The Mazor X is indicated for precise positioning of surgical implants during general spinal and brain surgery. It may be used in either open or minimally invasive or percutaneous procedures. Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic projections from standard C-Arms into volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects.
Device Story
Mazor X is a stereotaxic surgical guidance system for spine and brain procedures. It accepts inputs from pre-operative CT scans, intra-operative 3D scans, or external 3D imaging systems. The system enables surgical planning and virtual placement of instruments/implants; it then guides the surgeon to deploy accessories at precise anatomical locations. Used in OR settings by surgeons. The system includes 3D imaging capabilities that process 2D fluoroscopic projections from standard C-Arms into volumetric 3D images for visualization of high-contrast objects. It assists in accurate implant positioning, potentially improving surgical precision and patient outcomes.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including software validation per IEC 62304 and FDA guidance, and measurement comparison testing between the X Align module and Surgimap.
Technological Characteristics
Stereotaxic instrument and radiological image processing system. Features 3D imaging capability for processing 2D fluoroscopic projections into volumetric 3D images. Software-driven guidance system. Validated per IEC 62304 standards.
Indications for Use
Indicated for patients undergoing general spinal and brain surgery requiring precise positioning of surgical instruments or implants. Applicable for open, minimally invasive, or percutaneous procedures.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
K200935 — Mazor X · Mazor Robotics , Ltd. · May 8, 2020
K203005 — Mazor X · Mazor Robotics , Ltd. · Oct 27, 2020
K152041 — Renaissance X System · Mazor Robotics , Ltd. · Nov 17, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, featuring three abstract human profiles connected by flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 4, 2017
Mazor Robotics Ltd. % Ahava Stein Regulatory Affairs Consultant A. Stein - Regulatory Affairs Consulting Ltd. 20 Hatas St., (Beit Hapaamon, Suite 102) Kfar Saba. IL 4442520
Re: K163221
Trade/Device Name: Mazor X Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, HAW, LLZ Dated: March 8, 2017 Received: March 13, 2017
Dear Ahava Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name Mazor X
K163221
vice Name
zor X
Indications for Use (Describe)
The Mazor X is indicated for precise positioning of surgical implants during general spinal and brain surgery. It may be used in either open or minimally invasive or percutaneous procedures.
Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic projections from standard C-Arms into volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
## K163221
## (Premarket Notification [510(k)] Number)
### 1. Submitter Information
| Manufacturer Name and Address | Mazor Robotics Ltd. |
|-------------------------------|-----------------------------------------------|
| | PO Box 3104, |
| | 7 HaEshel St., |
| | Caesarea Park South 3088900, |
| | Israel |
| Official Correspondent | Ahava Stein and Moshe Rosenberg |
| | A. Stein - Regulatory Affairs Consulting Ltd. |
| | 20 Hata'as St. (Beit Hapaamon, Suite 102) |
| | Kfar Saba 4442520, |
| | Israel |
- 2. Date Prepared: November 13, 2016
- 3. Device Name Mazor X
| Proprietary Name: | Mazor X |
|--------------------------|------------------------------------------------------------------------------------------------|
| Common Name: | Combination of:<br>1. Stereotaxic instrument; and<br>2. System, Image Processing, Radiological |
| FDA Classification Name: | 21 CFR 882.4560; Stereotaxic instrument |
| FDA Classification: | Class II, Product Code OLO, HAW and LLZ |
### 4. Predicate Devices
The Mazor X is substantially equivalent to the following devices:
| Manufacturer | Device | 510(k) | Date Cleared |
|---------------------|----------------------|---------|--------------------|
| Mazor Robotics Ltd. | Renaissance X System | K152041 | November 17, 2015 |
| Nemaris Inc. | Surgimap | K111019 | September 30, 2011 |
{4}------------------------------------------------
### 5. Device Description
The Mazor X enables the surgeon to precisely position surgical instruments and/or implants (in spinal and brain surgery). The Mazor X enables guidance for spine and brain procedures and intra-operative 3D image processing capabilities. The planning of the surgical procedure and virtual placement of surgical instruments and/or implants (e.g., a screw) can be achieved through pre-operation planning based on the patient's CT scan or intra-operative planning based on Mazor X 3D Scan image or on a 3D image uploaded from an external 3D image acquiring system. The Mazor X enables accurate deployment of surgical accessories in the precise anatomical location according to predefined planning. With the imaging capabilities of the system, the user can also visualize the implants on the patients CT. The Mazor X is a device modification of the original Renaissance X System cleared in 510(k) K152041.
#### 6. Indications for Use
The Mazor X is indicated for precise positioning of surgical instruments or spinal implants during general spinal and brain surgery. It may be used in either open or minimally invasive or percutaneous procedures.
Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic projections from standard C-Arms into volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects.
### Performance Standards 7.
There are no performance standards under the Federal Food, Drug and Cosmetic Act, for the Mazor X.
### 8. Performance Testing
The following Performance tests were performed on the Mazor X:
- Software validation testing in accordance with the FDA Guidance for the Premarket . Submissions for Software Contained in Medical Devices (May 11, 2005) and the IEC
{5}------------------------------------------------
62304 standard for Medical Device Software - Software Life Cycle Processes. The software validation tests demonstrate that the modified software version meets its design requirements.
- Measurement Comparison testing comparison testing of the X Align module and the . Surgimap measurement calculations.
### Technological Characteristics Compared to Predicate Device 9.
The differences in the software do not adversely affect the safety, effectiveness or performance of the Mazor X. The Mazor X was tested according to the aforementioned validation and performance tests (detailed in section 8) and found compliant.
The technological characteristics, e.g., overall design, materials, mechanism of action, mode of operation, performance characteristics, etc., and the indications for use of the Mazor X are substantially equivalent to the predicate devices cited above.
#### 10. Conclusion
The performance testing and comparison to the predicate devices demonstrate that the Mazor X is as safe, as effective and performs as well as the legally marketed Renaissance X System and Surgimap predicate devices. Therefore, the Mazor X is substantially equivalent to the predicate devices and may be cleared for marketing.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.