CoreLink Navigation Instruments

K214059 · Corelink, LLC · OLO · Jan 19, 2022 · Neurology

Device Facts

Record IDK214059
Device NameCoreLink Navigation Instruments
ApplicantCorelink, LLC
Product CodeOLO · Neurology
Decision DateJan 19, 2022
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.4560
Device ClassClass 2
AttributesTherapeutic

Intended Use

CoreLink Navigation Instruments are intended to be used in the preparation and placement of Tiger OCT screw implants during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System S8 (V1.2.20), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

Device Story

CoreLink Navigation Instruments are non-sterile, reusable stainless steel manual surgical instruments; designed to interface with Medtronic StealthStation® System S8 for spinal surgery. Instruments assist surgeons in precisely locating anatomical structures during preparation and placement of Tiger OCT screw implants. Used in open or minimally invasive procedures. Surgeon operates instruments while tracking via StealthStation® navigation system. Output is visual guidance on navigation system display, allowing surgeon to correlate instrument position with patient anatomy. Benefits include increased precision in screw placement, potentially reducing surgical complications.

Clinical Evidence

Bench testing only. Confirmatory positional accuracy testing conducted per ASTM F2554 using Medtronic StealthStation® System S8 (V1.2.0). Engineering analysis, including tolerance stack analysis, performed to demonstrate accuracy.

Technological Characteristics

Non-sterile, reusable manual surgical instruments; stainless steel construction. Designed for use with Medtronic StealthStation® System S8. Positional accuracy testing performed per ASTM F2554.

Indications for Use

Indicated for use in spinal surgery for preparation and placement of Tiger OCT screw implants in open or minimally invasive procedures; intended for patients requiring stereotactic surgery where anatomical structures can be referenced to CT, MR, fluoroscopy, or digitized landmarks.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The logos are displayed side-by-side. January 19, 2022 CoreLink, LLC % Nathan Wright Engineer & Regulatory Specialist Empirical Testing Corp. 4628 Northpark Drive Colorado Springs, Colorado 80918 Re: K214059 Trade/Device Name: CoreLink Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: December 23, 2021 Received: December 27, 2021 ## Dear Nathan Wright: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K214059 Device Name CoreLink Navigation Instruments #### Indications for Use (Describe) CoreLink Navigation Instruments are intended to be used in the preparation and placement of Tiger OCT screw implants during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System S8 (V1.2.20), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | Submitter's Name: | CoreLink, LLC | |----------------------------|------------------------------------------------------------------------------------------| | Submitter's Address: | 2072 Fenton Logistics Park<br>St. Louis, Missouri 63026 | | Submitter's Telephone: | 888-349-7808 | | Contact Person: | Nathan Wright MS<br>Empirical Testing Corp.<br>719-351-0248<br>nwright@empiricaltech.com | | | Image: Empirical Testing Corp. Logo | | Date Summary was Prepared: | December 23, 2021 | | Trade or Proprietary Name: | CoreLink Navigation Instruments | | Common or Usual Name: | Orthopedic Stereotaxic Instruments | | Classification: | Class II per 21 CFR §882.4560 | | Product Code: | OLO | | Classification Panel: | Orthopedic - Spinal Devices (DHT6B) | # K214059 510(K) SUMMARY ## DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION: CoreLink Navigation Instruments are non-sterile, re-usable manual surgical instruments made from stainless steel. These instruments are designed to interface with the Medtronic StealthStation® Navigation System to assist surgeons in precisely locating anatomical structures. The purpose of this submission is to add instruments to the previously cleared CoreLink Navigation Instruments set to offer compatibility with CoreLink OCT pedicle screws. # INDICATIONS FOR USE CoreLink Navigation Instruments are intended to be used in the preparation and placement of Tiger OCT Screw implants during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System S8 (V1.2.0), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. # TECHNOLOGICAL CHARACTERISTICS The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are identical between the subject and predicates: - . Indications for use - Materials of manufacture ● - . Principles of operation {4}------------------------------------------------ | 510k<br>Number | Trade or Proprietary or Model Name | Manufacturer | Predicate<br>Type | |----------------|-------------------------------------------------------------|-----------------------------------|-------------------| | K212825 | CoreLink Navigation Instruments | CoreLink, LLC | Primary | | K181111 | Navigated INFINITY™ Instruments | Medtronic Sofamor Danek USA, Inc. | Additional | | K200863 | Tiger Occipital-Cervical-Thoracic Spinal<br>Fixation System | CoreLink, LLC | Reference | ### Table 5-1 Predicate Devices # PERFORMANCE DATA In support of this Special 510(k) submission, CoreLink, LLC has conducted confirmatory positional accuracy testing per ASTM F2554 with Medtronic StealthStation® System S8 Version 1.2.0 (1.2.0-20) as well as engineering analysis including tolerance stack analysis to demonstrate that the system modifications provide adequate and substantially equivalent accuracy for their intended use. ## CONCLUSION The overall technology characteristics and performance data lead to the conclusion that the CoreLink Navigation Instruments are substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...