i-STAT CHEM8+ cartridge with the i-STAT 1 System

K183678 · Abbott Point of Care, Inc. · CGA · Feb 7, 2020 · Clinical Chemistry

Device Facts

Record IDK183678
Device Namei-STAT CHEM8+ cartridge with the i-STAT 1 System
ApplicantAbbott Point of Care, Inc.
Product CodeCGA · Clinical Chemistry
Decision DateFeb 7, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The i-STAT CHEM8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of glucose and creatinine in arterial or venous whole blood in point of care or clinical laboratory settings. Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Device Story

The i-STAT CHEM8+ is a single-use, disposable cartridge used with the handheld i-STAT 1 Analyzer. It accepts 2-3 drops of arterial or venous whole blood. The system performs amperometric measurement of oxidized hydrogen peroxide produced by glucose oxidase (for glucose) and creatinine amidinohydrolase (for creatinine) activity. The analyzer automatically moves fluid across sensors, performs internal quality checks, and provides quantitative results in approximately 2 minutes. Used by trained medical professionals at the point of care or in clinical labs. The output assists clinicians in diagnosing and monitoring metabolic and renal conditions. The system includes accessories like a downloader/recharger, electronic simulator, and printer.

Clinical Evidence

No clinical trials were performed. Evidence consists of analytical bench testing, including precision/reproducibility studies (n=80-320 per analyte), linearity (polynomial regression), interference testing (CLSI EP07), and method comparison against predicate devices (n=180-185 samples). Method comparison showed high correlation (r=0.99-1.00) for both analytes across the reportable ranges.

Technological Characteristics

Single-use disposable cartridge containing sensors, fluid pouch, and waste chamber. Amperometric sensing principle using enzymatic reactions (glucose oxidase for glucose; creatinine amidohydrolase/creatinine amidinohydrolase/sarcosine oxidase for creatinine). Dimensions/materials not specified. Standalone operation with i-STAT 1 Analyzer. Traceable to NIST SRM 967 (creatinine) and SRM 965 (glucose).

Indications for Use

Indicated for in vitro quantification of glucose and creatinine in arterial or venous whole blood for patients requiring diagnosis, monitoring, or treatment of carbohydrate metabolism disorders (e.g., diabetes, hypoglycemia) or renal diseases/dialysis monitoring. For use in point-of-care or clinical laboratory settings by trained professionals.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. February 7, 2020 Abbott Point of Care Inc. Susan Tibedo Director, Regulatory Affairs 400 College Road East Princeton, NJ 08540 Re: K183678 Trade/Device Name: i-STAT CHEM8+ cartridge with the i-STAT 1 System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: CGA, CGL Dated: January 9, 2020 Received: January 10, 2020 #### Dear Susan Tibedo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Kellie B. Kelm, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K183678 Device Name i-STAT CHEM8+ cartridge with the i-STAT 1 System #### Indications for Use (Describe) The i-STAT CHEM8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of glucose, and creatinine in arterial or venous whole blood in point of care or clinical laboratory settings. Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, and pancreatic islet cell carcinoma. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary The information in this 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92. | 1. Submitter Information | | |--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Owner | Abbott Point of Care Inc.<br>400 College Road East<br>Princeton, NJ 08540 | | Contact | Primary: Susan Tibedo<br>Director Regulatory Affairs<br>susan.tibedo@abbott.com<br>Phone: 609-454-9360<br><br>Secondary: Maria Figueroa<br>Manager Regulatory Affairs<br>maria.l.figueroa@abbott.com<br>Phone: 609-454-9271 | | Date Prepared | February 5, 2020 | | 510(k) Number | k183678 | ## 2. Device Information Proprietary Name i-STAT CHEM8+ cartridge with i-STAT 1 System Common Name Chemistry test, analyzer, handheld | Product<br>code | Device Classification<br>name | Regulation<br>Number | Class | Panel | |-----------------|-----------------------------------------------|----------------------|-------|--------------------| | CGA | Glucose Oxidase, Glucose | 862.1345 | II | Clinical Chemistry | | CGL | Electrode, Ion Based<br>Enzymatic, Creatinine | 862.1225 | II | Clinical Chemistry | {4}------------------------------------------------ #### 3. Predicate Device | Proprietary | SYNCHRON Systems Glucose Reagent on UniCel DxC | |-------------|------------------------------------------------| | Name | 600/800 SYNCHRON Clinical System | SYNCHRON Systems Creatinine Reagent on UniCel DxC 600/800 SYNCHRON Clinical System #### 510(k) Number K042291 | Product<br>code | Device Classification<br>name | Regulation<br>Number | Class | Panel | |-----------------|----------------------------------------------|----------------------|-------|--------------------| | CGA | Glucose Oxidase, Glucose | 862.1345 | II | Clinical Chemistry | | CGX | Alkaline Picrate,<br>Colorimetry, Creatinine | 862.1225 | II | Clinical Chemistry | #### 4. Device Description The i-STAT CHEM8+ test cartridge contains test reagents to analyze whole blood at the point of care or in the clinical laboratory for glucose and creatinine. The test is contained in a single-use, disposable cartridge. Cartridges require two to three drops of whole blood which are typically applied to the cartridge using a transfer device. The i-STAT 1 Analyzer is a handheld, in vitro diagnostic analytical device designed to run only i-STAT test cartridges. The instrument interacts with the cartridge to move fluid across the sensors and generate a quantitative result (within approximately 2 minutes). The i-STAT 1 System is comprised of the i-STAT 1 analyzer, the i-STAT test cartridges and accessories (i-STAT 1 Downloader/Recharger, electronic simulator and portable printer). The system is designed for use by trained medical professionals at the patient point of care or in the clinical laboratory and is for prescription use only. #### 5. Intended Use Statement The i-STAT CHEM8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of glucose and creatinine in arterial or venous whole blood in point of care or clinical laboratory settings. Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. {5}------------------------------------------------ | Feature or<br>Characteristic | Predicate | Candidate | |------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | SYNCHRON Systems<br>Glucose Reagent on UniCel<br>DxC 600/800 SYNCHRON<br>Clinical System (K042291)<br>The glucose test system is intended<br>for the quantitative determination of<br>glucose concentration in human<br>serum, plasma, urine or<br>cerebrospinal fluid (CSF).<br><br>Glucose measurements are used in<br>the diagnosis and treatment of<br>carbohydrate metabolism disorders<br>including diabetes mellitus, neonatal<br>hypoglycemia, idiopathic<br>hypoglycemia, and pancreatic islet<br>cell carcinoma. | Glucose Test with i-STAT 1 System<br><br>The i-STAT CHEM8+ cartridge with the<br>i-STAT 1 System is intended for use in<br>the <i>in vitro</i> quantification of glucose and<br>creatinine in arterial or venous whole<br>blood in point of care or clinical<br>laboratory settings.<br><br>Glucose measurements are used in the<br>diagnosis, monitoring, and treatment of<br>carbohydrate metabolism disorders<br>including, but not limited to, diabetes<br>mellitus, neonatal hypoglycemia,<br>idiopathic hypoglycemia, and pancreatic<br>islet cell carcinoma. | | Reportable Range | 0.2 – 33.3 mmol/L (Serum, Plasma)<br>0 – 600 mg/dL<br>up to 1200 mg/dL by sample<br>dilution | 1.1 – 38.9 mmol/L (mEq/L)<br>20 – 700 mg/dL<br>0.20 – 7.00 g/L | | Sample Type | Serum, plasma, urine, CSF | Arterial or venous whole blood | | Sample Volume | 0.5 mL (500 µL) | 95 µL | | Sample<br>Preparation | Sample tubes prepared and then<br>processed within analyzer | Ready to Use | | Traceability | NIST SRM 917 | NIST SRM 965 | | Calibration | Must be conducted every 24 hours<br>and with each new reagent | 1-point on-board contained within the<br>cartridge | | Time to Test<br>(Sample Stability) | Serum or plasma: Within 8 hours at<br>room temperature, or up to 48 hours<br>if stored at +2°C to +8°C<br><br>Urine: Within 2 hours of collection | Heparinized samples: within 30 minutes of<br>collection | | Principle of<br>Measurement | O2 depletion by glucose oxidase<br>activity | Amperometric measurement of oxidized<br>hydrogen peroxide produced by glucose<br>oxidase activity | | Reagent Format | Reagent handling system, stored<br>within analyzer | Cartridge | | Reagent Storage<br>and Stability | Room temperature | 2°C to 8°C (35-46°F) | | Analyzer Type | Floor Model | Handheld | ### 6. Summary Comparison of Technological Characteristics {6}------------------------------------------------ | Similarities and Differences: System (Test and Instrument): Creatinine (CREA) | | | |-------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Feature or<br>Characteristic | Predicate<br>SYNCHRON Systems<br>Creatinine Reagent on UniCel<br>DxC 600/800 SYNCHRON<br>Clinical System (K042291) | Candidate<br>CREA Test with i-STAT 1 System | | Intended Use | The creatinine test system is intended<br>for the quantitative determination of<br>creatinine in human serum and urine.<br>Measurements of creatinine are used<br>in the diagnosis and treatment of<br>renal disease. Serum creatinine<br>measurements prove useful in<br>evaluation of kidney glomerular<br>function and in monitoring renal<br>dialysis. | The i-STAT CHEM8+ cartridge with the<br><i>in vitro</i> quantification of glucose and<br>creatinine in arterial or venous whole<br>blood in point of care or clinical<br>laboratory settings.<br>Creatinine measurements are used in the<br>diagnosis and treatment of renal diseases,<br>in monitoring renal dialysis, and as a<br>calculation basis for measuring other<br>urine analytes. | | Reportable Range | 0.1 – 25.0 mg/dL (Serum) | 0.2-20.0 mg/dL | | Sample Type | Serum, urine | Arterial or venous whole blood | | Sample Volume | 0.5 mL (500 $\mu$ L) | 95 $\mu$ L | | Sample<br>Preparation | Sample tubes prepared and then<br>processed within analyzer | Ready to Use | | Traceability | Isotope Dilution Mass Spectroscopy | NIST SRM 967 | | Calibration | Must be conducted every 24 hours<br>and with each new reagent | 1-point on-board contained within the<br>cartridge | | Time to Test<br>(Sample Stability) | Serum or plasma: Within 8 hours at<br>room temperature, or up to 48 hours<br>if stored at +2°C to +8°C<br><br>Urine: Within 2 hours of collection | Heparinized samples: within 30 minutes<br>of collection | | Principle of<br>Measurement | Colorimetric (520 nm) measurement<br>of creatinine-alkaline picrate<br>complex formation | Amperometric measurement of oxidized<br>hydrogen peroxide produced by<br>creatinine amidinohydrolase activity | | Reagent Format | Reagent handling system, stored<br>within analyzer | Cartridge | | Reagent Storage<br>and Stability | Room temperature | 2°C to 8°C (35-46°F) | | Analyzer Type | Floor Model | Handheld | {7}------------------------------------------------ ## 7. Performance Characteristics #### Analytical Performance - a. Precision ## Precision 20 days (aqueous materials) The precision of the i-STAT Glucose and Creatinine tests on the i-STAT 1 Wireless Analyzer was evaluated using 5 levels of aqueous materials. This 20-day multi-day precision testing was based on CLSI document EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline-Third Edition. The study was conducted using multiple instruments and one test cartridge lot over 20 days at one site. Total precision ('within-laboratory', Sr), within-run, (Sr), between-run, (Sr) and between-day, (Saa) were estimated for each level. The results of the 20-day precision study are shown in Table 1. | Table 1: 20-Day Precision of i-STAT Glucose and Creatinine tests on the i-STAT 1<br>Analyzer | | | | | | | | | | | | |----------------------------------------------------------------------------------------------|----------------|----|-------|-------|------------|------------|------------|-------------|-------------|-------------|-------------| | i-STAT<br>Test | Fluid<br>Level | N | Mean | Total | | Within-run | | Between-run | | Between-day | | | | | | | ST | CVT<br>(%) | Sr | CVr<br>(%) | Srr | CVrr<br>(%) | Sdd | CVdd<br>(%) | | Creatinine<br>(mg/dL) | CV L1 | 81 | 15.90 | 0.337 | 2.1 | 0.321 | 2.0 | 0.096 | 0.6 | 0.030 | 0.2 | | | CV L2 | 81 | 4.23 | 0.101 | 2.4 | 0.093 | 2.2 | 0.029 | 0.7 | 0.027 | 0.6 | | | CV L3 | 81 | 1.69 | 0.035 | 2.1 | 0.033 | 2.0 | 0.008 | 0.5 | 0.007 | 0.4 | | | CV L4 | 80 | 0.51 | 0.029 | 5.7 | 0.027 | 5.3 | 0.008 | 1.6 | 0.006 | 1.2 | | | CV L5 | 80 | 0.18 | 0.028 | 15.6 | 0.026 | 14.4 | 0.007 | 3.9 | 0.007 | 3.9 | | Glucose<br>(mg/dL) | CV L1 | 80 | 26.7 | 0.49 | 1.8 | 0.47 | 1.8 | 0.13 | 0.5 | 0.10 | 0.4 | | | CV L2 | 81 | 40.9 | 0.52 | 1.3 | 0.46 | 1.1 | 0.16 | 0.4 | 0.18 | 0.4 | | | CV L3 | 81 | 123.0 | 0.47 | 0.4 | 0.43 | 0.3 | 0.13 | 0.1 | 0.12 | 0.1 | | | CV L4 | 80 | 286.5 | 1.40 | 0.5 | 1.25 | 0.4 | 0.55 | 0.2 | 0.33 | 0.1 | | | CV L5 | 80 | 608.1 | 5.56 | 0.9 | 5.21 | 0.9 | 1.46 | 0.2 | 1.31 | 0.2 | #### Precision (Whole Blood) The whole blood precision of the i-STAT Glucose and Creatinine Tests on the i-STAT 1 Analyzer were evaluated using venous whole blood (native or altered) samples targeted to four levels within the test reportable range. One test cartridge lot was used across 3 point of care sites. At each site, each sample was tested 3 times on each of 7 i-STAT 1 Wireless Analyzers (total of 21 test results per sample per instrument). The results of the whole blood precision using the i-STAT 1 Wireless Analyzer are shown in Table 2. {8}------------------------------------------------ | Table 2: Whole Blood Precision Results – i-STAT 1 System | | | | | | | | | | | |----------------------------------------------------------|-----------------|------|----|-------|-------|------|-------|----------------|------|-------------| | | Within-Analyzer | | | Total | | | | | | | | i-STAT Test | Level | Site | N | Mean | SD | %CV | SD | SD 95% CI | %CV | %CV 95% CI | | | | 01 | 21 | 95.3 | 0.98 | 1.0 | 0.98 | 0.81 to 1.39 | 1.0 | 0.8 to 1.5 | | | | 01 | 21 | 72.3 | 1.23 | 1.7 | 1.23 | 1.03 to 1.75 | 1.7 | 1.4 to 2.4 | | | | 02 | 21 | 95.1 | 0.90 | 0.9 | 0.90 | 0.75 to 1.27 | 0.9 | 0.8 to 1.3 | | | 30-110 | 02 | 21 | 95.5 | 0.69 | 0.7 | 0.69 | 0.57 to 0.98 | 0.7 | 0.6 to 1.0 | | | | 02 | 21 | 80.0 | 0.38 | 0.5 | 0.38 | 0.31 to 0.63 | 0.5 | 0.4 to 0.8 | | | | 03 | 21 | 101.3 | 0.76 | 0.8 | 0.76 | 0.61 to 1.17 | 0.8 | 0.6 to 1.2 | | | | 03 | 21 | 87.8 | 0.58 | 0.7 | 0.58 | 0.47 to 0.86 | 0.7 | 0.5 to 1.0 | | | | 03 | 21 | 98.9 | 0.58 | 0.6 | 0.63 | 0.50 to 1.07 | 0.6 | 0.5 to 1.1 | | Glucose<br>(mg/dL) | | 01 | 21 | 148.2 | 0.62 | 0.4 | 0.62 | 0.51 to 0.88 | 0.4 | 0.3 to 0.6 | | | 111-150 | 02 | 20 | 143.0 | 0.85 | 0.6 | 0.85 | 0.70 to 1.22 | 0.6 | 0.5 to 0.9 | | | | 02 | 14 | 143.3 | 1.25 | 0.9 | 1.25 | 1.06 to 2.03 | 0.9 | 0.7 to 1.4 | | | | 03 | 21 | 142.2 | 0.79 | 0.6 | 0.79 | 0.65 to 1.12 | 0.6 | 0.5 to 0.8 | | | | 01 | 21 | 385.7 | 2.38 | 0.6 | 2.98 | 2.33 to 5.40 | 0.8 | 0.6 to 1.4 | | | 151-400 | 02 | 21 | 318.0 | 3.25 | 1.0 | 3.25 | 2.70 to 4.62 | 1.0 | 0.8 to 1.5 | | | | 03 | 21 | 151.8 | 1.02 | 0.7 | 1.02 | 0.83 to 1.57 | 0.7 | 0.5 to 1.0 | | | | 01 | 21 | 618.4 | 7.95 | 1.3 | 7.95 | 6.54 to 11.67 | 1.3 | 1.1 to 1.9 | | | 401-700 | 02 | 21 | 444.2 | 2.23 | 0.5 | 2.23 | 1.85 to 3.14 | 0.5 | 0.4 to 0.7 | | | | 03 | 21 | 582.0 | 2.82 | 0.5 | 2.93 | 2.33 to 4.84 | 0.5 | 0.4 to 0.8 | | | | 01 | 21 | 0.82 | 0.038 | 4.6 | 0.044 | 0.035 to 0.078 | 5.4 | 4.3 to 9.5 | | | | 01 | 21 | 0.60 | 0.038 | 6.3 | 0.038 | 0.031 to 0.053 | 6.3 | 5.2 to 8.8 | | | <1 | 01 | 21 | 0.52 | 0.049 | 9.4 | 0.049 | 0.040 to 0.070 | 9.4 | 7.7 to 13.5 | | | | 02 | 20 | 0.96 | 0.049 | 5.1 | 0.051 | 0.041 to 0.086 | 5.3 | 4.3 to 9.0 | | | | 02 | 21 | 0.96 | 0.053 | 5.5 | 0.053 | 0.044 to 0.080 | 5.5 | 4.6 to 8.3 | | | | 02 | 21 | 0.56 | 0.058 | 10.4 | 0.058 | 0.048 to 0.082 | 10.4 | 8.6 to 14.6 | | | | 02 | 21 | 0.87 | 0.044 | 5.1 | 0.049 | 0.039 to 0.084 | 5.6 | 4.5 to 9.7 | | | | 03 | 21 | 0.81 | 0.031 | 3.8 | 0.031 | 0.025 to 0.048 | 3.8 | 3.1 to 5.9 | | | | 03 | 21 | 0.70 | 0.031 | 4.4 | 0.032 | 0.025 to 0.052 | 4.6 | 3.6 to 7.4 | | | | 03 | 21 | 0.59 | 0.031 | 5.3 | 0.031 | 0.025 to 0.048 | 5.3 | 4.2 to 8.1 | | Creatinine<br>(mg/dL) | 1-15 | 03 | 21 | 1.23 | 0.049 | 4.0 | 0.049 | 0.039 to 0.078 | 4.0 | 3.2 to 6.3 | | | | 03 | 21 | 1.17 | 0.049 | 4.2 | 0.049 | 0.040 to 0.074 | 4.2 | 3.4 to 6.3 | | | 1.5-2.0 | 01 | 21 | 1.53 | 0.049 | 3.2 | 0.049 | 0.040 to 0.074 | 3.2 | 2.6 to 4.8 | | | | 02 | 14 | 1.83 | 0.053 | 2.9 | 0.053 | 0.044 to 0.090 | 2.9 | 2.4 to 4.9 | | | | 02 | 21 | 1.97 | 0.062 | 3.1 | 0.062 | 0.051 to 0.092 | 3.1 | 2.6 to 4.7 | | | | 03 | 21 | 1.70 | 0.058 | 3.4 | 0.058 | 0.047 to 0.088 | 3.4 | 2.8 to 5.2 | | | 5.0-7.0 | 01 | 21 | 5.62 | 0.172 | 3.1 | 0.172 | 0.142 to 0.246 | 3.1 | 2.5 to 4.4 | | | | 02 | 21 | 6.31 | 0.246 | 3.9 | 0.246 | 0.202 to 0.366 | 3.9 | 3.2 to 5.8 | | | | 03 | 21 | 5.30 | 0.072 | 1.4 | 0.072 | 0.059 to 0.107 | 1.4 | 1.1 to 2.0 | | | 7.0-12 | 02 | 21 | 9.47 | 0.127 | 1.3 | 0.155 | 0.121 to 0.277 | 1.6 | 1.3 to 2.9 | | | | 01 | 21 | 14.37 | 0.388 | 2.7 | 0.388 | 0.321 to 0.561 | 2.7 | 2.2 to 3.9 | | | >12 | 02 | 21 | 14.90 | 0.515 | 3.5 | 0.515 | 0.427 to 0.732 | 3.5 | 2.9 to 4.9 | | | | 03 | 21 | 14.30 | 0.558 | 3.9 | 0.558 | 0.462 to 0.795 | 3.9 | 3.2 to 5.6 | {9}------------------------------------------------ ### b. Linearity The study was designed based on CLSI EP06-A: Evaluation of the linearity of quantitative measurement procedures. The linearity of the i-STAT Glucose and Creatinine tests on the i-STAT 1 Analyzer were evaluated by preparing whole blood samples of varying analyte levels that spanned the reportable range of the test. The best fitting regression model was the third order model. The absolute degree of nonlinearity results met the acceptance criteria for each of the levels tested for each analyte. Therefore, the i-STAT Glucose and Creatinine test demonstrated linearity over the reportable range as shown in Table 3. Regression summary of the Glucose and Creatinine response versus the concentration of the whole blood samples of varying analyte levels is also provided in Table 3. | Table 3: Regression Summary for the i-STAT tests on the i-STAT 1 Analyzer | | | | | | |---------------------------------------------------------------------------|------------------|--------------|--------|-----------|--------| | i-STAT Test | Reportable Range | Range Tested | Slope | Intercept | R2 | | Glucose (mg/dL) | 20-700 | 17 - 620 | 0.9794 | -1.603 | 0.9991 | | Creatinine (mg/dL) | 0.20 – 20.0 | 0.13 - 16.2 | 1.021 | 0.0763 | 0.9978 | # c. Limit of Quantitation (LoO) The study was based on the CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline-Second Edition. The LoO of the i-STAT Glucose and Creatinine tests were evaluated on the i-STAT 1 Analyzer using whole blood that was altered to low glucose (< 20 mg/dL) and low creatinine (< 0.2 mg/dL) concentrations and two test cartridge lots. The LoQs for the i-STAT Glucose and Creatinine tests were determined to be 12 mg/dL and 0.10 mg/dL respectively, which is below the lower limit of the reportable range for each of the i-STAT Glucose and Creatinine tests as shown in Table 3. # d. Limit of Blank and Detection (LoB/LoD) The study was based on CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline-Second Edition. The LoB/LoD of the i-STAT glucose and creatinine tests were evaluated on the i-STAT 1 Analyzer using whole blood that was altered to "blank" analyte (glucose or creatinine) concentration for LoB testing and two "low" analyte (glucose or creatinine) concentrations for LoD testing. The LoB and LoD were determined based on the maximal LoB or LoD value obtained for each lot tested. The LoB and LoD for i-STAT Creatinine and Glucose tests on the i-STAT 1 Analyzer were determined as shown in the Table 4 below. | Table 4: Summary of LoB and LoD Results Per Test (mg/dL) | | | |----------------------------------------------------------|------|------| | i-STAT Test | LoB | LoD | | Creatinine | 0.05 | 0.10 | | Glucose | 1 | 2 | {10}------------------------------------------------ #### Hematocrit Sensitivity to the i-STAT Glucose Test e. A hematocrit sensitivity study was performed for the i-STAT Glucose test to assess the effect of hematocrit. Three hematocrit levels (low, mid and high) were evaluated across four (4) glucose levels across the reportable range of the test. The results of the study demonstrated that the i-STAT Glucose test performs equivalently at different hematocrit levels when tested on the i-STAT 1 analyzer. # f. Oxygen Sensitivity The effect on oxygen on the i-STAT Creatinine and Glucose tests was evaluated with high and low ranges of oxygen. The equivalency between the high and low conditions was determined by a two-sample equivalence test using the analyte allowable error (Ea) relative to the analyte control mean/median results. The study demonstrated that the i-STAT Creatinine and Glucose tests are insensitive to oxygen changes between 20 to >500 mmHg for creatinine and 25 to >500 mmHg for glucose. # g. Altitude The results of the study show equivalent creatinine results at elevations up to 6367 feet and for glucose up to 9523 feet. The correlation coefficient and slope results from the Passing-Bablok regression analysis met the acceptance criteria as summarized as shown in Table below. | Test | Average Measured<br>Altitude | Results | |------------|------------------------------|----------------------| | Creatinine | 6367 feet | r 1.00<br>Slope 1.13 | | Glucose | 9523 feet | r 1.00<br>Slope 1.00 | {11}------------------------------------------------ #### h. Interference The interference performance of the i-STAT Glucose and Creatinine tests on the i-STAT 1 Analyzer was evaluated using whole blood samples based on CLSI EP07 ED3: Interference Testing in Clinical Chemistry, Third Edition. The effect of each substance was evaluated by comparing the performance of a control sample, spiked with blank solvent solution, with the test results from a sample spiked with the potentially interfering substance at the toxic/pathological concentration based on CLSI EP37 ED1: Supplemental Tables for Interference Testing in Clinical Chemistry, First Edition, as applicable. A substance was identified as an interferent if the difference between the control and test samples was outside of the allowed error (Ea) for the i-STAT test. Table 6 and Table 7 contain the lists of potentially interfering substances tested for the i-STAT Glucose and Creatinine tests and the interference results. | Table 6: Summary of Substance Tested and Interference Results for the i-STAT Glucose<br>test | | | | | |----------------------------------------------------------------------------------------------|--------------------|---------|--------------|------------------------------------| | Substance | Test Concentration | | Interference | Interference | | | mmol/L | mg/dL | (Yes/No) | Results | | Acetaldehyde | 0.045a | 0.2 | No | | | Acetaminophen | 1.03 | 15.6 | No | | | Lithium Acetoacetate | 2.0 | 20 | No | | | N-Acetyl-L-Cysteine | 0.92 | 15.0 | No | | | Ammonium Chloride | 2.0a | 10.70 | No | | | L-Ascorbic Acid | 0.298 | 5.25 | No | | | Bilirubin | 0.684 | 40 | No | | | Lithium Bromide | 37.5a | 325.69 | Yes | Decreased results<br>≥ 11.8 mmol/L | | Cholesterol | 10.3 | 400 | No | | | Creatinine | 1.326 | 15 | No | | | Dopamine Hydrochloride | 4.06 µmol/L | 0.0621 | No | | | Ethanol | 130 | 600 | No | | | Lithium Fluoride | 0.0632 | 0.12 | No | | | Formaldehyde | 0.133a | 0.399 | No | | | Fructose | 1 | 18 | No | | | Galactose | 3.33 | 60 | No | | | Gentamicin Sulfate | 0.0628 | 3 | No | | | Glucosamine Hydrochloride | 0.030a | 0.647 | No | | | Glutathione, reduced | 3 | 3 mEq/L | No | | | Glycolic Acid | 10.0a | 76.05 | No | | #### i-STAT Glucose Test a No CLSI EP37 test concentration available. The molecular weight of the substance tested was used to convert the test concentration from mmol/L to mg/dL. The molecular weight of each substance could vary depending on the form chosen. {12}------------------------------------------------ | test | | | | | |-----------------------|--------------------|----------|--------------|------------------------------------| | Substance | Test Concentration | | Interference | Interference | | | mmol/L | mg/dL | (Yes/No) | Results | | Guaifenesin | 0.0227 | 0.45 | No | | | Hemoglobin | 10 g/L | 1000 | No | | | Sodium Heparin | 3.30 U/mL | 330 U/dL | No | | | β-Hydroxybutyric Acid | 6.0aa | 62.47 | No | | | Hydroxyurea | 0.405 | 3.08 | Yes | Increased results<br>≥ 0.08 mmol/L | | Ibuprofen | 1.06 | 21.9 | No | | | Isoniazid | 0.438 | 6 | No | | | Lithium Lactate | 10 | 90 | No | | | Mannose | 1.0a | 18.02 | No | | | Maltose | 10.5 | 360 | No | | | pH | 8.0 pH units | N/A | No | | | Lithium Pyruvate | 0.570 | 5 | No | | | Lithium Salicylate | 0.207 | 2.86 | No | | | Lithium Thiocyanate | 0.898 | 5.22 | No | | | Sodium Thiosulfate | 16.7a | 264.04 | No | | | Triglyceride | 16.94 | 1500 | No | | | Uric Acid | 1.4 | 23.5 | No | | | Xylose | 3a | 45.04 | No | | | Gentisic Acid | 0.0973 | 1.50 | No | | | Table 6: Summary of Substance Tested and Interference Results for the i-STAT Glucose #### i-STAT Creatinine Test | Table 7: Summary of Substance Tested and Interference Results for the i-STAT<br>Creatinine test | | | | | | |-------------------------------------------------------------------------------------------------|--------------------|--------|--------------|------------------------------------|--| | Substance | Test Concentration | | Interference | Interference | | | | mmol/L | mg/dL | (Yes/No) | Results | | | Acetaldehyde | 0.045a | 0.2 | No | | | | Acetaminophen | 1.03 | 15.6 | No | | | | N-Acetyl-L-Cysteine | 0.92 | 15.0 | No | | | | L-Ascorbic Acid | 0.298 | 5.25 | No | | | | β-Hydroxybutyric Acid | 6.0a | 62.47 | No | | | | Bicarbonate | 35.0 | 294.0 | No | | | | Bilirubin | 0.684 | 40 | No | | | | Lithium Bromide | 37.5a | 325.69 | Yes | Increased results<br>≥ 18.3 mmol/L | | | Calcium Chloride | 5.0 | 20 | No | | | | Creatine | 0.382a | 5.01 | No | | | | Dopamine Hydrochloride | 4.06 umol/L | 0.0621 | No | | | | Formaldehyde | 0.133a | 0.399 | No | | | a No CLSI EP37 test concentration available. The molecular weight of the substance tested was used to convert the test concentration from mmol/L to mg/dL. The molecular weight of each substance could vary depending on the form chosen. {13}------------------------------------------------ | Table 7: Summary of Substance Tested and Interference Results for the I-STAT Creatinine test | | | | | |----------------------------------------------------------------------------------------------|--------------------|--------|-----------------------|------------------------------------| | Substance | Test Concentration | | Interference (Yes/No) | Interference Results | | | mmol/L | mg/dL | | | | Glycolic Acid | 10.0a | 76.05 | No | | | Hemoglobin | 10 g/L | 1000 | No | | | Hydroxyurea | 0.405 | 3.08 | Yes | Increased results<br>≥ 0.03 mmol/L | | Lithium Lactate | 10 | 90 | No | | | Methyldopa | 107 $ μ $ mol/L | 2.25 | No | | | pH | 8.0 pH units | N/A | No | | | Lithium Pyruvate | 0.570 | 5 | No | | | Lithium Salicylate | 0.207 | 2.86 | No | | | Sodium Thiosulfate | 16.7a | 264.04 | No | | | Triglyceride | 16.94 | 1500 | No | | | Uric Acid | 1.4 | 23.5 | No | | | Table 7: Summary of Substance Tested and Interference Results for the i-STAT ## Comparison Study #### Method Comparison with Predicate Device i. Method comparison was demonstrated in a study comparing the i-STAT Glucose and Creatinine test performance on the i-STAT 1 to the Beckman DxC. The study was based on CLSI guideline EP09c-ED3. Venous and arterial blood specimens were evaluated and analyzed on the i-STAT 1 against plasma specimens on the Beckman DxC. A Passing-Bablok linear regression analysis was performed using the first replicate result from the i-STAT 1 versus the singlicate result of the comparative method. The i-STAT System automatically runs a comprehensive set of quality checks of both the analyzer and cartridge performance each time a sample is tested. This internal quality system will suppress results by generating a Quality Check Code (OCC) if the analyzer, cartridge or sample does not meet certain internal specifications. When a QCC occurs, a single code number, the type of problem and the next step to be taken will be displayed on the i-STAT Analyzer. The failure rate for a single cartridge due to OCCs may be as high as 4%. The rate of failure for two consecutive cartridges due to QCCs may be as high as 1.7%. | Table 5: Method Comparison Results | | | | | | |------------------------------------|-----|-------|-----------|------|--| | i-STAT Test | N | Slope | Intercept | r | | | Glucose | 185 | 0.98 | 0.00 | 1.00 | | | Creatinine | 180 | 1.043 | -0.062 | 1.00 | | #### 8. Conclusion The results of these studies demonstrate that performance of the i-STAT CHEM8+ Glucose and Creatinine tests with the i-STAT 1 System are substantially equivalent to the comparative method.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...