i-STAT CHEM8+ cartridge with the i-STAT 1 System

K210958 · Abbott Point of Care, Inc. · CGA · Aug 13, 2021 · Clinical Chemistry

Device Facts

Record IDK210958
Device Namei-STAT CHEM8+ cartridge with the i-STAT 1 System
ApplicantAbbott Point of Care, Inc.
Product CodeCGA · Clinical Chemistry
Decision DateAug 13, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The i-STAT CHEM8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of sodium, potassium, chloride, ionized calcium, glucose, blood urea nitrogen, creatinine, hematocrit, and total carbon dioxide in arterial or venous whole blood in point of care or clinical laboratory settings. Sodium measurements are used for monitoring electrolyte imbalances. Potassium measurements are used in the diagnosis and monitoring of diseases and clinical conditions that manifest high and low potassium levels. Chloride measurements are primarily used in the diagnosis, monitoring, and treatment of electrolyte and metabolic disorders including, but not limited to, cystic fibrosis, diabetic acidosis, and hydration disorders. Ionized calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany. Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma. Blood urea nitrogen measurements are used for the diagnosis, monitoring, and treatment of certain renal and metabolic diseases. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. Hematocrit measurements can aid in the determination and monitoring of normal or abnormal total red cell volume status that can be associated with conditions including anemia and erythrocytosis. The i-STAT Hematocrit test has not been evaluated in neonates. Carbon dioxide measurements are used in the diagnosis, monitoring, and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.

Device Story

Handheld i-STAT 1 System uses single-use, disposable CHEM8+ cartridges for point-of-care or clinical lab blood analysis. User applies 2-3 drops of arterial or venous whole blood (with or without anticoagulant) to cartridge; i-STAT 1 analyzer moves fluid across sensors to generate quantitative results in ~2 minutes. System includes analyzer, cartridges, downloader/recharger, electronic simulator, and printer. Provides clinicians with rapid electrolyte, metabolic, and hematocrit data to inform diagnosis and treatment of renal, metabolic, and acid-base disorders. Benefits include immediate clinical decision-making at the patient's side.

Clinical Evidence

Bench testing only. Matrix equivalency study compared non-anticoagulated vs. heparin-anticoagulated venous and arterial whole blood samples (N=310-314 per analyte). Passing-Bablok regression analysis showed high correlation (r=0.93-0.999) across all analytes. Specimen stability study confirmed stability of non-anticoagulated samples for 3 minutes.

Technological Characteristics

Handheld in vitro diagnostic analyzer; single-use disposable cartridges with electrochemical sensors. Analyzes whole blood (arterial/venous). Connectivity via i-STAT 1 Downloader/Recharger. No changes to hardware or sensing principles from predicate.

Indications for Use

Indicated for in vitro quantification of sodium, potassium, chloride, ionized calcium, glucose, BUN, creatinine, hematocrit, and TCO2 in arterial or venous whole blood. Used in point-of-care or clinical laboratory settings by trained professionals. Hematocrit test not evaluated in neonates.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font. August 13, 2021 Abbott Point of Care Inc. Maria Figueroa Associate Director, Regulatory Affairs 400 College Road East Princeton, New Jersey 08540 Re: K210958 Trade/Device Name: i-STAT CHEM8+ cartridge with the i-STAT 1 System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: CGA, CGL, JGS, CEM, CGZ, JFP, CDS, JFL, JPI Dated: March 30, 2021 Received: March 31, 2021 Dear Maria Figueroa: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Marianela Perez-Torres, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) k210958 #### Device Name i-STAT CHEM8+ cartridge with the i-STAT 1 System #### Indications for Use (Describe) The i-STAT CHEM8+ cartridge with the i-STAT 1 System is in the in vitro quantification of sodium, potassium, chloride, ionized calcium, glucose, blood urea nitrogen, creatinine, hematocrit, and total carbon dioxide in arterial or venous whole blood in point of care or clinical laboratory settings. Sodium measurements are used for monitoring electrolyte imbalances. Potassium measurements are used in the diagnosis and clinical conditions that manifest high and low potassium levels. Chloride measurements are primarily used in the diagnosis, monitoring, and treatment of electrolyte and metabolic disorders including, but not limited to, cystic fibrosis, diabetic acidosis, and hydration disorders. Ionized calcium measurements are used in the diagnosis and treatment of parathyroid disease, chronic renal disease and tetany. Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, and pancreatic islet cell carcinoma. Blood urea nitrogen measurements are used for the diagnosis, and treatment of certain renal and metabolic diseases. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. Hematoorit measurements can aid in the determination and monitoring of normal total red cell volume status that can be associated with conditions including anemia and erythrocytosis. The i-STAT Hematocrit test has not been evaluated in neonates. Carbon dioxide measurements are used in the diagnosis, monitoring, and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary The information in this 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92. ## 1. Submitter Information | Owner | Abbott Point of Care Inc.<br>400 College Road East<br>Princeton, NJ 08540 | |---------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact | Primary: Maria Figueroa<br>Associate Director Regulatory Affairs<br>Phone: 609-454-9271<br><br>Secondary contact person for all communications:<br>Secondary: Jaquelyn Gesumaria<br>Senior Specialist Regulatory Affairs<br>Phone: 609-454-9384 | Date Prepared August 12, 2021 #### Device Information 2. Proprietary Name i-STAT CHEM8+ cartridge with the i-STAT 1 System Chemistry test, analyzer, handheld Common Name | Product<br>code | Device Classification<br>name | Regulation<br>Number | Class | Panel | |-----------------|-----------------------------------------------|----------------------|-------|--------------------| | CGA | Glucose Oxidase,<br>Glucose | 862.1345 | II | Clinical Chemistry | | CGL | Electrode, Ion Based<br>Enzymatic, Creatinine | 862.1225 | II | Clinical Chemistry | | JGS | Electrode, Ion Specific,<br>Sodium | 862.1665 | II | Clinical Chemistry | | CEM | Electrode, Ion Specific,<br>Potassium | 862.1600 | II | Clinical Chemistry | | CGZ | Electrode, Ion Specific<br>Chloride | 862.1170 | II | Clinical Chemistry | | CDS | Electrode, Ion Specific,<br>Urea Nitrogen | 862.1770 | II | Clinical Chemistry | {4}------------------------------------------------ | Product<br>code | Device Classification<br>name | Regulation<br>Number | Class | Panel | |-----------------|--------------------------------------------|----------------------|-------|--------------------| | JFL | Bicarbonate, Carbon<br>Dioxide Test System | 862.1160 | II | Clinical Chemistry | | JFP | Electrode, Ion Specific,<br>Calcium | 862.1145 | II | Clinical Chemistry | | JPI | Device, Hematocrit<br>Measuring | 862.6400 | II | Hematology | #### Predicate Device 3. Proprietary Name i-STAT CHEM8+ cartridge with the i-STAT 1 System | 510(k) Numbers | | K183678 | |----------------|--|---------| | | | K183688 | | | | K191298 | | | | K191360 | | | | K183680 | | 510(k) Number | Product code | Device Classification name | Regulation Number | Class | Panel | |---------------|--------------|--------------------------------------------|-------------------|-------|--------------------| | K183678 | CGA | Glucose Oxidase, Glucose | 862.1345 | II | Clinical Chemistry | | | CGL | Electrode, Ion Based Enzymatic, Creatinine | 862.1225 | II | Clinical Chemistry | | K183688 | JGS | Electrode, Ion Specific, Sodium | 862.1665 | II | Clinical Chemistry | | | CEM | Electrode, Ion Specific, Potassium | 862.1600 | II | Clinical Chemistry | | | CGZ | Electrode, Ion Specific Chloride | 862.1170 | II | Clinical Chemistry | | | CDS | Electrode, Ion Specific, Urea Nitrogen | 862.1770 | II | Clinical Chemistry | | K191298 | JFL | Bicarbonate, Carbon Dioxide Test System | 862.1160 | II | Clinical Chemistry | | K191360 | JFP | Electrode, Ion Specific, Calcium | 862.1145 | II | Clinical Chemistry | | K183680 | JPI | Device, Hematocrit Measuring | 862.6400 | II | Hematology | # 4. Device Description The i-STAT CHEM8+ test cartridge contains test reagents to analyze whole blood at the point of care or in the clinical laboratory for sodium (Na), potassium (K), chloride (CI), ionized calcium (iCa), glucose (Glu), blood urea nitrogen (BUN), creatinine (Crea), hematocrit (Hct), and total carbon dioxide (TCO2). The test is contained in a single-use, disposable cartridge. Cartridges require two to three drops of whole blood which are typically applied to the cartridge using a transfer device. {5}------------------------------------------------ The i-STAT 1 Analyzer is a handheld, in vitro diagnostic analytical device designed to run only i-STAT test cartridges. The instrument interacts with the cartridge to move fluid across the sensors and generate a quantitative result (within approximately 2 minutes). The i-STAT 1 System is comprised of the i-STAT 1 analyzer, the i-STAT test cartridges and accessories (i-STAT 1 Downloader/Recharger, electronic simulator and portable printer). The system is designed for use by trained medical professionals at the patient point of care or in the clinical laboratory and is for prescription use only. ## 5. Intended Use Statement The i-STAT CHEM8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of sodium, potassium, chloride, ionized calcium, glucose, blood urea nitrogen, creatinine, hematocrit, and total carbon dioxide in arterial or venous whole blood in point of care or clinical laboratory settings. Sodium measurements are used for monitoring electrolyte imbalances. Potassium measurements are used in the diagnosis and monitoring of diseases and clinical conditions that manifest high and low potassium levels. Chloride measurements are primarily used in the diagnosis, monitoring, and treatment of electrolyte and metabolic disorders including, but not limited to, cystic fibrosis, diabetic acidosis, and hydration disorders. Ionized calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany. Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma. Blood urea nitrogen measurements are used for the diagnosis, monitoring, and treatment of certain renal and metabolic diseases. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. Hematocrit measurements can aid in the determination and monitoring of normal or abnormal total red cell volume status that can be associated with conditions including anemia and erythrocytosis. The i-STAT Hematocrit test has not been evaluated in neonates. Carbon dioxide measurements are used in the diagnosis, monitoring, and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. {6}------------------------------------------------ # 6. Reason for Submission Addition of anticoagulant free (non-anticoagulant) whole blood matrix to previously cleared i-STAT CHEM8+ cartridge with i-STAT 1 System under k183678, k183688, k191298, k191360 and k183680. | Similarities and Differences | | | |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Feature or<br>Characteristic | Candidate Device: i-STAT CHEM8+<br>cartridge with the i-STAT 1 System | Predicate Device: i-STAT CHEM8+<br>cartridge with the i-STAT 1 System<br>(K183678, K183680, K183688,<br>K191298, K191360) | | Intended Use | Same | The i-STAT CHEM8+ cartridge with the i-<br>STAT 1 System is intended for use in the in<br>vitro quantification of sodium, potassium,<br>chloride, ionized calcium, glucose, blood<br>urea nitrogen, creatinine, hematocrit, and<br>total carbon dioxide in arterial or venous<br>whole blood in point of care or clinical<br>laboratory settings. | | | | Sodium measurements are used for<br>monitoring electrolyte imbalances. | | | | Potassium measurements are used in the<br>diagnosis and monitoring of diseases and<br>clinical conditions that manifest high and low<br>potassium levels. | | | | Chloride measurements are primarily used in<br>the diagnosis, monitoring, and treatment of<br>electrolyte and metabolic disorders<br>including, but not limited to, cystic fibrosis,<br>diabetic acidosis, and hydration disorders. | | | | Ionized calcium measurements are used in<br>the diagnosis and treatment of parathyroid<br>disease, a variety of bone diseases,<br>chronic renal disease and tetany. | | | | Glucose measurements are used in the<br>diagnosis, monitoring, and treatment of<br>carbohydrate metabolism disorders<br>including, but not limited to, diabetes<br>mellitus, neonatal hypoglycemia, idiopathic<br>hypoglycemia, and pancreatic islet cell<br>carcinoma. | | | | Blood urea nitrogen measurements are<br>used for the diagnosis, monitoring, and<br>treatment of certain renal and metabolic<br>diseases. | | | | Creatinine measurements are used in the<br>diagnosis and treatment of renal diseases, | | Similarities and Differences | | | | Feature or<br>Characteristic | Candidate Device: i-STAT CHEM8+<br>cartridge with the i-STAT 1 System | Predicate Device: i-STAT CHEM8+<br>cartridge with the i-STAT 1 System<br>(K183678, K183680, K183688,<br>K191298, K191360) | | | | calculation basis for measuring other urine<br>analytes. | | | | Hematocrit measurements can aid in the<br>determination and monitoring of normal or<br>abnormal total red cell volume status that<br>can be associated with conditions including<br>anemia and erythrocytosis. The i-STAT<br>Hematocrit test has not been evaluated in<br>neonates.<br><br>Carbon dioxide measurements are used in<br>the diagnosis, monitoring, and treatment of<br>numerous potentially serious disorders<br>associated with changes in body acid-base<br>balance. | | Reportable Range | Same | Reportable Range:<br>Sodium 100-180 mmol/L<br>Potassium 2.0-9.0 mmol/L<br>Chloride 65-140 mmol/L<br>Ionized 0.25-2.50 mmol/L<br>Calcium<br>Glucose 20-700 mg/dL<br>BUN/Urea 3-140 mg/dL<br>Creatinine 0.2-20.0 mg/dL<br>Hematocrit 15-75 %PCV<br>TCO2 5-50 mmol/L | | Sample Type | Same | Arterial or venous whole blood | | Sample Volume | Same | 95 μL | | Sample<br>Preparation | Same | Ready to Use | | Sample Collection | With balanced heparin anticoagulant With lithium heparin anticoagulant Without anticoagulant | With balanced heparin anticoagulant With lithium heparin anticoagulant | | Time<br>Testing/Sample<br>Stability<br>(time from<br>collection to<br>sample fill) | <b>With anticoagulant:</b><br>TCO2, iCa within 10 minutes<br>Na, K, Cl, Glu,<br>BUN, Crea, Hct within 30 minutes<br><b>Without anticoagulant:</b><br>TCO2, iCa, Na, K,<br>Cl, Glu, BUN,<br>Crea, Hct within 3 minutes | <b>With anticoagulant:</b><br>TCO2, iCa within 10 minutes<br>Na, K, Cl, Glu,<br>BUN, Crea, Hct within 30 minutes | ## 7. Summary Comparison of Technological Characteristics {7}------------------------------------------------ {8}------------------------------------------------ ## 8. Performance Characteristics #### Analytical Performance #### a. Precision ## Precision 20 days (Aqueous Materials) Precision of the i-STAT Sodium, Potassium, Chloride, Ionized Calcium, Glucose, Blood Urea Nitrogen, Creatinine, Hematocrit, and Total Carbon Dioxide tests in the i-STAT CHEM8+ cartridge with the i-STAT 1 System was previously reviewed under: > K183678 for Glucose and Creatinine K183680 for Hematocrit K186388 for Sodium, Potassium, Chloride and BUN K191298 for TCO2 K191360 for iCa ## Precision (Whole Blood) Whole blood precision of the i-STAT Sodium, Potassium, Chloride, Ionized Calcium, Glucose, Blood Urea Nitrogen, Creatinine, Hematocrit, and Total Carbon Dioxide tests in the i-STAT CHEM8+ cartridge with the i-STAT 1 System using whole blood venous and arterial specimens collected with lithium heparin was previously reviewed under: > K183678 for Glucose and Creatinine K183680 for Hematocrit K186388 for Sodium, Potassium, Chloride and BUN K191298 for TCO2 K191360 for iCa ## b. Linearity Linearity of the i-STAT Sodium, Potassium, Chloride, Ionized Calcium, Glucose, Blood Urea Nitrogen, Creatinine, Hematocrit, and Total Carbon Dioxide tests in the i-STAT CHEM8+ cartridge with the i-STAT 1 System was previously reviewed under: > K183678 for Glucose and Creatinine K183680 for Hematocrit K186388 for Sodium, Potassium, Chloride and BUN K191298 for TCO2 K191360 for iCa # c. Limit of Quantitation (LoO) LoQ of the i-STAT Sodium, Potassium, Chloride, Ionized Calcium, Glucose, Blood Urea Nitrogen, Creatinine, Hematocrit, and Total Carbon Dioxide tests in the i-STAT CHEM8+ cartridge with the i-STAT 1 System was previously reviewed under: {9}------------------------------------------------ K183678 for Glucose and Creatinine K183680 for Hematocrit K186388 for Sodium, Potassium, Chloride and BUN K191298 for TCO2 K191360 for iCa # d. Limit of Blank and Detection (LoB/LoD) LoB and LoD of the i-STAT Ionized Calcium, Glucose, Creatinine, and Hematocrit tests in the i-STAT CHEM8+ cartridge with the i-STAT 1 System was previously reviewed under: > K183678 for Glucose and Creatinine K183680 for Hematocrit K191360 for iCa # e. Interference The performance of the i-STAT Sodium, Potassium, Chloride, Ionized Calcium, Glucose, Blood Urea Nitrogen, Creatinine, Hematocrit, and Total Carbon Dioxide tests in the i-STAT CHEM8+ cartridge with the i-STAT 1 System in the presence of potentially interfering substances was previously reviewed under: > K183678 for Glucose and Creatinine K183680 for Hematocrit K186388 for Sodium, Potassium, Chloride and BUN K191298 for TCO2 K191360 for iCa # Comparison Study # f. Method Comparison of Candidate versus Predicate Device Method comparison of the i-STAT Sodium, Potassium, Chloride, Ionized Calcium, Glucose, Blood Urea Nitrogen, Creatinine, Hematocrit, and Total Carbon Dioxide tests in the i-STAT CHEM8+ cartridge with the i-STAT 1 System using whole blood venous and arterial specimens collected with lithium heparin was previously reviewed under: K183678 for Glucose and Creatinine K183680 for Hematocrit K186388 for Sodium, Potassium, Chloride and BUN K191298 for TCO2 K191360 for iCa {10}------------------------------------------------ # g. Matrix Equivalence A matrix equivalence study was conducted at three (3) point of care sites to assess the performance of the i-STAT Sodium, Potassium, Chloride, Ionized Calcium, Glucose, Blood urea nitrogen, Creatinine, Hematocrit, and Total Carbon Dioxide tests in the i-STAT CHEM8+ cartridge on the i-STAT 1 System using non-anticoagulated venous and arterial whole blood samples. The matrix equivalence of each analyte in the i-STAT CHEM8+ cartridge was assessed by comparing samples collected without anticoagulant (candidate) to samples collected with balanced heparin or lithium heparin anticoagulant (primary sample). A Passing-Bablok linear regression analysis was performed using the first replicate result from the candidate (y-axis) versus the mean result from the primary sample (x-axis). The regression analysis results are summarized in the table below. | Analyte | N | Units | Candidate<br>Range | Primary Sample<br>Range | r | Slope | Intercept | |---------|-----|--------|--------------------|-------------------------|------|-------|-----------| | Na | 314 | mmol/L | 110 - 174 | 111 - 173 | 0.99 | 1.00 | 0.50 | | K | 313 | mmol/L | 2.2 - 7.7 | 2.2 - 7.5 | 0.96 | 1.00 | 0.00 | | Cl | 314 | mmol/L | 76 - 136 | 79 - 137 | 0.98 | 1.00 | -0.50 | | iCa | 314 | mmol/L | 0.41 - 2.48 | 0.71 - 2.28 | 0.85 | 1.04 | -0.04 | | Glu | 313 | mg/dL | 29 - 663 | 35 - 660 | 1.00 | 1.01 | -0.63 | | BUN | 310 | mg/dL | 4 - 120 | 4 - 118 | 1.00 | 1.00 | 0.00 | | Crea | 312 | mg/dL | 0.2 - 19.4 | 0.2 - 19.4 | 1.00 | 1.00 | 0.00 | | Hct | 311 | %PCV | 16 - 75 | 16 - 73 | 0.99 | 1.00 | 0.46 | | TCO2 | 273 | mmol/L | 9 - 42 | 11 - 41 | 0.95 | 1.00 | 0.00 | The results of the study support the addition of whole blood samples collected without anticoagulant for use with the i-STAT CHEM8+ cartridge on the i-STAT 1 System. # 9. Conclusion The results of the study demonstrate that performance of the i-STAT Sodium, Potassium, Chloride, Ionized Calcium, Glucose, Blood Urea Nitrogen, Creatinine, Hematocrit, and Total Carbon Dioxide tests in the i-STAT CHEM8+ cartridge with the i-STAT 1 System are substantially equivalent when tested using the anticoagulated and nonanticoagulated sample type.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...