K223710 · Abbott Point of Care, Inc. · CGA · Jul 28, 2023 · Clinical Chemistry
Device Facts
Record ID
K223710
Device Name
i-STAT CG8+ cartridge with the i-STAT 1 System
Applicant
Abbott Point of Care, Inc.
Product Code
CGA · Clinical Chemistry
Decision Date
Jul 28, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The i-STAT CG8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of glucose in arterial, venous, or capillary whole blood in point of care or clinical laboratory settings. Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma. The i-STAT 1 Analyzer is intended for use in the in vitro quantification of various analytes in whole blood or plasma in point of care or clinical laboratory settings. Analyzers and cartridges should be used by healthcare professionals trained and certified to use the system and should be used according to the facility's policies and procedures. The i-STAT System is for in vitro diagnostics use. Caution: Federal law restricts this device to sale by or on the order of a licensed practitioner.
Device Story
i-STAT 1 System is a handheld, point-of-care diagnostic platform using single-use disposable cartridges. Input: 2-3 drops of whole blood (arterial, venous, or capillary) applied to cartridge. Operation: Cartridge contains sensors, fluid pack, and channels; instrument moves sample across sensors via mechanical features. Principle: Amperometric measurement; voltage applied to electrodes; current generated by reduction of hydrogen peroxide (produced by catalytic oxidation of glucose via glucose oxidase) is proportional to glucose concentration. Output: Quantitative glucose result displayed on analyzer screen. Used by trained healthcare professionals in clinical or point-of-care settings. Output informs diagnosis, monitoring, and treatment of carbohydrate metabolism disorders. Benefits: Rapid, bedside quantitative analysis.
Clinical Evidence
Bench testing only. Precision studies (CLSI EP05-A3) across multiple sites and lots showed total CVs < 2%. Linearity (CLSI EP06) confirmed 20–700 mg/dL range. Interference testing (CLSI EP07) identified bromide, hydroxyurea, and isoniazid as interferents. Method comparison (CLSI EP09c) against predicate (n=339) and epoc system (n=208) showed high correlation (r=1.00 and r=0.99, respectively). Matrix comparison (CLSI EP35) supported use of lithium heparin and non-anticoagulated samples.
Technological Characteristics
Handheld, point-of-care electrochemical analyzer. Cartridge-based reagent format with glucose oxidase-based amperometric peroxide detection. Requires 95 µL sample volume. Connectivity via i-STAT 1 System accessories (Downloader/Recharger). Storage: 2-8°C (refrigerated) or 18-30°C (room temp for 2 months).
Indications for Use
Indicated for in vitro quantification of glucose in arterial, venous, or capillary whole blood for patients with carbohydrate metabolism disorders, including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma. Contraindicated for capillary whole blood use in patients with severe hypotension, shock, hyperosmolar-hyperglycemia, or severe dehydration.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
i-STAT CHEM8+ cartridge with the i-STAT 1 System (K210958)
Reference Devices
epoc Blood Analysis System
Related Devices
K223755 — i-STAT G cartridge with the i-STAT 1 System · Abbott Point of Care, Inc. · Sep 11, 2023
K183678 — i-STAT CHEM8+ cartridge with the i-STAT 1 System · Abbott Point of Care, Inc. · Feb 7, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of the HHS logo is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 28, 2023
Abbott Point of Care Inc. Jacquelyn Gesumaria Principal Specialist Regulatory Affairs 400 College Road East Princeton, New Jersey 08540
Re: K223710
Trade/Device Name: i-STAT CG8+ cartridge with the i-STAT 1 System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: CGA, JJE Dated: June 28, 2023 Received: June 28, 2023
#### Dear Jacquelyn Gesumaria:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Paula V. Caposino -S
Paula Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) k223710
Device Name
i-STAT CG8+ cartridge with the i-STAT 1 System
#### Indications for Use (Describe)
The i-STAT CG8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of glucose in arterial, venous, or capillary whole blood in point of care or clinical laboratory settings.
Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, and pancreatic islet cell carcinoma.
The i-STAT 1 Analyzer is intended for use in the in vitro quantification of various analytes in whole blood or plasma in point of care or clinical laboratory settings. Analyzers and cartridges should be used by healthcare professionals trained and certified to use the system and should be used according to the facility's policies and procedures.
The i-STAT System is for in vitro diagnostics use. Caution: Federal law restricts this device to sale by or on the order of a licensed practitioner.
Type of Use (Select one or both, as applicable)
| <span style="padding-right: 5px;"> <input checked="true" type="checkbox"/> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------------------------------------------------------------------------------------------------------------------------|
| <span style="padding-right: 5px;"> <input type="checkbox"/> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Abbott logo. On the left is a blue stylized letter A. To the right of the letter is the word "Abbott" in black, bold font.
# 510(k) SUMMARY
The information in this 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
### I. SUBMITTER INFORMATION
| Owner | Abbott Point of Care Inc.<br>400 College Road East<br>Princeton, NJ 08540 |
|---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact | Primary: Jacquelyn Gesumaria<br>Principal Specialist Regulatory Affairs<br>Phone: 609-454-9384<br><br>Secondary: Mojgan Soleimani<br>Associate Director Regulatory Affairs<br>Phone: 613-295-0932 |
| Date Prepared | June 28, 2023 |
### II. DEVICE INFORMATION
| Proprietary Name | i-STAT CG8+ cartridge with the i-STAT 1 System |
|------------------|------------------------------------------------|
| Common Name | Chemistry test, analyzer, handheld |
| 510(k) Number | K223710 |
| Product Code | Device Classification<br>Name | Regulation<br>Number | Class | Panel |
|--------------|-------------------------------|----------------------|-------|--------------------|
| CGA | Glucose Oxidase,<br>Glucose | 862.1345 | II | Clinical Chemistry |
## III. PREDICATE DEVICE
K210958 (K183678) 510(k) Number
| Product<br>Code | Device Classification<br>Name | Regulation<br>Number | Class | Panel |
|-----------------|-------------------------------|----------------------|-------|--------------------|
| CGA | Glucose Oxidase, Glucose | 862.1345 | II | Clinical Chemistry |
{4}------------------------------------------------
### IV. DEVICE DESCRIPTION
The i-STAT CG8+ cartridge is used with the i-STAT 1 analyzer as part of the i-STAT 1 System to measure glucose (Glu) in arterial, venous or capillary whole blood.
The i-STAT 1 System is an in vitro diagnostic (IVD) medical device intended for the quantitative determination of various clinical chemistry tests contained within i-STAT cartridges using whole blood. The i-STAT 1 System consists of a portable blood analyzer (i-STAT 1 analyzer), single-use disposable test cartridges (i-STAT cartridges), liquid quality control and calibration verification materials, and accessories (i-STAT 1 Downloader/Recharger, i-STAT Electronic Simulator and i-STAT 1 Printer). The i-STAT 1 System, including the i-STAT CG8+ cartridge, is designed for use by trained medical professionals in point of care or clinical laboratory settings and is for prescription use only.
The i-STAT CG8+ cartridge contains the required sensors, a fluid pack (calibrant pouch), a sample entry well and closure, fluid channels, waste chamber, and the necessary mechanical features for controlled fluid movement within cartridge. The i-STAT cartridge format allows all the tests in the cartridge to be performed simultaneously. All the test steps and fluid movement occur within the i-STAT CG8+ cartridge. Cartridges require two to three drops of whole blood, which are typically applied to the cartridge using a transfer device, by the trained user before the cartridge is placed within the analyzer.
The i-STAT 1 analyzer is a handheld, in vitro diagnostic analytical device designed to run only i-STAT test cartridges. The instrument interacts with the i-STAT CG8+ cartridge to move fluid across the sensors and generate a quantitative result (within approximately 2 minutes).
# V. INTENDED USE STATEMENT
The i-STAT CG8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of glucose in arterial, venous, or capillary whole blood in point of care or clinical laboratory settings.
Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.
The i-STAT 1 Analyzer is intended for use in the in vitro quantification of various analytes in whole blood or plasma in point of care or clinical laboratory settings. Analyzers and cartridges should be used by healthcare professionals trained and certified to use the system and should be used according to the facility's policies and procedures.
The i-STAT System is for in vitro diagnostics use. Caution: Federal law restricts this device to sale by or on the order of a licensed practitioner.
{5}------------------------------------------------
# VI. SUMMARY COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
| | Table 1: Similarities and Differences (Test and Instrument) | | | | | | | | | |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|
| Feature or<br>Characteristic | Candidate Devices:<br><i>i-STAT CG8+</i> cartridge with the <i>i-STAT 1</i><br>System, <i>i-STAT</i> Glucose test | Predicate Device:<br><i>i-STAT CHEM8+</i> cartridge with the <i>i-STAT 1</i><br>System, <i>i-STAT</i> Glucose test (K210958) | | | | | | | | |
| Intended Use | The <i>i-STAT CG8+</i> cartridge with the<br><i>i-STAT 1 System</i> is intended for use in the<br><i>in vitro</i> quantification of glucose in arterial,<br>venous or capillary whole blood in point of<br>care or clinical laboratory settings.<br>Glucose measurements are used in the<br>diagnosis, monitoring, and treatment of<br>carbohydrate metabolism disorders<br>including, but not limited to, diabetes<br>mellitus, neonatal hypoglycemia,<br>idiopathic hypoglycemia, and pancreatic<br>islet cell carcinoma. | The <i>i-STAT CHEM8+</i> cartridge with the<br><i>i-STAT 1 System</i> is intended for use in the<br><i>in vitro</i> quantification of sodium,<br>potassium, chloride, ionized calcium,<br>glucose, blood urea nitrogen, creatinine,<br>hematocrit, and total carbon dioxide in<br>arterial or venous whole blood in point of<br>care or clinical laboratory settings.<br>Glucose measurements are used in the<br>diagnosis, monitoring, and treatment of<br>carbohydrate metabolism disorders<br>including, but not limited to, diabetes<br>mellitus, neonatal hypoglycemia, idiopathic<br>hypoglycemia, and pancreatic islet cell<br>carcinoma. | | | | | | | | |
| Device<br>Classification | Same | Class II | | | | | | | | |
| Product Code | Same | CGA | | | | | | | | |
| Regulation<br>No. | Same | 862.1345 | | | | | | | | |
| Reportable<br>Range | Same | 1.1 – 38.9 mmol/L<br>20 – 700 mg/dL<br>0.20 - 7.00 g/L | | | | | | | | |
| Sample Type | Arterial, venous, or capillary whole blood | Arterial and venous whole blood | | | | | | | | |
| Sample<br>Volume | Same | 95 µL | | | | | | | | |
| Sample<br>Preparation | Same | Ready to Use | | | | | | | | |
| Sample<br>collection | Same | • Without anticoagulant<br>• With balanced heparin anticoagulant or<br>lithium heparin anticoagulant | | | | | | | | |
| Traceability | Same | NIST SRM965 | | | | | | | | |
| Calibration | Same | 1-point on-board contained within<br>cartridge | | | | | | | | |
{6}------------------------------------------------
| Table 1: Similarities and Differences (Test and Instrument) | | |
|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Feature or Characteristic | Candidate Devices:<br>i-STAT CG8+ cartridge with the i-STAT 1 System, i-STAT Glucose test | Predicate Device:<br>i-STAT CHEM8+ cartridge with the i-STAT 1 System, i-STAT Glucose test (K210958) |
| Time to Test<br>(Sample Stability) | Without anticoagulant:<br>Arterial and venous within 3 minutes<br>With anticoagulant:<br>Capillary within 3 minutes<br>Arterial and venous within 30 minutes | Without anticoagulant:<br>Arterial and venous within 3 minutes<br>With anticoagulant:<br>Arterial and venous within 30 minutes |
| Principle of Measurement | Same | Glu: Glucose oxidase based amperometric peroxide detection |
| Reagent Format | Same | Cartridge |
| Reagent Storage and Stability | Refrigerated at 2 to 8°C (35 to 46°F) until expiration date<br>Room Temperature at 18-30°C (64-86°F) for 2 months | Refrigerated at 2 to 8°C (35 to 46°F) until expiration date<br>Room Temperature at 18-30°C (64-86°F) for 14 days |
| Analyzer Type | Same | Handheld |
#### PERFORMANCE CHARACTERISTICS VII.
#### A. Analytical Performance
#### a. Precision/Reproducibility:
- i. Precision 20 days (Aqueous materials)
The precision of the i-STAT Glucose (Glu) test in the i-STAT CG8+ cartridge on the i-STAT 1 System was evaluated using five (5) levels of aqueous material. This 20day multi-day precision testing was based on CLSI document EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition. The study was conducted using multiple analyzers and two (2) test cartridge lots over 20 days at one site. Repeatability, between-run, between-day, and within-laboratory precision were estimated for each level. The results of the 20-day precision study for the i-STAT CG8+ cartridge on the i-STAT 1 System are shown in Table 2.
| Table 2: Results of 20-Day Precision of the i-STAT CG8+ Cartridge on the i-STAT 1 Analyzer | | | | | | | | | | | |
|--------------------------------------------------------------------------------------------|----------------|----|-------|---------------|------|-------------|------|-------------|------|-----------------------|------|
| Test<br>(units) | Fluid<br>Level | N | Mean | Repeatability | | Between-run | | Between-day | | Within-<br>Laboratory | |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Glu<br>(mg/dL) | CV L1 | 80 | 28.4 | 0.36 | 1.25 | 0.20 | 0.69 | 0.24 | 0.85 | 0.47 | 1.66 |
| | CV L2 | 80 | 42.0 | 0.36 | 0.87 | 0.29 | 0.70 | 0.23 | 0.56 | 0.52 | 1.25 |
| | CV L3 | 80 | 124.7 | 0.66 | 0.53 | 0.55 | 0.44 | 0.52 | 0.42 | 1.00 | 0.80 |
| | CV L4 | 80 | 279.4 | 1.47 | 0.53 | 1.79 | 0.64 | 1.26 | 0.45 | 2.63 | 0.94 |
| | CV L5 | 80 | 576.8 | 4.70 | 0.81 | 2.35 | 0.41 | 2.91 | 0.50 | 6.01 | 1.04 |
{7}------------------------------------------------
#### ii. Multi-site and operator-to-operator precision (Aqueous materials)
Multi-day precision testing was performed at three (3) sites using a panel of aqueous solutions containing five (5) levels of glucose. At each site, each level was tested once a day by two (2) operators for five (5) days on six (6) i-STAT 1 Analyzers using i-STAT CG8+ cartridges. Within-run, between-day, betweenoperator and within-site (total) variance components were calculated by site. These components were also calculated for all sites combined and provided in the Table 3 below.
{8}------------------------------------------------
| Table 3: Multi-Day Precision of the i-STAT CG8+ Cartridge on the i-STAT 1 Analyzer | | | | | | | | | | | | | | | |
|------------------------------------------------------------------------------------|----------------|----|-------|------------|------|-------------|------|------------------|------|---------------------|------|--------------|------|---------|------|
| Test<br>(units) | Fluid<br>Level | N | Mean | Within-Run | | Between-Day | | Between-Operator | | Within-Site (Total) | | Between-Site | | Overall | |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Glu<br>(mg/dL) | CV L1 | 91 | 28.1 | 0.39 | 1.40 | 0.16 | 0.55 | 0.08 | 0.29 | 0.43 | 1.54 | 0.11 | 0.41 | 0.45 | 1.59 |
| | CV L2 | 90 | 41.4 | 0.50 | 1.21 | 0.12 | 0.30 | 0.17 | 0.41 | 0.54 | 1.32 | 0.00 | 0.00 | 0.54 | 1.32 |
| | CV L3 | 96 | 124.5 | 0.64 | 0.52 | 0.47 | 0.38 | 0.08 | 0.06 | 0.8 | 0.64 | 0.00 | 0.00 | 0.80 | 0.64 |
| | CV L4 | 90 | 280.3 | 1.42 | 0.51 | 0.33 | 0.12 | 0.22 | 0.08 | 1.47 | 0.53 | 0.44 | 0.16 | 1.54 | 0.55 |
| | CV L5 | 90 | 578.4 | 4.25 | 0.73 | 1.78 | 0.31 | 0.00 | 0.00 | 4.61 | 0.80 | 0.00 | 0.00 | 4.61 | 0.80 |
{9}------------------------------------------------
#### iii. Precision (Whole Blood)
Whole blood precision of the i-STAT Glucose test in the i-STAT CG8+ cartridge on the i-STAT 1 System was evaluated using arterial, venous, and capillary whole blood specimens collected with lithium heparin. The whole blood precision was assessed using the duplicate test results collected across multiple point of care sites. The mean values for each sample were divided into subintervals for each sample type across the reportable range for the i-STAT Glucose test. The results are summarized in Table 4.
| Table 4: Whole Blood Precision of arterial, venous, and capillary whole blood for the i-STAT<br>Glucose test in the i-STAT CG8+ cartridge on the i-STAT 1 Analyzer | | | | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|--------------|-----|-------|------|------|
| Test<br>(units) | Sample Type | Sample Range | N | Mean | SD | %CV |
| Glu<br>(mg/dL) | Venous Whole Blood | 20-90 | 29 | 73.9 | 0.86 | 1.17 |
| | | >90-150 | 102 | 111.7 | 1.08 | 0.96 |
| | | >150-250 | 27 | 173.4 | 1.82 | 1.05 |
| | | >250-400 | 13 | 308.5 | 2.08 | 0.67 |
| | | >400-700 | 9 | 544.3 | 7.92 | 1.46 |
| Glu<br>(mg/dL) | Arterial Whole Blood | 20-90 | 5 | 80.8 | 0.77 | 0.96 |
| | | >90-150 | 105 | 113.6 | 0.75 | 0.66 |
| | | >150-250 | 35 | 178.0 | 1.55 | 0.87 |
| | | >250-400 | 8 | 280.7 | 2.19 | 0.78 |
| Glu<br>(mg/dL) | Capillary Whole Blood | 20-90 | 32 | 77.6 | 1.08 | 1.39 |
| | | >90-150 | 107 | 108.3 | 2.34 | 2.16 |
| | | >150-700 | 15 | 203.8 | 2.74 | 1.34 |
#### b. Linearity/assay reportable range:
#### i. Linearity
The study was designed based on CLSI EP06-Ed2: Evaluation of the Linearity of Quantitative Measurement Procedures - Second Edition.
The linearity of the i-STAT Glucose test in the i-STAT CG8+ cartridge with the i-STAT 1 System was evaluated by preparing whole blood samples of varying glucose levels. The i-STAT Glucose test in the i-STAT CG8+ cartridge demonstrated linearity over the reportable range for each i-STAT test. Regression summary of the response for the i-STAT Glucose test versus the concentration of the whole blood samples of varying glucose levels is provided in Table 5.
| Table 5: Regression Summary for the i-STAT Glucose test in the i-STAT CG8+ Cartridge on the<br>i-STAT 1 Analyzer | | | | | | |
|------------------------------------------------------------------------------------------------------------------|-------|------------------|--------------|-------|-----------|--------|
| Test | Units | Reportable Range | Range Tested | Slope | Intercept | R2 |
| Glu | mg/dL | 20 – 700 | 17.1 – 795.4 | 0.994 | -1.385 | 0.9993 |
#### c. Detection Limit
#### i. Limit of Quantitation (LoQ)
The study was based on the CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition.
The LoQ of the i-STAT Glucose test in the i-STAT CG8+ cartridge was evaluated on the i-STAT 1 analyzer using two (2) i-STAT CG8+ cartridge lots and whole blood
{10}------------------------------------------------
that was altered to a low glucose level. The LoQ for the i-STAT Glucose test in the i-STAT CG8+ cartridge was determined to be at or below the lower limit of the reportable range for the i-STAT Glucose test as shown in Table 6.
| Table 6: Summary of LoQ Results for i-STAT Glucose Test in the i-STAT CG8+<br>Cartridge | | |
|-----------------------------------------------------------------------------------------|----------------------------------------|----------------|
| Test (units) | Lower limit of the<br>reportable range | Determined LoQ |
| Glu (mg/dL) | 20 | 17 |
#### ii. Limit of Blank and Detection (LoB/LoD)
The study was based on CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition.
The LoB and LoD of the i-STAT Glucose (Glu) test in the i-STAT CG8+ cartridge was evaluated on the i-STAT 1 analyzer using two (2) i-STAT CG8+ cartridge lots. Whole blood was altered to a blank glucose level for LoB testing. Whole blood was altered to two (2) low levels of glucose for LoD testing.
The LoB and LoD were determined based on the maximal LoB or LoD value obtained for each lot tested.
The determined LoB and LoD for the i-STAT Glucose test in the i-STAT CG8+ cartridge on the i-STAT 1 analyzer is shown in the Table 7.
| Table 7: Summary of LoB and LoD Results | | |
|-----------------------------------------|-----|-----|
| Test (units) | LoB | LoD |
| Glu (mg/dL) | 0.2 | 0.9 |
#### d. Analytical Specificity
#### i. Interference
The study was based on CLSI EP07-ED3: Interference Testing in Clinical Chemistry, Third Edition.
The interference performance of the i-STAT Glucose test in the i-STAT CG8+ cartridge on the i-STAT 1 analyzer with the i-STAT 1 System was evaluated using whole blood samples based on CLSI EP07-ED3: Interference Testing in Clinical Chemistry, Third Edition. The effect of each substance was evaluated by comparing the performance of a control sample, spiked with blank solvent solution, with the test results from a test sample spiked with the potentially interfering substance at the toxic/pathological concentration based on CLSI EP37-ED1: Supplemental Tables for Interference Testing in Clinical Chemistry, First Edition, as applicable. A substance was identified as an interferent if the difference between the control and test samples was outside of the allowable error (+Ea) for the i-STAT Glucose test. For an identified interferent, a dose-response was performed to determine the degree of interference as a function of the substance concentration.
Table 8 contains the lists of potentially interfering substances tested and the interference results for the i-STAT CG8+ cartridge.
{11}------------------------------------------------
| Table 8: Potentially Interfering Substances and Test Concentrations for the i-STAT Glucose test in<br>the i-STAT CG8+ Cartridge | | | | | | |
|---------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|--------------------------------|----------------|--------------------------|------------------------------------|--|
| | Test Concentration | | | | | |
| Substance 1 | mmol/L<br>(unless<br>specified) | mg/dL<br>(unless<br>specified) | i-STAT<br>Test | Interference<br>(Yes/No) | Comments | |
| Acetaldehyde 2 | 0.045 | 0.2 | Glu | No | | |
| Acetaminophen | 1.03 | 15.6 | Glu | No | | |
| Acetoacetate<br>(Lithium<br>Acetoacetate) | 2.0 | 20 | Glu | No | | |
| Acetyl Cysteine (N-<br>Acetyl-L-Cysteine) | 0.92 | 15 | Glu | No | | |
| Ammonium<br>(Ammonium<br>Chloride) 2 | 2.0 | 10.7 | Glu | No | | |
| Ascorbic Acid (L-<br>Ascorbic Acid) | 0.298 | 5.25 | Glu | No | | |
| β-Hydroxybutyric<br>Acid 2 | 6.0 | 62.46 | Glu | No | | |
| Bilirubin | 0.684 | 40 | Glu | No | | |
| | 2.5 | 21.7 | Glu | No | | |
| Bromide 2<br>(Lithium Bromide) | 37.5 | 325.7 | Glu | Yes | Use Another<br>Method | |
| Cholesterol | 10.3 | 400 | Glu | No | | |
| Creatinine | 1.326 | 15 | Glu | No | | |
| Dopamine<br>(Dopamine<br>Hydrochloride) | 4.06<br>μmol/L | 0.0621 | Glu | No | | |
| Ethanol | 130 | 600 | Glu | No | | |
| Fluoride (Lithium<br>Fluoride) | 0.0632 | 0.12 | Glu | No | | |
| Formaldehyde 2 | 0.133 | 0.399 | Glu | No | | |
| Fructose | 1 | 18 | Glu | No | | |
| Galactose | 3.33 | 60 | Glu | No | | |
| Gentamicin<br>(Gentamicin Sulfate) | 0.0628 | 3 | Glu | No | | |
| Gentisic Acid | 0.0973 | 1.5 | Glu | No | | |
| Glucosamine<br>(Glucosamine<br>Hydrochloride) 2 | 0.030 | 0.647 | Glu | No | | |
| Glutathione, reduced 3 | 3 | 3 mEq/L | Glu | No | | |
| Glycolic Acid 2 | 10.0 | 76.05 | Glu | No | | |
| Guaifenesin | 0.0227 | 0.45 | Glu | No | | |
| Hemoglobin | 10 g/L | 1000 | Glu | No | | |
| Heparin (Sodium<br>Heparin) | 3.30 U/mL | 330 U/dL | Glu | No | | |
| Hydroxyurea | 0.405 | 3.08 | Glu | Yes | Increased results ≥<br>0.08 mmol/L | |
| Ibuprofen | 1.06 | 21.9 | Glu | No | | |
| Table 8: Potentially Interfering Substances and Test Concentrations for the i-STAT Glucose test in | | | | | | |
| the i-STAT CG8+ Cartridge | | | | | | |
| Substance 1 | Test Concentration<br>mmol/L<br>(unless<br>specified) | mg/dL<br>(unless<br>specified) | i-STAT<br>Test | Interference<br>(Yes/No) | Comments | |
| Intralipid 20% | N/A | 2891 | Glu | No | Increased results > | |
| Isoniazid | 0.438 | 6 | Glu | Yes | Increased results ≥<br>0.29 mmol/L | |
| Lactate (Lithium<br>Lactate) | 10 | 90 | Glu | No | | |
| Maltose | 10.5 | 360 | Glu | No | | |
| Mannose 2 | 1 | 18.02 | Glu | No | | |
| Nithiodote (Sodium<br>Thiosulfate) 2 | 16.7 | 264.04 | Glu | No | | |
| pH | 8.0 pH<br>units | N/A | Glu | No | | |
| Pyruvate (Lithium<br>Pyruvate) | 0.570 | 5 | Glu | No | | |
| Salicylate (Lithium<br>Salicylate) | 0.207 | 2.86 | Glu | No | | |
| Thiocyanate (Lithium<br>Thiocyanate) | 0.898 | 5.22 | Glu | No | | |
| Triglyceride | 16.94 | 1500 | Glu | No | | |
| Uric Acid | 1.4 | 23.5 | Glu | No | | |
| Xylose 2 | 3 | 45.04 | Glu | No | | |
<sup>1</sup> The compound tested to evaluate the interfering substance is presented in parenthesis.
² The test concentration for this substance is not included in CLSI guideline EP37 1st edition.
{12}------------------------------------------------
#### ii. Other sensitivity studies
#### 1) Oxygen Sensitivity Study
The effect of oxygen on the i-STAT Glucose test in the i-STAT CG8+ cartridge on the i-STAT 1 System was evaluated with low and high oxygen levels using whole blood samples altered to four (4) glucose levels the reportable range of the i-STAT Glucose test. The equivalency between the high and low oxygen conditions was determined if the 95% confidence interval (CI) of the difference in means (or medians) was within the allowable error (±Ea). The study demonstrated that the i-STAT Glucose test in the i-STAT CG8+ cartridge on the i-STAT 1 System perform is insensitive to oxygen levels between 20 and 503 mmHg.
#### 2) Hematocrit Sensitivity
A hematocrit sensitivity study was conducted to evaluate the performance of the i-STAT Glucose test in the i-STAT CG8+ cartridge with the i-STAT 1 Sustem to assess the effect of hematocrit. Three (3) hematocrit levels (low, mid and high) were evaluated at four (4) glucose levels across the reportable range of the i-STAT Glucose test in the i-STAT CG8+ cartridge. The hematocrit sensitivity at each glucose level was assessed by comparing the results at the low and high hematocrit level to the mid hematocrit level. Equivalency was assessed by determining whether the difference between the low and high hematocrit level and the mid hematocrit level was within the allowable error (±Ea). The study demonstrated that i-STAT Glucose test in the i-STAT CG8+ cartridge with the i-STAT 1 System is insensitive to hematocrit levels between 15% to 75% packed cell volume (PCV).
{13}------------------------------------------------
## 3) Altitude
The performance of the i-STAT Glucose test in the i-STAT CG8+ cartridge on the i-STAT 1 analyzer at an altitude of approximately 10,000 feet above sea level was evaluated using whole blood samples at relevant glucose levels across the reportable range for the i-STAT Glucose test. The glucose test results obtained from the i-STAT CG8+ cartridges (candidate device) were compared to the glucose test results obtained from the i-STAT CHEM8+ (blue) cartridges on the i-STAT 1 analyzer (comparator device). Passing-Bablok regression analyses between the first replicate of the candidate device (y-axis) and mean of the comparator device (x-axis) were performed based on the CLSI EP09c: Measurement Procedure Comparison and Bias Estimation using Patient Samples, 3rd ed. The results of the correlation coefficient and slope met acceptance criteria and demonstrated equivalent performance between the candidate and comparator conditions at approximately 10,000 feet above sea level. The results are summarized in Table 9.
| Table 9: Summary of Altitude Study Results for the i-STAT CG8+ Cartridge | | | | |
|--------------------------------------------------------------------------|-----------------------------|----------------|-------|----------------|
| Test | Correlation Coefficient (r) | | Slope | |
| | r | 95% CI | Slope | 95% CI |
| Glu | 1.00 | 1.000 to 1.000 | 0.96 | 0.957 to 0.971 |
### B. Comparison Studies
#### a. Method Comparison with predicate device
Method comparison for arterial, venous, and capillary whole blood specimens on the i-STAT CG8+ cartridge with the i-STAT 1 System was demonstrated in studies based on CLSI EP09c-ED3: Measurement Procedure Comparison and Bias Estimation Using Patient Samples - Third Edition.
Lithium heparin venous and arterial whole blood specimens collected across multiple point of care sites were evaluated using i-STAT CG8+ cartridges on the i-STAT 1 analyzer against whole blood specimens tested on a comparative method. The first replicate glucose result from the i-STAT 1 analyzer was compared to the mean glucose result from the comparative method.
Two (2) capillary whole blood specimens collected from skin puncture with balanced heparin capillary tubes from each study subject across multiple point of care sites were evaluated and analyzed in singlicate on the i-STAT 1 analyzer against the comparative method. A Passing-Bablok linear regression analysis for glucose was performed using the singlicate result from the i-STAT 1 analyzer versus the singlicate result of the comparative method.
The venous, arterial, and capillary whole blood data were pooled, and a Passing-Bablok linear regression analysis was performed using the results from the i-STAT CG8+ cartridges on the i-STAT 1 analyzer versus the comparative method results.
Method comparison results for arterial, venous, and capillary whole blood specimens are shown in Table 10. In the table, N is the number of specimens in the data set, and r is the correlation coefficient.
| Table 10: Method Comparison Results for i-STAT CG8+ Cartridge with i-STAT 1 System | | | | | | |
|------------------------------------------------------------------------------------|--------------------|-------------------------------|-----|-------|-----------|------|
| Test | Comparative Method | | N | Slope | Intercept | r |
| | Arterial/Venous | Capillary | | | | |
| Glu | i-STAT CHEM8+ | epoc Blood Analysis<br>System | 547 | 0.98 | 1.62 | 1.00 |
{14}------------------------------------------------
### b. Matrix Equivalence
A matrix equivalence study was conducted to evaluate the performance of the i-STAT Glucose test in the i-STAT CG8+ cartridge on the i-STAT 1 System using nonanticoagulated venous and arterial whole blood specimens. The study design and analysis method were based on recommendations from the Clinical and Laboratory Standards Institute (CLSI) guideline EP35: Assessment of Equivalence or Suitability of Specimen Types for Medical Laboratory Measurement Procedures, 1st ed. The matrix equivalence of the i-STAT Glucose test in the i-STAT CG8+ cartridge was assessed by comparing arterial or venous whole blood specimens collected without anticoagulant (candidate specimen type) to samples collected with balanced heparin or lithium heparin anticoagulant (primary specimen type). Each specimen was tested in duplicate using two (2) i-STAT CG8+ cartridges with two (2) i-STAT 1 analyzers. A Passing-Bablok linear regression analysis was performed using the first replicate result from the candidate (y-axis) versus the mean result from the primary specimen (x-axis). The regression analysis results are summarized in Table 11. In the table, N is the number of specimens in the data set, and r is the correlation coefficient.
| Table 11: Matrix Equivalence Results | | | | | | |
|--------------------------------------|-----|-----------------------------|---------------------------|------|-------|-----------|
| Test (units) | N | Candidate Specimen<br>Range | Primary Specimen<br>Range | r | Slope | Intercept |
| Glu (mg/dL) | 297 | 39-688 | 39-671 | 1.00 | 1.00 | 0.00 |
#### CONCLUSION VIII.
The results of these studies demonstrate that performance of the i-STAT Glucose test in the i-STAT CG8+ cartridge with the i-STAT 1 System are substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.