Shinetell PlusTM Digital Early Pregnancy Test

K251053 · Hangzhou AllTest Biotech Co., Ltd. · LCX · Jul 15, 2025 · Clinical Chemistry

Device Facts

Record IDK251053
Device NameShinetell PlusTM Digital Early Pregnancy Test
ApplicantHangzhou AllTest Biotech Co., Ltd.
Product CodeLCX · Clinical Chemistry
Decision DateJul 15, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1155
Device ClassClass 2

Intended Use

Shinetell Plus™ Digital Early Pregnancy Test is intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period.

Device Story

Lateral flow immunoassay for qualitative hCG detection in urine; dip or midstream sampling. Urine sample flows across test strip; hCG binds to anti-β-hCG antibody-colloidal gold conjugate; complex captured by immobilized anti-α-HCG antibody at test line. LED light source detects colored line formation; digital processor converts signal to 'Pregnant' or 'Not Pregnant' result on LCD screen within 3 minutes. Invalid samples display '?'. Used by lay users in home setting; provides immediate visual result to assist in pregnancy status determination.

Clinical Evidence

Clinical performance evaluated via 200-subject method comparison (100 dip, 100 midstream) against predicate, showing 100% concordance. Early pregnancy detection study (n=65) demonstrated detection rates from 4.6% at 8 days before expected period to 100% at 1 day after. Lay user study (n=200) showed 100% agreement between lay user and professional results. Specificity study (n=900) across pre-, peri-, and post-menopausal cohorts showed 0% false positive rate. Analytical sensitivity confirmed at 10 mIU/mL.

Technological Characteristics

Lateral flow immunochromatographic assay. Materials: plastic housing, absorbent tip, test strip. Sensing: LED light source and photodiode for optical reflection measurement. Sensitivity: 10 mIU/mL. Format: midstream or dip. Calibration: WHO 5th International Standard (NIBSC 07/364).

Indications for Use

Indicated for qualitative detection of hCG in urine to aid early pregnancy detection in women, including as early as 6 days before the missed period. Intended for OTC use.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION July 15, 2025 Hangzhou AllTest Biotech Co., Ltd. % Jenny Xia Director LSI International Inc. 504 E Diamond Ave., Suite H Gaithersburg, Maryland 20877 Re: K251053 Trade/Device Name: Shinetell Plus™ Digital Early Pregnancy Test Regulation Number: 21 CFR 862.1155 Regulation Name: Human Chorionic Gonadotropin (HCG) Test System Regulatory Class: Class II Product Code: LCX Dated: June 3, 2025 Received: June 3, 2025 Dear Jenny Xia: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251053 - Jenny Xia Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K251053 - Jenny Xia Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, JOSEPH A. KOTAREK -5 Digitally signed by JOSEPH A. KOTAREK -5 Date: 2025.07.15 10:23:07 04'00" Joseph Kotarek Branch Chief, Toxicology Branch Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K251053 | | | Device Name Shinetell PlusTM Digital Early Pregnancy Test | | | Indications for Use (Describe) Shinetell PlusTM Digital Early Pregnancy Test is intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period. | | | Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} 510(k) SUMMARY K251053 1. Date: June 29, 2025 1. Submitter: Hangzhou AllTest Biotech Co., Ltd. No. 550 Yinhai Street Hangzhou, China 2. Contact person: Jenny Xia LSI International Inc. 504 East Diamond Ave., Suite H Gaithersburg, MD 20877 Telephone: 301-525-6856 Fax: 301-916-6213 Email: jxia@lsi-consulting.org 3. Device Name: Shinetell Plus™ Digital Early Pregnancy Test Classification: Class II Product Code: LCX CFR: 21 CFR 862.1155 4. Predicate Devices: Wondfo One Step HCG Urine Pregnancy Test Midstream, Wondfo One Step HCG Urine Pregnancy Test Strip, Wondfo One Step HCG Urine Pregnancy Test Cassette, K150022 5. Intended Use Shinetell Plus™ Digital Early Pregnancy Test is intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period. 6. Device Description Shinetell Plus™ Digital Early Pregnancy Test is used for in vitro qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine, and is designed to be tested in dip or midstream mode. The test device consists of a single test strip assembled in a plastic housing, with an absorbent tip. The device is in a ready-to-use format. Shinetell Plus™ Digital Early Pregnancy Test uses lateral flow immunoassay and light reflection for the detection of the HCG in urine specimens. The test would detect the light intensity by using the LED as the light source. After that, the result can be displayed on the display screen. {5} 2 # 7. Substantial Equivalence Information | Similarities | | | | --- | --- | --- | | Item | Candidate device | Predicate device | | Intended use | A rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period. | Same | | Specimen | Urine | Urine | | Assay technical | Immunochromatographic assay | Immunochromatographic assay | | Sensitivity | 10 mIU/mL | 10 mIU/mL | | Results | Qualitative | Qualitative | | Target user | Over the counter use | Over the counter use | | Differences | | | | Item | Device | Predicate | | Readout | Digital/LCD screen | Visual Testing Line | | Time to result | 3 minutes | 5 minutes | | Format | Midstream | Strip, cassette, midstream | # 8. Test Principle Human chorionic gonadotropin (HCG) is a glycoprotein produced by the placenta during pregnancy. Shinetell Plus™ Digital Early Pregnancy Test uses lateral flow immunoassay and light reflection for the detection of the HCG in urine specimens. After the appropriate urine sample is added to the absorbent tip, the clock symbol will appear and blink on the Display Screen after urine is applied, indicating the test is working. The HCG in urine specimen will react with the anti-β-HCG antibody-colloidal gold conjugate and form a compound. As the liquid flows to the Test area of the test strip, the compound will be captured by the anti-α-HCG antibody immobilized on the Test area, then a colored line will be formed on the Test area. The test would detect the light intensity by using the LED as the light source. The {6} background of the LED lighted zone will be darker due to the color of colloidal gold conjugate, and the intensity of the reflected light received by the photodiode decreases significantly. By detecting the signal parameter of the reflected light intensity, which is decreasing significantly, the test can determine that the sample is added properly. The LED light will irradiate the test strip and then be reflected to the photodiode and the photodiode can respond to the different light intensity value by testing the intensity of reflected light. In approximately 3 minutes, the result will be shown on the display screen. If test value is over the preset threshold, the test result is positive and "Pregnant" is displayed on the screen. Otherwise, the result is negative and "Not Pregnant" is displayed on the screen. If the sample is added improperly, the background color of the LED lighted zone does not decrease significantly. By detecting the signal parameter of the reflected light intensity, which isn't decreasing significantly, the test can determine that the sample is added improperly, and the test will be invalid and “?” is displayed on the screen ## 9. Performance Characteristics ## A. Analytical performance ## a. Precision/Reproducibility/Sensitivity Negative female urine was spiked with hCG standard (Traceable to the 5th WHO) to hCG concentrations of 0, 3, 5, 7.5, 8, 10, 25 and 50 mIU/mL. Each sample was tested by both dip and midstream methods in 10 replicates per day for 5 days for each device lot. Total of three device lots were tested. Tests were performed by three different operators for each sample concentration. The results are summarized in the table below: Midstream Testing | hCG Concentration (mIU/mL) | Lot 1 | | Lot 2 | | Lot 3 | | Total result | | % Negative | % Positive | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | - | + | - | + | - | + | - | + | | | | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100 | 0 | | 3 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100 | 0 | | 5 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100 | 0 | | 7.5 | 35 | 15 | 35 | 15 | 36 | 14 | 106 | 44 | 70.7 | 29.3 | | 8 | 25 | 25 | 24 | 26 | 24 | 26 | 73 | 77 | 48.7 | 51.3 | | 10 | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100 | | 25 | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100 | | 50 | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100 | Dip Testing {7} | hCG Concentration (mIU/mL) | Lot 1 | | Lot 2 | | Lot 3 | | Total result | | % Negative | % Positive | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | - | + | - | + | - | + | - | + | | | | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100 | 0 | | 3 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100 | 0 | | 5 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100 | 0 | | 7.5 | 35 | 15 | 36 | 14 | 36 | 14 | 107 | 43 | 71.3 | 28.7 | | 8 | 24 | 26 | 24 | 26 | 24 | 26 | 72 | 78 | 48 | 52 | | 10 | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100 | | 25 | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100 | | 50 | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100 | Shinetell Plus™ Digital Early Pregnancy Test exhibited reproducible results. Based on the above results, the sensitivity of Shinetell Plus™ Digital Early Pregnancy Test is demonstrated to be 10 mIU/mL. ## b. Linearity/assay reportable range: Linearity is not applicable since this is a qualitative test. ## c. Hook effect test: Negative urine samples were spiked with varying hCG concentrations (6,250 mIU/mL, 12,500 mIU/mL, 25,000 mIU/mL, 50,000 mIU/mL, 100,000 mIU/mL, 200,000 mIU/mL and 500,000 mIU/mL). All tested concentrations gave a positive result. The results demonstrated that no hook effect was observed at hCG concentration up to 500,000 mIU/mL. ## d. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: Shinetell Plus™ Digital Early Pregnancy Test is calibrated against reference material traceable to WHO International Standard 5th edition, NIBSC code 07/364. ## Stability: Products in sealed foil pouch are stable for 24 months at 35.6-86°F, based on the real time stability study. ## e. Specificity and cross reactivity To evaluate specificity, 300 urine samples were collected from healthy, non-pregnant female in pre-menopausal (ages 18~40 years old), peri-menopausal (41~55 years old) and post-menopausal (>55 years old) groups. 100 people for each age group. Both dip and midstream testing are evaluated. No false positive results were observed for any of the age groups. {8} To evaluate cross-reactivity, negative and positive urine samples (0, 5 and 10 mIU/mL hCG) were spiked with potential cross reactants (1000 mIU/mL hLH, 1000 mIU/mL hFSH, 1000 μIU/mL hTSH). No cross-reactivity was observed at tested concentration. To evaluate the effect of the hCG $\beta$ -core fragment, Negative urine samples (0 and $5\mathrm{mIU/mL}$ hCG) and positive urine samples (10 and $20,000\mathrm{mIU/mL}$ hCG) were spiked with hCG $\beta$ -core fragment (hCG $\beta$ cf) at concentrations of 50,000 pmol/L, 125,000 pmol/L, 250,000pmol/L and 500,000pmol/L. The performance of Shinetell Plus™ Digital Early Pregnancy Test is not affected by hCG $\beta$ -core fragment concentrations up to 500,000 pmol/L. # f. Interfering substance To evaluate potential interferers with Shinetell Plus™ Digital Early Pregnancy Test, urine samples containing 0, 5 and $10\mathrm{mIU / mL}$ hCG were spiked with the interfering substance to obtain the certain desired test concentration. No interference effect was observed at the tested concentration shown in table below: | Substance | Concentration | | --- | --- | | Acetaminophen | 20 mg/dL | | Acetylsalicylic acid | 20 mg/dL | | Ascorbic acid | 20 mg/dL | | Atropine | 20 mg/dL | | Caffeine | 20 mg/dL | | Gentisic acid | 20 mg/dL | | Glucose | 2 g/dL | | Hemoglobin | 20 mg/dL | | Tetracycline | 20 mg/dL | | Ampicillin | 20 mg/dL | | Albumin | 20 mg/dL | | β-hydroxybutyrate | 2000 mg/dL | | Ephedrine | 20 mg/dL | | Phenylpropanolamine | 20 mg/dL | | Phenothiazine | 20 mg/dL | | EDTA | 80 mg/dL | | Salicylic Acid | 20 mg/dL | | Benzoylecgonine | 10 mg/dL | | Cannabinol | 10 mg/dL | | Codeine | 6ug/dL | | Ethanol | 1.0% | | Bilirubin | 2mg/dL | | Pregnanediol | 1500μg/dL | | Thiophene | 20 mg/dL | | Ketone | 20 mg/dL | | Lipid | 20 mg/dL | | Lipoprotein | 20 mg/dL | | Lipoprotein | 20 mg/dL | | Lipoprotein | 20 mg/dL | | Lipoprotein | 20 mg/dL | | Lipoprotein | 20 mg/dL | | Lipoprotein | 20 mg | {9} To evaluate the effect of urine pH on the results of Shinetell Plus™ Digital Early Pregnancy Test, urine samples containing 0, 5 and 10 mIU/mL hCG were tested at pH values of 4, 5, 6, 7, 8 and 9. The results indicated that urine pH ranges between 4 and 9 does not affect the performance of Shinetell Plus™ Digital Early Pregnancy Test. To evaluate the effect of urine density on the results of Shinetell Plus™ Digital Early Pregnancy Test, urine samples containing 0, 5 and 10 mIU/mL hCG were tested at density values of 1.000, 1.005, 1.011, 1.015, 1.019, 1.024, 1.029 and 1.035. The results indicated that urine with a relative density of 1.000 to 1.035 does not affect the performance of Shinetell Plus™ Digital Early Pregnancy Test. ## B. Method comparison study Method comparison with predicate device The performance of the new device was compared to the predicate test. Urine samples were collected from 200 women presenting to test for pregnancy. All of these women were suspected to be pregnant in the early stage of less than 5 weeks. All samples were tested with candidate and predicate devices at three POC sites. Dip Testing | Pregnancy Test (Shinetell Plus™) | Predicate Test (Wondfo test) | | Total | | --- | --- | --- | --- | | | Positive (+) | Negative (-) | | | Positive (+) | 49 | 0 | 49 | | Negative (-) | 0 | 51 | 51 | | Total | 49 | 51 | 100 | Midstream Testing | Pregnancy Test (Shinetell Plus™) | Predicate Test (Wondfo test) | | Total | | --- | --- | --- | --- | | | Positive (+) | Negative (-) | | | Positive (+) | 48 | 0 | 48 | | Negative (-) | 0 | 52 | 52 | | Total | 48 | 52 | 100 | The conformity between Shinetell Plus™ Digital Early Pregnancy Test and the predicate device is 100%. ## C. Lay person study 200 women's individual pregnancy status was self-tested using either in-stream method or dip method. Individuals with varying educational and occupational {10} backgrounds from three sites were chosen for the study. Each subject tested her own urine sample using the device according to the package insert and provided a sample for professional testing. Summary | In-Stream Method | Professional Result | | Total | | | --- | --- | --- | --- | --- | | | | Positive | | Negative | | Lay user Result | Positive | 48 | 0 | 48 | | | Negative | 0 | 52 | 52 | | Total | | 48 | 52 | 100 | | Dip Method | Professional Result | | Total | | | --- | --- | --- | --- | --- | | | | Positive | | Negative | | Lay user Result | Positive | 49 | 0 | 49 | | | Negative | 0 | 51 | 51 | | Total | | 49 | 51 | 100 | From the above tables, the lay person results showed 100% positive and 100% negative conformity with the professional results. Spiked urine samples were also tested by lay person. Urine samples were prepared at 5mIU/ml, 7.5mIU/ml, 8mIU/ml and 10mIU/ml hCG concentrations by spiking hCG into negative pooled urine specimens. Each sample was aliquoted into individual containers and blind-labeled. These samples were tested by 200 lay persons. Lay person vs Professional | hCG Concentrations | Lay person results | | Professional results | | Percent Agreement | | --- | --- | --- | --- | --- | --- | | | No. of Negative | No. of Positive | No. of Negative | No. of Positive | | | 5 mIU/ml | 100 | 0 | 100 | 0 | 100% | | 7.5 mIU/ml | 71 | 29 | 69 | 31 | 98% | | 8 mIU/ml | 50 | 50 | 52 | 48 | 98% | | 10 mIU/ml | 0 | 100 | 0 | 100 | 100% | Each lay person was given a questionnaire to assess the readability of the labeling. The results of the questionnaire reflected that the consumers found the test easy to use and that they did not have trouble understanding the labeling and interpreting {11} the results. ## D. Early Pregnancy Test Study In this study, total 650 urine samples from 65 characterized cycle segments of conceptive cycles were collected from 65 pregnant women. All samples were masked and randomized. Each sample was tested both in-stream and dip methods using three lots of the device. The new device detected 67.7% positive hCG five days before the Expected Menstrual Period (EMP), and 100% positive hCG on the day of EMP. No differences were observed between different test methods. The following table is the summary of the data. | Day relative to Expected Menstrual Period(EMP) | Number of Positive | Number of Negative | Number of Total | % Positive | | --- | --- | --- | --- | --- | | EMP-8 | 3 | 62 | 65 | 4.6% | | EMP-7 | 8 | 57 | 65 | 12.3% | | EMP-6 | 23 | 42 | 65 | 35.3% | | EMP-5 | 44 | 21 | 65 | 67.7% | | EMP-4 | 56 | 9 | 65 | 86.2% | | EMP-3 | 62 | 3 | 65 | 95.4% | | EMP-2 | 64 | 1 | 65 | 98.5% | | EMP-1 | 65 | 0 | 65 | 100.0% | | EMP | 65 | 0 | 65 | 100.0% | | EMP+1 | 65 | 0 | 65 | 100.0% | ## 10. Conclusion Based on the test principle and performance characteristics of the device including precision, cut-off, interference, specificity, method comparison and lay-user studies of the device, it's concluded that Shinetell Plus™ Digital Early Pregnancy Test is substantially equivalent to the predicate.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...