K970707 · Selfcare, Inc. · CGA · Jun 20, 1997 · Clinical Chemistry
Device Facts
Record ID
K970707
Device Name
ELECT II BLOOD GLUCOSE MONITORING SYSTEM
Applicant
Selfcare, Inc.
Product Code
CGA · Clinical Chemistry
Decision Date
Jun 20, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Intended Use
The Elect II Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. The Elect II System is intended for use outside the body (in vitro diagnostic use) by diabetics at home as an aid to monitor the effectiveness of diabetes control.
Device Story
System consists of glucose reagent test strip, amperometric meter, and quality control solution. User inserts test strip into meter; applies fingerstick capillary blood sample to strip. Meter measures electrical current generated by glucose oxidation; current is proportional to glucose concentration. Result displayed in 15 seconds. System calibrated to plasma-equivalent values. Used by diabetics at home for self-monitoring of blood glucose levels to aid in diabetes management.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and methodology comparison to predicate devices.
Technological Characteristics
Amperometric glucose meter; glucose oxidase enzymatic biosensor; electrochemical detection. System includes reagent test strips and quality control solution. Calibrated to plasma-equivalent values. 15-second measurement time.
Indications for Use
Indicated for quantitative measurement of glucose in fresh capillary whole blood for diabetics at home to monitor diabetes control effectiveness.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Precision Q.I.D. Blood Glucose Monitoring System (K945887)
Related Devices
K961985 — ELECT BLOOD GLUCOSE MONITORIG SYSTEM · Selfcare, Inc. · Sep 5, 1996
K062829 — DIACHEX BLOOD GLUCOSE MONITORING SYSTEM · Tyson Bioresearch, Inc. · Oct 10, 2006
K091765 — EBCARRYON BLOOD GLUCOSE MONITORING SYSTEM · Visgeneer, Inc. · Jan 19, 2010
Submission Summary (Full Text)
{0}
K970707
JUN 20 1997
# 510(k) Summary
**Manufacturer:** Selfcare, Inc.
200 Prospect Street
Waltham, MA 02154
Telephone: (617) 647-3900
Fax: (617) 647-3939
**Contact Person:** Carol Adiletto
**Date:** February 25, 1997
**Device Name:** Elect II Blood Glucose Monitoring System
**Common Name:** Blood glucose meter and reagent test strips for blood glucose
**Classification (meter/strips):**
Class II as per 21 CFR Section 862.1345
Glucose test system; Code 75 CGA
(glucose oxidase, glucose)
Classification panel - Clinical Chemistry
**Classification (control):**
Class I as per 21 CFR Section 862.1660,
Code 75 JJX (singe analyte control)
**Classification (lancets):**
Class I as per 21 CFR Section 878.4800,
Code 79 FMK (lancet, blood)
**Performance Standards:** None established under section 514
**Equivalent Devices:**
Elect Blood Glucose Monitoring System
(Selfcare, Inc. K961985)
Precision Q.I.D. Blood Glucose Monitoring System
(Medisense, Inc. K945887)
80
{1}
81
# Product Description:
The Elect II Blood Glucose Monitoring System comprises a glucose reagent test strip, a amperometric meter and a quality control solution. The user inserts a test strip in the Elect II Meter and applies a small drop of blood from a fingerstick to the sample point on the strip. The meter measures the electrical current that is generated which is proportional to the concentration of glucose present in the blood sample. The blood glucose result is displayed in 15 seconds. The system is calibrated to give a result equivalent to a plasma value, the measurement used by most clinical laboratories.
# Intended Use:
The Elect II Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. The Elect II System is intended for use outside the body (in vitro diagnostic use) by diabetics at home as an aid to monitor effectiveness of diabetes control.
# Substantial Equivalence:
The Elect II Blood Glucose Monitoring System is similar in technological characteristics, methodology and intended use to the predicate devices listed above. The Elect II System utilizes similar biosensor methodology based on a standard enzymatic methodology employing the enzyme glucose oxidase coupled to an electrochemical detection system to measure glucose levels in whole blood.
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
JUN 20 1997
Carol A. Adiletto, M.S.
Clinical Director
SelfCare, Inc.
200 Prospect Street
Waltham, Massachusetts 02154
Re: K970707
Elect II Blood Glucose Monitoring System
Regulatory Class: II
Product Code: CGA
Dated: February 25, 1997
Received: February 26, 1997
Dear Ms. Adiletto:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}
Selfcare, Inc.
Elect II Blood Glucose Monitoring System 510(k) Submission
Confidential
# Indications for Use Form
510(k) Number (if known): K970707
Device Name: Elect II Blood Glucose Monitoring System
Indication for Use:
The Elect II Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. The Elect II System is intended for use outside the body (in vitro diagnostic use) by diabetics at home as an aid to monitor the effectiveness of diabetes control.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
Prescription Use ☑ OR Over-The-Counter Use ☐
04
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.