology Co., Ltd.
Product Code
LCX · Clinical Chemistry
Decision Date
Apr 12, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Indications for Use
The hCG One Step Pregnancy Test Strip Rx is intended for qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in early detection of pregnancy. For in Vitro Diagnostic Use Only. For prescription use. The hCG One Step Pregnancy Test Midstream Rx is intended for qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in early detection of pregnancy. For in Vitro Diagnostic Use Only. For prescription use. The HCG One Step Pregnancy Test Strip OTC is intended for qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in early detection of pregnancy. For in Vitro Diagnostic Use Only. For over-the-counter use. The HCG One Step Pregnancy Test Midstream OTC is intended for qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in early detection of pregnancy. For in Vitro Diagnostic Use Only. For over-the-counter use.
Device Story
Rapid, one-step lateral flow immunoassay for qualitative hCG detection in urine; aids early pregnancy detection. Device formats: test strip or midstream (plastic card with absorbent tip). User applies urine via dip (strip/midstream) or direct stream (midstream). Principle: hCG in urine reacts with anti-βHCG antibody-colloidal gold conjugate; captured by immobilized anti-αHCG antibody on nitrocellulose membrane. Results read visually: two colored lines (C and T) indicate positive; one line (C) indicates negative; no C line indicates invalid. Used in home or point-of-care settings by patients or healthcare professionals. Output provides binary pregnancy status (pregnant/not pregnant) to inform clinical decision-making.
Clinical Evidence
Bench testing and lay-user study. Precision/reproducibility study (n=90 per concentration) confirmed 25 mIU/mL sensitivity. Method comparison study (n=330) against predicate showed 100% accuracy, sensitivity, and specificity. Lay-user study (n=330) demonstrated 100% agreement between lay-user and professional results; Flesch-Kincaid grade level 7 readability.
Technological Characteristics
Lateral flow immunoassay. Components: nitrocellulose membrane, sample pad, monoclonal alpha and beta HCG antibodies. Format: strip or midstream. Storage: 2-30°C. Traceability: WHO 6th International Standard (NIBSC 18/244). Qualitative visual readout.
Indications for Use
Indicated for qualitative detection of hCG in urine to aid in early pregnancy detection. Intended for both prescription and over-the-counter use.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K233624
B Applicant
Hangzhou Aichek Medical Technology Co., Ltd.
C Proprietary and Established Names
HCG One Step Pregnancy Test Strip OTC, HCG One Step Pregnancy Test Strip Rx, HCG One Step Pregnancy Test Midstream OTC, HCG One Step Pregnancy Test Midstream Rx
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LCX | Class II | 21 CFR 862.1155 - Human Chorionic Gonadotropin (HCG) Test System | CH - Clinical Chemistry |
| JHI | Class II | 21 CFR 862.1155 - Human chorionic gonadotropin (HCG) test system | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
New device
B Measurand:
Human chorionic gonadotropin (hCG)
C Type of Test:
Qualitative chromatographic immunoassay
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The HCG One Step Pregnancy Test Strip OTC is intended for qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in early detection of pregnancy. For in Vitro Diagnostic Use Only. For over-the-counter use.
The HCG One Step Pregnancy Test Strip Rx is intended for qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in early detection of pregnancy. For in Vitro Diagnostic Use Only. For prescription use.
The HCG One Step Pregnancy Test Midstream OTC is intended for qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in early detection of pregnancy. For in Vitro Diagnostic Use Only. For over-the-counter use.
The HCG One Step Pregnancy Test Midstream Rx is intended for qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in early detection of pregnancy. For in Vitro Diagnostic Use Only. For prescription use.
C Special Conditions for Use Statement(s):
Rx and OTC
D Special Instrument Requirements:
None
IV Device/System Characteristics:
A Device Description:
The HCG One Step Pregnancy Test Strip OTC, HCG One Step Pregnancy Test Strip Rx, HCG One Step Pregnancy Test Midstream OTC and HCG One Step Pregnancy Test Midstream Rx are lateral flow immunoassays for the qualitative detection of human chorionic gonadotropin (hCG). The tests are available in two formats: 1) The midstream format, which consists of a single test strip assembled in a plastic card, with an absorbent tip, and 2) the strip format, which is a single test strip. The strip format tests (HCG One Step Pregnancy Test Strip OTC and HCG One Step Pregnancy Test Strip Rx) are exactly the same and the midstream format tests (HCG One Step Pregnancy Test Midstream OTC and HCG One Step Pregnancy Test Midstream Rx) are exactly the same.
B Principle of Operation:
The tests are rapid, one-step lateral flow immunoassays that utilize a combination of antibodies including a monoclonal hCG antibody to qualitatively detect hCG in human urine samples. The tests come in two formats, the midstream format and the strip format. Testing is conducted by immersing the absorbent tip in urine (midstream format and strip format) or urinating on the absorbent tip (midstream format) and obtaining the result from the colored lines.
K233624 - Page 2 of 9
{2}
After application of the urine specimen, the hCG within the urine reacts with the anti-βhCG antibody-colloidal gold conjugate to form a compound. The compound is captured by the anti-αhCG antibody immobilized on the test area, then a colored line will form in the test line region. A colored line will always develop in the control line region if sufficient sample volume has been applied to the test strip. The test result is shown in the result window and read visually 3-10 minutes after urine application. Two distinct colored lines, one line in the control line region and another in the test line region, indicate a positive test result (pregnant). Absence of a colored line in the control line region and only a colored line in the control line region indicates a negative test result (not pregnant). Absence of a colored line in the test line region, even in the presence of a colored line in the test line region, indicates an invalid test result.
V Substantial Equivalence Information:
A Predicate Device Name(s): One Step HCG Urine Pregnancy Test
B Predicate 510(k) Number(s): K043443
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K233624 | K043443 |
| --- | --- | --- |
| Device Trade Name | HCG One Step Pregnancy Test Strip Rx
HCG One Step Pregnancy Test Strip OTC
HCG One Step Pregnancy Test Midstream Rx
HCG One Step Pregnancy Test Midstream OTC | One Step HCG Urine Pregnancy Test |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | For qualitative detection of hCG in urine to aid in the determination of pregnancy. | Same |
| Product Usage | For over-the-counter use and Prescription use. | Same |
| Specimen Type | Urine | Same |
| Test Principle | Chromatographic immunoassay | Same |
| Analytical Sensitivity | 25 mIU/mL | Same |
| General Device Characteristic Differences | | |
| Read Time | 3-10 minutes | 3-5 minutes |
| Traceability | WHO 6th International Standard | WHO 3rd International Standard |
K233624 - Page 3 of 9
{3}
VI Standards/Guidance Documents Referenced:
None
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Urine samples from non-pregnant females (negative) were spiked with the hCG WHO 6th IS reference material to provide seven urine samples with hCG concentrations of 0, 12.5, 18.75, 25, 37.5, 50 and 100 mIU/mL. Each sample was tested using 3 lots of each device (i.e., the strip format and the midstream format of the test). Testing on the midstream format was performed using both the dip sampling method and midstream sampling method. Tests were performed in replicates of 2 per day, at 3 sites, by 3 operators, over 15 days for each sample concentration. A total of 90 replicates were tested at each hCG concentration for each format of the device. The sensitivity was determined to be 25mIU/mL hCG. The results are summarized in the tables below:
Summary Precision Results Using the Strip Format (dip sampling method)
| hCG Concentration (mIU/mL) | Lot 1 | | Lot 2 | | Lot 3 | | Total Result | % Positive | % Negative | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Site 1 | | Site 2 | | Site 3 | | | | | |
| | Operator 1 | | Operator 2 | | Operator 3 | | | | | |
| | + | - | + | - | + | - | + | | | - |
| 0 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0.0% | 100.0% |
| 12.5 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0.0% | 100.0% |
| 18.75 | 11 | 19 | 12 | 18 | 11 | 19 | 34 | 56 | 37.8% | 62.2% |
| 25 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100.0% | 0.0% |
| 37.5 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100.0% | 0.0% |
| 50 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100.0% | 0.0% |
| 100 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100.0% | 0.0% |
Summary Precision Results Using the Midstream Format (dip sampling method)
| hCG Concentration (mIU/mL) | Lot 1 | | Lot 2 | | Lot 3 | | Total Result | % Positive | % Negative | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Site 1 | | Site 2 | | Site 3 | | | | | |
| | Operator 1 | | Operator 2 | | Operator 3 | | | | | |
| | + | - | + | - | + | - | + | | | - |
| 0 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0.0% | 100.0% |
| 12.5 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0.0% | 100.0% |
| 18.75 | 13 | 17 | 12 | 18 | 10 | 20 | 35 | 55 | 38.9% | 61.1% |
| 25 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100.0% | 0.0% |
| 37.5 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100.0% | 0.0% |
| 50 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100.0% | 0.0% |
| 100 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100.0% | 0.0% |
K233624 - Page 4 of 9
{4}
Summary Precision Results Using the Midstream Format (midstream sampling method)
| hCG Concentration (mIU/mL) | Lot 1 | | Lot 2 | | Lot 3 | | Total Result | % Positive | % Negative | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Site 1 | | Site 2 | | Site 3 | | | | | |
| | Operator 1 | | Operator 2 | | Operator 3 | | | | | |
| | + | - | + | - | + | - | + | | | - |
| 0 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0.0% | 100.0% |
| 12.5 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0.0% | 100.0% |
| 18.75 | 13 | 17 | 12 | 18 | 10 | 20 | 35 | 55 | 38.9% | 61.1% |
| 25 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100.0% | 0.0% |
| 37.5 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100.0% | 0.0% |
| 50 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100.0% | 0.0% |
| 100 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100.0% | 0.0% |
2. Linearity:
Linearity is not applicable since this is a qualitative test.
3. Analytical Specificity/Interference:
Cross-Reactivity
The candidate devices were tested for potential cross-reactivity from luteinizing hormone (LH), follicle-stimulating hormone (FSH), and thyroid-stimulating hormone (TSH). A pooled human urine sample was used to prepare samples with hCG levels of 0 mIU/mL, 25 mIU/mL and 100 mIU/mL that were then spiked with each cross reactant up to the following concentrations: 750 mIU/mL LH, 1000 mIU/mL FSH or 1 mIU/mL TSH. Spiked samples were tested in replicates of 15 using 3 lots of the candidate device for each format (5 replicates per lot). The results demonstrated no cross reactivity with FSH at 1000mIU/mL, and TSH at 1 mIU/mL. The highest concentration of LH at which no cross-reactivity was observed was 500 mIU/mL.
Interference Study
To evaluate potential interference, urine samples from non-pregnant females were used to prepare hCG negative (0 mIU/mL hCG) and positive (25 mIU/mL hCG) samples that were then spiked with final concentrations of the potential interferents listed in table below. For each substance, 15 replicates were tested on 3 device lots (5 replicates tested by each lot). The results demonstrated no interference from substances at the concentrations shown in the table below.
K233624 - Page 5 of 9
{5}
Interfering substances and concentrations
| Interfering Substance | Concentration | Interfering Substance | Concentration |
| --- | --- | --- | --- |
| Acetaminophen | 20mg/dL | Hydroxybutyric acid | 2000mg/dL |
| Acetone | 20mg/dL | Ibuprofen | 40mg/dL |
| Acetylsalicylic acid | 20mg/dL | L-ephedrine hydrochloride | 20mg/dL |
| Albumin | 2000mg/dL | Methadone | 20mg/dL |
| Ampicillin | 20mg/dL | Morphin | 10mg/dL |
| Ascorbic acid | 20mg/dL | Nicotine | 10mg/dL |
| Atropine | 20mg/dL | Phenylpropanolamine | 20mg/dL |
| Bilirubin | 2mg/dL | Proephedrine hydrochloride | 20mg/dL |
| Caffeine | 20mg/dL | Salicylic acid | 20mg/dL |
| Cannabinol | 10mg/dL | Tetracycline | 20mg/dL |
| Gentisic acid | 20mg/dL | Uric acid | 20mg/dL |
| Glucose | 2000mg/dL | Ethanol | 1% |
| Hemoglobin | 250mg/dL | Vitamin B | 800μg/mL |
| Blood | 15% | | |
## Effect of hCG β-core fragment
Urine samples from non-pregnant females were used to prepare samples with hCG levels of 0 mIU/mL (negative samples) and 25 mIU/mL (positive samples). The positive and negative samples then were spiked with hCG β-core fragment at concentrations of 62,500, 125,000, 250,000, 500,000, 1,000,000 pmol/L. All samples were tested in 15 replicates on 3 device lots (5 replicates per lot). The results demonstrated that the candidate devices are not affected by concentrations of hCG β-core fragment up to 1,000,000 pmol/L.
## Effect of urine pH
To evaluate potential interference from changes in urine pH, urine samples containing 0 mIU/mL and 25 mIU/mL hCG were tested in 15 replicates on 3 lots of device (5 replicates per lot) using samples at pH values of 4.0, 5.0, 6.0, 7.0, 8.0 and 9.0. The results demonstrated that samples within the pH range of 4.0 to 9.0 do not interfere with either positive or negative results from the devices.
## Effect of urine specific gravity
To evaluate potential interference from changes in specific gravity, urine samples containing 0 mIU/mL and 25 mIU/mL hCG were adjusted to specific gravities of 1.000, 1.015, 1.018, 1.020, 1.022, 1.026, 1.031, 1.035 and tested in 15 replicates on 3 lots of the device (5 replicates per lot). The results demonstrated that changes in specific gravity, ranging from 1.000-1.035, do not interfere with either positive or negative results from the devices.
## High dose hook effect study
To evaluate high dose hook effect, negative urine sample spiked with hCG at 25 mIU/mL, 100 mIU/mL, 500 mIU/mL, 1,000 mIU/mL, 100,000 mIU/mL, 500,000 mIU/mL, 1,000,000 mIU/mL, and 2,000,000 mIU/mL. The samples were tested on 3 lots of devices (5 replicates
K233624 - Page 6 of 9
{6}
tested by each lot). The results showed no hook effect for hCG concentrations up to 2,000,000 mIU/mL.
4. Assay Reportable Range:
Not applicable. This is a qualitative test.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The tests are calibrated against reference material traceable to WHO International Standard 6th edition, NIBSC code 18/244.
6. Detection Limit:
The detection limit was determined in the precision study (see Section VII.A.1 above).
7. Assay Cut-Off:
The cutoff for a positive test is 25 mIU/mL. See Precision/Reproducibility section above (Section VII.A.1).
B Comparison Studies:
1. Method Comparison with Predicate Device:
The performance of the two device formats (strip, midstream) of the test was compared to the predicate device. Testing was performed by laboratory professionals at three sites with urine samples from 330 women between the age of 18-45 who suspect they are in the early stages of pregnancy. All pregnant women were in the early stages of pregnancy (less than 5 weeks). All samples were tested with candidate and predicate devices at three sites by 3 different professionals (a total of 9 different professionals conducted the study). The results are summarized below.
Summary of Method Comparison Results Using the Strip Format (dip sampling method)
| | Predicate device | | Total | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | |
| Candidate device | Positive | 58 | 0 | 58 |
| | Negative | 0 | 52 | 52 |
| Total | | 58 | 52 | 110 |
Summary of Method Comparison Results Using Midstream Format (dip sampling method)
| | Predicate device | | Total | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | |
| Candidate device | Positive | 59 | 0 | 59 |
| | Negative | 0 | 52 | 52 |
| Total | | 59 | 52 | 111 |
K233624 - Page 7 of 9
{7}
Summary of Method Comparison Results Using Midstream Format (midstream sampling method)
| | Predicate device | | Total | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | |
| Candidate device | Positive | 58 | 0 | 58 |
| | Negative | 0 | 51 | 51 |
| Total | | 58 | 51 | 109 |
The data shows that the agreement with the predicate device was 100%.
2. **Matrix Comparison:**
Not applicable. This device is for testing with human urine only.
C **Clinical Studies:**
1. **Clinical Sensitivity:**
Not applicable.
2. **Clinical Specificity:**
Not applicable.
3. **Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):**
**Lay-User Study**
A lay user study was conducted at 3 sites with 330 women who suspect they are in the early stages of pregnancy with diverse educational and occupational backgrounds who were between the ages of 18 and 45. This included 110 lay-users using the strip format of the test and 220 lay-users using the midstream format (111 users in dip sampling method and 109 in midstream sampling method). The lay users tested their own urine sample and provided a sample for professional testing. The data shows that the agreement between lay-user results and professional results was 100%. The results are summarized below.
Summary of Lay User Testing of Strip Format
| Candidate device | Professional test results | | Total | |
| --- | --- | --- | --- | --- |
| | | Positive | | Negative |
| Lay user test results | Positive | 58 | 0 | 58 |
| | Negative | 0 | 52 | 52 |
| Total | | 58 | 52 | 110 |
K233624 - Page 8 of 9
{8}
Summary of Lay User Testing of Midstream Format (dip sampling method)
| Candidate device | Professional test results | | Total | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | |
| Lay user test results | Positive | 59 | 0 | 59 |
| | Negative | 0 | 52 | 52 |
| Total | | 59 | 52 | 111 |
Summary of Lay User Testing of Midstream (midstream sampling method)
| Candidate device | Professional test results | | Total | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | |
| Lay user test results | Positive | 58 | 0 | 58 |
| | Negative | 0 | 58 | 58 |
| Total | | 58 | 59 | 109 |
Ease of use of the candidate devices was assessed through a questionnaire that was completed at the end of the study. The questionnaire results indicated that lay-users found the tests easy to use, the results clear and easy to read and the instructions for use easy to understand. A Flesch-Kincaid reading analysis was performed on each OTC package insert and the score demonstrated a reading Grade Level of 7.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
Not applicable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K233624 - Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.