CardioChek Plus Test System

K151530 · Polymer Technology Systems, Inc. · CGA · Jul 7, 2015 · Clinical Chemistry

Device Facts

Record IDK151530
Device NameCardioChek Plus Test System
ApplicantPolymer Technology Systems, Inc.
Product CodeCGA · Clinical Chemistry
Decision DateJul 7, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The CardioChek Plus Test System is a small portable analyzer and test strip system intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only. The tests strips are for the quantitative determination of glucose, total cholesterol, HDL (high density lipoprotein) cholesterol and triglycerides in venous whole blood and capillary whole blood from the fingertip. A Chol/HDL ratio and estimated values for LDL (low density lipoprotein) cholesterol and non-HDL cholesterol are calculated by the CardioChek Plus analyzer. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Device Story

Portable analyzer system for professional healthcare settings; uses analyte-specific test strips (glucose, lipids) and ChekMate strips for system verification. Input: venous or capillary whole blood applied to test strips. Operation: analyzer measures reflectance/optical signals from strips; calculates Chol/HDL ratio and estimates LDL/non-HDL cholesterol. Output: quantitative blood analyte concentrations displayed to clinician. Use: point-of-care testing by healthcare professionals. Benefit: rapid diagnostic assessment of metabolic and lipid disorders; ChekMate strips verify optics, calibration, and result handling algorithms without replacing liquid quality control.

Clinical Evidence

Bench testing only. Intermediate precision evaluation performed over 20 days per CLSI EP5-A2; within-run, between-run, between-day, and total % CV for all results were <3.0%. Stability studies support 18-month shelf life and 500-use re-use stability for ChekMate strips.

Technological Characteristics

Portable analyzer; optical sensing system. Includes dry test strips for glucose and lipids. ChekMate strips used for system verification of optics, calibration, and algorithms. Connectivity/software details not specified. Sterilization not applicable to analyzer; strips are single-use.

Indications for Use

Indicated for professional healthcare settings for quantitative determination of glucose, total cholesterol, HDL cholesterol, and triglycerides in venous or capillary whole blood. Used for diagnosis/treatment of lipid, lipoprotein, and carbohydrate metabolism disorders, including diabetes mellitus, atherosclerosis, liver/renal diseases, and hypoglycemia.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is placed to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 30, 2015 POLYMER TECHNOLOGY SYSTEMS, INC. MEGAN BURNS REGULATORY AFFAIRS ASSOCIATE 7736 ZIONSVILLE ROAD INDIANAPOLIS IN 46268 Re: K151530 Trade/Device Name: CardioChek Plus Test System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: CGA, CHH, JGY, LBR Dated: June 5, 2015 Received: June 8, 2015 Dear Ms. Megan Burns: This letter corrects our substantially equivalent letter of July 7, 2015. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the {1}------------------------------------------------ electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Courtney H. Lias -S Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K151530 Device Name CardioChek Plus Test System #### Indications for Use (Describe) The CardioChek Plus Test System is a small portable analyzer and test strip system intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only. The tests strips are for the quantitative determination of glucose, total cholesterol, HDL (high density lipoprotein) cholesterol and triglycerides in venous whole blood from the fingertip. A Chol/HDL ratio and estimated values for LDL (low density lipoprotein) cholesterol and non-HDL cholesterol are calculated by the CardioChek Plus analyzer. · Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. · HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. · Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellius, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders. · Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|---------------------------------------------| | × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### Section 5: 510(k) Summary #### 1. Applicant Information: Polymer Technology Systems, Inc. 7736 Zionsville Road Indianapolis, IN 46268 Contact Person: Megan A. Burns, RAC Phone: 317-870-5610 Fax: 317-870-5608 #### 2. Date of Preparation: July 2, 2015 #### 3. Trade Name: CardioChek Plus Test System #### 4. Device Description: The CardioChek Plus Test System includes the CardioChek Plus professional analyzer and analyte-specific test strips (PTS Panels eGLU test strips, PTS Panels Glucose test strips and PTS Panels Lipid Panel test strips). The CardioChek Plus Test System is modified in this submission by the addition of the CardioChek ChekMate strips. CardioChek ChekMate strips are dry strips that mimic the use of the PTS PANELS test strips to check the CardioChek analyzer system optics, calibration and result handling algorithms. ChekMate strips should not be used as a substitute for liquid quality control materials. There is no change to any of the test strips in any of the above named systems. ## 5. Classification Names: | Glucose Test System | | |--------------------------|-------------------------------------------------------------------| | Panel: | Clinical Chemistry 75 | | Product Codes: | CGA | | Device Classification: | Class 2 | | Regulation: | 21 CFR 862.1345 | | Cholesterol Test System | | | Panel: | Clinical Chemistry 75 | | Product Codes: | CHH | | Device Classification: | Class I [meets limitation of exemption as per 21 CFR 862.9(c)(4)] | | Regulation: | 21 CFR 862.1175 | | Triglyceride Test System | | | Panel: | Clinical Chemistry 75 | | Product Codes: | JGY | | Device Classification: | Class I [meets limitation of exemption as per 21 CFR 862.9(c)(4)] | | Regulation: | 21 CFR 862.1705 | {4}------------------------------------------------ | Lipoprotein Test System | | |-------------------------|-------------------------------------------------------------------| | Panel: | Clinical Chemistry 75 | | Product Codes: | LBR, MRR | | Device Classification: | Class I [meets limitation of exemption as per 21 CFR 862.9(c)(4)] | | Regulation: | 21 CFR 862.1475 | ## 6. Intended Use The CardioChek Plus Test System is a small portable analyzer and test strip system intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only. The tests strips are for the quantitative determination of glucose, total cholesterol, HDL (high density lipoprotein) cholesterol and triglycerides in venous whole blood and capillary whole blood from the fingertip. A Chol/HDL ratio and estimated values for LDL (low density lipoprotein) cholesterol and non-HDL cholesterol are calculated by the CardioChek Plus analyzer. - Cholesterol measurements are used in the diagnosis and treatment of ● disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. - . HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. - . Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders. - Glucose measurements are used in the diagnosis and treatment of . carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. ## 7. Reason for 510(k): Device Modification ## 8. Predicate Device Information Device Trade Name: CardioChek Plus Test System Device Company: Polymer Technology Systems, Inc. (PTS) k140068 510(k) number: Comparison to Predicate: The modified system is identical to the unmodified system with the exception of the addition of the CardioChek ChekMate strips. {5}------------------------------------------------ ## 9. Performance Characteristics The strips for each system are identical to the previously cleared strips. The CardioChek Plus Test System is unmodified compared to the predicate device with the exception of the addition of the CardioChek ChekMate strips. ## CardioChek ChekMate Strips Precision: An Intermediate Precision evaluation was performed over a 20 day period as described in CLSI EP5-A2 Evaluation of Precision Performance of Quantitative Measurement Methods. Three lots of ChekMate strips were evaluated on the CardioChek Plus using two (2) analyzers for each lot. Within run, between run, between day and total % CV for all results were less than 3.0%. ## CardioChek ChekMate Strips Value Assignment: Previously reported in k142302 ## CardioChek ChekMate Strips Stability: Previously reported in k142302 Real-time Stability: The study supports an 18 month stability claim. ## Re-use Stability: The study supports the claimed re-use stability of 500 uses for the ChekMate strips. ## 10. Conclusion: The performance characteristics above demonstrate that the modification of the above named test system by the addition of the CardioChek ChekMate strips is substantially equivalent to the unmodified system.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...