ADVIA CHEMISTRY GLUCOSE OXIDASE ASSAY AND ADVIA CHEMISTRY GLUCOSE HEXOKINASE
Applicant
Bayer Healthcare, Diagnostics Division
Product Code
CGA · Clinical Chemistry
Decision Date
Sep 29, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Bayer ADVIA 1650 Glucose Oxidase and Glucose Hexokinase II assays are in vitro diagnostic devices for use in the quantitative determination of glucose levels in human cerebrospinal fluid (CSF), serum, plasma, and urine on the ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia and insulin overdose.
Device Story
The Bayer ADVIA Chemistry Glucose assays (Hexokinase and Oxidase) are in vitro diagnostic reagents used on the ADVIA 1650 Chemistry system. The device processes human CSF, serum, plasma, or urine samples. The Hexokinase assay uses a multi-component liquid reagent system (ATP, NAD, hexokinase, G6PDH) to catalyze glucose phosphorylation, resulting in NADH production proportional to glucose concentration. The Glucose Oxidase assay uses a single reagent (glucose oxidase, peroxidase, phenol, indicator) to produce a red-violet quinoneimine dye. Both assays measure absorbance changes via spectrophotometry. Used in clinical laboratory settings by trained personnel, the output is a quantitative glucose concentration value. This data assists clinicians in diagnosing and managing carbohydrate metabolism disorders. The device provides a diagnostic tool for monitoring glucose levels across multiple matrices.
Clinical Evidence
Bench testing only. Precision studies followed NCCLS EP5-A guidelines using CSF quality controls over 12 days (n=42-43), showing total imprecision CVs of 2.7-3.4%. Linearity was confirmed with ≤5% deviation across reportable ranges (0-700 mg/dL for Hexokinase; 0-750 mg/dL for Oxidase). Method comparison against the predicate using 55-56 patient CSF samples yielded R-values of 0.987 (Hexokinase) and 0.994 (Oxidase), with slopes of 1.026 and 0.973 respectively.
Technological Characteristics
In vitro diagnostic reagent assays for automated chemistry analyzer. Glucose Oxidase and Hexokinase II enzymatic methods. Analytical ranges: 0-750 mg/dL (Oxidase) and 0-700 mg/dL (Hexokinase).
Indications for Use
Indicated for quantitative glucose determination in human CSF, serum, plasma, and urine to diagnose and treat carbohydrate metabolism disorders, including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and insulin overdose.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K062239 — GLUCOSE HK NEW FORMULATION TEST SYSTEM · Roche Diagnostics Corp. · Sep 11, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
## SEP 2 9 2004
### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Glucose methods for ADVIA® 1650TM
This summary of 510(k) safety and effectiveness information is being submitted in accordance This summary of 910(th) sales) and 1 and 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K642015
#### 1. Intended Use
The Bayer ADVIA 1650 Glucose Oxidase and Glucose Hexokinase II assays are in vitro The Dayer AD TIA 1000 Of to occasion in the sure glucose levels in human cerebrospinal fluid (CSF), serum, plasma, and urine on the ADVIA® 1650 Chemistry System. Such measurements (CDF), seruin, phalma, and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypogleemia, idiopathic hypoglycemia and insulin overdose.
#### 2. Predicate Device
| Product Name | Reagent<br>Part # | Calibrator Part # |
|--------------------------|-------------------|-------------------|
| Roche Glucose Hexokinase | 20763020 | 10759350 |
#### Device / Method 3.
| Product Name | Reagent<br>Part #/ BAN # | Calibrator Part<br>#/ BAN # |
|-----------------------------------------|--------------------------|-----------------------------|
| Bayer ADVIA® 1650<br>Glucose Hexokinase | B01-4597-01/<br>04005536 | T03-1291-62/<br>09784096 |
| Bayer ADVIA® 1650<br>Glucose Oxidase | B01-4130-01/<br>04903429 | T03-1291-62/<br>09784096 |
Imprecision
| ADVIA 1650 Glucose<br>Oxidase | | ADVIA 1650 Glucose<br>Hexokinase II | | Roche Glucose Hexokinase | |
|-------------------------------|---------------------|-------------------------------------|---------------------|--------------------------|---------------------|
| Level<br>(mg/dL) | Within-run<br>CV(%) | Level<br>(mg/dL) | Within-run<br>CV(%) | Level<br>(mg/dL) | Within-run<br>CV(%) |
| 58.0 | 1 | 60.2 | 1.2 | 31 | 1.6 |
| 34.2 | 1.1 | 36.9 | 1.7 | 59 | 1.8 |
{1}------------------------------------------------
| Specimen<br>Type | Comparison<br>System (x) | N | Regression<br>Equation | Syx | r | Sample Range<br>mg/dL |
|-------------------------|--------------------------|----|------------------------|-----|-------|-----------------------|
| Glucose<br>(Oxidase) | Roche<br>(On Integra) | 56 | $Y = 0.97x + 4.47$ | 7.6 | 0.994 | 37-530 |
| Glucose<br>(Hexokinase) | Roche<br>(On Integra) | 55 | $Y = 1.03x - 1.25$ | 6.4 | 0.987 | 39-214 |
## Correlation (Y=ADVIA 1650, X=comparison system)
Interfering Substances
Interfering substances are not expected in cerebrospinal fluid samples.
Analytical Range
Glu-HKII: 0- 700 mg/dL Glu-Ox: 0- 750 mg/dL
Carter
7/20/04
Date
Andres Holle Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the symbol. The logo is black and white.
Public Health Service
SEP 2 9 2004
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Andres Holle Manager, Regulatory Affairs Bayer Healthcare LLC. Diagnostics Division 511 Benedict Avenue Tarrytown, NY 10591
k042015 Re:
R042019
Trade/Device Name: Glucose Hexokinase II assay for the ADVIA® 1650 Chemistry System Glucose Oxidase assay for the ADVIA® 1650 Chemistry System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: CGA, CFR Dated: July 15, 2004 Received: August 5, 2004
Dear Mr. Holle:
We have reviewed your Section 510(k) premarket notification of intent to market the device w & nowe and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The 1 ou may, diereleve, marrer of the Act include requirements for annual registration, listing of general controls proficiouring practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device it may or subject to back of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA can oo louits in there announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be devised alat termination that your device complies with other requirements of the Act that I Drederal statutes and regulations administered by other Federal agencies. You must or any I odoral statutes and regirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
#### Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) I ms lotter will and will your to be finding of substantial equivalence of your device to a legally premaince noticate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, r fou don't on the promotion and advertising of your device, please contact the Office of of quebelle on the preside Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Jean M. Cooper, MS, DVM.
Jean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): KOL 2015
Device Name: Glucose Hexokinase II assay for the ADVIA® 1650 Chemistry System
Indications For Use:
The Bayer ADVIA® 1650 Glucose Hexokinase II assay is an in vitro diagnosic device The Bayer ADVAN® TUSO Glaoooo Hoxokhabo in human cerebrospinal fluid (CSF), for use in the first for use in the quantitutive actornination on the ADVIA® 1650 Chemistry system. serum, plasma (illinian nopalin), the diagnosis and treatment of carbohydrate Such measurements are about the claghester the clayments, idiopathic hypoglycemia and insulin overdose.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Albert Cert
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and
510(k) k042015
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Glucose Oxidase assay for the ADVIA® 1650 Chemistry System
Indications For Use:
The Bayer ADVIA® 1650 Glucose Oxidase assay is an in vitro diagnostic device for use The Bayer ADVIA® 7000 Glucose in human cerebrospinal fluid (CSF), serum,
in the quantitative determination of glucose in human cerebrospinal fluid (CSF), serum, in the quantitative determination of groups in haman seems . Such measurements are
plasma and urine on the ADVIA® 1650 Chemistry system. Such measurements are plasma and urne on the ADVIA® 1000 Onehabelism disorders including
used in the diagnosis and treatment of carbohydrate and insulin used in the diagnosis and treathent of Sarbonyards in the months and insulin overdose.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Alberto Lutz
Division Sign-Off
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K042015
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.