CardioChek Plus Test System, CardioChek Home Test System

K162282 · Polymer Technology Systems, Inc. D/B/A Pts Diagnostics · CGA · Dec 22, 2016 · Clinical Chemistry

Device Facts

Record IDK162282
Device NameCardioChek Plus Test System, CardioChek Home Test System
ApplicantPolymer Technology Systems, Inc. D/B/A Pts Diagnostics
Product CodeCGA · Clinical Chemistry
Decision DateDec 22, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The CardioChek Plus Test System is intended for multiple patient use in professional healthcare settings. The CardioChek Home Test System is intended for single-patient use.

Device Story

CardioChek Plus/Home Test System; point-of-care diagnostic analyzer; measures glucose, cholesterol, HDL from blood samples. Modifications include wireless/wired communication, software solutions, printer connectivity, MEMo chip update, battery change, multi-language support, and updated cleaning/disinfection instructions. Operated by healthcare professionals or patients; provides quantitative test results for clinical decision-making. System robustness validated for 11,000 cleaning/disinfection cycles using Super Sani-Cloth. Device maintains fundamental scientific technology of previously cleared predicates.

Clinical Evidence

Bench testing only. Verification and validation activities confirmed that the modified analyzers meet all predetermined acceptance criteria and perform as well as the predicate devices.

Technological Characteristics

Reflectance photometry; enzymatic dry-strip chemistry; lot-specific MEMo chip calibration; 4 AA battery power; wired/wireless connectivity; USB printer support; multi-language software; larger form factor than predicates.

Indications for Use

Indicated for quantitative determination of total cholesterol, HDL cholesterol, and glucose in capillary whole blood (Home system) or venous/capillary whole blood (Plus system). Home system intended for single-person use; Plus system intended for multiple-patient use in professional healthcare settings. Used for diagnosis/treatment of lipid, lipoprotein, and carbohydrate metabolism disorders, atherosclerosis, and liver/renal diseases.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a seal for the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 22, 2016 ### POLYMER TECHNOLOGY SYSTEMS, INC. D/B/A PTS DIAGNOSTICS MARGO ENRIGHT DIRECTOR OF REGULATORY AND CLINICAL AFFAIRS 7736 ZIONSVILLE ROAD INDIANAPOLIS IN 46268 Re: k162282 Trade/Device Name: Cardiochek Plus Test System, Cardiochek Home Test System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: CGA. NBW. CHH. LBR Dated: August 11, 2016 Received: August 15, 2016 Dear Ms. Enright: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Kellie B. Kelm -S for: Courtney H. Lias Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number (if known) K162282 Device Name CardioChek Home Test System #### Indications for Use (Describe) The CardioChek Home Test System (consisting of the CardioChek Home analyzer and CardioChek Home Chol+HDL+Glu test strips) is for the quantitative determination of total cholesterol, HDL (high density lipoprotein) cholesterol and glucose in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only. - Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. - HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes ● mellitus), atherosclerosis, and various liver and renal diseases. - Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders ● including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. A Chol/HDL ratio is calculated by the CardioChek Home analyzer. Type of Use (Select one or both, as applicable) _ Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {3}------------------------------------------------ 510(k) Number (if known) K162282 Device Name CardioChek Plus Test System #### Indications for Use (Describe) The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels Chol+HDL+Glu test strips) is for the quantitative determination of total cholesterol. HDL (high density lipoprotein) cholesterol and glucose in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only. - Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. - HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. - Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. A Chol/HDL ratio is calculated by the CardioChek Plus analyzer. | Type of Use (Select one or both, as applicable) | | | | | | | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|----------------------------------------------|--------------------------------------------------------------------------------------------------|-------------|---------------------------------------------| | <table><tr><td><table><tr><td><div> <span> <span style="text-decoration: line-through;">X</span> </span> </div></td><td>Prescription Use (Part 21 CFR 801 Subpart D)</td></tr></table></td></tr><tr><td><table><tr><td><div></div></td><td>Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table></td></tr></table> | <table><tr><td><div> <span> <span style="text-decoration: line-through;">X</span> </span> </div></td><td>Prescription Use (Part 21 CFR 801 Subpart D)</td></tr></table> | <div> <span> <span style="text-decoration: line-through;">X</span> </span> </div> | Prescription Use (Part 21 CFR 801 Subpart D) | <table><tr><td><div></div></td><td>Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <div></div> | Over-The-Counter Use (21 CFR 801 Subpart C) | | <table><tr><td><div> <span> <span style="text-decoration: line-through;">X</span> </span> </div></td><td>Prescription Use (Part 21 CFR 801 Subpart D)</td></tr></table> | <div> <span> <span style="text-decoration: line-through;">X</span> </span> </div> | Prescription Use (Part 21 CFR 801 Subpart D) | | | | | | <div> <span> <span style="text-decoration: line-through;">X</span> </span> </div> | Prescription Use (Part 21 CFR 801 Subpart D) | | | | | | | <table><tr><td><div></div></td><td>Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <div></div> | Over-The-Counter Use (21 CFR 801 Subpart C) | | | | | | <div></div> | Over-The-Counter Use (21 CFR 801 Subpart C) | | | | | | ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) page 4 {4}------------------------------------------------ 510(k) Number (if known) K162282 Device Name CardioChek Home Test System Indications for Use (Describe) The CardioChek Home Test System (consisting of the CardioChek Home analyzer and CardioChek Home Chol +HDL test strips) is for the quantitative determination of total cholesterol and HDL (high density lipoprotein) cholesterol in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only. - Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol ● in the blood and lipid and lipoprotein metabolism disorders. - HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes ● mellitus), atherosclerosis, and various liver and renal diseases. A Chol/HDL ratio is calculated by the CardioChek Home analyzer. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|------------------------------------------------| | | ☐ Prescription Use (Part 21 CFR 801 Subpart D) | | ☑ | Over-The-Counter Use (21 CFR 801 Subpart C) | ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Page 5 {5}------------------------------------------------ 510(k) Number (if known) #### K162282 Device Name CardioChek Plus Test System #### Indications for Use (Describe) The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels Chol+HDL test strips) is for the quantitative determination of total cholesterol and HDL (high density lipoprotein) cholesterol in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only. - Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. - HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. A Chol/HDL ratio is calculated by the CardioChek Plus analyzer. Type of Use (Select one or both, as applicable) | <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | |-------------------------------------------------------------------------------------------------| | □ Over-The-Counter Use (21 CFR 801 Subpart C) | ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Page 6 {6}------------------------------------------------ 510(k) Number (if known) K162282 Device Name CardioChek Home Test System #### Indications for Use (Describe) The CardioChek Home Test System (consisting of the CardioChek Home analyzer and CardioChek Home Chol+Glu test strips) is for the quantitative determination of total cholesterol and glucose in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only. - Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol ● in the blood and lipid and lipoprotein metabolism disorders. - Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. | Type of Use (Select one or both, as applicable) | | | |-------------------------------------------------|------------------------------------------------|--| | | ☐ Prescription Use (Part 21 CFR 801 Subpart D) | | | | ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | | ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {7}------------------------------------------------ 510(k) Number (if known) #### K162282 Device Name CardioChek Plus Test System #### Indications for Use (Describe) The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels Chol+Glu test strips) is for the quantitative determination of total cholesterol and glucose in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only. - Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. - Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------| | | <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {8}------------------------------------------------ ### SECTION 5: 510(k) SUMMARY- k162282 This summary of safety and effectiveness information is submitted in compliance with 21CFR 807.92. December 9. 2016 ### 1. Submitter Information/FacilityAddress: Polymer Technology Systems, Inc. 7736 Zionsville Road Indianapolis, IN 46268 2. Contact Person: MargoEnright, RAC Phone Number: 317-870-5610 x1012 email: menright@ptsdiagnostics.com ### 3. Trade Names: - CardioChek Plus Test System - CardioChek Home Test System Components: - CardioChek Plus analyzer - CardioChek Home analyzer - PTS Panels Chol+HDL+Glu test strips - PTS Panels Chol+HDL test strips - PTS Panels Chol+Glu test strips - CardioChek Home Chol+HDL+Glu test strips - CardioChek Home Chol+HDL test strips - CardioChek Home Chol+Glu test strips ### 4. Regulatory Information CardioChek Plus Test System | Product Code | Classification | Regulation Section | Panel | |--------------|---------------------------------------------------------------------|-------------------------------------------------------|-------------------| | CGA | Class II | 21 CFR 862.1345<br>Glucose test system | Chemistry<br>(75) | | CHH | Class I, meets the<br>limitation of exemption 21<br>CFR 862.9(c)(4) | 21 CFR 862.1175<br>Cholesterol (Total)<br>test system | Chemistry<br>(75) | | LBR | Class I, meets the<br>limitation of exemption 21<br>CFR 862.9(c)(4) | 21 CFR 862.1475<br>Lipoprotein test<br>system | Chemistry<br>(75) | {9}------------------------------------------------ CardioChek Plus Home Test System | Product Code | Classification | Regulation Section | Panel | |--------------|---------------------------------------------------------------------|-------------------------------------------------------|-------------------| | NBW | Class II | 21 CFR 862.1345<br>Glucose test system | Chemistry<br>(75) | | CGA | Class II | 21 CFR 862.1345<br>Glucose test system | Chemistry<br>(75) | | CHH | Class I, meets the<br>limitation of exemption 21<br>CFR 862.9(c)(4) | 21 CFR 862.1175<br>Cholesterol (Total)<br>test system | Chemistry<br>(75) | | LBR | Class I, meets the<br>limitation of exemption 21<br>CFR 862.9(c)(4) | 21 CFR 862.1475<br>Lipoprotein test<br>system | Chemistry<br>(75) | ### 5. Device Description: The PTS Panels Chol+HDL+Glu, Chol+HDL and Chol+Glu test strips and the CardioChek Home Chol+HDL+Glu, Chol+HDL and Chol+Glu test strips are used with the CardioChek Plus and CardioChek Home analyzers to measure total cholesterol, HDL cholesterol and glucose in whole blood. The test strips utilize enzymatic methods on a dry strip that is read by reflectance photometry. These test strips are for in vitro diagnostic use only. ### 6. Intended Use: The CardioChek Home Test System (consisting of the CardioChek Home analyzer and CardioChek Home Chol+HDL+Glu test strips) is for the quantitative determination of total cholesterol, HDL (high density lipoprotein) cholesterol and glucose in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only. · Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. · HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. · Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. A Chol/HDL ratio is calculated by the CardioChek Home analyzer. {10}------------------------------------------------ The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels Chol+HDL+Glu test strips) is for the quantitative determination of total cholesterol, HDL (high density lipoprotein) cholesterol and glucose in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only. · Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. · HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. · Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. A Chol/HDL ratio is calculated by the CardioChek Plus analyzer. The CardioChek Home Test System consisting of the CardioChek Home analyzer and CardioChek Home Chol+HDL test strips) is for the quantitative determination of total cholesterol and HDL (high density lipoprotein) cholesterol in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only. - · Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. - HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. A Chol/HDL ratio is calculated by the CardioChek Home analyzer. {11}------------------------------------------------ The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels Chol+HDL test strips) is for the quantitative determination of total cholesterol and HDL (high density lipoprotein) cholesterol in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only. > • Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. · HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. A Chol/HDL ratio is calculated by the CardioChek Plus analyzer. The CardioChek Home Test System (consisting of the CardioChek Home analyzer and CardioChek Home Chol+Glu test strips) is for the quantitative determination of total cholesterol and glucose in capillary whole blood from the fingertip and is intended to be used by a single person and should not be shared. This system is for in vitro diagnostic use only. > • Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. > · Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. The CardioChek Plus Test System (consisting of the CardioChek Plus analyzer and PTS Panels Chol+Glu test strips) is for the quantitative determination of total cholesterol and glucose in venous whole blood and capillary whole blood from the fingertip and is intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only. > · Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. > · Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. {12}------------------------------------------------ ### 7. Reason for Special510(k): Device Modifications The CardioChek Plus Test System and the CardioChek Home Test System are modifications of the original devices (cleared in k071507, k071593 and k041750) to include the following: - Wireless communications capability (professional system) ● - Software solutions capability ● - Printer connectivity capability ● - MEMo chip appearance ● - Battery type - Multiple language software capability - Wired PC Communication ● - Analyzer dimensions ● The CardioChek Plus Test System and the CardioChek Home Test System include the PTS Panels test strips for use with the CardioChek Plus analyzer and the CardioChek Home test strips for use with the CardioChek Home analyzer. The test strips are unmodified from K041750, K071507 and K071593. The three test strips are: Chol+HDL+Glu, Chol+HDL and Chol+Glu. The current trade names for each are: - a. PTS Panels Chol+HDL+Glu test strips - b. PTS Panels Chol+HDL test strips - c. PTS Panels Chol+Glu test strips - d. CardioChek Home Chol+HDL+Glu test strips - e. CardioChek Home Chol+HDL test strips - f. CardioChek Home Chol+Glu test strips ### 8. Statement of Substantial Equivalence The CardioChek Plus Test System and CardioChek Home Test System are substantially equivalent to the predicates (k071507, k071593 and k041750). ## 9. Predicate Device Information ### Predicates Name: PTS Panels Chol+HDL+Glu Test Strips Device Company: Polymer Technology Systems, Inc. (PTS) 510(k) Number: K071507 Name: PTS Panels Chol+HDL Test Strips Device Company: Polymer Technology Systems, Inc. (PTS) 510(k) Number: K071593 {13}------------------------------------------------ Name: PTS Panels Chol+Glu Test Strips Device Company: Polymer Technology Systems, Inc. (PTS) 510(k) Number: K041750 # 10. Similarities and Differences ### Similarities - Both systems employ reflectance photometry. ● - Both systems provide results that correlate to reference methods. - Both test systems use the same test strips. - Both systems require a lot specific memory chip for result calculation, which are included in ● the same package with the test strips ### Differences - The software for the CardioChek Plus and CardioChek Home analyzers have multiple . language options. - The CardioChek Plus and CardioChek Home analyzers provide: ● - Software solutions capability - - Wired PC communication capability - - Printer connectivity capability via a USB - - Wireless communication capability (professional analyzer) - - . The MEMo chip appearance is changed to make it more aesthetically pleasing. (The printed circuit board inside is the unmodified.) - Analyzer outer case is larger. ● - The modified analyzers use 4 AA batteries; the BioScanner Plus uses 2 AAA ● batteries. ### 11. Verification and Validation Summary As required by the risk analysis, all verification and validation activities were performed and the results demonstrated that the predetermined acceptance criteria were met. ### 12. Conclusion The intended use of the device is unchanged. The CardioChek Plus and CardioChek Home test systems are as safe and effective and perform as well as the predicate devices: k071507, k071593 and k041750.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...