Stat Profile Prime Plus Analyzer System

K180340 · Nova Biomedical Corporation · CGA · Sep 26, 2018 · Clinical Chemistry

Device Facts

Record IDK180340
Device NameStat Profile Prime Plus Analyzer System
ApplicantNova Biomedical Corporation
Product CodeCGA · Clinical Chemistry
Decision DateSep 26, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The Stat Profile® Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings for quantitative determination of glucose, creatinine, and blood urea nitrogen in heparinized arterial and venous whole blood. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell tumor. Creatinine measurements are used in the diagnosis and treatment of certain renal conditions and for monitoring adequacy of dialysis. Blood urea nitrogen (BUN) measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.

Device Story

Stat Profile Prime Plus Analyzer is a benchtop blood chemistry analyzer for hospital laboratory use. It accepts lithium heparinized arterial and venous whole blood samples (135 µL) from syringes, open tubes, or small cups. The device utilizes modular sensor cartridges (Primary and Auxiliary) to measure Glucose, Creatinine, and BUN. Glucose and Creatinine are measured via enzymatic amperometric sensors (platinum anode at 0.70V); BUN is measured via an ammonium ion-selective electrode using the Nernst equation. The system includes an onboard Quality Management System (QMS) for continuous monitoring. Results are displayed on a 10.1-inch color touch screen and can be printed via an integrated thermal printer. The device aids clinicians in diagnosing and treating renal conditions, metabolic diseases, and carbohydrate metabolism disturbances. It is designed for low-maintenance operation in clinical settings.

Clinical Evidence

Bench testing only. Performance verification included method comparison studies against the predicate analyzer, within-run and run-to-run precision studies (20 replicates per run), linearity testing across the analytical measurement range, and interference screening per CLSI EP7-A2. Results demonstrated good correlation, linearity, and precision meeting all pre-defined acceptance criteria.

Technological Characteristics

Materials: TPE tubing (Pharmed BPT). Sensing: Amperometric (glucose, creatinine) and potentiometric (BUN) sensors. Energy: Electrical. Form factor: Benchtop analyzer with 10.1" WXGA touch screen. Connectivity: Integrated 1D/2D barcode scanner. Software: Embedded firmware for sensor configuration and data processing. Traceability: NIST SRM 967a (creatinine), 912a (urea), 917a (glucose).

Indications for Use

Indicated for healthcare professionals in clinical laboratory settings for quantitative determination of glucose, creatinine, and BUN in heparinized arterial and venous whole blood. Not for point-of-care use.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 26, 2018 Nova Biomedical Corporation Cesidio Tempesta Senior Regulatory Affairs Specialist 200 Prospect St. Waltham, MA 02454 Re: K180340 Trade/Device Name: Stat Profile Prime Plus Analyzer System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: CGA, CDS, CGL Dated: August 24, 2018 Received: August 27, 2018 Dear Cesidio Tempesta: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Kellie B. Kelm -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number K180340 #### Device Name #### Stat Profile® Prime Plus Analyzer System #### Indications for Use (Describe) The Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings for quantitative determination of Glucose, Creatinine, and Blood Urea Nitrogen, in heparinized arterial and venous whole blood. | Glucose (Glu) | Measurements are used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell tumor. | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Creatinine (Creat) | Measurements are used in the diagnosis and treatment of certain renal conditions and for monitoring adequacy of dialysis. | | Blood Urea Nitrogen (BUN) | Measurements are used in the diagnosis and treatment of certain renal and metabolic diseases. | | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <div style="display:flex; align-items:center;"> <div style="margin-right:5px;">X</div> <div>Prescription Use (Part 21 CFR 801 Subpart D)</div> </div> | |-------------------------------------------------------------------------------------------------------------------------------------------------------| | <div style="display:flex; align-items:center;"> <div style="margin-right:5px;"></div> <div>Over-The-Counter Use (21 CFR 801 Subpart C)</div> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. #### This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary K180340 | 510(K) Owner: | Nova Biomedical Corporation | |----------------------|-------------------------------------------------| | Registration Number: | 1219029 | | Address: | 200 Prospect St.<br>Waltham, MA 02454 | | Phone: | 781-894-0800 | | Fax Number: | 784-891-4806 | | Contact Person: | Cesidio Tempesta, Regulatory Affairs Specialist | | Date Prepared: | September 25, 2018 | Proprietary Name: Stat Profile Prime Plus Analyzer System Common or Usual Name: Blood Gas Analyzer #### Classification Name: Multiple | Classification Name | Regulation # | Class | Product Code | |---------------------------|--------------|-------|--------------| | Glucose test system | 862.1345 | II | CGA | | Creatinine test system | 862.1225 | II | CGL | | Urea Nitrogen Test System | 862.1770 | II | CDS | #### Product Code: CGA #### Predicate Device: K110648 - Stat Profile pHOx Ultra Analyzer System #### Device Description: The Stat Profile Prime Plus Analyzer System is designed to be a low cost, low maintenance analyzer for the hospital laboratory setting. It consists of the analyzer, sensor cartridges, and thermal paper for an onboard printer. Optionally, it provides for reading of barcode labels (such as operator badges and data sheets). The system architecture and user interface for this proposed device is based on the previously cleared Stat Profile Prime CCS Analyzer System (K131703). The primary predicate for this proposed device is the Stat Profile pHOx Ultra Analyzer System (K110648). The Stat Profile Prime Plus Analyzer has slots to accommodate two sensor cartridges (Primary and Auxiliary). The analyzer will determine the configuration of the system by detecting which sensor cards are installed. The reporting of CO-Oximeter parameters (or not reporting them) will also be determined by the selection of the Sensor Cards: #### Primary Sensor Card Port: There are two options for the primary sensor card: - . Primary Sensor Card 1 shall enable and report the following listed analytes: - o PO2, PCO2, pH, Hct, tHb, Na, Cl, K, iCa, iMg, Glu, SO2, O2Hb, COHb, MetHb, HHb - . Primary Sensor Card 2 shall enable and report the following listed analytes: - o PO2, PCO2, pH, Hct, tHb, Na, Cl, K, iCa, iMg, Glu, SO2 {4}------------------------------------------------ #### Auxiliary Sensor Card Port: The reporting of Creatinine and BUN parameters (Or not reporting them) shall be determined by the selection of the Auxiliary Sensor Card - Auxiliary Sensor Card 1 enables Creatinine and BUN parameters ● - . Auxiliary Sensor Card 2 is a "dummy" sensor card, and will not report any parameters. Similar to the primary predicate device, the Stat Profile Prime Plus Analyzer is a blood gas/cooximetry/electrolyte/chemistry and hematology analyzer with an enhanced test menu and multiple quality control options. Both traditional internal and external quality control will be used, as well as an on-board Quality Management System (QMS), an electronic monitoring approach that insures the analyzer is working properly at all times. The Stat Profile Prime Plus Analyzer accepts samples from syringes, open tubes, and small cups. The minimum sample size for analysis is 135 µL. Sample collection, preparation and application to the analyzer are the same as for the previously cleared predicate. The end user can select which analytes are to be tested in the panel. #### Stat Profile Prime Plus Analyzer System Components: The Stat Profile Prime Plus Analyzer System is comprised of the following components. - . Stat Profile Prime Plus Analyzer System - Primary Sensor Cartridge ● - . Auxiliary Sensor Cartridge - Stat Profile Prime Plus Auto-Cartridge Quality Control Pack ● - . Stat Profile Prime Plus Calibrator Cartridge - Stat Profile Prime Plus External Ampuled Control ● - IFU/Labeling ● #### Sample Types: The Stat Profile Prime Plus Analyzer System accepts lithium heparinized arterial and venous whole blood. #### Measured Parameters: The Stat Profile Prime Plus Analyzer measures: - . Glucose (Glu) - . Creatinine (Creat) - . Blood Urea Nitrogen (BUN) #### Calculated Parameters: The following parameters are calculated by the Prime Plus Analyzer based on results of the directly measured parameters. - BUN/Creat Ratio (BUN/Creat) ● - Blood Osmolality (OSM) ● - Estimated Glomerular Filtration Rate (eGFR) ● #### Intended Use: The Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings for quantitative determination of Glucose, Creatinine, and Blood Urea Nitrogen, in heparinized arterial and venous whole blood. | Glucose (Glu) | Measurements are used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell tumor. | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Creatinine (Creat) | Measurements are used in the diagnosis and treatment of certain renal conditions and for monitoring adequacy of dialysis. | | Blood Urea Nitrogen (BUN) | Measurements are used in the diagnosis and treatment of certain renal and metabolic diseases. | {5}------------------------------------------------ #### Summary of the Technological Characteristics: The Stat Profile Prime Plus Analyzer is substantially equivalent to the previously cleared for market Stat Profile pHOx Ultra Analyzer System (K110648) in intended use. It uses the same sensor technology and measurement algorithms, and the formulations of the internal and external controls and the calibration cartridge are the same for the tested parameters. The External Control solutions for use with the Stat Profile Prime Plus Analyzer are substantially equivalent to those cleared for use with the predicate Stat Profile pHOx Ultra Analyzer System (K110648). See Table 1 for comparison of the predicate and proposed devices. #### Principles of Measurement: #### Glucose: Glucose measurement is based on the level of H₂O₂ produced during the enzymatic reaction between glucose and oxygen molecules in the presence of the glucose oxidase enzyme. At a constant potential of 0.70 volts, electroactive H3O2 is oxidized at the surface of the platinum anode. The current generated by the flow of electrons at the platinum electrode is proportional to the glucose concentration of the sample. #### Creatinine: The Prime Plus Creatinine sensor uses 3 enzymes. These 3 enzymes catalyze the conversion of Creatinine, ultimately forming formaldehyde, glycine, and hydrogen peroxide. At a constant potential of 0.70 volts, electroactive H2O2 is oxidized at the surface of the platinum anode. The current generated by the flow of electrons at the surface of the platinum electrode is proportional to the Creatinine concentration of the sample. #### BUN: The Prime Plus Analyzer uses urease, which has been chemically bonded to a membrane, to catalyze the conversion of urea present in the sample to ammonia and CO). At the pH of the sample, ammonia converts predominantly to the ammonium ion. An ammonium ion selective electrode is used to detect the ammonium formed by the above reactions. This measurement is then related to the concentration of urea present in the original sample via the Nernst equation. #### Summary of Performance Testing: Performance testing was completed to demonstrate that the Stat Profile Prime Plus Analyzer is substantially equivalent in performance, safety and efficacy to the Stat Profile pHOx Ultra Analyzer System. The performance testing included: #### Method Comparison Studies: A study was performed to compare the Stat Profile Prime Plus to the Nova Stat Profile pHOx Ultra analyzer to assess the equivalence of the analyzers in the measurement of Glucose, Creatinine, and BUN in heparinized whole blood in a clinical laboratory setting. The blood comparison data for Glucose, Creatinine, and BUN, for the Stat Profile Prime Plus analyzers meet the acceptance criteria. #### Precision/Reproducibility - Within Run and Run to Run Studies: Within Run and Run to Run precision was evaluated by replication studies performed on three Stat Profile Prime Plus analyzers. Within Run Precision testing consisted of one run of each of the following sample types and levels was performed, 20 replicates per run: - . Stat Profile Prime Plus Internal Controls: Levels 1-5 - Stat Profile Prime Plus Ampuled Controls: Levels 1-5 - Two whole bloods, sampled from syringes To assess Run to Run Precision for whole blood, triplicate analyses were performed on a single whole blood sample in ten separate runs during a single day. The systems were recalibrated before each triplicate run. {6}------------------------------------------------ The precision data for all parameters meet the within run imprecision specifications for the Stat Profile Prime Plus analyzers. The precision data for all parameters meet the between analyzer run to run imprecision specifications for the Stat Profile Prime Plus analyzers. #### Linearity Testing: The study assessed the linearity of all parameters to establish and/or verify the Analytical Measurement Range (AMR) for the Stat Profile Prime Plus Analyzer on whole blood. The linearity was measured using method comparison. The evaluation of the linear range included lower and upper limits of the AMR and various medical decision limits. All Stat Profile Prime Plus results were compared to the reference analyzer and/or the product specifications. The linearity comparison data for all parameters for the Stat Profile Prime Plus analyzers shows good correlation and linearity to the reference analyzers across the claimed measurement range for all parameters and met the acceptance criteria. #### Specificity / Interference Testing: The purpose of this study was to identify substances that may interfere with the Stat Profile Prime Plus sensors. If interference was identified, a dose response study was performed to determine the concentration where the interfering substance may alter results. This Interference Testing study was performed using whole blood collected in lithium heparin vacutainers. The possible interfering substances were tested at two analyte concentrations. Many substances were screened as potential interferents through analysis on the Prime Plus analyzers. Substance screening was completed according to the CLSI EP7-A2 quideline. The interfering substances identified during the screening process are listed in the table below: | Parameter | Interfering Substance | Concentration of interfering<br>substance | Interference | |------------|-----------------------|-------------------------------------------|--------------------------| | Glucose | Hydroxyurea | 0.2 mg/dL | No interference observed | | | | 0.4 mg/dL | Bias of 19.2% | | | Oxalate | 125 mg/dL | No interference observed | | | | 250 mg/dL | Bias of -13.9% | | | Thiocyanate | 3.4 mmol/L | No interference observed | | | | 5.1 mmol/L | Bias of 13.5% | | Creatinine | Hydroxyurea | 0.06 mg/dL | No interference observed | | | | 0.08 mg/dL | Bias of 15.5% | | | Thiocyanate | 1.7 mmol/L | No interference observed | | | | 3.4 mmol/L | Bias of 42.9% | #### Conclusion: The results of software validation and performance verification testing confirmed that the Stat Profile Prime Plus Analyzer is safe and effective for its intended purpose and that the Stat Profile Prime Plus Analyzer System is substantially equivalent to that of the predicate Stat Profile pHOx Ultra Analyzer System (K110648). {7}------------------------------------------------ ## Table 1: Comparison of Predicate and Proposed devices | Characteristic | Predicate: K110648 - Stat Profile pHOx Ultra Analyzer | Proposed: Stat Profile Prime Plus Analyzer | |------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indication For Use | The Stat Profile pHOx Ultra Analyzer without CO-Oximeter is intended for in vitro<br>diagnostic use by health care professionals and/or point-of-care usage in the<br>quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), Hemoglobin<br>(Hb) in heparinized whole blood; Na+, K'-, Cl-, Ca++, Mg++, Glucose (Glu), Lactate<br>(Lac), BUN (Urea), and Creatinine (Creat) in heparinized whole blood, serum, or<br>plasma.<br><br>The Stat Profile pHOx Ultra Analyzer with CO-Oximeter is intended for in vitro | The Stat Profile Prime Plus Analyzer System is indicated for use by<br>healthcare professionals in clinical laboratory settings for quantitative<br>determination of Glucose, Creatinine, and Blood Urea Nitrogen, in<br>heparinized arterial and venous whole blood. | | | diagnostic use by health care professionals and for point-of-care usage in the<br>quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), total<br>Hemoglobin (tHb), Oxyhemoglobin (O2Hb), Carboxyhemoglobin (COHb),<br>Methemoglobin (MetHb), Deoxyhemoglobin (HHb), and total bilirubin (tBil) in<br>heparinized whole blood; Na+, K'-, Cl-, Ca++, Mg++, Glucose (Glu), Lactate (Lac),<br>BUN (Urea), and Creatinine (Creat) in Heparinized whole blood, serum, or plasma.<br>Total Bilirubin (tBil) was not evaluated on neonatal samples. | | | Acceptable Samples | | | | Sample Types | Sodium or lithium heparinized whole blood, serum, or plasma samples from syringes,<br>open tubes, small cups, and capillary tubes. | Lithium heparin whole blood from syringes, open tubes, and small cups. | | Sample Volumes | 60-200µL (dependent on panel selected) | 135µL | | Measurement Range | | | | Glu | 15 - 500 mg/dL | 15 - 500 mg/dL | | Creat | 0.2 - 20.0 mg/dL | 0.2 - 12.0 mg/dL | | BUN | 3 - 100 mg/dL | 3 - 100 mg/dL | | Principles of<br>Measurement | | | | Glu | Enzymatic sensor | Same | | Creat | Impedance sensor | Same | | BUN | Enzymatic sensor | Same | | | | | | Touch Screen | 12.1" LCD, 1024x768 pixel, Resistive Touch | 10.1" WXGA 1280 x 800 color touch screen | | Menu | Fully configurable test menu based on available sensors | Same | | Bar Code Scanner | Internal Integrated 1D/2D | Same | | Printer | 2" Roll, Thermal Transfer | Same | | Pump | Peristaltic Pump w/ Pressure Plate, TPE Tubing (Pharmed BPT) | Same | | Analog Board | Precision low level analog front end w/ amperometric and potentiometric amplifiers,<br>air detector circuitry and temperature control circuitry | Same |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...