Mazor X

K200935 · Mazor Robotics , Ltd. · OLO · May 8, 2020 · Neurology

Device Facts

Record IDK200935
Device NameMazor X
ApplicantMazor Robotics , Ltd.
Product CodeOLO · Neurology
Decision DateMay 8, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.4560
Device ClassClass 2

Intended Use

The Mazor X is indicated for precise positioning of surgical implants during general spinal surgery. It may be used in open or minimally invasive or percutaneous procedures. Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic prom standard C-Arms into volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects. The Mazor X navigation tracks the position of instruments, during spinal surgery, in relation to the surgical anatomy and identifies this position on diagnostic or intraoperative images of a patient.

Device Story

Mazor X is a stereotaxic surgical guidance system for spine procedures. It integrates intra-operative 3D image processing with surgical navigation. Input data includes 2D fluoroscopic projections from standard C-Arms, pre-operative CT scans, or external 3D images. The system processes these inputs to generate volumetric 3D images of high-contrast objects. During surgery, the system tracks the position of instruments relative to patient anatomy, displaying this on diagnostic or intra-operative images. Surgeons use the system for pre-operative or intra-operative planning and virtual placement of implants (e.g., screws). The device enables precise deployment of surgical accessories and implants at predefined anatomical locations. Used in clinical settings by surgeons, the system provides visual guidance that assists in accurate instrument positioning, potentially improving surgical outcomes and reducing procedural complexity.

Clinical Evidence

No clinical data. Performance was established through bench testing, including software validation (per FDA guidance and IEC 62304) and hardware verification of modified components against established specifications.

Technological Characteristics

Stereotaxic instrument and image processing system. Components include a workstation, bed frame, and bone mount interface. Software-based 3D image reconstruction from 2D fluoroscopic projections. Connectivity supports external 3D image acquisition. Software validation follows IEC 62304 standards.

Indications for Use

Indicated for precise positioning of surgical instruments or spinal implants during general spinal surgery in open, minimally invasive, or percutaneous procedures.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 8, 2020 Mazor Robotics Ltd. Shiran Conforti RA Manager 5 Shacham St. North Industrial Park Caesarea, 3079567 Israel Re: K200935 Trade/Device Name: Mazor X Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, LLZ Dated: April 7, 2020 Received: April 8, 2020 Dear Shiran Conforti: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200935 Device Name Mazor X # Indications for Use (Describe) The Mazor X is indicated for precise positioning of surgical implants during general spinal surgery. It may be used in open or minimally invasive or percutaneous procedures. Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic prom standard C-Arms into volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects. The Mazor X navigation tracks the position of instruments, during spinal surgery, in relation to the surgical anatomy and identifies this position on diagnostic or intraoperative images of a patient. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # SUMMARY OF SAFETY AND EFFECTIVENESS K200935 (Premarket Notification [510(k)] Number) ## 1. Submitter Information | Manufacturer Name and Address | Official Correspondent | |-------------------------------|------------------------------| | Mazor Robotics Ltd. | Shiran Conforti | | PO Box 3104, | Mazor Robotics Ltd. | | 5 Shacham St., | PO Box 3104, | | Caesarea Park North 3079567, | 5 Shacham St., | | Israel | Caesarea Park North 3079567, | | | Israel | ### 2. Date Prepared: April 07, 2020 - 3. Device Name Mazor X | Proprietary Name: | Mazor X | |--------------------------|------------------------------------------------------------------------------------------------| | Common Name: | Combination of:<br>1. Stereotaxic instrument; and<br>2. System, Image Processing, Radiological | | FDA Classification Name: | 21 CFR 882.4560; Stereotaxic instrument | # FDA Classification: Class II, Product Code OLO and LLZ ## 4. Predicate Devices The Mazor X is substantially equivalent to the following device: | Manufacturer | Device | 510(k) | Date Cleared | |---------------------|------------------------------------------|---------|------------------| | Mazor Robotics Ltd. | Mazor X System (Mazor X Stealth Edition) | K182077 | November 2, 2018 | {4}------------------------------------------------ ### ട. Device Description The modified Mazor X hosts guidance for spine procedures and intra-operative 3D image processing capabilities. It enables the surgeon to precisely position surgical instruments and/or implants. The planning of the surgical procedure and virtual placement of surgical instruments and/or implants (e.g., a screw) can be achieved through pre-operation planning based on the patient's CT scan or intra-operative planning based on Mazor X 3D Scan image or on a 3D image uploaded from an external 3D image acquiring system. The Mazor X enables accurate deployment of surgical accessories in the precise anatomical location according to predefined planning. With the imaging capabilities of the system, the user can also visualize the implants on the patients CT. The Mazor X is a device modification of the original Mazor X System cleared in 510(k) K182077. #### 6. Indications for Use The Mazor X is indicated for precise positioning of surgical instruments or spinal implants during general spinal surgery. It may be used in open or minimally invasive or percutaneous procedures. Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic projections from standard C Arms into volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects. The Mazor X navigation tracks the position of instruments, during spinal surgery, in relation to the surgical anatomy and identifies this position on diagnostic or intraoperative images of a patient. #### 7. Performance Standards There are no performance standards under the Federal Food, Drug and Cosmetic Act, for the Mazor X. {5}------------------------------------------------ #### 8. Performance Testing The following Performance tests were performed on the modified Mazor X system: - · Software validation testing in accordance with the FDA Guidance for the Premarket Submissions for Software Contained in Medical Devices (May 11, 2005) and the IEC 62304 Standard for Medical Device Software - Software Life Cycle Processes. The software validation tests demonstrate that the Mazor X software version meets the design requirements. Test cases were designed for testing procedure simplicity, system startup, security, user interfaces, diagnostics and error handling, performance and robustness, installation, and database. - · Hardware changes verification: the system parts and accessories that were modified were tested to verify that they meet the requirements. The test protocols that were used are identical to the test protocols used to verify and validated the same parts for the cleared Mazor X. #### 9. Technological Characteristics Compared to Predicate Device The device modifications included modified labeling, modified software with minor software changes and a slightly modified Mazor X System (minor hardware changes). The software changes included SW optimization within the established specifications, enhanced functionalities as well as screen enhancements (graphical enhancements and enhanced information presented on screen). The modified Mazor X workstation, Bed Frame and Mazor X Bone Mount interface - are very similar to the cleared Mazor X workstation Bed Frame and Mazor X Bone Mount interface with some minor technical and design improvements. The modifications do not adversely affect the safety, effectiveness and performance of the Mazor X system. The Mazor X system was tested according to the aforementioned validation and performance tests and found compliant. The technological characteristics, e.g., overall design, materials, mechanism of action, mode of operation, performance characteristics, etc., of the Mazor X system are substantially equivalent to the predicate device cited above. {6}------------------------------------------------ ## Conclusion 10. The performance testing and comparison to the predicate device demonstrate that the Mazor X system is as safe, as effective and performs as well as the legally marketed Mazor X System predicate device. Therefore, the Mazor X system is substantially equivalent to the Mazor X System cleared under K182077.