ALERE CHOLESTECH LDX ANALYZER, ALERE CHOLESTECH LDX LIPID PROFILE - GLU CASSETTE

K120615 · Alere San Diego Inc. Dba Biosite,Innovacon,Hemosen · CGA · May 15, 2013 · Clinical Chemistry

Device Facts

Record IDK120615
Device NameALERE CHOLESTECH LDX ANALYZER, ALERE CHOLESTECH LDX LIPID PROFILE - GLU CASSETTE
ApplicantAlere San Diego Inc. Dba Biosite,Innovacon,Hemosen
Product CodeCGA · Clinical Chemistry
Decision DateMay 15, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The Alere Cholestech LDX® System is a small, portable analyzer and test cassette system. The System is for in vitro diagnostic use only. The Lipid Profile•GLU Cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides and glucose in whole blood. A TC/HDL (total cholesterol/HDL cholesterol ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are calculated by the Alere Cholestech LDX® Analyzer. - Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. - HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. - Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders. - Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Device Story

Portable analyzer uses reflectance photometry to measure color changes on reagent pads of test cassettes; inputs are whole blood samples (fingerstick or venous). Device separates plasma from blood cells; enzymatic reactions produce purple-colored quinoneimine dye proportional to analyte concentration. Analyzer reads calibration data from magnetic stripe on cassette; converts reflectance readings to concentration values. Modification adds humidity sensor to ROM pack; software applies humidity correction factors from lookup table to assay results. Used in clinical settings; operated by healthcare professionals. Output displayed on LCD; assists clinicians in diagnosing lipid/metabolic disorders and monitoring treatment efficacy.

Clinical Evidence

Bench testing only. Sponsor performed humidity interference validation studies comparing modified device (ROM v3.41) against predicate (v3.30) across 20-80% relative humidity range. Precision studies confirmed performance meets predicate acceptance criteria.

Technological Characteristics

Reflectance photometry; enzymatic assay methodology. System includes portable analyzer and test cassettes with magnetic stripe for calibration. Modification adds humidity sensor to ROM pack and software-based humidity correction lookup table. Analyte measurement via enzymatic colorimetric reaction (cholesterol oxidase, glycerol kinase, glucose oxidase).

Indications for Use

Indicated for quantitative determination of total cholesterol, HDL, triglycerides, and glucose in whole blood for patients requiring diagnosis/treatment of lipid metabolism disorders, atherosclerosis, liver/renal disease, diabetes mellitus, nephrosis, endocrine disorders, neonatal/idiopathic hypoglycemia, or pancreatic islet cell carcinoma.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 5. 510(K) SUMMARY 510(k) number: ___ k120615 Date Prepared: 15 April 2013 | Submitter: | Alere San Diego, Inc. | Contact: | Edward Brehm | |------------|------------------------------------------------|----------|----------------------------| | Address: | 9975 Summers Ridge Road<br>San Diego, CA 92121 | | Regulatory Affairs Manager | | | | Email: | ed.brehm@alere.com | | Phone: | 858.805.3071 | Phone: | 858.805.3071 | | Fax: | 858.695.7100 | Fax: | 858.695.7100 | | Common Name (Device Type):<br>Alere Cholestech LDX® Lipid Profile · GLU<br>Cassette<br>Alere Cholestech LDX® Analyzer | |-----------------------------------------------------------------------------------------------------------------------| | | | | | Glucose:<br>Class:<br>Regulation number:<br>Product Code:<br>Panel: | II<br>21 CFR 862.1345<br>CGA<br>Clinical Chemistry | |-----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------| | Lipids:<br>Class:<br>Regulation number:<br>Product Code(s):<br>Panel: | I (meets limitation for exemption<br>per 21 CFR 862.9(c)(4) and (9)<br>21 CFR 862.1175, 862.1475, 862.1705<br>CHH, LBS, JGY<br>Clinical Chemistry | | Analyzer:<br>Class:<br>Regulation number:<br>Product Code:<br>Panel: | I (exempt)<br>21 CFR 862.2160<br>JJE<br>Clinical Chemistry | | Original 510(k) Submissions: | Clearance Date: | |---------------------------------------|-----------------| | K901900 – LDX Lipid Monitoring System | July 24, 1990 | | K932727 – Lipid Profile•GLU Cassette | Nov 9, 1993 | # MAY 1 5 2013 {1}------------------------------------------------ #### 5.1. Intended Use / Indications for Use: The Alere Cholestech LDX® System is a small, portable analyzer and test cassette system is for in vitro diagnostic use only. The Lipid Profile GLU Cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides and glucose in whole blood. A TC/HDL (total cholesterol/HDL cholesterol ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are calculated by the Alere Cholestech LDX® Analyzer. - . Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. - . HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. - . Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders. - Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism . disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. ### 5.2. Summary of Changes: The revision of the software is being changed from revision v3.41, which incorporates a humidity sensor as part of the ROM pack. This sensor measures the ambient humidity and applies a small correction factor from a lookup table to the result from the assay algorithm. ### 5.3. Substantial Equivalence: Version v3.41 of the Alere Cholestech LDX® Analyzer ROM pack software is substantially equivalent to Alere Cholestech LDX® Analyzer ROM pack software version v3.30. The change is invisible to the user. Analytical results when the analyzer is operated between 40% and 60% relative humidity are unchanged and no correction factor is required. In more extreme cases, when the ambient humidity is between 20% RH and 40% RH, or between 60% RH and 80%RH, a small correction factor is applied. #### 5.4. List of Similarities: The Intended use is unchanged The Indications for use is unchanged The analytical performance has been returned to its original intent The manufacturing process is unchanged ## 5.5. List of Differences: Software version v3.41 contains a humidity sensor which measures the ambient humidity and applies a small correction factor to the analytical results based on a lookup table. #### 5.6. Conclusion: Performance testing demonstrates that the software upgrade from revision v3.30 to v3.41 is substantially equivalent. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is composed of three stylized, curved lines that appear to be interconnected. The text is in all caps and reads "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA". ## Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 15, 2013 Alere San Diego, Inc. C/O Edward Brehm, Ph.D. 9975 Summers Ridge Road SAN DIEGO CA 92121 Re: K120615 Trade/Device Name: Alere Cholestech LDX® Lipid Profile • GLU Cassette Alere Cholestech LDX® Analyzer Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: CGA, CHH, LBS, JGY, JJE Dated: May 10, 2013 Received: May 13, 2013 ## Dear Dr. Brehm: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2—Dr. Brehm If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. ## Carol C. Benson -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use ## 510(k) Number (if known): k120615 ## Device Name: Alere Cholestech LDX® Analyzer Alere Cholestech LDX® Lipid Profile●GLU Cassette ## Intended Use / Indications for Use: The Alere Cholestech LDX® System is a small, portable analyzer and test cassette system. The System is for in vitro diagnostic use only. The Lipid Profile GLU Cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides and glucose in whole blood. A TC/HDL (total cholesterol/HDL cholesterol ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are calculated by the Alere Cholestech LDX® Analyzer. - Cholesterol measurements are used in the diagnosis and treatment of disorders . involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. - HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid . disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. - Triglyceride measurements are used in the diagnosis and treatment of patients with . diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders. - Clucose measurements are used in the diagnosis and treatment of carbohydrate . metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. Prescription Use .. X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Ruth A. Chester S. Division Sign-Off Office of In Vitro Diagnostics and Radiological Health 510(k) k120615
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...