K203549 · Nova Biomedical Corporation · CGA · Oct 18, 2022 · Clinical Chemistry
Device Facts
Record ID
K203549
Device Name
Nova Primary Glucose Analyzer System
Applicant
Nova Biomedical Corporation
Product Code
CGA · Clinical Chemistry
Decision Date
Oct 18, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Nova Primary Glucose Analyzer System is indicated for in vitro diagnostic use by healthcare professionals in clinical laboratory settings for the quantitative determination of Glucose in lithium heparinized venous whole blood and plasma. The measurement of Glucose is used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell carcinoma.
Device Story
Nova Primary Glucose Analyzer is a portable laboratory instrument for quantitative glucose measurement in lithium heparinized venous whole blood or plasma. Device accepts 25 microliter samples via syringes, collection tubes, or cups. Principle of operation: glucose oxidase enzymatic reaction produces H2O2; platinum anode oxidizes H2O2 at 0.70V constant potential; generated current is proportional to glucose concentration. For whole blood, an integrated impedance-based conductivity detector automatically adjusts results to provide plasma-equivalent values. Operated by trained laboratory personnel in clinical settings. Output displayed on PCAP color touchscreen; results printed via onboard thermal printer. Device aids clinicians in diagnosing and managing carbohydrate metabolism disorders. Benefits include automated hematocrit correction for whole blood samples and modern user interface compared to legacy systems.
Clinical Evidence
Bench testing only. Precision evaluated with QC, venous whole blood, and plasma samples (n=20 replicates; run-to-run n=240/80). Linearity confirmed across 20–900 mg/dL range using 10 levels. Interference testing performed for 30+ substances. Method comparison against YSI 2300 (n=174 whole blood, n=170 plasma) showed high correlation (r > 0.99). No clinical studies performed.
Technological Characteristics
Glucose oxidase enzymatic sensor; platinum anode electrode; impedance-based conductivity detector for hematocrit correction. Dimensions: 19.0 x 19.0 x 6.0 inches. Power: 100-240 VAC, 47-63 Hz. Interface: PCAP color touchscreen. Connectivity: Optional wireless barcode scanner/keyboard via internal USB dongle. Traceable to isotope dilution mass spectrometry.
Indications for Use
Indicated for healthcare professionals in clinical laboratory settings for quantitative glucose determination in lithium heparinized venous whole blood and plasma; used for diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
YSI Model 2300 Stat Glucose And L-lactate Analyzer (K891480)
K063821 — NOVA STATSTRIP GLUCOSE HOSPITAL METER SYSTEM · Nova Biomedical Corp. · Jan 24, 2007
K070960 — NOVA STATSTRIP XPRESS GLUCOSE HOSPITAL METER SYSTEM · Nova Biomedical Corp. · Jul 11, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 18, 2022
Nova Biomedical Corporation Robert Zinck Senior Manager Regulatory Affairs 200 Prospect Street Waltham, MA 02454
Re: K203549
Trade/Device Name: Nova Primary Glucose Analyzer System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: CGA Dated: July 29, 2022 Received: August 2, 2022
### Dear Robert Zinck:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Paula Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203549
#### Device Name
Nova Primary Glucose Analyzer System
#### Indications for Use (Describe)
The Nova Primary Glucose Analyzer System is indicated for in vitro diagnostic use by healthcare professionals in clinical laboratory setting for the quantitative determination of Glucose in lithium heparinized venous whole blood and plasma.
The measurement of Glucose is used in the diagnosis and treatment of carbohydrate metabolism distuding diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell carcinoma.
Type of Use (Select one or both, as applicable)
| <span></span> | <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|--------------------------------------------------------------------------------------|
| <span></span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K203549
| 510(K) Owner: | Nova Biomedical Corporation |
|----------------------|-------------------------------------------------|
| Registration Number: | 1219029 |
| Address: | 200 Prospect St.<br>Waltham, MA 02454 |
| Phone: | 781-894-0800 |
| Fax Number: | 784-891-4806 |
| Contact Person: | Robert Zinck, Senior Manager Regulatory Affairs |
| Date Prepared: | October 14, 2022 |
Proprietary Name: Nova Primary Glucose Analyzer System
#### Common or Usual Name: Blood Glucose Analyzer
#### Classification Name:
| Regulation section | Classification | Product<br>code |
|---------------------------------------|----------------|-----------------|
| 21 CFR § 864.1345 Glucose Test System | Class II | CGA |
## Product Code:
CGA
#### Predicate Device:
k891480 -YSI 2300 Stat Glucose and L-Lactate Analyzer
#### Device Description:
The Nova Primary Glucose Analyzer System is a small, portable laboratory glucose analyzer that measures blood glucose levels in lithium heparinized whole blood or plasma utilizing a glucose oxidase based sensor and membrane/cap assembly.
The Nova Primary Glucose Analyzer accepts samples from syringes, blood collection tubes, microcentrifuge tubes, and sample cups. The sample size for analysis is 25 microliters (aspirated volume).
#### Nova Primary System Components:
The Nova Primary Glucose Analyzer System is comprised of the following components.
- Nova Primary Glucose Analyzer ●
- . Nova Primary Glucose Sensor
- . Nova Primary Glucose Membrane
- Nova Primary Calibrator Cartridge ●
- Optional Barcode Scanner .
- . IFU/Labeling
#### Sample Types:
The Nova Primary Glucose Analyzer System accepts lithium heparinized venous whole blood and plasma.
{4}------------------------------------------------
#### Intended Use:
Please see Indication for Use.
#### Indication for Use:
The Nova Primary Glucose Analyzer System is indicated for in vitro diagnostic use by healthcare professionals in clinical laboratory settings for the quantitative determination of Glucose in lithium heparinized venous whole blood and plasma.
The measurement of Glucose is used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypodycemia, and idiopathic hypodlycemia, and pancreatic islet cell carcinoma.
#### Summary of the Technological Characteristics:
The Nova Primary Glucose Analyzer was designed to be substantially equivalent in performance to the predicate device, incorporating modern technology updates in the analyzer platform.
The Nova Primary Glucose Analyzer utilizes a discrete, easily replaceable glucose sensor and membrane/cap assembly. Glucose measurement is conducted on diluted venous whole blood or plasma samples and is based on the enzymatic reaction between glucose and oxygen molecules in the presence of the glucose oxidase enzyme. All glucose measurement algorithms are intended to report plasma equivalent glucose results that are substantially equivalent to the predicate device plasma glucose result, and are traceable to isotope dilution mass spectrometry. For whole blood specimens, the Nova Primary Glucose Analyzer utilizes an impedance-based conductivity detector to automatically adjust glucose measurement and provide the plasma equivalent glucose value based on the conductivity of the sample.
#### Principle of Measurement:
Glucose measurement is based on the level of H2O2 produced during the enzymatic reaction between glucose and oxygen molecules in the presence of the dlucose enzyme. At a constant potential of 0.70 volts, electroactive HzO2 is oxidized at the surface of the platinum anode. The current generated by the flow of electrons at the surface of the platinum electrode is proportional to the glucose concentration of the sample.
#### Summary of Performance Testing:
Testing was completed to show that the Nova Primary Glucose Analyzer demonstrates substantial equivalence to the YSI 2300 Stat Glucose and L-Lactate Analyzer. The performance testing included:
- Method Comparison Studies ●
- Precision/Reproducibility Studies .
- Linearity Testing
- Specificity / Interference Testing ●
- Detection Limit .
- Shelf Life Stability Testing .
The results of the performance testing confirmed that the Nova Primary Glucose Analyzer demonstrates substantial equivalence to the YSI 2300 Stat Glucose and L-Lactate Analyzer.
#### Conclusion:
The results of software validation and performance verification testing confirmed that the Nova Primary Glucose Analyzer System demonstrates substantial equivalence to the YSI 2300 Stat Glucose and L-Lactate Analyzer (k891480 - predicate device).
{5}------------------------------------------------
| | Table 1: Comparison of Predicate and Proposed Devices |
|--|-------------------------------------------------------|
|--|-------------------------------------------------------|
| Characteristic | Predicate: | Proposed: |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended/Indications<br>for Use | The YSI 2300 STAT PLUS Glucose and<br>Lactate Analyzer is a laboratory instrument<br>intended for use in clinical care and sports<br>medicine applications. It provides quick<br>measurements of glucose in whole blood,<br>plasma or serum; and of L-lactate in whole<br>blood, plasma, or cerebrospinal fluid (CSF).<br>In whole blood or plasma, glucose and L-<br>lactate can be measured simultaneously. | The Nova Primary Glucose Analyzer System<br>is indicated for <i>in vitro</i> diagnostic use by<br>healthcare professionals in clinical laboratory<br>settings for the quantitative determination of<br>Glucose in lithium heparinized venous whole<br>blood and plasma.<br>The measurement of Glucose is used in the<br>diagnosis and treatment of carbohydrate<br>metabolism disturbances including diabetes<br>mellitus, neonatal hypoglycemia, and<br>idiopathic hypoglycemia, and pancreatic islet<br>cell carcinoma. |
| Intended Users | Trained laboratory personnel | Same |
| Measured Tests | Glucose and Lactate | Glucose |
| Glucose Enzyme | Glucose Oxidase | Same |
| Acceptable Sample<br>Types | Whole Blood, Plasma, Serum | Lithium Heparin Whole Blood or Plasma |
| Sample Size<br>Requirement | 25 microliters (aspirated volume) | Same |
| Physical Dimensions | 10.0 x 14.0 x 14.0 inches | 19.0 x 19.0 x 6.0 inches |
| Instrument Weight | 25 pounds | Less than 30 pounds including the calibrator<br>cartridge |
| Sample Tray | Available Option | Not available |
| Bar Code Scanner | Not available | Optional, wireless bar code scanner via<br>internal USB dongle |
| Keyboard | Not available | Optional, wireless keyboard via internal USB<br>dongle |
| Power Supply | 110-120 VAC, 220-240 VAC<br>50–60 Hz<br>50 Watts nominal | 100-240 VAC<br>47-63 Hz<br>Less than 50 Watts |
| Automatic Whole<br>Blood Hematocrit<br>Correction | No - Hematocrit is independently measured<br>on a whole blood specimen then manually<br>entered into YSI 2300 to calculate equivalent<br>plasma glucose value. | Yes - Contains an impedance-based<br>Conductivity Detector. Whole blood samples<br>are automatically adjusted to provide plasma<br>equivalent glucose values based on the<br>conductivity of the sample. |
| User Interface<br>Display | Alpha Numeric Liquid Crystal Display with 20<br>membrane keypad | PCAP Color Touchscreen Display with icon<br>based Graphical User Interface |
| Printer | Yes onboard 2 ¼" thermal printer | Yes, onboard 2" thermal printer |
| Communication<br>Protocol | RS-232 serial port | Not available |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.